K180535 · GE Healthcare · IYN · Mar 29, 2018 · Radiology
Device Facts
Record ID
K180535
Device Name
Voluson P6 and Voluson P8
Applicant
GE Healthcare
Product Code
IYN · Radiology
Decision Date
Mar 29, 2018
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1550
Device Class
Class 2
Attributes
Pediatric
Indications for Use
Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: Fetal/OB; Abdominal (including renal and GYN/pelvic); Pediatric; Small Organ (breast, testes, thyroid, salivary gland, lymph nodes, pediatic and neonatal patients); Neonatal and Adult Cephalic; Cardiac (adult and pediatric); Peripheral Vascular (PV); Musculo-skeletal Conventional and Superficial; Transrectal (Including Urology and Prostate) (TR); Transvaginal (TV).
Device Story
Mobile ultrasound console with keyboard, color LCD, and DICOM networking; utilizes linear, curved linear, and matrix phased array transducers (mechanical/electronic scanning). Inputs: acoustic signals from human tissue. Processing: real-time 3D/4D acquisition, image analysis, and reporting. Outputs: diagnostic ultrasound images and fluid flow analysis displayed for clinician review. Used in clinical settings by physicians/technicians for diagnostic imaging and biopsy guidance. Benefits: non-invasive visualization of internal anatomy and blood flow to support clinical decision-making. Includes software features: IOTA Simple Rules for ovarian tumor assessment, SonoMetrium for endometrial cavity 3D analysis, and Coded Contrast imaging.
Clinical Evidence
No clinical data required. Substantial equivalence supported by bench testing, including acoustic output measurement, biocompatibility, cleaning/disinfection effectiveness, and thermal/electrical/electromagnetic/mechanical safety testing per AAMI/ANSI/IEC/ISO standards.
Technological Characteristics
Mobile console with color LCD; supports linear, curved linear, and matrix phased array transducers. Connectivity: DICOM. Software: IOTA Simple Rules, SonoMetrium, Coded Contrast. Safety standards: AAMI/ANSI ES60601-1, IEC60601-1-2, IEC60601-2-37, ISO10993-1, NEMA UD 2, ISO14971.
Indications for Use
Indicated for diagnostic ultrasound imaging and fluid flow analysis in fetal/OB, abdominal, pediatric, small organ, neonatal/adult cephalic, cardiac, peripheral vascular, musculoskeletal, transrectal, and transvaginal applications. Includes biopsy guidance.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
Voluson P Series; Voluson P6/P8 Diagnostic Ultrasound System (K160162)
Reference Devices
Voluson E Series; E6/E8/E10 Diagnostic Ultrasound System (K172342)
Voluson E Series; E6/E8/E10 Diagnostic Ultrasound System (K162269)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 29, 2018
GE Healthcare % Mr. Bryan Behn Regulatory Affairs Director 9900 Innovation Drive WAUWATOSA WI 53226
Re: K180535
Trade/Device Name: Voluson P6 and Voluson P8 Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, ITX Dated: February 26, 2018 Received: February 28, 2018
Dear Mr. Behn:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael D'Hara For
Robert Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'g' and 'e' intertwined in a stylized script, enclosed within a circular border. The border has decorative flourishes or embellishments around the edge, giving it a classic or vintage appearance. The logo is presented in grayscale.
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known) K180535
Device Name Voluson P6 and Voluson P8
#### Indications for Use (Describe)
The device is a general-purpose ultrasound system. Specific clinical applications remain the same as previously cleared: Fetal/OB; Abdominal (including renal and GYN/pelvic); Pediatric; Small Organ (breast, testes, thyroid, salivary gland, lymph nodes, pediatic and neonatal patients); Neonatal and Adult Cephalic; Cardiac (adult and pediatric); Peripheral Vascular (PV); Musculo-skeletal Conventional and Superficial; Transrectal (Including Urology and Prostate) (TR); Transvaginal (TV).
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (7/17)
PSC Publishing Services (301) 443-6740 EF
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined in a stylized script, enclosed within a circular border. The logo is presented in a gray color scheme, with the letters and border appearing in a lighter shade of gray against a slightly darker gray background.
### Indications for Use Forms
The following forms represent indications with clinical applications and exam types along with the modes of operation for the Voluson P6 and Voluson P8 systems. Combinations identified "P" for the system represents those previously cleared in K160162. This modification does not add to indications to the previously cleared system level.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' intertwined and enclosed within a circular border. The logo is presented in a gray color scheme, with the letters and border appearing in a lighter shade of gray against a slightly darker gray background.
