K180567 · Medical Components, Inc. (dba MedComp) · LJS · Apr 12, 2018 · General Hospital
Device Facts
Record ID
K180567
Device Name
C3 Wave System
Applicant
Medical Components, Inc. (dba MedComp)
Product Code
LJS · General Hospital
Decision Date
Apr 12, 2018
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.5970
Device Class
Class 2
Indications for Use
C3 Wave System is intended to provide real time tip location information of a central venous catheter by utilization of ECG to observe P-wave changes as the tip approaches the right atrium of the heart via the superior vena cava.
Device Story
C3 Wave System assists in PICC placement by providing real-time catheter tip location. System uses three ECG leads on patient chest and a third lead generated by switching from RA to PICC stylet. ECG waveform is wirelessly transmitted to a tablet for operator viewing. As catheter approaches heart atrium, P-wave amplitude changes are visualized. Operated by skilled nurses or physicians in hospital/clinical settings. Output allows clinicians to confirm catheter tip location in lower third of Superior Vena Cava, potentially reducing need for confirmatory X-ray or fluoroscopy.
Clinical Evidence
No clinical data. Bench testing only, including biocompatibility (ISO 10993-5, -10), sterilization validation (ANSI/AAMI/ISO 11135-1), and packaging integrity (ASTM F1929-15, ASTM F1140/F1140M-13).
Technological Characteristics
System includes iPad, C3 Wave Hub, ECG leads, and remote. Sterile accessory packs sterilized via Ethylene Oxide (EO). Packaging validated per ASTM F1929-15 and ASTM F1140/F1140M-13. Biocompatibility per ISO 10993-1. Wireless transmission of ECG data to tablet interface.
Indications for Use
Indicated for positioning of Peripherally Inserted Central Catheters (PICC) in adult patients as an alternative to chest X-ray or fluoroscopy. Not contraindicated but limited in patients with atrial fibrillation, atrial flutter, severe tachycardia, pacemaker-driven rhythm, or COPD.
Regulatory Classification
Identification
A percutaneous, implanted, long-term intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings, such as luer hubs, and accessories that facilitate the placement of the device. The device allows for repeated access to the vascular system for long-term use of 30 days or more, and it is intended for administration of fluids, medications, and nutrients; the sampling of blood; and monitoring blood pressure and temperature. The device may be constructed of metal, rubber, plastic, composite materials, or any combination of these materials and may be of single or multiple lumen design.
Special Controls
*Classification.* Class II (special controls) Guidance Document: “Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters.”
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is the Department of Health and Human Services logo, and on the right, there is the FDA logo. The FDA logo is in blue and consists of the letters "FDA" followed by the words "U.S. Food & Drug Administration".
April 12, 2018
Medical Components, Inc. (dba Medcomp) Courtney Nix Regulatory Affairs Manager, North America and EU 1499 Delp Drive Harleysville, Pennsylvania 19438
Re: K180567
Trade/Device Name: C3 Wave System Regulation Number: 21 CFR 880.5970 Regulation Name: Percutaneous, Implanted, Long-Term Intravascular Catheter Regulatory Class: Class II Product Code: LJS Dated: March 12, 2018 Received: March 15, 2018
Dear Courtney Nix:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Tina Kiang -
s
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K180567
Device Name
C3 Wave System
#### Indications for Use (Describe)
The C3 Wave System is indicated for use in the positioning of Peripherally Inserted Central catheters (PICC). The C3 Wave provides real-time catheter tip location by displaying changes in the patient's cardiac electrical activity. The C3 Wave is indicated for use as an alternative method to chest X-ray or fluoroscopy confirmation of PICC tip placement in adult patients.
Note: Limiting, but not contrindicated, situations for this technique are patients where cardiac rhythms may change presentation of the P-Wave: Atrial fibrillation Atrial flutter Severe tachycardia Pacemaker-Driven Rhythm Chronic obstructive pulmonary disease (COPD)
Such patients are easily identified prior to PICC insertion. Use of additional method is necessary to confirm catheter tip location.
