The Ultrasonic Doppler uses continuous-wave Doppler to detect fetal heart rate and play the fetal heart sound from early gestation thru delivery for singleton pregnancies. The Ultrasonic Doppler is intended for use by trained healthcare professionals only in a clinical setting.
Device Story
Hand-held device for non-invasive fetal heart rate monitoring; utilizes continuous-wave Doppler ultrasound. System comprises main unit (display, speaker, battery) and wired probe (transducer with two angled semi-circle crystals). Probe transmits 2 MHz ultrasonic signal; receives reflected continuous wave signal; detects Doppler shift. Main unit processes signal to display fetal heart rate on OLED screen and output audio via speaker. Used by trained healthcare professionals in clinical settings. Provides real-time fetal heart rate data to assist clinicians in monitoring fetal well-being.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including biocompatibility (ISO 10993-1, -5, -10), electrical safety (IEC 60601-1, -1-2, -1-11), and acoustic output measurement (IEC 60601-2-37, NEMA UD2).
Indicated for detection of fetal heart rate and fetal heart sounds in singleton pregnancies from early gestation (12 weeks) through delivery. Intended for use by trained healthcare professionals in clinical settings.
Regulatory Classification
Identification
A fetal ultrasonic monitor is a device designed to transmit and receive ultrasonic energy into and from the pregnant woman, usually by means of continuous wave (doppler) echoscopy. The device is used to represent some physiological condition or characteristic in a measured value over a period of time (e.g., perinatal monitoring during labor) or in an immediately perceptible form (e.g., use of the ultrasonic stethoscope). This generic type of device may include the following accessories: signal analysis and display equipment, electronic interfaces for other equipment, patient and equipment supports, and component parts. This generic type of device does not include devices used to image some relatively unchanging physiological structure or interpret a physiological condition, but does include devices which may be set to alarm automatically at a predetermined threshold value.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services (HHS) and the Food and Drug Administration (FDA). The HHS logo is on the left, and the FDA logo is on the right. The FDA logo includes the FDA acronym in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.
July 19, 2018
Shenzhen IMDK Medical Technology Co., Ltd. Yuan Xia Administrative Director C Zone, 10F, Building16, Yuanshan Industrial B Area Gongming Street, Guangming District Shenzhen, 518106 CHINA
Re: K180419
> Trade/Device Name: Ultrasonic Doppler (Model YM-2T8) Regulation Number: 21 CFR 884.2660 Regulation Name: Fetal ultrasonic monitor and accessories Regulatory Class: Class II Product Code: KNG Dated: June 8, 2018 Received: June 21, 2018
Dear Yuan Xia:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Sharon M. Andrews -S
for
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K180419
Device Name Ultrasonic Doppler Model YM-2T8
### Indications for Use (Describe)
The Ultrasonic Doppler uses continuous-wave Doppler to detect fetal heart rate and play the fetal heart sound from early gestation thru delivery for singleton pregnancies. The Ultrasonic Doppler is intended for use by trained healthcare professionals only in a clinical setting.
| Type of Use (Select one or both, as applicable) | | |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|
| <table border="0"><tr><td><span style="font-size:16px;"><b>☑</b></span> Prescription Use (Part 21 CFR 801 Subpart D)</td></tr><tr><td><span style="font-size:16px;">□</span> Over-The-Counter Use (21 CFR 801 Subpart C)</td></tr></table> | <span style="font-size:16px;"><b>☑</b></span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size:16px;">□</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
| <span style="font-size:16px;"><b>☑</b></span> Prescription Use (Part 21 CFR 801 Subpart D) | | |
| <span style="font-size:16px;">□</span> Over-The-Counter Use (21 CFR 801 Subpart C) | | |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary K180419
#### 1. Submitter
| Name and Address: | Shenzhen IMDK Medical Technology Co., Ltd. | |
|-------------------|------------------------------------------------------------------------|--|
| | C Zone, 10F, Building 16, Yuanshan Industrial B Area, Gongming Street, | |
| | Guangming District, 518106, P.R. China | |
| | Telephone: +86 13662694320 | |
| Contact Person: | Yuan Xia | |
| Date Prepared: | July 5, 2018 | |
#### 2. Device Information
| Trade Name: | Ultrasonic Doppler |
|------------------------|------------------------------------------|
| Model: | YM-2T8 |
| Classification Name: | Fetal ultrasonic monitor and accessories |
| Regulation Number: | 21 CFR 884.2660 |
| Device Classification: | Class II |
| Product Code: | KNG |
| Review Panel: | Obstetrics/Gynecology |
#### 3. Predicate Device
| 510(k) Number: | K153475 |
|----------------|---------------------------------|
| Device Name: | SD5 Ultrasonic Tabletop Doppler |
| Manufacturer: | Edan Instruments, Inc. |
The predicate device has not been subject to a design-related recall.
