K180585 · Tanses Technologies, Inc. · NHN · May 11, 2018 · Physical Medicine
Device Facts
Record ID
K180585
Device Name
Collagentex Rx-6
Applicant
Tanses Technologies, Inc.
Product Code
NHN · Physical Medicine
Decision Date
May 11, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5500
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Collagentex RX-6 model is intended to emit light energy in the near infrared spectrum to provide non heating adjunctive use for temporary relief of minor chronic neck and shoulder pain of musculoskeletal origin.
Device Story
Collagentex Rx-6 is a floor-mounted, stand-alone infrared lamp system for clinical use. It emits non-collimated, non-pulsing polychromatic light (580-1500 nm) via six 500W quartz halogen lamp assemblies. The device features two metal vertical towers with an adjustable horizontal lamp module; height is controlled via a wired remote. An oscillation system pivots the module 15 degrees for coverage. Operators position the device over a massage table, ensuring a minimum 8-inch distance from the patient. The control console includes a timer, emergency stop, and mechanical hour counter. Treatment is administered by a clinician; the device automatically shuts off after the set time, followed by a 4-minute cool-down. The device provides non-heating adjunctive therapy to temporarily relieve minor chronic musculoskeletal neck and shoulder pain.
Clinical Evidence
Bench testing only. The device demonstrated compliance with IEC 60601-1-2 (EMC), NRTL, CSA, and ANSI electrical safety standards, and ANSI/IESNA RP-27.1-05 and RP-27.3-07 photobiological safety standards for the device and eye protection.
Technological Characteristics
Quartz halogen lamp source (500W); 580-1500nm wavelength; 44.2 mW/sqcm intensity. Metal construction. Dimensions: 96" x 27.68" x 96". Powered by 220/240 VAC. Non-contact device. Standards: ISO 13485/14971:2007 (risk), IEC 60601-1-2 (EMC), IESNA/ANSI RP-27.1-05/RP-27.3-07 (photobiological). No software-based algorithm.
Indications for Use
Indicated for patients experiencing minor chronic neck and shoulder pain of musculoskeletal origin requiring non-heating adjunctive light therapy.
Regulatory Classification
Identification
An infrared lamp is a device intended for medical purposes that emits energy at infrared frequencies (approximately 700 nanometers to 50,000 nanometers) to provide topical heating.
Special Controls
*Classification.* Class II (special controls). The device, when it is an infrared therapeutic heating lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
May 11, 2018
Tanses Technologies Inc. Kirk Kiremitci President 4450 Highway 13 Fabreville, H7R 6E9 Canada
Re: K180585
Trade/Device Name: Collagentex Rx-6 Regulation Number: 21 CFR 890.5500 Regulation Name: Infrared Lamp Regulatory Class: Class II Product Code: NHN Dated: February 1, 2018 Received: March 5, 2018
Dear Kirk Kiremitci:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely. Jennifer R. Stevenson -For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K180585
Device Name Collagentex Rx-6
#### Indications for Use (Describe)
Collagentex RX-6 model is intended to emit light energy in the near infrared spectrum to provide non heating adjunctive use for temporary relief of minor chronic neck and shoulder pain of musculoskeletal origin.
Type of Use (Select one or both, as applicable)
| <span style="text-decoration: overline;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|------------------------------------------------------------------------------------------------|
| <span style="text-decoration: overline;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
### Date Prepared
February 1, 2018
#### Submitter Information
Tanses Technologies inc. 4450 Highway 13 Laval, Quebec H7R 6E9 Canada Tel: 450-622-4004 Fax: 450-622-1540 Contact name: Kirk Kiremitci Email: kirk@tanses.com
#### Device Information
| Device Trade Name: | Collagentex |
|---------------------|------------------------------------------------------|
| Device Model Name: | RX-6 |
| Common /Usual Name: | Near Infrared Lamp |
| Requlation Name: | Lamp, non heating, for adjunctive use in pain relief |
| Requlatory Class: | Class II |
| Requlation Number: | 21 CFR 890.5500 |
| Product Code: | NHN |
| Panel: | Physical Medicine |
#### Predicate Device
| Submitter: | Tanses Technologies Inc. Canada |
|--------------------|---------------------------------|
| Manufacturer: | Tanses Technologies Inc. |
| Trade Name: | Collagentex |
| Model: | RX-1 |
| Regulation Number: | 21 CFR 890.5500 |
| Product Code: | NHN |
| 510(k): | K162022 |
#### Device Description
The Collagentex Rx-6 is a stand alone light device which emits polychromatic visible to near infrared light in the 580 to 1500 nanometer range. The light is non-collimating and non-pulsing. There are a total of 6 exposure areas, each set to receive 44.2miliWatt/sqcm or 2.6 Joule/minute at the minimum distance of 8 inches.