### Diagnostic Ultrasound Indications for Use Form
Voluson P6 / Voluson P8 Ultrasound System
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | | | | |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|--------------------|---------------------|----------------|------------------|--|--|
| Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>[Notes] | | |
| Ophthalmic | | | | | | | | | | | | | |
| Fetal / Obstetrics[7] | P | P | P | P | P | P | P | P | P | P | [5,6] | | |
| Abdominal[1] | P | P | P | P | P | P | P | P | P | P | [5,6] | | |
| Pediatric | P | P | P | P | P | P | P | P | P | P | [6] | | |
| Small Organ[2] | P | P | P | P | P | P | P | P | P | P | [6] | | |
| Neonatal Cephalic | P | P | P | P | P | P | P | P | P | P | [6] | | |
| Adult Cephalic | P | P | P | P | P | P | P | P | P | P | [6] | | |
| Cardiac[3] | P | P | P | P | P | P | P | P | P | P | [6] | | |
| Peripheral Vascular | P | P | P | P | P | P | P | P | P | P | [6] | | |
| Musculo-skeletal Conventional | P | P | P | P | P | P | P | P | P | P | [5,6] | | |
| Musculo-skeletal Superficial | P | P | P | P | P | P | P | P | P | P | [6] | | |
| Other | | | | | | | | | | | | | |
| Exam Type, Means of Access | | | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | | | |
| Transrectal[8] | P | P | P | P | P | P | P | P | P | P | [5,6] | | |
| Transvaginal | P | P | P | P | P | P | P | P | P | P | [5,6] | | |
| Transuretheral | | | | | | | | | | | | | |
| Intraoperative | | | | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | | | |
N = new indication; P = previously cleared by FDA
[1] Abdominal includes renal, GYN/Pelvic Notes:
[2] Small organ includes breast, testes, thyroid, salivary gland, lymph nodes, pediatric and neonatal patients
[3] Cardiac is Adult and Pediatric
[5] 3D/4D Imaging Mode
[6] Includes imaging of guidance of biopsy (2D/3D/4D)
[7] Includes infertility monitoring of follicle development
[8] Includes urology/prostate
[*] Combined modes are B/M, B/Color M, B/PWD, B/Color/PWD, B/PWD
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" in a stylized, cursive font, enclosed within a circle. The circle has a decorative border with small, rounded elements. The logo is presented in a grayscale color scheme.
### Diagnostic Ultrasound Indications for Use Form
#### Voluson P6 / Voluson P8 with IC9-RS Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | | | |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|--------------------|---------------------|----------------|------------------|--|
| Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>[Notes] | |
| Ophthalmic | | | | | | | | | | | | |
| Fetal / Obstetrics[7] | P | P | P | | P | P | P | P | P | P | [ 6] | |
| Abdominal[1] | P | P | P | | P | P | P | P | P | P | [ 6] | |
| Pediatric | | | | | | | | | | | | |
| Small Organ[2] | | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | | |
| Cardiac[3] | | | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | | | |
| Musculo-skeletal Conventional | | | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | | | |
| Other | | | | | | | | | | | | |
| Exam Type, Means of Access | | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | | |
| Transrectal[8] | P | P | P | | P | P | P | P | P | P | [ 6] | |
| Transvaginal[10] | P | P | P | | P | P | P | P | P | P | [ 6] | |
| Transuretheral | | | | | | | | | | | | |
| Intraoperative | | | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | | |
N = new indication; P = previously cleared by FDA
[1] Abdominal includes renal, GYN/Pelvic Notes:
[2] Small organ includes breast, testes, thyroid, salivary gland, lymph nodes, pediatric and neonatal patients
[3] Cardiac is Adult and Pediatric
[5] 3D/4D Imaging Mode
[6] Includes imaging of guidance of biopsy (2D/3D/4D)
[7] Includes infertility monitoring of follicle development
[8] Includes urology/prostate
[*] Combined modes are B/M, B/Color M, B/PWD, B/Color/PWD, B/PWD
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' intertwined within a circular border. The logo is presented in a gray color scheme, with the letters and border appearing in a lighter shade of gray against a slightly darker background.
