K180605 · Bionix Development Corporation · KNT · Oct 24, 2018 · Gastroenterology, Urology
Device Facts
Record ID
K180605
Device Name
GastroFlush
Applicant
Bionix Development Corporation
Product Code
KNT · Gastroenterology, Urology
Decision Date
Oct 24, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5980
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
GastroFlush is indicated for use in clearing occlusions/clogs from enteral feeding tubes placed for nutrition and/or decompression in adult or adolescent patients that have tube size 8Fr-24Fr and tube length no longer than 42 inches. It is intended to be used by or under the direction of a licensed physician.
Device Story
GastroFlush is a device for clearing occlusions in enteral feeding tubes (G-tubes, NG tubes). It connects to a standard jet lavage unit to receive a pressurized pulsatile water stream. The device uses a system of one-way valves to ensure unidirectional flow into the clogged tube and a pressure relief valve to prevent overpressure, directing excess water to an overflow basin. A secondary port allows for patency checks using a syringe. The device utilizes ENFit connectors (ISO 80369-3) to prevent misconnections. It is used in clinical or home settings by or under the direction of a physician. By safely dislodging clogs with water pressure, the device restores tube function, avoiding the need for tube replacement and benefiting the patient by maintaining continuous access for nutrition or decompression.
Clinical Evidence
No clinical testing was performed. Bench testing only. Testing included tube clearance efficacy (n=30 per group, 100% success rate at 95% C/R), leak-free pressure testing (up to 30 PSI), and overpressure valve testing (mean opening pressure 12.71 PSI).
Technological Characteristics
Device consists of fluid conduits, one-way valves, and a pressure relief valve. Connectors comply with ISO 80369-3 (ENFit). Operates via external jet lavage unit (water source). Meets AAMI/ANSI ID54:1996/(R) 2012 standards. Single-patient use.
Indications for Use
Indicated for clearing occlusions/clogs from enteral feeding tubes (8Fr-24Fr, max 42 inches) used for nutrition or decompression in adult or adolescent patients. Intended for use by or under the direction of a licensed physician.
Regulatory Classification
Identification
A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).
Special Controls
*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.
Predicate Devices
TubeClear In-Patient Tube Clearing System (K121571)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.
October 24, 2018
Bionix Development Corporation James Huttner Vice President, Product Development 5154 Enterprise Blvd. Toledo, Ohio 43612
Re: K180605
Trade/Device Name: GastroFlush Regulation Number: 21 CFR 876.5980 Regulation Name: Gastrointestinal Tube And Accessories Regulatory Class: Class II Product Code: KNT, PIF Dated: September 12, 2018 Received: September 18, 2018
Dear James Huttner:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Daniel G. Walter Jr -S
for
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K180605
Device Name GastroFlush
Indications for Use (Describe)
GastroFlush is indicated for use in clearing occlusions/clogs from enteral feeding tubes placed for nutrition and/or decompression in adult or adolescent patients that have tube size 8Fr-24Fr and tube length no longer than 42 inches.
It is intended to be used by or under the direction of a licensed physician.
| Type of Use (Select one or both, as applicable) | | | | | |
|-------------------------------------------------|---------------------------------------------|--|--|--|--|
| X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) | | | | |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
K180605 page 1 of 7
# 510(K) Summary:
Image /page/3/Picture/2 description: The image shows the logo for Bionix. The logo consists of a blue circle with a white "B" inside, followed by the word "BIONIX" in blue, sans-serif font. A small registered trademark symbol is located to the right of the "X".
Corporate Office 5154 Enterprise Blvd., Toledo, Ohio 43612 Phone 419.727.8421 • Fax 419.727.8426
October 12, 2018
2.
The following 510(k) summary has been prepared pursuant to requirements specified in 21CFR 807.92
#### 1. Submitter Information:
| a. | Applicant: | Bionix Development Corporation<br>5154 Enterprise Blvd.<br>Toledo, Ohio 43612 | | |
|--------------|----------------------|-------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|--|
| b. | Contact: | | James Huttner M.D., Ph.D.<br>Vice President, New Product Development<br>Phone: (419) 727-8421<br>Fax: (419) 727-4430<br>Email: jhuttner@bionix.com | |
| Device Name: | | | | |
| a. | Trade Name: | GastroFlush | | |
| b. | Common Name: | Enteral Tube Un-Clogger | | |
| c. | Classification Name: | Gastrointestinal tube and accessories | | |
| d. | Regulation Numer: | 876.5980 | | |
| e. | Product Code(s) : | KNT; PIF | | |
{4}------------------------------------------------
#### 3. Indications for Use:
GastroFlush is indicated for use in clearing occlusions/clogs from enteral feeding tubes placed for nutrition and/or decompression in adult or adolescent patients that have tube size 8Fr-24Fr and tube length no longer than 42 inches.
It is intended to be used by or under the direction of a licensed physician.
