Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs
K180476 · Comfort Rubber Gloves Industries Sdn. Bhd. · LZC · May 15, 2018 · General Hospital
Device Facts
Record ID
K180476
Device Name
Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs
Applicant
Comfort Rubber Gloves Industries Sdn. Bhd.
Product Code
LZC · General Hospital
Decision Date
May 15, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6250
Device Class
Class 1
Indications for Use
The Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs Labeling Claim (Green) is a specialty medical glove which is a disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between examiner and patient. In addition, these gloves are worn to protect the wearer against exposure to chemotherapy drugs.
Device Story
Disposable, non-sterile, powder-free nitrile examination glove; intended for use by healthcare professionals to prevent cross-contamination and provide barrier protection against chemotherapy drugs. Device functions as a physical barrier. Performance validated via ASTM D6978-05 permeation testing for specific chemotherapy agents. Used in clinical settings; worn by examiners. Provides protection against chemical exposure; breakthrough times vary by drug agent.
Clinical Evidence
No clinical data. Bench testing only. Performance validated via ASTM D6319-10(2015) for physical properties/dimensions, ASTM D6124-06(2011) for powder content, ASTM D5151 for watertightness, and ASTM D6978-05 for chemotherapy drug permeation. Biocompatibility testing performed per ISO 10993-10 (irritation/sensitization) and ISO 10993-5 (cytotoxicity/systemic toxicity).
Technological Characteristics
Material: Nitrile. Form factor: Disposable, powder-free, non-sterile examination glove. Standards: ASTM D6319-10(2015) (dimensions/physical properties), ASTM D6124-06(2011) (powder content), ASTM D6978-05 (chemotherapy permeation), ISO 10993 (biocompatibility). Dimensions: Min 240mm length, min 0.05mm thickness. Color: Green.
Indications for Use
Indicated for use as a disposable medical glove worn on the examiner's hand or finger to prevent cross-contamination between examiner and patient and to provide protection against exposure to specific chemotherapy drugs.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Predicate Devices
Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs Labeling Claim (Blue) (K162405)
Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs Labeling Claim (White) (K162405)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 15, 2018
Comfort Rubber Gloves Industries Sdn. Bhd. Chan Yew Men QA & QMS Manager Lot 821, Jalan Matang 34750 Matang, Perak Malyasia
Re: K180476
Trade/Device Name: Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs Labeling Claim (Green) Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: Class I Product Code: LZC, LZA Dated: February 12, 2018 Received: February 22, 2018
Dear Chan Yew Men:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Geeta K. Pamidimukkala -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
510(k) Number (if known)
### K180476
### Device Name
Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs Labeling Claim (Green)
Indications for Use (Describe)
The Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs Labeling Claim (Green) is a specialty medical glove which is a disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between examiner and patient. In addition, these gloves are worn to protect the wearer against exposure to chemotherapy drugs. Tested for use chemotherapy drugs. Tested drugs are as follows:
| | Average Breakthrough Detection Time (minutes) |
|-------------------------------------|-----------------------------------------------|
| Cisplatin 1.0 mg/ml | ≥ 240 |
| Cyclophosphamide (Cytoxan) 20 mg/ml | ≥ 240 |
| Dacarbazine (DTIC) 10.0 mg/ml | ≥ 240 |
| Doxorubicin Hydrochloride 2.0 mg/ml | ≥ 240 |
| Etoposide (Toposar) 20.0 mg/ml | ≥ 240 |
| Fluorouracil 50.0 mg/ml | ≥ 240 |
| Paclitaxel (Taxol) 6.0 mg/ml | ≥ 240 |
Please note that the following drugs have extremely low permeation time for
*Carmustine (BCNU) 3.3mg.ml - 23.4 (mins)
*Thiotepa 10.0 mg/ml - 16.2 (mins)
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
### ATTACHMENT 2
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
{3}------------------------------------------------
## 510(k) SUMMARY
### POWDER FREE NITRILE EXAMINATION GLOVES TESTED FOR USE WITH CHEMOTHERAPY DRUGS LABELING CLAIM (GREEN)
#### 1.0 Submitter:
| Name | Comfort Rubber Gloves Industries Sdn. Bhd. |
|----------------------|-----------------------------------------------------------------------|
| Address | Lot 821, Jalan Matang,<br>34750 Matang, Perak, Malaysia.<br>Malaysia. |
| Phone No. | 605-847 2777 |
| Fax No. | 605-847 9108 |
| Contact Person | Chan Yew Men (Mr.) |
| Date of Preparation: | April 17, 2018 |
#### 2.0 Name of the Device
Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs Labeling Claim (Green)
| Common Name | : | Patient Examination Gloves |
|---------------------|---|------------------------------------------------------------------------------------------------------------------------------------------------|
| Classification Name | : | Patient Examination Gloves (21 CFR 880.6250 product<br>code LZA)<br>Patient Examination Gloves Specialty (21 CFR 880.6250<br>product code LZC) |
| 510(K) Number | : | K180476 |
#### 3.0 Identification of The Legally Marketed Devices That equivalency is claimed:
Predicate
Device Name : Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs Labeling Claim (Blue) Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs Labeling Claim (White)
- Company : Comfort Rubber Gloves Industries Sdn. Bhd.
