K180581 · Feilo Sylvania Germany GmbH · LEJ · Jun 22, 2018 · General, Plastic Surgery
Device Facts
Record ID
K180581
Device Name
UV Lamps for Sunlamp Products
Applicant
Feilo Sylvania Germany GmbH
Product Code
LEJ · General, Plastic Surgery
Decision Date
Jun 22, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4635
Device Class
Class 2
Indications for Use
This ultraviolet lamp is intended for use in sunlamp products for tanning of human skin.
Device Story
Component UV sunlamps; low-pressure, mercury rare gas discharge devices; tubular glass envelope with internal fluorescent phosphor coating. Operates via external ballast supply; emits UV radiation in 200-710nm range (primarily 320-400nm UVA, with some 260-320nm UVB). Used as replacement components in sunlamp tanning products. Manufacturer controls construction via QMS and testing to ensure radiation limits per 21 CFR 1040.20. Benefits patient by providing controlled UV exposure for tanning.
Clinical Evidence
Bench testing only. Compliance with IEC/EN 61195:2014, IEC/EN 60355-2-27:2013, EN60081:1998, EN61228:2008, EN62471:2008, and EN50581:2012. Spectral analysis performed to verify irradiance compliance with 21 CFR 1040.20 radiation limits.
Technological Characteristics
Low-pressure mercury rare gas discharge lamps; tubular glass envelope with internal fluorescent phosphor coating. Powered by external ballast. Wavelength range 200-710nm. Complies with IEC/EN 61195, IEC/EN 60355-2-27, EN60081, EN61228, EN62471, and 21 CFR 1040.20. Manufactured under ISO 9001 QMS.
Indications for Use
Indicated for tanning of human skin using sunlamp products.
Regulatory Classification
Identification
A sunlamp product is any device designed to incorporate one or more ultraviolet (UV) lamps intended for irradiation of any part of the living human body, by UV radiation with wavelengths in air between 200 and 400 nanometers, to induce skin tanning. This definition includes tanning beds and tanning booths. A UV lamp intended for use in sunlamp products is any lamp that produces UV radiation in the wavelength interval of 200 to 400 nanometers in air.
Special Controls
*Classification.* Class II (special controls). The special controls for sunlamp products and UV lamps intended for use in sunlamp products are:(1) Conduct performance testing that demonstrates the following:
(i) Device meets appropriate output performance specifications such as wavelengths, energy density, and lamp life; and
(ii) Device's safety features, such as timers to limit UV exposure and alarms, function properly.
(2) Demonstrate that device is mechanically safe to prevent user injury.
(3) Demonstrate software verification, validation, and hazard analysis.
(4) Demonstrate that device is biocompatible.
(5) Demonstrate that device is electrically safe and electromagnetically compatible in its intended use environment.
(6)
*Labeling* —(i)*Sunlamp products.* (A) The warning statement below must appear on all sunlamp products and must be placed in a black box. This statement must be permanently affixed or inscribed on the product when fully assembled for use so as to be legible and readily accessible to view by the person who will be exposed to UV radiation immediately before the use of the product. It shall be of sufficient durability to remain legible throughout the expected lifetime of the product. It shall appear on a part or panel displayed prominently under normal conditions of use so that it is readily accessible to view whether the tanning bed canopy (or tanning booth door) is open or closed when the person who will be exposed approaches the equipment and the text shall be at least 10 millimeters (height). Labeling on the device must include the following statement: Attention: This sunlamp product should not be used on persons under the age of 18 years. (B) Manufacturers shall provide validated instructions on cleaning and disinfection of sunlamp products between uses in the user instructions.
(ii)
*Sunlamp products and UV lamps intended for use in sunlamp products.* Manufacturers of sunlamp products and UV lamps intended for use in sunlamp products shall provide or cause to be provided in the user instructions, as well as all consumer-directed catalogs, specification sheets, descriptive brochures, and Web pages in which sunlamp products or UV lamps intended for use in sunlamp products are offered for sale, the following contraindication and warning statements:(A) “Contraindication: This product is contraindicated for use on persons under the age of 18 years.”
(B) “Contraindication: This product must not be used if skin lesions or open wounds are present.”
(C) “Warning: This product should not be used on individuals who have had skin cancer or have a family history of skin cancer.”
(D) “Warning: Persons repeatedly exposed to UV radiation should be regularly evaluated for skin cancer.”