Type of Use (Select one or both, as applicable)
| <span style="font-size:10pt">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> |
|------------------------------------------------------------------------------------|
| <span style="font-size:10pt">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| K180567 | | 510(k) SUMMARY | Special 510K |
|---------|------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------|--------------|
| | A. Submitter Information: | | |
| | Submitter: | Medical Components Inc.<br>(dba Medcomp®)<br>1499 Delp Drive<br>Harleysville, PA 19438<br>Tel: (215) 256-4201 x 2285<br>Fax: (215) 256-9191 | |
| | Registration Number: | 2518902 | |
| | Contact Person: | Courtney Nix<br>Regulatory Affairs Manager,<br>North America and Europe | |
| | Date of Preparation: | 03/02/2018 | |
| | B. Subject Device | | |
| | Trade Name: | C3 Wave System | |
| | Device: | Catheter, intravascular, Therapeutic Long-Term<br>Greater Than 30 Days | |
| | Product Code: | LJS | |
| | Regulation Description: | Percutaneous, implanted, long-term intravascular<br>catheter | |
| | C.F.R. Section: | 880.5970 | |
| | Class: | II | |
| | Regulation Medical<br>Specialty and Review<br>Panel: | General Hospital | |
| | C. Predicate Device | | |
| | 510(k) Number: | K170934 | |
| | 510(k) Holder: | Medcomp® | |
| | Trade Name: | C3 Wave System | |
| | Device: | Catheter, intravascular, Therapeutic Long-Term<br>Greater Than 30 Days | |
| | Product Code: | LJS | |
| | Regulation Description: | Percutaneous, implanted, long-term intravascular<br>catheter | |
| | C.F.R. Section: | 880.5970 | |
| | Class: | II | |
| | Regulation Medical<br>Specialty and Review<br>Panel: | General Hospital | |
Medcomp®: C3 Wave System
K180567 510(k) Summary
{4}------------------------------------------------
# D. Subject of Submission:
The purpose of this submission is to expand the accessory packs offerings for the C3 Wave System. The K170394 clearance includes three accessory packs. The submission is to add an accessory code that contains a large drape. As a result, the packaging system had to be revised to a larger header bag. All testing was completed and is the subject of this submission.
# E. Device Description:
C3 Wave is designed to provide a continuous display of electrocardiograph [ECG] waveform to be used as a guide in placement of peripherally-inserted central catheters [PICC] in the lower third of the Superior Vena Cava [SVC] of a patient. The principle for operation of this system uses three ECG leads placed on the patient's chest and generates a third ECG lead by switching from RA to PICC stylet. The ECG waveform is wirelessly transmitted to a tablet which allows the operator to view and record changes to the ECG waveform as the tip of the catheter approaches the heart. As the PICC catheter approaches the atrium of the heart, the P wave in the ECG waveform shows substantial changes. This system is designed to aid the visualization of changes in P wave amplitude.
The C3 Wave system must only be operated by a skilled nurse, physician, or trained medical professional who has been qualified in placement of PICC's and trained in the proper use of this device.
## F. Indications for Use:
The C3 Wave System is indicated for use in the positioning of Peripherally Inserted Central Catheters (PICC). The C3 Wave provides real-time catheter tip location information by displaying changes in the patient's cardiac electrical activity. The C3 Wave is indicated for use as an alternative method to chest X-ray or fluoroscopy confirmation of PICC tip placement in adult patients.
Note: Limiting, but not contraindicated, situations for this technique are patients where cardiac rhythms may change presentation of the P-Wave:
- -Atrial fibrillation
- -Atrial flutter
- Severe tachycardia
- -Pacemaker-Driven Rhythm
- -Chronic obstructive pulmonary disease (COPD)
Such patients are easily identified prior to PICC insertion. Use of additional method is necessary to confirm catheter tip location.
## G. Intended Use:
Medcomp®: C3 Wave System K180567 510(k) Summary
{5}------------------------------------------------
C3 Wave System is intended to provide real time tip location information of a central venous catheter by utilization of ECG to observe P-wave changes as the tip approaches the right atrium of the heart via the superior vena cava.
# H. Comparison to Predicate Devices:
| Device | Proposed Device:<br>C3 Wave System<br>(MRC3RD004) | Predicate Device:<br>C3 Wave System<br>(MRC3RD000) | Substantially<br>Equivalent<br>Comparison |
|------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------|