: K162405 510(K) No.
#### 4.0 Description of the Device:
The proposed device is the Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs Labeling Claim (Green). The proposed device is for single use and is provided non-sterile. The device meets the requirements for dimensions, physical properties, and thickness in ASTM D6319 - 10(2015) Standard Specification for Nitrile Examination Gloves for Medical Application. The device accomplishes its intended function by creating a
{4}------------------------------------------------
physical barrier between the examiner and patient and creates a barrier for the chemotherapy drugs listed in the Indications for Use.
#### 5.0 Indication for Use of the Device
The Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs Labeling Claim (Green) are a specialty medical glove which is a disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between examiner and patient. In addition, these gloves are worn to protect the wearer against exposure to chemotherapy drugs. Tested for use with chemotherapy drugs. Tested drugs are as follows:
| | Average Breakthrough Detection Time (minutes) |
|-------------------------------------|-----------------------------------------------|
| Cisplatin 1.0 mg/ml | ≥ 240 |
| Cyclophosphamide (Cytoxan) 20 mg/ml | ≥ 240 |
| Dacarbazine (DTIC) 10.0 mg/ml | ≥ 240 |
| Doxorubicin Hydrochloride 2.0 mg/ml | ≥ 240 |
| Etoposide (Toposar) 20.0 mg/ml | ≥ 240 |
| Fluorouracil 50.0 mg/ml | ≥ 240 |
| Paclitaxel (Taxol) 6.0 mg/ml | ≥ 240 |
Please note that the following drugs have extremely low permeation time for *Carmustine (BCNU) 3.3 mg/ml - 23.4 (mins) *Thiotepa 10.0 mg/ml - 16.2 (mins)
#### 6.0 Summary of the Technological Characteristics of the Device:
The Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs Labeling Claim (Green) are summarized with the following technological characteristics compared to ASTM D6319 - 10(2015) Standard Specification for Nitrile Examination Gloves for Medical Application or equivalent standards as shown in Table 1.
The chemotherapy claim is similar to the Predicate, which has a qloves thickness which complies with the ASTM standard ASTM D6319 - 10(2015): Standard Specification for Nitrile Examination Gloves for Medical Application.
