K180466 · Inari Medical, Inc. · QEW · May 17, 2018 · Cardiovascular
Device Facts
Record ID
K180466
Device Name
FlowTriever Retrieval/Aspiration System
Applicant
Inari Medical, Inc.
Product Code
QEW · Cardiovascular
Decision Date
May 17, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.5150
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The FlowTriever Retrieval/Aspiration System consists of the FlowTriever Catheter, Aspiration Guide Catheter, and Retraction Aspirator. The FlowTriever Retrieval/Aspiration System is indicated for: · The non-surgical removal of emboli and thrombi from blood vessels. - · Injection, infusion, and/or aspiration of contrast media and other fluids into or from a blood vessel. The Flow Triever Retrieval/Aspiration System is intended for use in the peripheral vasculature and for the treatment of pulmonary embolism.
Device Story
System for minimally invasive pulmonary embolism treatment; comprises Aspiration Guide Catheter, FlowTriever Catheter (6-18mm sizes), and hand-lever operated Retraction Aspirator. Physician inserts FlowTriever Catheter through Aspiration Guide Catheter to thrombus site; deploys self-expanding wireform disks to engage clot. Retraction Aspirator simultaneously aspirates fluids and retracts FlowTriever Catheter with captured thrombus into Aspiration Guide Catheter. Restores blood flow; reduces RV/LV ratio. Used in clinical settings; physician-operated under fluoroscopic guidance. Benefits include mechanical clot removal without surgery.
Clinical Evidence
FLARE study: prospective, single-arm, multicenter, 106 subjects. Primary safety endpoint: composite of major adverse events (device-related death, major bleeding, treatment-related adverse events) at 48 hours. Primary effectiveness endpoint: reduction in RV/LV ratio from baseline to 48 hours. Study met performance goals for primary endpoints. Acute safety/performance also evaluated in swine pulmonary artery model with histological assessment showing no significant findings.
Technological Characteristics
Single-use, over-the-wire catheter system. Components: Aspiration Guide Catheter, FlowTriever Catheter with self-expanding wireform disks, Retraction Aspirator. Materials biocompatibility per ISO 10993-1. Mechanical operation via hand-lever aspiration and manual retraction. Fluoroscopically visible. Sterilized.
Indications for Use
Indicated for non-surgical removal of emboli and thrombi from peripheral vasculature and treatment of pulmonary embolism in patients requiring fluid injection, infusion, or aspiration.
Regulatory Classification
Identification
An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is a seal with an abstract design, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The text is arranged in three lines, with "FDA" on the first line, "U.S. FOOD & DRUG" on the second line, and "ADMINISTRATION" on the third line. The logo is clean and professional, reflecting the FDA's role in regulating food and drugs in the United States.
May 17, 2018
Inari Medical, Inc. Eben Gordon Vice President, Regulatory Affairs & Quality Assurance 9272 Jeronimo Road Suite 124 Irvine, California 92618
Re: K180466
Trade/Device Name: FlowTriever Retrieval/Aspiration System Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy Catheter Regulatory Class: Class II Product Code: DXE Dated: March 7, 2018 Received: March 8, 2018
Dear Eben Gordon:
We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good
{1}------------------------------------------------
manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
## Nicole G. Ibrahim -S
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K180466
Device Name
FlowTriever Retrieval/Aspiration System
Indications for Use (Describe)
The Flow Triever Retrieval/Aspiration System consists of the FlowTriever Catheter, Aspiration Guide Catheter, and Retraction Aspirator. The FlowTriever Retrieval/Aspiration System is indicated for:
· The non-surgical removal of emboli and thrombi from blood vessels.
- · Injection, infusion, and/or aspiration of contrast media and other fluids into or from a blood vessel.
