3F Single Lumen and 4F Dual Lumen PowerPICC Provena Catherers; 3F Single Lumen and 4F Dual Lumen PowerPICC Provena Catherters with SOLO Valve Technology
K180548 · C.R. Bard, Inc (Bard Has Joined Bd) · LJS · Apr 26, 2018 · General Hospital
Device Facts
Record ID
K180548
Device Name
3F Single Lumen and 4F Dual Lumen PowerPICC Provena Catherers; 3F Single Lumen and 4F Dual Lumen PowerPICC Provena Catherters with SOLO Valve Technology
Applicant
C.R. Bard, Inc (Bard Has Joined Bd)
Product Code
LJS · General Hospital
Decision Date
Apr 26, 2018
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.5970
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The PowerPICC Provena Catheters and PowerPICC Provena Catheters with SOLO2 Valve Technology are intended for short- or long-term peripheral access to the central venous system for intravenous therapy and blood sampling.
Device Story
Peripherally inserted central catheter (PICC) providing vascular access for intravenous therapy, blood sampling, power injection of contrast media, and central venous pressure monitoring. Device consists of catheter shaft, extension legs, and optional proximal silicone SOLO2 valve. Used in clinical settings (IR, nursing) for short or long-term access. Inserted percutaneously via peel-away sheath introducer. Healthcare providers use the device to deliver fluids or draw blood; the valve (if present) manages fluid flow. Benefits include reliable vascular access and compatibility with power injection protocols.
Clinical Evidence
Bench testing only. Performance testing included assembly tensile testing (ISO 10555-1), dimensional characterization, leak decay testing, and hydraulic burst testing. Biocompatibility evaluation conducted per ISO 10993-1. No clinical data presented.
Technological Characteristics
Materials: Polyether Polyurethane (extension legs), Polycarbonate Polyurethane (shaft/junction), Silicone (SOLO2 valve). Design: Peripherally inserted central catheter, single or dual lumen, open-ended distal tip. Connectivity: None. Sterilization: Provided sterile. Standards: ISO 10555-1, ISO 10993-1, BS EN ISO 14971.
Indications for Use
Indicated for short or long-term peripheral access to the central venous system for intravenous therapy, blood sampling, power injection of contrast media, and central venous pressure monitoring. Recommended for use with 20 gauge or larger lumen for pressure monitoring.
Regulatory Classification
Identification
A percutaneous, implanted, long-term intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings, such as luer hubs, and accessories that facilitate the placement of the device. The device allows for repeated access to the vascular system for long-term use of 30 days or more, and it is intended for administration of fluids, medications, and nutrients; the sampling of blood; and monitoring blood pressure and temperature. The device may be constructed of metal, rubber, plastic, composite materials, or any combination of these materials and may be of single or multiple lumen design.
Special Controls
*Classification.* Class II (special controls) Guidance Document: “Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters.”
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
April 26, 2018
C.R. Bard, Inc (Bard has joined BD) Brvan Stone Associate Manager, Regulatory Affairs 605 North 5600 West Salt Lake City, Utah 84116
Re: K180548
Trade/Device Name: PowerPICC Provena Catheters PowerPICC Provena Catheters with SOLO2 Valve Technology Regulation Number: 21 CFR 880.5970 Regulation Name: Percutaneous, Implanted, Long-Term Intravascular Catheter Regulatory Class: Class II Product Code: LJS Dated: February 27, 2018 Received: March 1, 2018
Dear Bryan Stone:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good
{1}------------------------------------------------
manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
## Geeta K. Pamidimukkala -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
### K180548
#### Device Name
PowerPICC Provena Catheters with SOLO2 Valve Technology
Indications for Use (Describe)
The PowerPICC Provena Catheters with SOLO² Valve Technology are indicated for short or long-term peripheral access to the central venous system for intravenous therapy, blood sampling, power injection of contrast media, and allows for central venous pressure monitoring. For central venous pressure monitoring, it is recommended that a catheter lumen of 20 gauge or larger be used.
| Catheter Size | Maximum Flow Rate |
|------------------|-------------------|
| 3 F Single Lumen | 3 mL/sec |
| 4 F Dual Lumen | 5 mL/sec |
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------------|
| Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known)
### K180548
Device Name PowerPICC Provena Catheters
#### Indications for Use (Describe)
The PowerPICC Provena Catheters are indicated for short or long-term peripheral access to the central venous system for intravenous therapy, blood sampling, power injection of contrast media, and allows for central venous pressure monitoring. For central venous pressure monitoring, it is recommended that a catheter lumen of 20 gauge or larger be used.