| CHARACTERISTIC | STANDARDS | DEVICE PERFORMANCE | | Comparison |
|------------------------------|--------------------------------------------|--------------------------------------------------------------|--------------------------------------------------------------|------------|
| Manufacturer(s) | | Comfort Rubber<br>Gloves Industries<br>Sdn. Bhd | Comfort Rubber<br>Gloves Industries<br>Sdn. Bhd | Same |
| 510(k) number | | K162405 | K180476 | |
| Dimension | ASTM D6319 -<br>10(2015) | Length-Min 240mm<br>Thickness palm and<br>finger- Min 0.05mm | Length-Min 240mm<br>Thickness palm and<br>finger- Min 0.05mm | Same |
| Physical Properties | ASTM D6319 -<br>10(2015) | Meets | Meets | Same |
| Thickness - Finger<br>- Palm | ASTM D6319 -<br>10(2015) | Meets | Meets | Same |
| Powder Content | ASTM D6124 -<br>06(2011)<br>(≤ 2 mg/glove) | Meets | Meets | Same |
### Comparison of technological characteristics with the predicate device
{5}------------------------------------------------
| | CHARACTERISTIC | STANDARDS | DEVICE PERFORMANCE | | |
|--|--------------------------------------|---------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------|
| | | | Predicate | Current | Comparison |
| | | | Powder Free Nitrile<br>Examination Gloves<br>Tested for Use with<br>Chemotherapy<br>Drugs Labeling<br>Claim (Blue) | Powder Free Nitrile<br>Examination Gloves<br>Tested for Use with<br>Chemotherapy<br>Drugs Labeling<br>Claim (Green) | |
| | Chemotherapy Drug<br>Permeation Test | ASTM D6978-05 | | | |
| | Test Chemotherapy Drug | Concentration | Minimum Breakthrough Detection Time<br>(min) | | |
| | *Carmustine (BCNU) | 3.3 mg/ml | 18.2 | 23.4 | |
| | Cisplatin | 1.0 mg/ml | >240 | >240 | |
| | Cyclophosphamide (Cytoxan) | 20 mg/ml | >240 | >240 | |
| | Dacarbazine (DTIC) | 10.0 mg/ml | >240 | >240 | Similar - The |
| | Doxorubicin Hydrochloride | 2.0 mg/ml | >240 | >240 | chemotherapy |
| | Etoposide (Toposar) | 20.0 mg/ml | >240 | >240 | drugs tested |
| | Fluorouracil | 50.0 mg/ml | >240 | >240 | have similar |
| | Paclitaxel (Taxol) | 6.0 mg/ml | >240 | >240 | breakthrough |
| | *Thiotepa | 10.0 mg/ml | 57.3 | 16.2 | detection |
| | Warning Statement | | * WARNING:<br>Please note that the<br>following drugs have<br>extremely low<br>permeation times<br>Carmustine (BCNU):<br>18.2 minutes and<br>Thiotepa: 57.3<br>minutes. | * WARNING:<br>Please note that the<br>following drugs have<br>extremely low<br>permeation times<br>Carmustine (BCNU):<br>23.4 minutes and<br>Thiotepa: 16.2<br>minutes. | times; the<br>drugs with low<br>permeation<br>times are the<br>same. |
| | | DEVICE PERFORMANCE | | |
|--------------------------------------|---------------|---------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------|-------------------------------|
| CHARACTERISTIC | STANDARDS | Predicate | Current | Comparison |
| | | Powder Free Nitrile<br>Examination Gloves<br>Tested for Use with<br>Chemotherapy<br>Drugs Labeling<br>Claim (White) | Powder Free Nitrile<br>Examination Gloves<br>Tested for Use with<br>Chemotherapy<br>Drugs Labeling<br>Claim (Green) | |
| Chemotherapy Drug<br>Permeation Test | ASTM D6978-05 | | | |
| Test Chemotherapy Drug | Concentration | Minimum Breakthrough Detection Time<br>(min) | | |
| *Carmustine (BCNU) | 3.3 mg/ml | 24.0 | 23.4 | Similar - The<br>chemotherapy |
| Cisplatin | 1.0 mg/ml | >240 | >240 | chemotherapy |
| Cyclophosphamide (Cytoxan) | 20 mg/ml | >240 | >240 | drugs tested |
| Dacarbazine (DTIC) | 10.0 mg/ml | >240 | >240 | have similar |
| Doxorubicin Hydrochloride | 2.0 mg/ml | >240 | >240 | breakthrough |
| Etoposide (Toposar) | 20.0 mg/ml | >240 | >240 | detection |