The Flow Triever Retrieval/Aspiration System is intended for use in the peripheral vasculature and for the treatment of pulmonary embolism.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(K) SUMMARY
# K180466
| Date prepared | February 16, 2018 |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name | Inari Medical, Inc.<br>9272 Jeronimo Road, Suite 124<br>Irvine, CA 92618<br>949.600.8433 x114 |
| Contact person | Eben Gordon<br>Vice President, Regulatory Affairs & Quality Assurance |
| Trade name | FlowTriever Retrieval/Aspiration System |
| Common name | Embolectomy catheter |
| Regulation Name | Embolectomy catheter |
| Classification number | 21 CFR 870.5150 |
| Product code | DXE |
| Regulatory class | II |
| Predicate device | Infusion Aspiration Catheter System (K143563) |
| Reference device | EkoSonic Endovascular System (K140151) |
| Description | The FlowTriever Retrieval/Aspiration System is a single-use over-the-wire<br>catheter-based system for the minimally invasive treatment of pulmonary<br>embolism. The system is comprised of three main components packaged<br>separately:<br>● Aspiration Guide Catheter<br>● FlowTriever Catheter (available in 3 sizes: 6-10 mm, 11-14 mm, and 15-<br>18 mm)<br>● Retraction Aspirator<br>The FlowTriever Catheter is inserted through the Aspiration Guide Catheter and<br>advanced to the thrombus. Self-expanding wireform disks are deployed to engage<br>thrombus by retracting the outer delivery catheter. The hand-lever operated<br>Retraction Aspirator simultaneously aspirates fluids and retracts the FlowTriever<br>Catheter with thrombus into the Aspiration Guide Catheter to capture clot and<br>restore blood flow. |
| Indications for Use | The FlowTriever Retrieval/Aspiration System consists of the FlowTriever<br>Catheter, Aspiration Guide Catheter, and Retraction Aspirator. The FlowTriever<br>Retrieval/Aspiration System is indicated for:<br>The non-surgical removal of emboli and thrombi from blood vessels. Injection, infusion, and/or aspiration of contrast media and other fluids into or from a blood vessel. The FlowTriever Retrieval/Aspiration System is intended for use in the peripheral<br>vasculature and for the treatment of pulmonary embolism. |
| Summary of<br>substantial equivalence | The purpose of this submission is to expand the indications for use of the<br>FlowTriever/Aspiration System for the treatment of pulmonary embolism and<br>propose device design changes.<br><br>The FlowTriever Retrieval/Aspiration System and the predicate device have the<br>same intended use: removal of obstructing material (including emboli and<br>thrombi) from blood vessels by mechanical means.<br><br>The device design changes do not significantly change the underlying<br>technological characteristics and do not raise different questions of safety and<br>effectiveness compared to the predicate. |
| <b>Non-Clinical Testing</b><br>In alignment with the Design Failure Modes and Effects Analysis, verification and<br>validation testing was identified in support of the FlowTriever<br>Retrieval/Aspiration System. These tests included:<br>Pouch seal visual inspection and dye penetration Pouch peel, seal strength Visual and dimensional inspections Guidewire compatibility Snap fit, Dilator Luer to Guide Catheter hemostasis valve Deployment force of Wireform Catheter from Delivery Catheter Retraction force of Wireform Catheter into Delivery Catheter Retraction force of Wireform Catheter into Guide Catheter Leakage testing of Wireform Catheter with guidewire in place Guide Catheter/Infusion Wireform Catheter kink radius Liquid leakage under pressure Leakage testing, devices with Guide Catheter hemostasis valve Test of conical fittings with 6% Luer taper Air leakage into hub assembly during aspiration Retraction Aspirator retraction testing Tube Set Check Valve and vent plug inspection Tube set, leakage Tube Set check valve cracking pressure Determination of flow rate through catheter | |
{4}------------------------------------------------
- Corrosion resistance
- . Tube set tensile strength
{5}------------------------------------------------
- Particulate matter evaluation
- Burst pressure and burst testing ●
- Simulated Use Tracking and tensile, Delivery Catheter, Guide Catheter, Dilator, Wireform Catheter
- . Simulated Use tracking and torque
Biocompatibility testing in accordance with ISO 10993-1:
- MEM elution ●
- ISO Guinea Pig Maximization Sensitization
- ISO Intracutaneous Irritation
- Acute systemic toxicity
- Hemolysis, direct contact and extract method ●
- Complement Activation, C3a and SC5b-9 Assay ●
- Unactivated Partial Thromboplastin Time (UPTT) Test
Acute evaluation of the safety and performance of the FlowTriever Retrieval/Aspiration System was successfully performed in swine pulmonary arteries. Specifically the ability to place and deploy the device, the fluoroscopic visibility and the ability to infuse fluid through the catheters was evaluated, and all acceptance criteria were successfully met. In addition, the histological evaluation of the vessel specimens after use of the test devices was performed and demonstrated no significant gross or histopathological findings in treated vessels.
### Clinical Testing
Inari Medical conducted the FlowTriever Pulmonary Embolectomy Clinical ("FLARE") Study, a prospective, single-arm, multicenter study, to evaluate the safety and effectiveness of the FlowTriever System for the treatment of acute pulmonary embolism (PE). A total of 106 subjects were enrolled at 18 investigational sites. The primary safety endpoint was the composite of major adverse events (MAE), defined as device-related death, major bleeding and treatment-related adverse events (clinical deterioration, pulmonary vascular injury, or cardiac injury) measured at 48 hours. The primary effectiveness endpoint was reduction in RV/LV ratio from baseline to 48 hours. The study met the performance goals for the primary study endpoints. Refer to the Instruction for Use for detailed clinical study information.
### Conclusion
Compared to the predicate device, the expanded indication and device modifications to the FlowTriever Retrieval/Aspiration System do not change the intended use and the technological differences do not raise different questions of safety and effectiveness. Based on the results of pre-clinical testing and the FLARE Clinical Study, the data supports that the FlowTriever Retrieval/Aspiration System is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.