For the fabrication of crowns, bridges and structures for implant supported provisional removable denture and appliance prosthetics. -Provisional anterior and posterior crowns & bridges - Implant and abutment supported prosthetics. - Partial, complete and hybrid denture prosthetics (base and teeth) - Removable appliances (splint)
Device Story
Harvest Dental Polymer Blocks are industrially polymerized PMMA milling discs/blocks. Used by dental laboratories/clinicians with CAD/CAM milling systems to fabricate provisional dental prosthetics (crowns, bridges, dentures, splints). Device serves as raw material for subtractive manufacturing. Output is a custom-milled dental restoration. Benefits include patient access to durable, biocompatible provisional prosthetics during treatment planning or long-term restoration phases. Device is used in clinical/laboratory settings; operator is a dental professional.
Clinical Evidence
Bench testing only. Device tested for physical properties per ISO 10477 and ISO 20795-1, including flexural strength (>90 MPa), water absorption (0.024 mg/mm³), water solubility (0.0000 mg/mm³), and residual monomer content (<1.1%). Biocompatibility testing performed per ISO 10993-5 (cytotoxicity), ISO 10993-10 (sensitization), and ISO 10993-10 Annex B3 (oral mucosal irritation). All results passed.
Technological Characteristics
Material: Polymethyl methacrylate (PMMA) with <1.1% coloring oxides. Form factor: Milling discs or blocks. Standards: ISO 10477 (polymer-based crown/bridge materials), ISO 20795-1, ISO 22112 (polymer teeth), ISO 10993 (biocompatibility). Connectivity: None (standalone material).
Indications for Use
Indicated for fabrication of provisional anterior and posterior crowns, bridges, implant/abutment supported prosthetics, partial/complete/hybrid denture prosthetics, and removable appliances (splints).
Regulatory Classification
Identification
A temporary crown and bridge resin is a device composed of a material, such as polymethylmethacrylate, intended to make a temporary prosthesis, such as a crown or bridge, for use until a permanent restoration is fabricated.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.
Harvest Dental Products, LLC Sasha Der Avanessian Owner 905 Columbia Street Brea, California 92821
December 12, 2018
Re: K180578
Trade/Device Name: Harvest Dental Polymer Blocks Regulation Number: 21 CFR 872.3770 Regulation Name: Temporary Crown And Bridge Resin Regulatory Class: Class II Product Code: EBG, EBI, MQC Dated: September 14, 2018 Received: September 14, 2018
Dear Sasha Der Avanessian:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrl/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely, Digitally signed by Mary S. Mary S. F Runner - S3 rvically -5:
> Runner -S33 - Date: 2018.12.12 14:33:57
> Runner -S33 - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K180578
Device Name Harvest Dental Polymer Blocks
Indications for Use (Describe)
Indications for Use:
For the fabrication of crowns, bridges and structures for implant supported provisional removable denture and appliance prosthetics.
-Provisional anterior and posterior crowns & bridges
- Implant and abutment supported prosthetics.
- Partial, complete and hybrid denture prosthetics (base and teeth) - Removable appliances (splint)
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Harvest Dental Laboratory Products. The logo features two green leaves above the words "HarvestDental" in a gray sans-serif font. Below the company name are the words "LABORATORY PRODUCTS" in a smaller, gray sans-serif font.
# K180578 510(k) Summary
This Summary of 510K safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.
| 510K Submitter | Harvest Dental Products, LLC<br>905 Columbia St, Brea, CA 92821<br>Tel: 714-674-7400<br>hello@harvestdental.com |
|-----------------------------|-----------------------------------------------------------------------------------------------------------------|
| Contact person: | Sasha Der Avanessian<br>President & CEO<br>Tel: 714-674-7400<br>sasha@harvestdental.com |
| Date Summary was prepared | 12/11/18 |
| Trade Name of device: | Harvest Dental Polymer Blocks |
| Common name: | Crown & Bridge, Temporary, Resin |
| Classification name: | TEMPORARY CROWN AND BRIDGE RESIN. (21 CFR 872.3770 - Product code EBG) |
| Classification Product Code | EBG |
| Panel | Dental |
| Classification | Class II |
| Predicate Device | K150432<br>Union Dental S.A<br>Idodentine (Dental Polymer Blank) |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Harvest Dental Laboratory Products. The logo features two green leaves above the text "HarvestDental" in gray. Below the text is the phrase "LABORATORY PRODUCTS" in a smaller, gray font.