#### 4. Substantial Equivalence Device(s):
The GastroFlush enteral tube un-clogger is similar in intended use and function to the following devices used to unclog plugged enteral tubes such as G-tubes (gastrostomy tubes). J-tubes (jejunostomy tubes), nasogastric feeding tubes, and nasogastric sumps or decompression tubes:
- a. TubeClear In-Patient Tube Clearing System, manufactured and legally marketed by Actuated Medical, Bellefonte, PA. This device is listed under regulation number 876.5980, Tubes, Gastrointestinal and Accessories, and is classified as a Class II device. This device has been cleared by the FDA under K121571.
- b. DeClogger, manufactured and legally marketed by Bionix Development Corporation, Toledo, Ohio. This device is listed under regulation number 876.5980, Tubes, Gastrointestinal and Accessories, and is classified as a Class II device. This device has been cleared by the FDA under K905164.
#### 5. Device Description
The GastroFlush device uses a pulsatile stream of pressurized water to gently dislodge and flush clogs from obstructed G-tubes, NG tubes and other tubes placed for the purposes of enteral nutrition and/or gastrointestinal decompression in the clinical or home setting. A series of connected one-way valves is used to ensure that the pressurized water flows in only the correct direction. Pressure relief valves with specified "cracking pressure" are used to ensure that overpressure of the system is avoided. Lastly, specialty connectors including the use of the new ENFit enteral feeding tube connectors conforming to ISO 80369-3 ensure against the inadvertent misconnection of the GastroFlush device to the clogged enteral tube, or the use of Luer syringes and/or similar devices in the attempt to unclog the enteral tube.
A schematic representation of the GastroFlush device is shown below to better describe the directional water flow of the device.
{5}------------------------------------------------
Image /page/5/Figure/1 description: The image shows a schematic diagram of a water input system. The system includes a water input, a specialty connector, a one-way valve, and a female ENFit connector that leads to an enteral tube. Additionally, there is a pressure relief valve and a one-way valve that connects to a male ENFit connector for syringe access to check patency, and an overflow outlet.
Figure 1 Schematic Drawing Showing GastroFlush Water Flow
From the diagram, water flows in a unidirectional fashion through the GastroFlush and into the connected clogged enteral tube. Directional water flow is maintained using a one-way valve (Valve 1) that permits water flow only towards the enteral tube and prevents back-flow of potentially contaminated water.
To prevent potentially hazardous high-pressure that could result from attempts to clear a clogged enteral tube, a safety outlet conduit is provided with an integrated pressure relief valve. The cracking pressure of this valve is set high enough to allow sufficient pressure to unclog the enteral tube, but well below the burst pressure of the enteral tube, the GastroFlush tubing, or any of its components. Water exiting through the pressure relief valve is directed to an overflow basin.
To test the enteral tube for patency after performing the de-clogging procedure, a port is provided to allow a flush of water from a syringe. A second one-way valve (Valve 2) maintains unidirectional flow through this tube and prevents back-flow and/or high-pressure water exiting this port.
Specialty connectors have been used throughout the GastroFlush to ensure proper directional connection to the water delivery unit. In accordance with the new ISO standard 80369-3, ENFit connectors are used to connect the GastroFlush device to the enteral tube, and to allow an ISO
{6}------------------------------------------------
80369-3 ENFit compatible syringe to connect to the patency check port. This prevents the inadvertent connection of a syringe with a Luer or other fitting to the GastroFlush.
A pressurized pulsatile stream of water is supplied to the GastroFlush by a standard jet lavage unit. A specialty connector (Specialty Connector 1) is used to ensure proper connection of the GastroFlush to the water delivery unit, as each jet lavage model has a different, proprietary connector. This connector (Specialty Connector 1) is of a type not compatible with ENFit, Luer, or other commonly used medical connectors.
A 2-D CAD drawing of the GastroFlush is shown in Figure 2, and the various components are again labeled.
Image /page/6/Figure/4 description: The image shows a diagram of a fluid system with several components labeled. The system includes a "Specialty Connector 1", a "Pressure Relief Valve", and a "One-Way Valve 1". Additionally, there is an "ENFit Connector 1", a "One-Way Valve 2", and an "ENFit Connector 2", all connected to a reservoir labeled "To Overflow".
Figure 2 GastroFlush Assembly
### 6. Comparison to Predicate Devices:
The GastroFlush enteral tube un-clogger is similar in function and intended use to the TubeClear In-Patient Tube Clearing System, manufactured and legally marketed by Actuated Medical, Bellefonte, PA. The GastroFlush is also similar in function and intended use to the DeClogger, manufactured and legally marketed by Bionix Development Corporation, Toledo, Ohio.