### Diagnostic Ultrasound Indications for Use Form
### Voluson P6 / Voluson P8 with 4C-RS Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | | | |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|--------------------|---------------------|----------------|------------------|--|
| Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>[Notes] | |
| Ophthalmic | | | | | | | | | | | | |
| Fetal / Obstetrics[7] | P | P | P | | P | P | P | P | P | P | [6] | |
| Abdominal[1] | P | P | P | | P | P | P | P | P | P | [6] | |
| Pediatric | P | P | P | | P | P | P | P | P | P | [6] | |
| Small Organ[2] | | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | | |
| Cardiac[3] | | | | | | | | | | | | |
| Peripheral Vascular | P | P | P | | P | P | P | P | P | P | [6] | |
| Musculo-skeletal Conventional | P | P | P | | P | P | P | P | P | P | [6] | |
| Musculo-skeletal Superficial | | | | | | | | | | | | |
| Other | | | | | | | | | | | | |
| Exam Type, Means of Access | | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | | |
| Transrectal[8] | | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | | |
| Transuretheral | | | | | | | | | | | | |
| Intraoperative | | | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | | |
N = new indication; P = previously cleared by FDA
Notes: [1] Abdominal includes renal, GYN/Pelvic
[2] Small organ includes breast, testes, thyroid, salivary gland, lymph nodes, pediatric and neonatal patients
[3] Cardiac is Adult and Pediatric.
[5] 3D/4D Imaging Mode.
[6] Includes imaging of guidance of biopsy (2D/3D/4D).
[7] Includes infertility monitoring of follicle development.
[8] Includes urology/prostate.
[*] Combined modes are B/M, B/Color M, B/PWD, B/Color/PWD, B/PWD
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" in a stylized script, enclosed within a circular frame. The frame has a decorative, swirling pattern around the letters. The logo is presented in a grayscale color scheme.
### Diagnostic Ultrasound Indications for Use Form
#### Voluson P6 / Voluson P8 with E8C-RS Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | | | |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|--------------------|---------------------|----------------|------------------|--|
| Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>[Notes] | |
| Ophthalmic | | | | | | | | | | | | |
| Fetal / Obstetrics[7] | P | P | P | | P | P | P | P | P | P | [6] | |
| Abdominal[1] | P | P | P | | P | P | P | P | P | P | [6] | |
| Pediatric | | | | | | | | | | | | |
| Small Organ[2] | | | | | | | | | | | | |
| Neonatal Cephalic | P | P | P | | P | P | P | P | P | P | [6] | |
| Adult Cephalic | | | | | | | | | | | | |
| Cardiac[3] | | | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | | | |
| Musculo-skeletal Conventional | | | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | | | |
| Other | | | | | | | | | | | | |
| Exam Type, Means of Access | | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | | |
| Transrectal[8] | P | P | P | | P | P | P | P | P | P | [6] | |
| Transvaginal | P | P | P | | P | P | P | P | P | P | [6] | |
| Transuretheral | | | | | | | | | | | | |
| Intraoperative | | | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | | |
N = new indication; P = previously cleared by FDA
Notes: [1] Abdominal includes renal, GYN/Pelvic
[2] Small organ includes breast, testes, thyroid, salivary gland, lymph nodes, pediatric and neonatal patients
[3] Cardiac is Adult and Pediatric
[5] 3D/4D Imaging Mode
[6] Includes imaging of guidance of biopsy (2D/3D/4D)
[7] Includes infertility monitoring of follicle development
[8] Includes urology/prostate
[*] Combined modes are B/M, B/Color M, B/PWD, B/Color/PWD, B/PWD
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" in a stylized script, enclosed within a circular frame. The frame has a decorative, swirling pattern around the letters. The logo is presented in a grayscale color scheme.
### Diagnostic Ultrasound Indications for Use Form
#### Voluson P6 / Voluson P8 with 12L-RS Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | | | |
|----------------------------------------------------|-------------------|---|------------|------------|---------------|-----------------|---------------|-----------------|------------------|-------------|---------------|--|
| Clinical Application<br>Anatomy/Region of Interest | B | M | PW Doppler | CW Doppler | Color Doppler | Color M Doppler | Power Doppler | Combined Modes* | Harmonic Imaging | Coded Pulse | Other [Notes] | |
| Ophthalmic | | | | | | | | | | | | |
| Fetal / Obstetrics[7] | | | | | | | | | | | | |
| Abdominal[1] | | | | | | | | | | | | |
| Pediatric | P | P | P | | P | P | P | P | P | P | [6] | |
| Small Organ[2] | P | P | P | | P | P | P | P | P | P | [6] | |
| Neonatal Cephalic | | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | | |
| Cardiac[3] | | | | | | | | | | | | |
| Peripheral Vascular | P | P | P | | P | P | P | P | P | P | [6] | |
| Musculo-skeletal Conventional | P | P | P | | P | P | P | P | P | P | [6] | |
| Musculo-skeletal Superficial | P | P | P | | P | P | P | P | P | P | [6] | |
| Other | | | | | | | | | | | | |
| Exam Type, Means of Access | | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | | |
| Transrectal[8] | | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | | |
| Transuretheral | | | | | | | | | | | | |
| Intraoperative | | | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | | |
N = new indication; P = previously cleared by FDA
Notes: [1] Abdominal includes renal, GYN/Pelvic
[2] Small organ includes breast, testes, thyroid, salivary gland, lymph nodes, pediatric and neonatal patients
[3] Cardiac is Adult and Pediatric.