# Indications for Use:
For the fabrication of crowns, bridges and structures for implant supported provisional removable denture and appliance prosthetics.
- Provisional anterior and posterior crowns & bridges
- Implant and abutment supported prosthetics.
- Partial, complete and hybrid denture prosthetics (base and teeth)
- Removable appliances (splint)
### Device Description:
Harvest Dental Polymer Blocks are industrially polymerized, pre-colored or clear dental milling discs and blocks designed for milled fabrication of temporary anterior and posterior crowns and bridges (tooth-colored variants) or bite splints and dental CAD/CAM systems.
#### Technological characteristics
The technologic characteristics are highly similar as demonstrated in performance testing and in chemical composition; both devices are composed primarily of polymethylmethacrylate, while the amount and percentage of color oxides in the submission device varies from the predicate.
#### Comparison of Required Technology Characteristics
The following table shows a summary of the technological characteristics of DD medical polymers compared to the predicate device.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Harvest Dental Laboratory Products. The logo features two green leaves above the text "HarvestDental" in gray. Below "HarvestDental" is the text "LABORATORY PRODUCTS" also in gray.
| Feature | Submission device | Predicate device |
|----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade name | Harvest Dental Polymer Blank | Idodentine Disc |
| 510 (K) | K180578 | K150432 |
| Product code | EBG | EBG |
| Regulatory Class | CLASS II | Class II |
| Manufacturer | Harvest Dental | Union Dental S.A |
| Indications for use | For the fabrication of crowns, bridges and structures for implant supported provisional removable denture and appliance prosthetics. - Provisional anterior and posterior crowns & bridges. - Implant and abutment supported prosthetics. - Partial, complete and hybrid denture prosthetics (base and teeth). - Removable appliances (splint) | Acrylic polymer blank particularly suitable for making removable or temporary dental structures such as crowns and bridges using milling technology using CAD/CAM. Indications - Temporary anterior and posterior crowns - Temporary anterior and posterior bridges with up to two adjacent pontics Maximum recommended usage period: 12 months - Removable structures for dentures (denture bases) - Removable structures for therapeutic restorations (bite splints or occlusal splints) |
| Technology | Blank for dental CAD/CAM | Blank for dental CAD/CAM |
| Shape | Disc or Block | Disc or Block |
| Shade | VITA-shades, clear, pink | VITA-shades, clear, pink |
| Raw Material | PMMA | PMMA |
| Chemical Compositions [Units] | | |
| Material Base | Polymethyl methacrylate | Polymethyl methacrylate |
| Coloring oxides | < 1,1% | Not specified |
| Physical Characteristics [Units] | | |
| Performance Testing | ISO 10477<br>ISO 20795 | Tested according to DIN EM ISO 20795-1 and ISO 10477 |
| Biocompatibility | ISO 10993<br>No cytotoxic | E.N ISO 10933-1, -5, |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Harvest Dental Laboratory Products. The logo features the word "Harvest" in gray, with the word "Dental" in green next to it. Above the words is a green leaf. Below the words "HarvestDental" are the words "LABORATORY PRODUCTS" in gray.
### Preclinical testing and standards Clinical and Non clinical performance
The biocompatibility test for "Harvest Dental Polymer Blocks" was made in the laboratory and all the results was:
| Standard | Conclusion | Traceability |
|---------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------|
| ISO 10993-10, 2010. | In the test conditions, the test<br>substance Harvest Dental<br>Polymer Blocks was classified<br>as no sensitizing. | Report Number:<br>RL15885/2018LLNA-B<br>Study Number:<br>15885/2018LLNA |
| ISO 10993-10, annex B3<br>2010. | Under the study conditions, the<br>test substance Harvest Dental<br>Polymer Blocks was classified as<br>non-irritant to oral mucose of<br>hamsters, according<br>to<br>ISO10993: 10, 2010. | Report Number:<br>RL15886/2018IMO-B<br>Study Number:<br>15886/2018IMO |
| ISO 10993-5, 2009 annex C | Under the conditions of this<br>study, the test substance<br>Harvest Dental Polymer Blocks<br>did not promote reduction of cell<br>viability higher than 30%.<br>Therefore the test substance<br>Harvest Dental Polymer Blocks<br>was not cytotoxic in this assay. | Report Number:<br>RL16016/2018CT-B<br>Study Number:<br>16016/2018CT |
This PMMA has been tested for biocompatibility and physical properties. This PMMA used has been rated as biocompatible and non-toxic and this PMMA material also met the applicable requirements of ISO 10447(crown & bridges material), and ISO 22112(polymer teeth) standards. The subject device has passed testing according to ISO 10477/2004: DENTISTRY - POLYMERBASED CROWN AND BRIDGE MATERIALS. This international standard specification, requirements and test methods for the polymer-based crown and bridge materials. The subject device also has passed testing according to ISO 10993-5 and has been rated as Non cytotoxic.