| Fluorouracil | 50.0 mg/ml | >240 | >240 | times; the |
| Paclitaxel (Taxol) | 6.0 mg/ml | >240 | >240 | drugs with low |
| *Thiotepa | 10.0 mg/ml | 54.9 | 16.2 | permeation |
{6}------------------------------------------------
| | Please note that the<br>following drugs have<br>extremely low<br>permeation times<br>Carmustine (BCNU):<br>24.0 minutes and<br>Thiotepa: 54.9<br>minutes. | Please note that the<br>following drugs have<br>extremely low<br>permeation times<br>Carmustine (BCNU):<br>23.4 minutes and<br>Thiotepa: 16.2<br>minutes. | same. |
|--|-----------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|-------|
|--|-----------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|-------|
| CHARACTERISTIC | STANDARDS | DEVICE PERFORMANCE | | Comparison |
|------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| | | Predicate | Current | |
| Biocompatibility:<br>Primary Skin Irritation | ISO 10993-<br>10:2010<br>Biological<br>evaluation of<br>medical devices --<br>Part 10: Tests for<br>irritation and skin<br>sensitization | Passes<br>Under the conditions<br>of the study, the<br>subject device is non-<br>irritating | Passes<br>Under the conditions<br>of the study, the<br>subject device is non-<br>irritating | Same |
| Biocompatibility:<br>Dermal Sensitization | ISO 10993-<br>10:2010<br>Biological<br>evaluation of<br>medical devices --<br>Part 10: Tests for<br>irritation and skin<br>sensitization | Passes<br>Under the conditions<br>of the study, the<br>subject device is non-<br>sensitization | Passes<br>Under the conditions<br>of the study, the<br>subject device is non-<br>sensitization | Same |
| Biocompatibility:<br>Cytotoxicity and acute<br>systemic toxicity | ISO 10993-5:2009<br>Biological<br>evaluation of<br>medical devices --<br>Part 5: Tests for <i>in vitro</i> cytotoxicity<br>ISO 10993-5:2017<br>Biological<br>evaluation of<br>medical devices --<br>Part 5: Tests for<br>systemic toxicity | N/A | Passes<br>Certificate of Analysis<br>for the color imparting<br>compound indicates<br>that the compound<br>meets technical<br>specifications, is non-<br>cytotoxic, and is not<br>systemically toxic. | Different |
| Watertight (1000ml) | 21 CFR 800.20<br>ASTM D5151 | Passes | Passes | Same |
| Indication for Use | | The Powder Free<br>Nitrile Examination<br>Gloves Tested for<br>Use with<br>Chemotherapy Drugs<br>Labeling Claim (Blue)<br>is a specialty medical<br>glove which is a<br>disposable device<br>intended for medical<br>purpose that is worn<br>on the examiner's<br>hand or finger to<br>prevent | The Powder Free<br>Nitrile Examination<br>Gloves Tested for<br>Use with<br>Chemotherapy Drugs<br>Labeling Claim<br>(Green) is a specialty<br>medical glove which<br>is a disposable<br>device intended for<br>medical purpose that<br>is worn on the<br>examiner's hand or<br>finger | Similar |
{7}------------------------------------------------
| contamination<br>between examiner<br>and patient. In<br>addition these gloves<br>are worn to protect<br>the wearer against<br>exposure to<br>chemotherapy drugs.<br>Tested<br>chemotherapy drugs<br>are as follows: | contamination<br>between examiner<br>and patient. In<br>addition these gloves<br>are worn to protect<br>the wearer against<br>exposure to<br>chemotherapy drugs.<br>Tested chemotherapy<br>drugs are as follows: | | examiner's hand or<br>finger to prevent<br>contamination<br>between examiner<br>and patient. In<br>addition these gloves<br>are worn to protect<br>the wearer against<br>exposure to<br>chemotherapy drugs.<br>Tested<br>chemotherapy drugs<br>are as follows: | |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Average<br>Breakthrough<br>Detection Time<br>(minutes) | Average<br>Breakthrough<br>Detection Time<br>(minutes) | | Average<br>Breakthrough<br>Detection Time<br>(minutes) | |