| Indications<br>for Use | The C3 Wave System<br>is indicated for use in<br>the positioning of<br>Peripherally Inserted<br>Central Catheters<br>(PICC). The C3 Wave<br>provides real-time<br>catheter tip location<br>information by<br>displaying changes in<br>the patient's cardiac<br>electrical activity. The<br>C3 Wave is indicated<br>for use as an<br>alternative method to<br>chest X-ray or<br>fluoroscopy<br>confirmation of PICC<br>tip placement in adult<br>patients.<br><br>Note: Limiting, but not<br>contraindicated,<br>situations for this<br>technique are patients<br>where cardiac rhythms<br>may change<br>presentation of the P-<br>Wave:<br>-Atrial fibrillation<br>-Atrial flutter- Severe<br>tachycardia<br>-Pacemaker-Driven<br>Rhythm<br>-Chronic obstructive<br>pulmonary disease<br>(COPD)<br>Such patients are | The C3 Wave System<br>is indicated for use in<br>the positioning of<br>Peripherally Inserted<br>Central Catheters<br>(PICC). The C3 Wave<br>provides real-time<br>catheter tip location<br>information by<br>displaying changes in<br>the patient's cardiac<br>electrical activity. The<br>C3 Wave is indicated<br>for use as an<br>alternative method to<br>chest X-ray or<br>fluoroscopy<br>confirmation of PICC<br>tip placement in adult<br>patients.<br><br>Note: Limiting, but not<br>contraindicated,<br>situations for this<br>technique are patients<br>where cardiac rhythms<br>may change<br>presentation of the P-<br>Wave:<br>-Atrial fibrillation<br>-Atrial flutter- Severe<br>tachycardia<br>-Pacemaker-Driven<br>Rhythm<br>-Chronic obstructive<br>pulmonary disease<br>(COPD)<br>Such patients are<br>easily identified prior | Unchanged |
| | to PICC insertion. Use<br>of additional method is<br>necessary to confirm<br>catheter tip location. | to PICC insertion. Use<br>of additional method is<br>necessary to confirm<br>catheter tip location. | |
| User/Setting | Prescription Use –<br>Hospital/Clinical<br>Setting | Prescription Use –<br>Hospital/Clinical<br>Setting | Unchanged |
| Patient/User<br>Interface | iPad | iPad | Unchanged |
| System | iPad, C3 Wave Hub,<br>ECG Leads, C3 Wave<br>Remote | iPad, C3 Wave Hub,<br>ECG Leads, C3 Wave<br>Remote | Unchanged |
| Overall<br>System<br>Components<br>(incl.<br>Accessory<br>Packs) | iPad, iPad Case,<br>Remote, Remote<br>Battery, Hub, Hub<br>Battery, Hub Power<br>Supply, Leads, Base,<br>ECG Clip Cable,<br>Remote Cover,<br>Electrodes, Skip Prep,<br>Drape | iPad, iPad Case,<br>Remote, Remote<br>Battery, Hub, Hub<br>Battery, Hub Power<br>Supply, Leads, Base,<br>ECG Clip Cable,<br>Remote Cover,<br>Electrodes, Skip Prep,<br>Drape | Equivalent; Change<br>to Dimension of<br>Drape |
| Sterile<br>Accessory<br>Pack<br>Components | ECG Clip Cable,<br>Remote Cover, and<br>Drape | ECG Clip Cable,<br>Remote Cover, and<br>Drape | Equivalent; Change<br>to Dimension of<br>Drape |
| Drape<br>Dimensions | 100" x 86" | 45" x 40" | Equivalent; Change<br>to Dimension of<br>Drape |
| Sterilization<br>Method | Sterile Accessory<br>Packs – EO | Sterile Accessory<br>Packs – EO | Unchanged |
| Packaging<br>Systems | 12.250"x18.500"<br>Tyvek Headerbag | 17.500"x10.750"<br>Tyvek Headerbag | Equivalent; Change<br>to Dimension of<br>Headerbag to<br>accommodate larger<br>drape |
| | CSR wrapped kit<br>placed inside a<br>Tyvek /LLDPE<br>headerbag and<br>shipped within a<br>corrugated shipping<br>carton | CSR wrapped kit<br>placed inside a<br>Tyvek /LLDPE<br>headerbag and<br>shipped within a<br>corrugated shipping<br>carton | |
| Testing<br>(Accessory<br>Packs) | ISTA 2A, Endotoxin,<br>EO Residuals, and<br>LAL | ISTA 2A, Endotoxin,<br>EO Residuals, and<br>LAL | Unchanged |
Table 6.1: 510(K) Summary: Design Comparison Matrix
{6}------------------------------------------------
#### Non-Clinical Testing l.
The Following non-clinical testing was completed on the modified accessory pack (larger drape and headerbag). Testing was performed as a result of the risk assessment and design control activities based on the subject device modifications from the cleared predicate.
Medcomp®: C3 Wave System K180567 510(k) Summary
{7}------------------------------------------------
## Biocompatibility:
Biocompatibility was performed for the device modifications per ISO 10993-1 for a skin contacting surface device, with limited exposure (i.e. less than 24 hours).
## CYTOTOXICITY:
ISO 10993-5: 2009; Biological Evaluation of Medical Devices – Part 5: Tests for In Vitro Cytotoxicity
#### SENSITIZATION:
ISO 10993-10; Biological Evaluation of Medical Devices – Part 10: Tests for Irritation and Skin Sensitization
#### IRRITATION:
ISO 10993-10; Biological Evaluation of Medical Devices – Part 10: Tests for Irritation and Skin Sensitization
#### Sterilization testing on the modified subject device:
The sterilization cycle was developed using the Overkill Approach defined in ANSI/AAMI/ISO 11135-1: 2014, Annex B. The Microbiological Performance Qualification was performed in accordance with ANSI/AAMI/ISO 11135-1: 2014, section 9.4.2 and the Physical Performance Qualification was performed in accordance with section 9.4.3.
The residuals for C3 ECG Cable Accessory Pack (MRC3RD004) were in accordance with AAMI/ANSI/ISO 10993-7:2008, Biological Evaluation of Medical Devices- Part 7: Ethylene Oxide Sterilization Residuals, the maximum level is 4/mg/device.
The packaging integrity of the pouch seal was validated with a dye test per ASTM F 1929-15.
The strength of the packaging seal was validated with a compressed air burst test per ASTM F1140/F1140M-13.
## J. Summary of Substantial Equivalence:
Based on the indications for use, design, and performance testing results, the modified subject device raises no new questions of safety or effectiveness compared to the predicate device and is substantially equivalent to the predicate device, C3 Wave System (K170934).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.