## Standards:
ISO 10477:2004 – Dentistry Polymer-Based crown and Bridge Materials ISO 22112:2005 - Polymer teeth ADA - 53 - Polymer-Based crown and Bridge Resin
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Harvest Dental Laboratory Products. The logo features two green leaves above the text "HarvestDental" in gray. Below "HarvestDental" is the text "LABORATORY PRODUCTS" in a smaller, gray font.
| Requirements acc. to DIN EN ISO 20795-1 / ISO 10477: | | | | | |
|------------------------------------------------------|------------------------------------|----------------------------------|--------------------------|------------------------------|---------------|
| Requirement | Required Value<br>(ISO<br>20795-1) | Required Value<br>(ISO<br>10477) | Value<br>Subm.<br>Device | Value<br>Predicate<br>Device | Passed/Failed |
| Flexural<br>strength | ≥ 65<br>[MPa] | ≥ 50<br>[MPa] | > 90<br>Mpa | 90 [MPa] | Passed |
| Water<br>absorption | ≤ 32<br>[µg/mm³] | ≤ 40<br>[µg/mm³] | 0.024<br>mg/mm³ | 23<br>[µg/mm³] | Passed |
| Water<br>solubility | ≤ 1.6<br>[µg/mm³] | ≤ 7.5<br>[µg/mm³] | 0.0000<br>mg/mm³ | 0.0<br>[µg/mm³] | Passed |
| Residual<br>monomer<br>content [%] | ≤ 2.2 % | - | < 1,1% | 1.4 % | Passed |
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Harvest Dental Laboratory Products. The logo features two green leaves above the text "HarvestDental" in gray. Below "HarvestDental" is the text "LABORATORY PRODUCTS" in a smaller, gray font.
# Substantial Equivalence:
The differences in indications for use are limited to the following wording differences as follows:-
| Submission Device | Predicate device | Differences | Discussion |
|----------------------------------------------------------------------------------------------------|---------------------------------------------------|--------------------------------------------------------|------------|
| Fabrication of crowns,<br>bridges | Same | | |
| | Same | | |
| Structures for implant<br>supported provisional<br>removable denture and<br>appliance prosthetics. | Temporary | The use of the word<br>provisional verses<br>Temporary | Note 1 |
| Provisional anterior and<br>posterior crowns & bridges. | Same | with up to two<br>adjacent<br>pontics | Note 2 |
| Implant and abutment<br>supported prosthetics. | Same | bite splints<br>or occlusal splints | Note 3 |
| - Partial, complete and<br>hybrid denture prosthetics<br>(base and teeth). | Same | | |
| | Maximum<br>recommended usage<br>period: 12 months | Submission doesn't<br>use these words | Note 1 |
| | Acrylic polymer blank | Submission<br>doesn't<br>use these words | Note 4 |
| | using milling<br>technology using<br>CAD/CAM. | Submission doesn't<br>use these words | Note 5 |
## Note 1
These two words can often be used interchangeably, however not necessarily meaning the same thing in true context. For example: -
The Predicate device uses "Temporary" meaning the device is expected to come to an end, in their case at 12 months.
The Submission device uses the word "Provisional" meaning that it will come to an end but doesn't mean that it has finished it's useful life.