(c)
*Performance standard.* Sunlamp products and UV lamps intended for use in sunlamp products are subject to the electronic product performance standard at § 1040.20 of this chapter.
Predicate Devices
Havells Sylvania Germany/SLI Lichtsysteme GmbH (Class I exempt)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
June 22, 2018
Feilo Sylvania Germany GmbH % Steve Mlcoch President NATS Corp. PO Box 1257 30 Northport Rd. Sound Beach, New York 11789
Re: K180581
Trade/Device Name: UV Lamps for Sunlamp Products Regulation Number: 21 CFR 878.4635 Regulation Name: Ultraviolet Lamp For Tanning Regulatory Class: Class II Product Code: LEJ Dated: February 1, 2018 Received: March 1, 2018
Dear Steve Mlcoch:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good
{1}------------------------------------------------
manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Jennifer R. Stevenson -S3
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
Expiration Date: 06/30/2020 See PRA Statement below.
Form Approved: OMB No. 0910-0120
510(k) Number (if known) K180581
#### Device Name
UV Lamps for Sunlamp Products
Indications for Use (Describe)
This ultraviolet lamp is intended for use in sunlamp products for tanning of human skin.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (7/17)
Page 1 of 1
PSC Publishing Scrvices (301) 443-4740 EF
{3}------------------------------------------------
# 510(K) SUMMARY K180581 February 1, 2018
# SUBMITTER INFORMATION:
Feilo Sylvania Germany GmbH Graf-Zeppelin-Str. 9 91056 Erlangen Germany
Contact: Achim Tonat Tel: 49-9131-793-242 Email:achim.tonat@feilosylvania
# APPLICANT/FDA AGENT/CORRESPONDING OFFICIAL INFORMATION:
North American Technical Services (NATS) Corp. PO Box 1257 Sound Beach, NY 11789 Email: natscorp@aol.com Contact: Stephen T. Mlcoch
# DEVICE NAME:
Common Name: Ultraviolet Sunlamps Proprietary Model: Component UV Sunlamp Model and Part Number Family per attachment page list
Classification: II Classification Code, Category: LEJ, 21CFR 878.4635, General & Plastic Surgery
# DESCRIPTION:
The component ultraviolet lamp model family or part numbers are for use in sunlamp products to tan human skin. They fall into a generic class of UV lamps that have the same basic technological features, characteristics, except for power rating, diameter and length, with the exact same intended use. The UV sunlamp model family are for tanning human skin use. They vary in brand ID and unique lamp ratings accordingly. These sunlamps fit into a wavelength range of 200-710nm and operate from the end product ballast supply.
There are individual component sunlamps listed in the exhibit page. They are classified as low pressure, mercury rare gas discharge devices and generally have the same construction and technical principals of operation as that of a common fluorescent lamp. The UV lamps have a tubular glass envelope with internal surface coated with fluorescent phosphor. The sealed lamps include the composition and electrical means to activate the UV light within the wavelength specification. The construction is controlled by production procedures and test methods defined in the QMS. The UV radiation is controlled by these methods to assure safe use.
# INTENDED USE:
The ultraviolet lamp is intended for use in sunlamp products for tanning human skin.
{4}------------------------------------------------
### PREDICATE DEVICES:
Many of the UV sunlamps listed in this 510K previously existed and were offered for sale as class I, 510K exempt medical devices before September 2, 2014. Feilo Sylvania has merged and taken ownership of the Havells Sylvania Germany factory and business including product CDRH submissions by SLI Lichtsysteme GmbH. The model part numbers and brand names added all sustain equivalent construction, characteristics and safe performance. Essentially the newest UV lamps are identical to the sunlamps made at the same location controlled by Havells and substantially equivalent to the other predicate devices cited. All sunlamps listed, now controlled by Feilo Sylvania Germany, have the same intended use, sustain the same technological characteristics and are validated as safe and effective using the control methods and standards testing applied to the designed construction and production.
The predicate devices identified are:
- 1. Havells Sylvania Germany/ SLI Lichtsysteme GmbH - These companies produced class I 510K exempt UV sunlamps prior to September 2, 2014 of constructions now owned and controlled by Feilo Sylvania Germany at the same Havells location in Erlangen, Germany
- K152273 Cosmedico Light, Inc., dated October 30, 2015 2.
- K151370 Light Sources Incorporated dated December 17, 2015 3.
