K180546 · Jet Medical, Inc. · LJS · Nov 21, 2018 · General Hospital
Device Facts
Record ID
K180546
Device Name
Jet Power Injectable PICC
Applicant
Jet Medical, Inc.
Product Code
LJS · General Hospital
Decision Date
Nov 21, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5970
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Jet Power Injectable PICC is indicated for short term or long term peripheral access to the central venous system for intravenous therapy and power injection of contrast media, and allows for central venous pressure monitoring. For blood sampling, infusion, or therapies, use a 4F or larger catheter.
Device Story
Jet Power Injectable PICC is a trimmable, peripherally inserted central catheter (PICC) used for short or long-term central venous access. It facilitates intravenous therapy, power injection of contrast media, and central venous pressure monitoring. The device is inserted via the basilic, median cubital, or cephalic vein using the Modified Seldinger or Seldinger technique, with the tip placed in the superior vena cava. It is intended for use in adults. The catheter is provided sterile and includes accessories for percutaneous introduction. Healthcare providers use the device to deliver fluids, medications, or contrast media, or to withdraw blood samples. The device benefits patients by providing reliable, long-term venous access for various clinical therapies.
Clinical Evidence
No clinical data. Substantial equivalence is supported by non-clinical bench testing, including liquid/air leakage, force at break, gravity flow, chemical exposure, static burst, power injection performance, accelerated aging, and comprehensive biocompatibility testing per ISO 10993 standards.
Technological Characteristics
Trimmable PICC made of biocompatible medical-grade materials. Available in 3F (single lumen), 4F (single lumen), and 5F (double lumen) configurations. Features depth markings every 1cm with numerical markings every 5cm. Sterilized via 100% Ethylene Oxide (EO) gas. Complies with EN ISO 10555-1 for intravascular catheters.
Indications for Use
Indicated for adult patients requiring short or long-term peripheral access to the central venous system for IV therapy, power injection of contrast media, and central venous pressure monitoring. 4F or larger catheters required for blood sampling, infusion, or therapies.
Regulatory Classification
Identification
A percutaneous, implanted, long-term intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings, such as luer hubs, and accessories that facilitate the placement of the device. The device allows for repeated access to the vascular system for long-term use of 30 days or more, and it is intended for administration of fluids, medications, and nutrients; the sampling of blood; and monitoring blood pressure and temperature. The device may be constructed of metal, rubber, plastic, composite materials, or any combination of these materials and may be of single or multiple lumen design.
Special Controls
*Classification.* Class II (special controls) Guidance Document: “Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters.”
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 21, 2018
Jet Medical, Inc. % Courtney Nix Regulatory Affairs Manager, North America and Europe Medical Components Inc. (dba Medcomp®) 1499 Delp Drive Harleysville, Pennsylvania 19438
Re: K180546
Trade/Device Name: Jet Power Injectable PICC Regulation Number: 21 CFR 880.5970 Regulation Name: Percutaneous, Implanted, Long-Term Intravascular Catheter Regulatory Class: Class II Product Code: LJS Dated: October 22, 2018 Received: October 23, 2018
Dear Courtney Nix:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Geeta K. Pamidimukkala -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control. and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
| | DEPARTMENT OF HEALTH AND HUMAN SERVICES<br>Food and Drug Administration | | Form Approved: OMB No. 0910-0120<br>Expiration Date: 06/30/2020 |
|-------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|----------------------------------------------------------------------|
| | <b>Indications for Use</b> | | See PRA Statement below. |
| 510(k) Number (if known) | K180546 | | |
| Device Name | Jet Power Injectable PICC | | |
| Indications for Use (Describe) | The Jet Power Injectable PICC is indicated for short term or long term peripheral access to the central venous system for<br>intravenous therapy and power injection of contrast media, and allows for central venous pressure monitoring. For blood<br>sampling, infusion, or therapies, use a 4F or larger catheter. | | |
| | The maximum recommended infusion rate varies by catheter. The maximum recommended infusion rate for the 3F Jet<br>Power Injectable PICC is 1cc/sec and 5.0cc/sec for the 4F and 5F Jet Power Injectable PICC. | | |
| | | | |
| | | | |
| | | | |
| | | | |
| Type of Use (Select one or both, as applicable) | <input type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) | | <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
| | CONTINUE ON A SEPARATE PAGE IF NEEDED. | | |
| | This section applies only to requirements of the Paperwork Reduction Act of 1995. | | |
| | <b>*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*</b> | | |
| | The burden time for this collection of information is estimated to average 79 hours per response, including the<br>time to review instructions, search existing data sources, gather and maintain the data needed and complete<br>and review the collection of information. Send comments regarding this burden estimate or any other aspect<br>of this information collection, including suggestions for reducing this burden, to: | | |
| | Department of Health and Human Services<br>Food and Drug Administration<br>Office of Chief Information Officer<br>Paperwork Reduction Act (PRA) Staff<br>PRAStaff@fda.hhs.gov | | |
| | "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of<br>information unless it displays a currently valid OMB number." | | |
| FORM FDA 3881 (7/17) | Page 1 of 1 | | PSC Publishing Services (301) 443-6740<br>EF |
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for medCOMP. The logo consists of a red geometric shape that resembles a stylized letter "G" or an arrow pointing upwards. Below the geometric shape, the word "medCOMP" is written in a combination of red and black letters.