| Catheter Size | Maximum Flow Rate |
|------------------|-------------------|
| 3 F Single Lumen | 3 mL/sec |
| 4 F Dual Lumen | 5 mL/sec |
| Type of Use (Select one or both, as applicable) |
|------------------------------------------------------------------------------------|
| <span style="font-size:10px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
| <span style="font-size:10px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
#### 510(k) Summary K180548
21 CFR 807.92(a)
| | Submitter Name: | Bard Access Systems, Inc. (Bard has joined BD) |
|-----------------------|---------------------------|--------------------------------------------------------------------------------------------------|
| | Submitter Address: | 605 North 5600 West<br>Salt Lake City, UT 84116 |
| General<br>Provisions | Contact Person: | Bryan Stone<br>Associate Manager, Regulatory Affairs |
| | Telephone Number: | (801) 522-5876 |
| | Fax Number: | (801) 522-5425 |
| | Date of Preparation: | April 19, 2018 |
| | Trade Name(s): | PowerPICC Provena Catheters<br>PowerPICC Provena Catheters with SOLO2 Valve Technology |
| Subject Devices | Common Name: | Catheter, Intravascular, Therapeutic, Long-term Greater than 30 days |
| | Classification Name: | Percutaneous, Implanted, Long-term intravascular catheter |
| | Product Code/Regulation: | LJS/21 CFR §880.5970 |
| | Class | 2 |
| | Classification Panel | General Hospital |
| | | Predicate Trade Name: |
| Predicate Devices | Classification Name: | Percutaneous, Implanted, Long-term intravascular catheter |
| | Product Code/Regulation: | LJS/21 CFR §880.5970 |
| | Premarket Notification #: | K162443 (PowerPICC Provena)<br>K162441 (PowerPICC Provena Catheters with SOLO2 Valve Technology) |
| | Manufacturer: | Bard Access Systems, Inc. |
| Reference Device | Reference Trade Name: | PowerPICC SOLO Catheter |
| | Classification Name: | Percutaneous, Implanted, Long-term intravascular catheter |
| | Product Code/Regulation: | LJS/21 CFR §880.5970 |
| | Premarket Notification: | K072230 |
| | Manufacturer: | Bard Access Systems, Inc. |
{5}------------------------------------------------
| Device<br>Description | The PowerPICC Provena Catheters and PowerPICC Provena Catheters with SOLO2 Valve Technology are sterile, single use<br>devices designed to provide access to the patient's vascular system. The devices are intended for short- or long-term use, as<br>clinically indicated, to sample blood and administer fluids intravenously. The catheters are capable of central venous pressure<br>monitoring, and can withstand power injection of contrast media. The catheters are peripherally inserted central catheters<br>(PICC) and utilize the same placement technique as the predicate device. The SOLO2 versions of the subject devices include<br>a silicone valve on the proximal end.<br>The subject devices are provided sterile in basic interventional radiology (IR) as well as basic, full, and max barrier nursing<br>PICC kits with legally marketed components to assist in the placement procedure. These kits are available in both standard<br>and small patient versions. | | | | | | | | | | | |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|--|--|-------------------------------------------------------------------------------------|--|--|--|--|
| Intended Use | The PowerPICC Provena Catheters and PowerPICC Provena Catheters with SOLO2 Valve Technology are intended for<br>short- or long-term peripheral access to the central venous system for intravenous therapy and blood sampling. | | | | | | | | | | | |
| Indications For<br>Use | The PowerPICC Provena Catheters are indicated for short or long-term peripheral access to the central venous system for<br>intravenous therapy, blood sampling, power injection of contrast media, and allows for central venous pressure monitoring.<br>For central venous pressure monitoring, it is recommended that a catheter lumen of 20 gauge or larger be used. | | | | | | | | | | | |
| | Catheter Size Maximum Flow Rate 3F Single Lumen 3 mL/sec 4F Dual Lumen 5 mL/sec | | | | | | | Catheter Size Maximum Flow Rate 3F Single Lumen 3 mL/sec 4F Dual Lumen 5 mL/sec | | | | |
| | The PowerPICC Provena Catheters with SOLO2 Valve Technology are indicated for short or long-term peripheral access to<br>the central venous system for intravenous therapy, blood sampling, power injection of contrast media, and allows for central<br>venous pressure monitoring. For central venous pressure monitoring, it is recommended that a catheter lumen of 20 gauge<br>or larger be used. | | | | | | | | | | | |
{6}------------------------------------------------
| Technological<br>Characteristics | Technological characteristics of the subject PowerPICC Provena Catheters and PowerPICC Provena Catheters with SOLO²<br>Valve Technology are substantially equivalent with respect to basic design, function and fundamental scientific technology to<br>those of the cited primary predicate devices.<br><br>Key modifications made to the subject devices when compared to the predicate devices are as follows:<br>• The material formulation of the extension leg has been changed from Polycarbonate Polyurethane to Polyether<br>Polyurethane.<br><br>The following table provides a comparison between the subject and predicate devices. | | | |