The listed UV sunlamps devices made by Feilo Sylvania at the previously owned Havells location, have the same control methods and compliance validation. Brand names added, with part number identifications, have the same technological features and intended use. Cosmedico Light and Light Sources have equivalent over the counter UV sunlamps with the same indications of use. Generally, the constructions are equivalent and are validated as safe and effective.
The fundamental comparison information about critical feature to the predicate devices cited are:
| FEATURE | FEILO SYLVANIA | HAYELLS, SLI | COSMEDICO | LIGHT SOURCES |
|-------------------|---------------------------------------|--------------------------|-----------------------------|------------------------|
| Factory: | Erlangen Germany | / Erlangen Germany | / Cosmedico | / Light Sources |
| Code/K | LEJ, K180581 | / LEJ 9-2-14 exempt | /LEJ, K152273 | /LEJ, K151370 |
| Indication: | UV light to tan skin | / UV light to tan skin | /UV light to tan skin | / UV light to tan skin |
| Biocompatibility: | no contact | / no contact | / no contact | / no contact |
| Power Source | Ballast | / Ballast | / Ballast | / Ballast |
| Tech Specs: | wavelength*<br>UVA, UVB<br>Florescent | / same specs* | / equivalent*<br>Florescent | / equivalent* |
| Radiation | 21CFR1040.20 | / Accession #s | /CDRH compliant | /CDRH compliant |
| | Havells, SLI records<br>F Sylvania | 0720172-00<br>1810251-00 | | |
*The standard visual range of fluorescent lamps is 380nm-760nm. The UV radiation range is 300nm - 400nm. These lamps emit their energy closest to the visible range 320nm - 400nm, with the emission spectrum range of 260nm - 320mm containing a relative amount of UVB radiation. Feilo Sylvania and all predicate companies comply to the same consensus and industry standards and have equivalent QMS control activity.
PERFORMANCE DATA AND DESIGN CONTROL ACTNITIES:
The Feilo Sylvania UV lamps are intended for use in sunlamp products for tanning of human skin comply with the consensus standards and specifications listed below and are therefore validated as safe for the intended use. The device sunlamps are tested, validated and verification procedures are in place to confirm design specifications. Compliance with the following mandatory and voluntary standards is a QMS requirement:
INTERNATIONAL CONSENSUS STANDARDS AND APPLIED SPECIFICATIONS:
- . IEC/EN 61195:2014 - Double Capped Florescent Lamps Safety Specifications
- . IEC/EN 60355-2-27:2013 - Particular Appliances Safety Requirements for Skin Exposure to UV and IR Radiation
- EN60081:1998 - Double Capped Fluorescent Lamps - Performance Specifications
- . EN61228:2008 - Fluorescent Ultraviolet Lamps Used for Tanning. Measurement and Specification Method
- . EN62471:2008- Photobiological Safety of Lamps and Lamp Systems
- . EN50581: 2012 - Technical Documentation for the Assessment of Electrical and Electronic with Respect to the Restriction of Hazardous Substances
{5}------------------------------------------------
# FDA PERFORMANCE REFERENCE SPECIFICATION:
- o 21CFR 1040.20 - Performance standard for sunlamp products and ultraviolet lamps intended for the use in sunlamp products
Specific performance testing (spectral analysis) data is submitted to CDRH for the component UV lamps with measurement of irradiance to verify compliance with radiation limits defined in 21CFR1040.20.
QUALITY MANAGEMENT SYSTEM COMPLIANCE:
- ISO9001 Quality Management System
Accordingly, factory procedures are established for production, assembly and testing. The control activity shows that there are no new questions of safety and effectiveness for the described UV Sunlamps. The design analysis and predicate comparison confirm the functional characteristics are the same to the predicate devices and raise no other safety or effectiveness issues. Inspection verification procedures assure performance concerns are compliant to QMS specifications.
# CONCLUSION:
The Feilo Sylvania UV lamp is intended for use in sunlamp products for tanning of human skin. Each are the same or substantially equivalent to the predicate device sunlamps. They have the same intended use. Havells had FDA/CDRH registration# 1000553841 and SLI had registration #1000533299. They also have a history of UV reporting acknowledged by many related Accession #s that are now part of the Feilo Sylvania compliance records and data. The key comparisons and analysis of the safety, indications, specific use, performance, design, construction, material composition, energy source, power output, technological propelties, skin tan treatments, compliance testing, previous production history and method of operation, the manufacturer Feilo Sylvania verified that no significant differences exist between these device sunlamps and the predicated devices listed. Therefore, substantial equivalence exists.
{6}------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.