Harleysville, PA 19438
Tel: 215-256-4201
Fax. 215-256-1787
www.medcompnet.com
| Jet Power Injectable PICC<br>Section 6 | 510(k) SUMMARY<br>K180546 | Traditional 510K |
|--------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------|
| A. Submitter Information | | |
| Submitter Name: | Jet Medical Inc | |
| Address: | 97 Main Street<br>Schwenksville, PA, 19473 USA | |
| Registration Number: | 3007345394 | |
| Contact Person: | Courtney Nix<br>Cnix@medcompnet.com<br>Medical Components Inc (dba Medcomp®)<br>Regulatory Affairs Manager, North America<br>and EU (Est. Registration Number: 2518902) | |
| Date of Preparation: | 19NOV2018 | |
| B. Subject Device | | |
| Trade Name:<br>Common Name:<br>Regulation Name: | Jet Power Injectable PICC<br>Catheter, Intravascular, Therapeutic, Long-<br>Term Greater Than 30 Days<br>Percutaneous, Implanted, Long-Term<br>Intravascular Catheter | |
| Product Code:<br>21 CFR Regulation:<br>Class:<br>Classification Panel: | LJS<br>880.5970<br>II<br>General Hospital | |
| C. Predicate Device | | |
| Predicate<br>Trade Name:<br>510(k) Number:<br>510(k) Holder:<br>Common Name: | PRO-PICC® CT<br>K091953<br>Medical Components Inc.<br>Catheter, Intravascular, Therapeutic, Long-<br>Term Greater Than 30 Days | |
| Regulation Name:<br>Product Code:<br>21 CFR Regulation:<br>Class:<br>Classification Panel: | Percutaneous, Implanted, Long-Term<br>Intravascular Catheter<br>LJS<br>880.5970<br>II<br>General Hospital | |
- D. Device Description:
A family of trimmable peripherally inserted central catheters made from specially formulated biocompatible medical grade materials. Catheters are packaged in a tray
Jet Power Injectable PICC Section 6: 510(k) Site Summary
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for medCOMP. The logo consists of a red geometric shape that resembles a stylized letter "M" or a series of interconnected arrows. Below the geometric shape, the word "medCOMP" is written in a combination of red and black letters. The "med" portion is in red, while the "COMP" portion is in black.
Harleysville, PA 19438
Tel: 215-256-4201
Fax. 215-256-1787
www.medcompnet.com
Jet Power Injectable PICC
with accessories necessary for a percutaneous microintroducer introduction (Modified Seldinger or Seldinger Technique).
# E. Indications For Use
The Jet Power Injectable PICC is indicated for short term or long term peripheral access to the central venous system for intravenous therapy and power injection of contrast media, and allows for central venous pressure monitoring. For blood sampling, infusion, or therapies, use a 4F or larger catheter.
The maximum recommended infusion rate varies by catheter. The maximum recommended infusion rate for the 3F Jet Power Injectable PICC is 1cc/sec and 5.0cc/sec for the 4F and 5F Jet Power Injectable PICC.
### F. Intended Use
The Jet Power Injectable PICC is indicated for short term or long term peripheral access to the central venous system for intravenous therapy and power injection of contrast media, and allows for central venous pressure monitoring. For blood sampling, infusion, or therapies, use a 4F or larger catheter.