{7}------------------------------------------------
The following table summarizes the comparison of Bionix Development Corporation GastroFlush to these predicate devices:
| Attribute | TubeClear In-Patient<br>Tube Clearing System | DeClogger | GastroFlush |
|----------------------------------------|-------------------------------------------------------------|-------------------------------------------------------------|-------------------------------------------------------------|
| Intended Use | Clear Occlusions from<br>Feeding and Decompression<br>Tubes | Clear Occlusions from<br>Feeding and Decompression<br>Tubes | Clear Occlusions from<br>Feeding and Decompression<br>Tubes |
| Portion of Device Impacting<br>Clog | Clearing Stem Wire Tip | DeClogger Screw Tip | Water Jet |
| Mechanism of Clog Disruption | Linear Reciprocating Motion<br>of Clearing Stem Wire Tip | Twisting Motion of Screw<br>Tip | Pulsation of Water Jet |
| Composition of Declogging<br>Mechanism | Metal Wire | Molded Plastic | Water |
| Power System | Electric Motor Creates<br>Reciprocating Motion | Manual Twisting Motion of<br>DeClogger Tip | Separate Water Pump<br>Creates Pulsatile Water<br>Stream |
| Single Patient Use | Yes | Yes | Yes |
#### Table 1 Comparison of Substantial Equivalence
#### 7. Performance Standards
- a. This device meets Recognized Consensus Standard: AAMI/ANSI ID54:1996/(R) 2012, "Enteral feeding set adapters and connectors".
- b. The following Industry Guidance Document has also been considered in the design of this device: "Safety Considerations to Mitigate the Risks of Misconnections with Smallbore Connectors Intended for Enteral Applications", document issued on February 11, 2015.
- c. ISO Standard 80369 (ENFit Connectors).
{8}------------------------------------------------
### 8. Bench Testing:
The GastroFlush enteral Tube un-clogger was subjected to testing for safety and efficacy as described in the section titled "Bench Testing". Properties tested were:
- a. Tube Clearance Testing
- b. Leak-Free Pressure Testing
- c. Overpressure Valve Testing
The results of these tests may be found in the section "Performance Testing-Bench". These results are summarized below.
- Tube Clearance Testing. Testing was performed on the GastroFlush device's ability to a. clear an in-vitro occlusion from enteral feeding tubes designated as "worst case" examples, as defined through discussions with FDA reviewers. These included examples of the smallest diameter/longest tubes and the largest diameter/longest tubes intended to be used with the GastroFlush. Testing rationale and protocol was chosen to allow determination of results at the 95% C/R (Confidence/Reliability) level for an attribute data test of Risk Level 1. Accordingly, 30 enteral feeding tubes from each group were chosen for testing.
Results of Tube Clearance Testing showed no failures in the ability of the GastroFlush to clear in-vitro occlusions from enteral feeding tubes in "worst case" scenarios at the 95% C/R level. Mean time to tube clearance was 2.20 seconds (smallest diameter/longest tube), and 1.79 seconds (largest diameter/longest tube).
These results show that the GastroFlush device will successfully un-clog occluded enteral feeding tubes at the 95% Confidence/Reliability level.
- b. Leak-Free Pressure Testing. Pressure testing was performed to ensure leak-free performance of the GastroFlush enteral tube un-clogger. Testing rationale and protocol was chosen to allow determination of results at the 95% C/R level for an attribute data test of Risk Level 1. Accordingly, 30 enteral feeding tubes from each group were chosen for testing.
Results of Leak-Free Pressure Testing showed no leakage in any of the GastroFlush/enteral tube assemblies at pressures up to 30 PSI (twice the opening pressure of the overpressure valve).
These results show that the GastroFlush device will operate leak-free during use at the 95% Confidence/Reliability level.
- Overpressure Valve Testing. Testing was performed to evaluate the ability of the C. overpressure valve used in the GastroFlush to properly open at the indicated opening pressure of 12.5 PSI, and to prevent system operating pressure from rising over 15 PSI.
{9}------------------------------------------------
Testing rationale and protocol was chosen to allow determination of results at the 95% C/R level for a variable data test of Risk Level 1. Accordingly, 15 sample valves from the production lot were randomly selected to undergo Overpressure Valve Testing.
Results of Overpressure Valve Testing showed the mean opening pressure of the overpressure valves to be 12.71 PSI (SD = 0.34). These results show that the overpressure valves used in the construction of the GastroFlush device will open within specifications and below 15 PSI at the 95% Confidence/Reliability level.
### 9. Clinical Testing:
No clinical testing was performed on the GastroFlush Enteral Tube Un-Clogger from Bionix Development Corporation.
#### 10. Human Factors Validation Testing:
Human factors validation testing was performed to identify any design or procedural elements that might interfere with proper and safe use of the device. These results are tabulated in the section "Human Factors Validation Testing" and are summarized below.
#### 11. Conclusion:
These results show that the GastroFlush enteral tube un-clogger from Bionix Development Corporation will perform safely and adequately for the intended use to clear occlusions and clogs from enteral tubes placed for the purposes of nutrition and/or decompression in the clinical or home setting.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.