[5] 3D/4D Imaging Mode.
[6] Includes imaging of guidance of biopsy (2D/3D/4D).
[7] Includes infertility monitoring of follicle development.
[8] Includes urology/prostate.
[*] Combined modes are B/M, B/Color M, B/PWD, B/Color/PWD, B/PWD
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' intertwined in a stylized script, enclosed within a circular frame. The frame has a decorative, swirling pattern around the letters. The logo is presented in a grayscale color scheme.
### Diagnostic Ultrasound Indications for Use Form
#### Voluson P6 / Voluson P8 with 3Sc-RS Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | | | |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|--------------------|---------------------|----------------|------------------|--|
| Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>[Notes] | |
| Ophthalmic | | | | | | | | | | | | |
| Fetal / Obstetrics[7] | P | P | P | P | P | P | P | P | P | P | [6] | |
| Abdominal[1] | P | P | P | P | P | P | P | P | P | P | [6] | |
| Pediatric | P | P | P | P | P | P | P | P | P | P | [6] | |
| Small Organ[2] | | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | | |
| Adult Cephalic | P | P | P | P | P | P | P | P | P | P | [6] | |
| Cardiac[3] | P | P | P | P | P | P | P | P | P | P | [6] | |
| Peripheral Vascular | | | | | | | | | | | | |
| Musculo-skeletal Conventional | | | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | | | |
| Other | | | | | | | | | | | | |
| Exam Type, Means of Access | | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | | |
| Transrectal[8] | | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | | |
| Transuretheral | | | | | | | | | | | | |
| Intraoperative | | | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | | |
N = new indication; P = previously cleared by FDA
Notes: [1] Abdominal includes renal, GYN/Pelvic
[2] Small organ includes breast, testes, thyroid, salivary gland, lymph nodes, pediatric and neonatal patients
[3] Cardiac is Adult and Pediatric.
[5] 3D/4D Imaging Mode.
[6] Includes imaging of guidance of biopsy (2D/3D/4D).
[7] Includes infertility monitoring of follicle development.
[8] Includes urology/prostate.
[*] Combined modes are B/M, B/Color M, B/PWD, B/Color/PWD, B/PWD
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' intertwined and enclosed within a circular frame. The frame has a decorative, swirling pattern around the letters. The logo is presented in a grayscale color scheme.
### Diagnostic Ultrasound Indications for Use Form
### Voluson P6 / Voluson P8 with RIC5-9A-RS Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | | |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|--------------------|---------------------|----------------|------------------|
| Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>[Notes) |
| Ophthalmic | | | | | | | | | | | |
| Fetal / Obstetrics[7] | P | P | P | | P | P | P | P | P | P | [ 5,6] |
| Abdominal[1] | P | P | P | | P | P | P | P | P | P | [ 5,6] |
| Pediatric | | | | | | | | | | | |
| Small Organ[2] | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac[3] | | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | | |
| Musculo-skeletal Conventional | | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | | |
| Other | | | | | | | | | | | |
| Exam Type, Means of Access | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal[8] | P | P | P | | P | P | P | P | P | P | [ 5,6] |
| Transvaginal | P | P | P | | P | P | P | P | P | P | [ 5,6] |
| Transuretheral | | | | | | | | | | | |
| Intraoperative | | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | |
N = new indication; P = previously cleared by FDA
Notes: [1] Abdominal includes renal, GYN/Pelvic
[2] Small organ includes breast, testes, thyroid, salivary gland, lymph nodes, pediatric and neonatal patients
[3] Cardiac is Adult and Pediatric.
[5] 3D/4D Imaging Mode.
[6] Includes imaging of guidance of biopsy (2D/3D/4D).
[7] Includes infertility monitoring of follicle development.
[8] Includes urology/prostate.