# G. Comparison to Predicate Device(s):
| Attribute | Jet Power<br>Injectable PICC<br>(Proposed Device) | Predicate Device<br>Pro-PICC® CT<br>(K091953) | Status |
|----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use | The Jet Power<br>Injectable PICC is<br>indicated for short<br>term or long term<br>peripheral access to<br>the central venous<br>system for<br>intravenous therapy<br>and power injection<br>of contrast media,<br>and allows for<br>central venous<br>pressure monitoring.<br>For blood sampling,<br>infusion, or<br>therapies, use a 4F<br>or larger catheter.<br><br>The maximum<br>recommended<br>infusion rate varies<br>by catheter. The<br>maximum | The Pro-PICC® CT is<br>indicated for short or<br>long term peripheral<br>access to the central<br>venous system for<br>intravenous therapy<br>and power injection of<br>contrast media, and<br>allows for central<br>venous pressure<br>monitoring, when a 20<br>gauge or larger lumen<br>is used. For blood<br>sampling, infusion, or<br>therapies, use a 4F or<br>larger catheter.<br><br>The maximum<br>recommended infusion<br>rate varies by catheter<br>French size and is<br>printed on the<br>catheter. | Substantially<br>equivalent; both<br>devices are<br>intended for short<br>or long term<br>peripheral access<br>to the central<br>venous system<br>for infusion, blood<br>sampling, and<br>injection of<br>contrast media. |
| Jet Power Injectable PICC | | | |
| | recommended<br>infusion rate for the<br>3F Jet Power<br>Injectable PICC is<br>1cc/sec and<br>5.0cc/sec for the 4F<br>and 5F Jet Power<br>Injectable PICC | | |
| Catheter OD | 3F Single Lumen<br>4F Single Lumen<br>5F Double Lumen | 4F Single Lumen<br>5F Single and Double<br>Lumen<br>6F Double Lumen | Substantially<br>Equivalent |
| Lengths | 3F Single Lumen,<br>4F Single Lumen,<br>and 5F Double<br>Lumen:<br>55CM | 4F Single Lumen:<br>50CM<br>5F Single Lumen<br>60 CM<br>5F and 6F Double<br>Lumen<br>55CM | Substantially<br>Equivalent |
| Duration of<br>Use | Short Term:<br>Greater than 24<br>hours but less than<br>or equal to 30 days)<br><br>Long Term:<br>Greater than 30<br>days | Short Term:<br>Greater than 24 hours<br>but less than or equal<br>30 days<br><br>Long Term:<br>Greater than 30 days | Identical |
| Means of<br>Insertion | Modified Seldinger<br>or Seldinger<br>Technique | Modified Seldinger or<br>Seldinger Technique | Identical |
| Vascular<br>Access<br>Placement | Basilic vein<br>Median Cubital Vein<br>Cephalic Vein | Basilic vein<br>Median Cubital Vein<br>Cephalic Vein | Identical |
| Preferred<br>Location | Basilic vein | Basilic Vein | Identical |
| Final<br>anatomical<br>location of Tip | Superior Vena Cava | Superior Vena Cava | Identical |
| Patient<br>Population | Adults | Adults | Identical |
| Material<br>Colors | 3F and 4F Single<br>Lumen<br><br>Tubing Clamp:<br>Purple<br>Extension:<br>Clear<br>Luer:<br>Red | 4F and 5F Single<br>Lumen<br><br>Luer:<br>Purple<br>Tubing Clamp:<br>Purple<br>Extension:<br>Translucent Purple | Substantially<br>Equivalent;<br>biocompatibility<br>testing |
| Jet Power Injectable PICC | | | |
| | Single Lumen | Hub | |
| | Tubing W/ Taper: | Purple | |
| | White | Lumen | |
| | Hub: | Light Purple | |
| | White | | |
| | | 5F and 6F Double<br>Lumen | |
| | 5F Double Lumen | | |
| | | | |
| | Tubing Clamp: | Tubing Clamp: | |
| | Purple | Purple | |
| | Extension: | Luer: | |
| | Clear | White | |
| | Luer: | Luer: | |
| | Natural | Purple | |
| | Luer: | Extension: | |
| | Red | Translucent Purple | |
| | Double Lumen | Hub: | |
| | Tubing W/ Taper: | Purple | |
| | White | Lumen | |
| | Hub: | Light purple | |
| | White | | |
| Depth | Depth Markings are<br>marked every 1cm.<br>Numerical markings<br>are placed every<br>5cm | Depth markings are<br>marked every 1cm.<br>Numerical markings<br>are placed every 5cm. | Identical |
| Catheter | Single Lumen<br>Double Lumen | Single Lumen<br>Double Lumen | Identical |
| Sterility | EO | EO | Identical |
# Table 6.1: 510(k) Summary Design Comparison Matrix
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for medCOMP. The logo consists of a red geometric shape that resembles a stylized letter "G" or a square with a curved top. Below the geometric shape, the word "medCOMP" is written in a combination of red and black letters. The "med" portion is in red, while the "COMP" portion is in black.
