The TruClear Elite Hysteroscope is used to permit viewing of the cervical canal and uterine cavity for the purpose of performing diagnostic and operative procedures.
Device Story
TruClear Elite Hysteroscopes are reusable, rigid stainless steel instruments for uterine cavity visualization and operative procedures. Devices feature rod-lens optics for visualization, optical fibers for illumination, and a working channel for instrument introduction. Inflow/outflow channels maintain uterine distension and visualization via fluid management systems or pressurized saline. Used in clinical settings by physicians; device is disassembled for cleaning and sterilization. Output is direct visualization of the uterine cavity, enabling diagnostic assessment and surgical intervention. Benefits include improved visualization and access for operative procedures.
Rigid hysteroscope; stainless steel, PEEK, sapphire, and optical glass construction. Rod-lens optical system; fiber optic illumination. Dimensions: 7.25mm (Plus) and 6mm (Mini) outer diameters. Inflow/outflow channels for continuous fluid flow. Compatible with ≤300W metal halide/xenon light sources. Electrical safety per IEC 60601-1; EMC per IEC 60601-2-18. Reusable design; disassembled for reprocessing.
Indications for Use
Indicated for visualization of the cervical canal and uterine cavity to perform diagnostic and operative procedures in patients requiring hysteroscopy.
Regulatory Classification
Identification
A hysteroscope is a device used to permit direct viewing of the cervical canal and the uterine cavity by a telescopic system introduced into the uterus through the cervix. It is used to perform diagnostic and surgical procedures other than sterilization. This generic type of device may include obturators and sheaths, instruments used through an operating channel, scope preheaters, light sources and cables, and component parts.
Predicate Devices
Smith & Nephew TRUCLEAR Operative Hysteroscope 8.0 (K013870)
Smith & Nephew TRUCLEAR Operative Hysteroscope 5C (K152143)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 29, 2018
Covidien Heather L. Harvey, MS, RAC Regulatory Affairs Specialist 60 Middletown Ave. North Haven, CT 06473
Re: K180496
> Trade/Device Name: TruClear™ Elite Hysteroscopes Regulation Number: 21 CFR§ 884.1690 Regulation Name: Hysteroscope and Accessories Regulatory Class: II Product Code: HIH Dated: July 27, 2018 Received: July 30, 2018
Dear Heather L. Harvey:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be avare that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Sharon M. Andrews -S
for
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K180496
Device Name TruClear™ Elite Hysteroscopes
Indications for Use (Describe)
The TruClear Elite Hysteroscope is used to permit viewing of the cervical canal and uterine cavity for the purpose of performing diagnostic and operative procedures.
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary K180496
#### Date Prepared:
February 23, 2018
#### Submitter:
Covidien 60 Middletown Avenue North Haven, CT 06473
#### Contact:
Harvev L Heather Regulatory Affairs Specialist Telephone: (781) 577-5871 (203) 492-5029 Fax: heather.l.harvey@medtronic.com Email:
#### Name of Device:
| Trade/Proprietary Name: | TruClear™ Elite Hysteroscopes |
|-------------------------|-----------------------------------------------|
| Common Name: | Hysteroscope |
| Class: | Class II |
| Regulation: | 21 CFR 884.1690, Hysteroscope and Accessories |
| Panel Code: | Obstetrics and Gynecology |
| Product Code: | HIH (Hysteroscope and accessories) |
#### Predicate Device:
| Trade/Proprietary Name: | Smith & Nephew TRUCLEAR Operative Hysteroscope 8.0 & | | | |
|---------------------------------------------------|------------------------------------------------------|--|--|--|
| Smith & Nephew TRUCLEAR Operative Hysteroscope 5C | | | | |
| Common Name: | Hysteroscope | | | |
| Classification Name: | Hysteroscope & Accessories, HIH, 21 CFR 884.1690 | | | |
| 510(k) Number: | K013870. K152143 | | | |
| Manufacturer: | Smith & Nephew, 150 Minuteman Drive, Andover, MA | | | |
The predicate devices have not been subject to a design related recall.
#### Device Description:
TruClear™ Elite Hysteroscopes are reusable instruments for use in visualizing the uterine cavity and performing diagnostic and operative procedures. The TruClear™ Elite Hysteroscopes are stainless steel instruments designed with a rod-lens optical channel for visualization, optical fibers for illumination, and a working channel to permit introduction of instrumentation.
The subject Truclear™ Elite Hysteroscopes Plus (7.25mm) and Mini (6mm) are modifications to the previously cleared Smith & Nephew Operative Hysteroscopes 8.0 (K013870) and 5C (K152143). Both the Truclear™ Elite Plus and Mini Hysteroscopes are rigid hysteroscopes with a slanted distal tip, rod lens optics and disposable proximal fluid seal.
The TruClear™ Elite Hysteroscopes are composed of the working channel insert and the hysteroscope body. The device is designed to be disassembled for access to interior lumens during cleaning and sterilization. Like the predicate devices, the subject devices have an inflow and outflow channel to provide continuous flow during the clinical procedure. The inflow channel passes through the hysteroscope body and provides operative fluid from a fluid management system or pressurized saline bag to distend the uterine cavity. The outflow channel which is contained within the working channel insert, clears the uterine cavity to maintain visualization. The outflow channel also acts as the instrument channel through which the tissue removal device is inserted.