|---------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------|
| | Attribute | Subject Devices – PowerPICC Provena<br>Catheters and PowerPICC Provena Catheters<br>With SOLO² Valve Technology | | |
| | Owner | Same | | |
| | Classification | Same | | |
| 510(k) Status | Subjects of this Premarket Notification | K162443 – Concurrence date October 25, 2016<br>K162441 – Concurrence date April 24, 2017 | | |
| | | | | |
| Indications for Use | Same | The PowerPICC Provena Catheters are indicated for short or long-term peripheral access to the central venous system for intravenous therapy, blood sampling, power injection of contrast media and allows for central venous pressure monitoring. For central venous pressure monitoring, it is recommended that a catheter lumen of 20 gauge or larger be used. | | |
| | | Catheter Size | Maximum Flow Rate | |
| | | 3F Single Lumen | 3 mL/sec | |
| | | 4F Dual Lumen | 5 mL/sec | |
| | | The PowerPICC Provena Catheters with SOLO2 Valve Technology are indicated for short or long-term peripheral access to the central venous system for intravenous therapy, blood sampling, power injection of contrast media, and allows for central venous pressure monitoring. For central venous pressure monitoring, it is recommended that a catheter lumen of 20 gauge or larger be used. | | |
| | | Catheter Size | Maximum Flow Rate | |
| | | 3F Single Lumen | 3 mL/sec | |
| | | 4F Dual Lumen | 5 mL/sec | |
| Commercial Name | Same | PowerPICC Provena Catheters<br>PowerPICC Provena Catheters with SOLO2 Valve Technology | | |
| Catheter | Same | 3F Single Lumen x 55 cm | | |
| Duration of Use | Same | Short (<30 days) or long-term (>30 days) | | |
| Means of<br>Insertion | Same | Percutaneous using a peel-away sheath<br>Introducer | | |
| Insertion Site | Same | Peripheral | | |
| Primary Device<br>Materials | Catheter Base Materials | Catheter Base Materials | | |
| | Shaft Tubing: | Shaft Tubing: | | |
| | Same | Polycarbonate Polyurethane | | |
| | Luer Connector: | Luer Connector: | | |
| | Same | Polyurethane | | |
| | Extension Legs: | Extension Legs: | | |
| | Polyether Polyurethane | Polycarbonate Polyurethane | | |
| | Junction: | Junction: | | |
| | Same | Polycarbonate Polyurethane (inner)<br>Polyether Polyurethane (outer) | | |
| | Valve (SOLO2 Only): | Valve (SOLO2 Only): | | |
| | Same | Silicone | | |
| Catheter<br>Proximal<br>Configuration | Same | Luer Connection<br>Luer Connection with Valve (SOLO2) | | |
| Catheter Distal<br>Configuration | Same | Open Ended | | |
| Number of<br>Lumens | Same | Single Lumen<br>Dual Lumen | | |
| | Power Injection<br>Maximum Flow Rate | Same | Catheter Size<br>3F Single Lumen | Maximum Flow Rate<br>3 mL/sec |
| | | | 4F Dual Lumen | 5 mL/sec |
| | Sterility | Same | Provided Sterile | |
| | Packaging<br>Configurations | Same | Both Standard and Small Patient versions of the following configurations: Basic Configuration Full Configuration Max Barrier Configuration IR Configuration | |
| | The technological differences listed above were evaluated using the same test requirements as the predicate devices, as defined in the Risk Assessment. Therefore, these differences in technological characteristics between the subject and predicate devices do not raise different questions of safety or effectiveness. | | | |
| Performance<br>Tests | Based on a risk analysis, the performance tests completed on the subject device were limited to those tests required to support a determination of substantial equivalence to the predicate device. In addition, when technological characteristics between the subject and predicate device were found to be identical, results of the performance testing conducted on the predicate device were applied to the subject device. The following table identifies the performance tests completed on the subject device based upon the specific modification to the extension leg material, including a test description and applicable standard associated with each test. | | | |
| | Verification /<br>Validation Method | Risk Acceptability Criteria (Acceptance Criteria of Test) | | |
| | Assembly Tensile<br>Testing | Test to demonstrate the peak tensile force of each test piece exceeds the minimum peak tensile force. ISO 10555-1:2013 – Sterile Single-Use Intravascular Catheters – Part 1: General requirements | | |
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
{10}------------------------------------------------
| | Dimensional<br>Characterization | Test to demonstrate that the new material formulation conforms correctly to design tolerances of the extension<br>legs.<br>BAS internal standards and procedures |
|------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Leak Decay<br>Testing | Testing performed to evaluate that the catheter assembly will not leak when the distal end of the catheter is<br>occluded.<br>BAS internal standards and procedures and ISO 10555-1:2013 – Sterile Single-Use Intravascular<br>Catheters - Part 1: General requirements |
| | Hydraulic Burst<br>Testing | Testing performed to evaluate that the catheter burst pressure exceeds the peak use pressure at maximum<br>flow conditions.<br>BAS internal standards and procedures and ISO 10555-1:2013 – Sterile Single-Use Intravascular<br>Catheters - Part 1: Genera…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.