[*] Combined modes are B/M, B/Color M, B/PWD, B/Color/PWD, B/PWD
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" in a stylized, cursive font, enclosed within a circular border. The border has small, decorative flourishes or droplets around its perimeter. The logo is presented in a grayscale color scheme.
### Diagnostic Ultrasound Indications for Use Form
#### Voluson P6 / Voluson P8 with RAB2-6-RS Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | | |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|--------------------|---------------------|----------------|------------------|
| Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>[Notes] |
| Ophthalmic | | | | | | | | | | | |
| Fetal / Obstetrics[7] | P | P | P | | P | P | P | P | P | P | [5,6] |
| Abdominal[1] | P | P | P | | P | P | P | P | P | P | [5,6] |
| Pediatric | | | | | | | | | | | |
| Small Organ[2] | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac[3] | | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | | |
| Musculo-skeletal Conventional | P | P | P | | P | P | P | P | P | P | [5,6] |
| Musculo-skeletal Superficial | | | | | | | | | | | |
| Other | | | | | | | | | | | |
| Exam Type, Means of Access | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal[8] | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Transuretheral | | | | | | | | | | | |
| Intraoperative | | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | |
N = new indication; P = previously cleared by FDA
Notes: [1] Abdominal includes renal, GYN/Pelvic
[2] Small organ includes breast, testes, thyroid, salivary gland, lymph nodes, pediatric and neonatal patients
[3] Cardiac is Adult and Pediatric.
[5] 3D/4D Imaging Mode.
[6] Includes imaging of guidance of biopsy (2D/3D/4D).
[7] Includes infertility monitoring of follicle development.
[8] Includes urology/prostate.
[*] Combined modes are B/M, B/Color M, B/PWD, B/Color/PWD, B/PWD
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
{12}------------------------------------------------
Image /page/12/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" in a stylized, cursive font, enclosed within a circular border. The background of the circle is filled with a solid color, while the letters and the border are in a contrasting color. The overall design is simple and recognizable, representing the brand identity of General Electric.
### 510(k) Summary
| | In accordance with 21 CFR 807.92 the following summary of information is provided: | | | | | | | |
|----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|--|--|--|
| Date: | February 26, 2018 | | | | | | | |
| Submitter: | GE Healthcare<br>9900 Innovation Dr<br>Wauwatosa, WI 53226 | | | | | | | |
| Primary Contact Person: | Bryan Behn<br>Regulatory Affairs Director<br>GE Healthcare<br>T:(262)247-5502<br>F:(414)918-8275 | | | | | | | |
| Secondary Contact Person: | Jiyeon Park<br>Regulatory Affairs Leader<br>GE Ultrasound Korea, Ltd.<br>T: +82 31 740 6307<br>F: +82 31 740 6431 | | | | | | | |
| Device:<br>Trade Name: | Voluson P6 / Voluson P8 | | | | | | | |
| Common/Usual Name: | Ultrasound system | | | | | | | |
| Classification Names:<br>Product Code: | Class II<br>Ultrasonic Pulsed Doppler Imaging System. 21CFR 892.1550 90-IYN<br>Ultrasonic Pulsed Echo Imaging System, 21CFR 892.1560, 90-IYO<br>Diagnostic Ultrasound Transducer, 21 CFR 892.1570, 90-ITX | | | | | | | |
| Main Predicate Device(s): | K160162 Voluson P Series; Voluson P6/P8 Diagnostic<br>Ultrasound System | | | | | | | |
| Reference Predicate: | K172342 Voluson E Series; E6/E8/E10 Diagnostic Ultrasound<br>System<br>K162269 Voluson E Series; E6/E8/E10 Diagnostic Ultrasound<br>System | | | | | | | |
| Device Description: | The systems are full-featured Track 3 ultrasound systems,<br>primarily for general radiology use and specialized for OB/GYN<br>with particular features for real-time 3D/4D acquisition. They<br>consist of a mobile console with keyboard control panel; color<br>LCD display, color video display and optional image storage and<br>printing devices. They provide high performance ultrasound<br>imaging and analysis and have comprehensive networking and<br>DICOM capability. They utilize a variety of linear, curved linear,<br>matrix phased array transducers including mechanical and<br>electronic scanning transducers, which provide accurate real-time | | | | | | | |
{13}------------------------------------------------
Image /page/13/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a circle. There are decorative flourishes around the circle, adding to the logo's classic design. The logo is presented in grayscale.
three-dimensional imaging supporting all standard acquisition modes.