The reason we chose the word "Provisional" is to establish that our device can actually last a lot longer than 12 months as indicated in our "Shelf life report" and our description of device, which shows the device can actually remain in place for 5 years. This is supported by the test data submitted in this review. Again this use of the word has no adverse impact on the safety or effective performance of the submitted device.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for Harvest Dental Laboratory Products. The logo features two green leaves above the words "HarvestDental" in gray. Below "HarvestDental" are the words "LABORATORY PRODUCTS" in a smaller, gray font.
# Note 2
Pontics are used to replace missing teeth, however the primary causes of complications in implant dentistry are related to biomechanics. For example, early loading failures especially in soft bone, when forces are greater than usual and/or implant sizes are shorter than 10mm.
The number of pontics for a prosthesis is related to strength of the abutments and the number of pontics with implants is often increased compared to natural teeth which can cause a problem.
In fixed prosthesis design, three adjacent pontics in the posterior reqions of the mouth are contraindicated with natural. The adjacent abutments are subjected to considerable additional force when they must support three missing teeth, especially in the posterior regions of the mouth.
Also the occlusal force to the five posterior teeth cannot be supported by two implants. The posterior reqions have greater bite forces than the anterior reqions. This is because typically the teeth in the posterior have less bone density, which puts an increased load on the anterior teeth
The additional forces distributed to the pontic abutments will cause them to flex under load. The greater the span the greater the flex on the pontic causing potential mechanical shear and tensile load failure on the abutments.
The chart below shows the additional flex placed on a pontic assuming a load of 25lbs and all other variables beinq equal.
| # of Pontics | Flex | Accept/Reject |
|--------------|------------------|---------------|
| 1 | 8 μ - 0.01 mm | Accept |
| 2 | 64 μ - 0.08 mm | Accept |
| 3 | 216 μ - 0.216 mm | Reject |
Flex is not so much an issue on natural teeth as they have a small degree of natural flex, apically and laterally in the area of the root. However an implant is more ridged due to its increased modulus of elasticity (Doesn't flex as much as a natural tooth).
Further more angled forces increase the stress and strain on an implant restoration especially in the maxillary anterior teeth. This can lead to porcelain fractures, abutment screws working loose and uncementing of the restoration causing even more stress on the restoration. More stress forces can cause the pontic to cantilever, creating an increased risk of crestal bone loss, failure of the implant or fracture of an abutment screw or implant body, especially in the posterior region due to greater bite force.
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for Harvest Dental Laboratory Products. The logo features two green leaves above the words "HarvestDental" in gray. Below "HarvestDental" are the words "LABORATORY PRODUCTS" in a smaller, gray font.
Pontics should be limited to 2 in proportion to the size of the premolars, typically between 13.5mm - 16mm. There are instances when a molar is one of the missing teeth to be replaced, which is between existing teeth creating a space between 10-15mm. In the instance when the space is greater than 15mm it is better for the doctor to use two pontics to replace the molar. Therefore when a 200 premolar and 136 molar are missing it would require the doctor to plan the replacement of 3 teeth rather than two as the span is related to the number of roots in the mandible and buccal roots in the maxilla.
By using 3 pontics or more in either a fixed prosthesis or implant restoration is contraindicative and a skilled restoration specialized would always design a restoration using a maximum of 2 pontics where possible or use more implants.
Therefore to either state or not state 2 pontic on the Substantial Equivalence indications has no impact on safety or effectiveness of the submitted device, as this is a decision of the specialist doctor when designing the Provisional anterior and posterior crowns & bridges.
## Note 3
We have deliberately left out the need to use the Polymer block for the use of bite splints or occlusal splints, even though this material can actually be used for both we will market a separate material to cover this need. This has no impact on the submitted device for safety or effectiveness.
#### Note 4
The use of the word "Acrylic Polymer Blank" is not required as this is already a given. This has no impact on the submitted device for safety or effectiveness.
#### Note 5
The use of the words "using milling technology using CAD/CAM is not required as this is also a qiven. This has no impact on the submitted device for safety or effectiveness.
#### Conclusion
Information provided in this 510(k) submission shows that the product is substantially equivalent to the Harvest Dental Polymer Block as predicate device Idodentine Blank.
Based on comparison of technology, including composition and performance testing, biocompatibility testing, and highly similar indications for use, Harvest Dental Polymer Blocks are equivalent to Idodentine (K150432).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.