{4}------------------------------------------------
Indications for Use:
The TruClear Elite Hysteroscope is used to permit viewing of the cervical canal and uterine cavity for the purpose of performing diagnostic and operative procedures. Predicate Device Comparison:
| | | | K180496 | K180496 |
|-------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device & Predicate Device(s): | K013870 | K152143 | Truclear Elite | Truclear Elite |
| | | | Hysteroscope | Hysteroscope |
| | | | Plus | Mini |
| Indication for Use Statement | The Smith &<br>Nephew<br>Operative<br>Hysteroscope<br>and Accessories<br>are used to<br>permit viewing<br>of the cervical<br>canal and<br>uterine cavity for<br>the purpose of<br>performing | The Smith &<br>Nephew<br>TRUCLEAR<br>Operative<br>Hysteroscope<br>5C and Sheath<br>5C are used to<br>permit viewing<br>of the cervical<br>canal and the<br>uterine cavity<br>for the purpose | The TruClear™<br>Elite<br>Hysteroscope<br>is used to<br>permit viewing<br>of the cervical<br>canal and<br>uterine cavity<br>for the purpose<br>of performing<br>diagnostic and | The TruClear™<br>Elite<br>Hysteroscope<br>is used to<br>permit viewing<br>of the cervical<br>canal and<br>uterine cavity<br>for the purpose<br>of performing<br>diagnostic and<br>operative |
| | diagnostic and<br>surgical<br>procedures | of performing<br>diagnostic and<br>surgical<br>procedures. | operative<br>procedures | procedures |
| Materials of<br>Construction | Stainless Steel,<br>PEEK, | Stainless Steel,<br>PEEK, | Stainless Steel,<br>PEEK, | Stainless Steel,<br>PEEK, |
| | Sapphire,<br>Optical Glass | Sapphire,<br>Optical Glass | Sapphire,<br>Optical Glass | Sapphire,<br>Optical Glass |
| Device Components | | | | |
| Dimensions | | | | |
| | | | | |
| Working Length (mm) | 191.5 (scope)<br>168 (sheath) | 219 (scope)<br>205 (sheath) | 201.1 | 201.1 |
| Overall Channel Length (mm) | 280 | 334 | Not provided | Not provided |
| Outer Diameter (mm) | 8.5 | 5.25 | 7.25 | 6 |
| Working Channel Diameter (mm) | [D shaped 5.5<br>by 5.4] | [D shaped 5.5<br>by 5.4] | 4.2 | 3.1 |
| Device Design | | | | |
| Distal Tip Shape | Flat | Slanted | Slanted | Slanted |
| Working Channel Closure | Stopcock | Stopcock | Seal | Seal |
| Compatible Light Source | | | | |
| Recommended Light source | Metal Halide<br>and Xenon | Metal Halide<br>and Xenon | Metal Halide<br>and Xenon | Metal Halide<br>and Xenon |
| Power Rating of Light source (watts) | ≤ 300 W | ≤ 300 W | ≤ 300 W | ≤ 300 W |
| Optics | | | | |
| Objective lens | | | | |
| Focal Length (mm) | 1.3 | 0.6 | 1.02 | 1.02 |
| Working Distance (mm) | 30 | 10 | 30 | 30 |
| Field of View (Degrees) | 80 | 85 | 80 | 80 |
| Direction of view (Degrees) | 0 | 0 | 0 | 0 |
| Optics | Rod lens | Fiber optic<br>(Image bundle) | Rod lens | Rod lens |
| Illumination fibers | | | | |
| Ratio of Luminous Energy Transmitted<br>to Energy Delivered | 23.5% (12%) | 15.6% (12.8%) | 23.5% | 23.5% |
| Image Transmission System | Not provided | Not provided | Rod lens | Rod lens |
| Eyepiece | | | | |
| Eyepiece Magnification | 21x | 37x | 19.4x | 19.4x |
| Image Quality | | | | |
| Resolution @ 30 mm Working<br>Distance (lp/mm) | 8.5 | 3 | 8 | 8 |
| Distortion | -26% barrel | -23% barrel | -22% barrel | -22% barrel |
{5}------------------------------------------------
The subject and predicate device have the same intended use.
The subject and predicate devices have different technological characteristics, including dimensions, device design, and optical characteristics. These differences do not raise different questions of safety and effectiveness.
### Summary of Performance Testing:
To further support a determination of substantial equivalence, non-clinical bench testing was conducted to support the subject device. The specific types of non-clinical testing conducted are listed below.
- 1. Insertion Force Test
- 2. Flow Performance Test
- Seal Performance Leak Test 3.
- 4. Image Quality Testing
- 5. Use Life Testing of Hysteroscopes
- Shelf Life of Hysteroscope Seal (Functional Performance and Package Integrity) 6.
- 7. Biocompatibility
- 8. Reprocessing Validation of Hysteroscope
- 9. Sterilization Validation of Hysteroscope Seal
- 10. Seal Integrity Testing
- 11. Distribution Testing
- 12. Electrical Safety Testing per IEC 60601-1:2005
- 13. EMC Testing per IEC 60601-2-18:2009
- 14. Usability Testing per IEC 62366-1:2015
#### Conclusion:
The TruClear™ Elite Hysteroscopes were found to be substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.