The device is a general-purpose ultrasound system. Specific Intended Use: clinical applications remain the same as previously cleared: Fetal/OB; Abdominal (including renal and GYN/pelvic); Pediatric; Small Organ (breast, testes, thyroid, salivary gland, lymph nodes, pediatric and neonatal patients); Neonatal and Adult Cephalic; Cardiac (adult and pediatric); Peripheral Vascular (PV); Musculo-skeletal Conventional and Superficial; Transrectal (Including Urology and Prostate) (TR); Transvaginal (TV).
The Voluson P Series (Voluson P6 / Voluson P8) employs the Technology: same fundamental scientific technology as its predicate devices.
#### Determination of Comparison to Predicates
Substantial Equivalence: The Voluson P6 / Voluson P8 is substantially equivalent to the predicate devices with regards to intended use, imaging capabilities, technologicial characteristics and safety and effectiveness.
- The systems are all intended for diagnostic ultrasound imaging and fluid flow analysis.
- The Voluson P Series and predicate Voluson P Series systems have the same clinical intended use.
- The Voluson P Series and predicate Voluson P Series ● systems have the same imaging modes.
- The Voluson P Series and predicate Voluson P Series ● systems transducers are identical.
- The Voluson P Series and predicate Voluson P Series systems have the same indications for use.
- The systems are manufactured with materials which have been evaluated and found to be safe for the intended use of the device.
- . The systems have acoustic power levels which are below the applicable FDA limits.
- The proposed Voluson P Series and predicate Voluson P ● Series systems have similar capability in terms of performing measurements, capturing digital images, reviewing and reporting studies.
{14}------------------------------------------------
Image /page/14/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" in a stylized, cursive font, enclosed within a circle. There are decorative flourishes around the circle, adding to the logo's classic design. The logo is presented in grayscale.
- The proposed Voluson P Series and predicate Voluson P ● series systems have been designed in compliance with approved electrical and physical safety standards.
- . The proposed Voluson P Series adds an improved version of existing software feature IOTA LR2 model called IOTA Simple Rules (previously cleared with K172342).
- The proposed Voluson P Series adds a 3D Analysis ● feature of the endometrial cavity called SonoMetrium. (previously cleared with K172342).
- The proposed Voluson P Series adde a software Coded ● Contrast (previously cleared with K172342).
### Summary of Non-Clinical Tests:
The device has been evaluated for acoustic output. biocompatibility, cleaning and disinfection effectiveness as well as thermal, electrical, electromagnetic, and mechanical safety, and has been found to conform to applicable medical device safety standards. The Voluson P6/P8 systems and its applications comply with voluntary standards:
- AAMI/ANSI ES60601-1, Medical Electrical Equipment -● Part 1: General Requirements for Safety, 2005/(R)2012 And A1:2012, C1:2009/(R)2012 And A2:2010/(R)2012 (Consolidated Text)
- IEC60601-1-2. Medical Electrical Equipment Part 1-2: ● General Requirements for Safety - Collateral Standard: Electromagnetic Compatibility Requirements and Tests 2007
- . IEC60601-2-37, Medical Electrical Equipment - Part 2-37: Particular Requirements for the Safety of Ultrasonic Medical Diagnostic and Monitoring Equipment, 2015
- ISO10993-1, Biological Evaluation of Medical Devices-● Part 1: Evaluation and Testing- Third Edition, 2009/(R)2013
- NEMA UD 2, Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment:2004 (R2009)
- . ISO14971, Application of risk management to medical devices: Second edition 2007
- NEMA PS 3.1 - 3.20 (2011), Digital Imaging and
{15}------------------------------------------------
Image /page/15/Picture/0 description: The image shows the logo for General Electric (GE). The logo is a circular emblem with the letters "GE" intertwined in the center. The letters are stylized and have a flowing, cursive appearance. The circle surrounding the letters has decorative flourishes or swirls around the perimeter.
Communications in Medicine (DICOM) Set. (Radiology)
The following quality assurance measures were applied to the development of the system:
- . Risk Analysis
- Requirements Reviews
- Design Reviews ●
- Testing on unit level (Module verification) ●
- Integration testing (System verification) ●
- Performance testing (Verification)
- Safety testing (Verification) ●
Transducer materials and other patient contact materials are biocompatible.
Summary of Clinical Tests:
The subject of this premarket submission, Voluson P6 / Voluson P8, did not require clinical studies to support substantial equivalence.
- Conclusion: GE Healthcare considers the Voluson P6 / Voluson P8 to be as safe, as effective, and performance is substantially equivalent to the predicate device(s).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.