- Harleysville, PA 19438
Tel: 215-256-4201
Fax. 215-256-1787
www.medcompnet.com
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for medCOMP. The logo consists of a red geometric shape that resembles a stylized house or arrow pointing upwards. Below the shape, the word "med" is written in red, followed by "COMP" in black.
Harleysville, PA 19438
Tel: 215-256-4201
Fax. 215-256-1787
www.medcompnet.com
### H. Bench/ Performance Data/ Non-Clinical Testing:
# Table 6.2: Performance Standards
| Bench Testing | Applicable<br>Performance<br>Standard | Standard Title | Revision/Date |
|----------------------------------------------------------------------------------------------------------|---------------------------------------|---------------------------------------------------------------------------------------------------------------------------|---------------|
| Liquid Leakage<br>Air Leakage | EN ISO 10555-1 | Intravascular<br>Catheter - Sterile<br>and Single - Use<br>Intravascular<br>Catheters – Part 1<br>General<br>Requirements | 2009 |
| Force at Break<br>Gravity Flow<br>Chemical Exposure<br>Max Static Burst<br>Power Injection<br>Elongation | | | |
| Accelerated Aging | ASTM F 1980 | Standard Guide for<br>Accelerated Aging<br>of Sterile Barrier<br>System for Medical<br>Devices | 2016 |
Jet Power Injectable PICC Section 6: 510(k) Site Summary 6-4
{7}------------------------------------------------
| | Jet Power Injectable PICC | |
|--|-----------------------------------------------|----------------------------------|
| | Suture Wing<br>Integrity | Passed Per Internal Requirements |
| | Pump Flow Rate | |
| | Priming Volume | |
| | Catheter Collapse<br>Test | |
| | Catheter Stiffness | |
| | Interaction Testing | |
| | Simulated Use | |
| | Tip Displacement<br>During Power<br>Injection | |
| | Power Injection<br>Flow Rate | |
| | Dynamic Failure | |
| | Power Injection<br>Simulation Testing | |
| | Cyclic Flexure /<br>Kink Test | |
All testing was conducted on dry catherers.
#### Biocompatibility: -
Biocompatibility was performed for the 5F Jet Power Injectable PICC per ISO 10993-1 for an externally communicating device with permanent contact with circulating blood (greater than 30 days of contact). Biocompatibility was performed on the final finished device. The biological end points include:
# Cytotoxicity
- ISO 10993-5 Biological Evaluation of Medical Devices Part 5: Tests for . In Vitro Cytotoxicity
# Sensitization
- ISO 10993-10 Biological Evaluation of Medical Devices Part 10: Tests . for irritation and Skin Sensitization
- OECD 406 Guidelines for Testing of Chemical, Test No. 406, Skin . Sensitization
#### Irritation
- ISO 10993-10 Biological Evaluation of Medical Devices Part 10: Tests . for Irritation and Skin Sensitization
# Acute Systemic Toxicity
- ISO 10993-11 Biological Evaluation of Medical Devices Part 11: Tests 0 for Systemic Toxicity
- United States Pharmacopeia 40, National Formulary 35 (USP), General . Chapter <151>, Pyrogen Test
# Subchronic Toxicity
Jet Power Injectable PICC Section 6: 510(k) Site Summary
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the MedComp logo. The logo consists of a red geometric shape that resembles a stylized house or building, with an opening in the center. Below the shape, the word "medCOMP" is written in a combination of red and black letters. The "med" part is in red, while the "COMP" part is in black.