| Cisplatin, 1.0 mg/ml -<br>≥ 240<br>Cyclophosphamide<br>(Cytoxan), 20.0<br>mg/ml - ≥ 240<br>Dacarbazine (DTIC),<br>10.0 mg/ml - ≥ 240<br>Doxorubicin<br>Hydrochloride, 2.0<br>mg/ml - ≥ 240<br>Etoposide (Toposar),<br>20.0 mg/ml - ≥ 240<br>Fluorouracil, 50.0<br>mg/ml - ≥ 240<br>Paclitaxel (Taxol), 6.0<br>mg/ml<br>- ≥ 240 | Cisplatin, 1.0 mg/ml -<br>≥ 240<br>Cyclophosphamide<br>(Cytoxan), 20.0<br>mg/ml - ≥ 240<br>Dacarbazine (DTIC),<br>10.0 mg/ml - ≥ 240<br>Doxorubicin<br>Hydrochloride, 2.0<br>mg/ml - ≥ 240<br>Etoposide (Toposar),<br>20.0 mg/ml - ≥ 240<br>Fluorouracil, 50.0<br>mg/ml - ≥ 240<br>Paclitaxel (Taxol), 6.0<br>mg/ml<br>- ≥ 240 | | Cisplatin, 1.0 mg/ml -<br>$\ge$ 240<br>Cyclophosphamide<br>(Cytoxan), 20.0<br>mg/ml - $\ge$ 240<br>Dacarbazine (DTIC),<br>10.0 mg/ml - $\ge$ 240<br>Doxorubicin<br>Hydrochloride, 2.0<br>mg/ml - $\ge$ 240<br>Etoposide (Toposar),<br>20.0 mg/ml - $\ge$ 240<br>Fluorouracil, 50.0<br>mg/ml - $\ge$ 240<br>Paclitaxel (Taxol), 6.0<br>mg/ml<br>- $\ge$ 240 | |
| Please note that the<br>following drugs have<br>extremely low<br>permeation time for: | Please note that the<br>following drugs have<br>extremely low<br>permeation time for: | | Please note that the<br>following drugs have<br>extremely low<br>permeation time for: | |
| Carmustine (BCNU)<br>3.3mg/ml - 18.2<br>(mins)<br>Thiotepa 10.0 mg/ml<br>- 57.3 (mins) | Carmustine (BCNU)<br>3.3mg/ml - 23.4<br>(mins)<br>Thiotepa 10.0 mg/ml<br>- 16.2 (mins) | | Carmustine (BCNU)<br>3.3mg/ml - 24.0<br>(mins)<br>Thiotepa 10.0 mg/ml<br>- 54.9 (mins) | |
| The Powder Free<br>Nitrile Examination<br>Gloves Tested for<br>Use with<br>Chemotherapy Drugs<br>Labeling Claim<br>(White) is a specialty<br>medical glove which<br>is a disposable<br>device intended for<br>medical purpose that<br>is worn on the | | | | |
{8}------------------------------------------------
| CHARACTERISTIC | STANDARDS | DEVICE PERFORMANCE | | |
|----------------|------------------------------------------------|-----------------------------------------|-----------------------------------------|------------|
| | | Predicate | Current | Comparison |
| Material | ASTM D6319 -<br>10(2015) | Nitrile | Nitrile | Same |
| Color | - | Blue or White | Green | Different |
| Size | Medical Glove<br>Guidance Manual<br>- Labeling | Extra Small<br>Small<br>Medium<br>Large | Extra Small<br>Small<br>Medium<br>Large | Same |
{9}------------------------------------------------
| | | Extra Large | Extra Large | |
|------------|------------------------------------------------|-------------|-------------|------|
| Single Use | Medical Glove<br>Guidance Manual<br>– Labeling | Single Use | Single Use | Same |
#### 7.0 Summary of Non-Clinical Performance Data
The device and the predicate share the same indication for use, same base polymeric material, same specifications for thickness and length, similar permeation rate for chemotherapy drugs, similar labeling properties, biocompatibility per ISO 10993-1, water tight test results, and are powder free.
#### 8.0 Summary of Clinical Performance Data
Clinical data is not needed for the intended use for this device.
#### 9.0 Conclusion
Based on the Indication for Use, technological characteristics, and non-clinical performance data, the Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs Labeling Claim (Green) (K180476) is as safe, as effective, and performs as well as the legally marketed predicate device (K162405).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.