The device measures 96 inches long by 27.68 inches wide by 96 inches tall and is made up of 3 main sections. There are 2 vertical towers positioned 84 inches apart. The towers are made of metal and house the main electrical components for the device. A control console which includes a set timer, emergency stop button, oscillating on/off switch and a mechanical hour counter is mounted on left positioned tower.
In between the towers, positioned horizontally is a lamp module. The lamp module is made up of 6 individual lamp assemblies placed inline. The lamps are all 500w quartz type and each cover an exposure area of 23cm diameter. The height of the lamp
{4}------------------------------------------------
module is adjustable via actuators positioned in the vertical columns. The actuators are controlled by a wired remote control. The device is to be mounted on the floor above a massage table provided by the device owner.
# Device Use
Once the user lays down on the massage table, the operator then adjusts the height of the lamp module using the remote control. The operator is instructed to ensure a minimum distance of 8 inches between the lamps and user. The operator then sets the desired treatment time and presses the start button to active the device. An oscillation system can be turned on which allows the light module to pivot side to side by 15 degrees. This allows for better overall coverage. At the end of the treatment time the device turns off and enters a 4 minute cool down mode.
# Intended Use
Collagentex RX-6 model is intended to emit light energy in the near infrared spectrum to provide non heating adjunctive use for temporary relief of minor chronic neck and shoulder pain of musculoskeletal origin.
# Technical Characteristics
| Characteristics | Collagentex Rx-6 | Collagentex Rx-1 |
|-----------------------------------------|------------------------------------------|------------------------------------------|
| Light source | Quartz Lamp | Quartz Lamp |
| Light Source Power | 500 W | 500 W |
| Exposure area | 23cm diameter * 6 modules | 23cm diameter |
| Emitted light wavelength<br>(effective) | 580nm to 1500nm | 580nm to 1500nm |
| Emitted light polarization | Not polarized | Not polarized |
| Indications of use | temporary pain relief | temporary pain relief |
| Product code | NHN | NHN |
| Emitted light intensity | 44.2 miliWatt/sqcm (2.6<br>Joule/minute) | 44.2 miliWatt/sqcm (2.6<br>Joule/minute) |
| Power supply | 220 VAC or 240 VAC | 120 VAC or 240 VAC |
| Design Principle | Quartz halogen lamp with a<br>filter | Quartz halogen lamp with a<br>filter |
| Skin Surface Temperature | 1.8C max. skin temp rise | 1.8C max. skin temp rise |
| Treatment Time | 20 minutes per site | 20 minutes |
| Patient Contact | No contact to human body | No contact to human body |
| Biocompatibility | No contact to human body | No contact to human body |
| Electrical Safety | NRTL, CSA, ANSI | NRTL, CSA, ANSI |
| Electromagnetic<br>Compatibility | 60601-2 (3rd ed.) | 60601-2 (3rd ed.) |
| Risk Analysis | According to ISO<br>13485/14971:2007 | According to ISO<br>13485/14971:2007 |
| Photobiological safety of<br>the device | IESNA/ANSI RP-27.1-05 | IESNA/ANSI RP-27.1-05 |
| Photobiological safety of<br>eyewear | IESNA/ANSI RP-27.3-07 | IESNA/ANSI RP-27.3-07 |
From the comparison form above, the subject device and the predicate device have the same indications for use, operation principle, skin temperature, wavelength spectrum, irradiation intensity and treatment time.
{5}------------------------------------------------
# Performance Data
The subject device has been tested and complied with the following voluntary recognizable standards IEC60601-1-2 3rd edition Electromagnetic compatibility requirements as well NRTL, CSA, ANSI electrical safety standards and ANSI/IESNA RP-27.1-05 and ANSI/IESNA RP-27.3-07 photobiological safety standards for the device and the eye protection.
# Conclusion
Collagentex RX-6 model is substantially equivalent to the predicate device in K162022. It uses the same technology, outputs the same wavelength spectrum, has the same irradiation intensity, treatment time and identical testing as the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.