Harlevsville, PA 19438
Tel: 215-256-4201
Fax. 215-256-1787
www.medcompnet.com
Jet Power Injectable PICC
- ISO 10993-11 Biological Evaluation of Medical Devices Part 11: Tests for 0 Systemic Toxicity
- OECD 407 Guidelines for the Testing of Chemicals, Test No. 407, Repeated . Dose 28-Day Oral Toxicity Study in Rodents
## Genotoxicity
- ISO 10993-3 Biological Evaluation of Medical Devices Part 3: Tests for . Genotoxicity, Carcinogenicity, and Reproductive Toxicity
- OECD 471 Guidelines for Testing of Chemicals. Test No.471. Bacterial . Reverse Mutation Test
- OECD 490 Guidelines for the Testing of Chemicals. Test No. 490. In Vitro . Mammalian Cell Gene Mutation Test Using the Thymidine Kinase Gene.
#### Implantation
- ISO 10993-6 Biological Evaluation of Medical Devices Part 6: Tests for o Local Effects After Implantation
- United States Pharmacopeia 40. National Formulary 35 (USP). General . Chapter < 88>, Biological Reactivity Tests, In Vivo, Implantation Test, Table 6: Evaluation of Encapsulation in the Implantation Test
### Hemocompatibility
- 0 ISO 10993-4 Biological Evaluation of Medical Devices - Part 4: Selection of Tests for Interactions with Blood
- 0 ASTM F756 Standard Practice for Assessment of Hemolytic Properties of Materials
## Further Testing
- European Pharmacopoeia 9th edition, section 2,2,24, Absorption . Spectrophotometry, Infrared (2017)
- ISO 10993-18 Biological Evaluation of Medical Devices Part 18: Chemical . Characterization of Materials
- United States Pharmacopeia 40, National Formulary 35 (USP), General ® Chapter <197>. Spectrophotometric Identification Tests, Infrared Absorption
#### J. Sterility
| Sterilization Validation: | ANSI/AAMI/ISO 11135-1:2014, AAMI TIR<br>28:2009(R)2013 |
|---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sterility Assurance Level | SAL is 10-6 |
| Sterilization Method: | The product is sterilized by 100% Ethylene<br>Oxide (EO) Gas in a fixed chamber |
| ETO Residual Level: | In accordance with AAMI/ANSI/ISO 10993-<br>7:2008, Biological Evaluation of Medical<br>Devices - Part 7: Ethylene Oxide Sterilization<br>Residuals, the maximum level is less than to 4<br>mg/device. |
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for medCOMP. The logo consists of a red geometric shape that resembles a stylized letter "M" or a medical symbol. Below the shape, the word "medCOMP" is written in a combination of red and black letters, with "med" in red and "COMP" in black.
Harleysville, PA 19438
Tel: 215-256-4201
Fax. 215-256-1787
www.medcompnet.com
Jet Power Injectable PICC
Pvrogen Test Method:
Tested in accordance with ANSI/AAMI ST72:2011 Bacterial Endotoxin-Test methods. routine monitoring, and alternatives to testing the product is Non-pvrogenic.
Endotoxin:
Tested in accordance with ANSI/AAMI ST72:2011 Bacterial Endotoxin-Test methods, routine monitoring, and alternatives to testing the maximum level is 20 EU/device.
#### K. Shelf Life
Shelf life testing was conducted per ASTM 1980F: 2016 Standard Guide for Accelerated Aaing of Sterile Barrier Systems for Medical devices as well as on the Jet Power Injectable PICC device using performance standard EN ISO 10555-1.
#### L. Summary of Substantial Equivalence:
In conclusion, the proposed device, Jet Power Injectable PICC, is considered substantially equivalent to the predicate devices, Pro-PICC® CT (K091953) as demonstrated through non-clinical testing performed.
The proposed device, Jet Power Injectable PICC, does not present any additional risks to patients or different considerations regarding equivalence than those presented by the predicate, Pro-PICC® CT (K091953). The proposed device. Jet Power Injectable PICC, and the predicate device, Pro-PICC® CT (K091953) are equivalent in terms of indications for use, intended use, specifications, anatomical location, biocompatibility, performance and labelling.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.