#### 4. Device Description
The Ultrasonic Doppler is a hand-held device for non-invasive measurement and display of the fetal heart beat utilizing the principle of Doppler shift of an ultrasound. The device consists of two main components, the main unit and probe. The main unit consists of the main board, power module, battery, speaker, and organic light-emitting diode screen. The wired probe consists of the transducer and includes two angled semi-circle ultrasonic crystals, one for transmission and one for reception. The ultrasonic signal is continuously transmitted at a frequency of 2 MHz. The reflected continuous wave signal is received by one of the crystals and then any detected Doppler shift is presented to the user.
The main unit is powered by two 1.5V AA alkaline batteries.
#### ട്. Indications for Use
The Ultrasonic Doppler uses continuous-wave Doppler to detect fetal heart beats, display fetal heart rate and play the fetal heart sound from early gestation thru delivery for singleton pregnancies. The Ultrasonic Doppler is intended for use by trained healthcare professionals only in a clinical setting.
{4}------------------------------------------------
#### Predicate Comparison 6.
The following table compares the Ultrasonic Doppler to the predicate device with respect to the indications for use and technological characteristics:
| Device | Subject Device<br>Ultrasonic Doppler<br>Model: YM-2T8 | Predicate Device<br>SD5 Ultrasonic Tabletop Doppler |
|---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K180419 | K153475 |
| Manufacturer | Shenzhen IMDK Medical Technology Co., Ltd. | EDAN Instruments, Inc. |
| Classification Name | Fetal ultrasonic monitor and accessories | Fetal ultrasonic monitor and accessories |
| Regulation Number | 844.2660 | 844.2660 |
| Device Class | Class II | Class II |
| Product Code | KNG | KNG |
| Indications for Use | The Ultrasonic Doppler uses<br>continuous-wave Doppler to detect<br>fetal heart beats, display fetal heart<br>rate and play the fetal heart sound<br>from early gestation thru delivery for<br>singleton pregnancies. The Ultrasonic<br>Doppler is intended for use by trained<br>healthcare professionals only in a<br>clinical setting. | The SD5 Ultrasonic TableTop Doppler<br>(hereinafter called "SD5") and SD6<br>Ultrasonic TableTop Doppler (hereinafter<br>called "SD6") are intended to be used by<br>health care professionals including<br>registered nurses, practical nurses,<br>midwives, ultrasound technicians, and<br>physician assistants, by prescription from<br>licensed physicians in hospitals, clinics and<br>private offices.<br><br>The 2 MHz and/or 3 MHz obstetrical probes<br>are indicated for the detection of fetal heart<br>rate from early gestation thru delivery and<br>as a general indication of fetal well being.<br>They can also be used to verify fetal heart<br>viability. |
| Gestational Age | 12 weeks | 12 weeks |
| Gestational Type | Singleton | Singleton |
| Technical Characteristics | | |
| Display Type | Digital Display; LCD | Digital Display; LCD |
| Power Suppler | Two 1.5V AA Alkaline Batteries | 100 V - 240 V~, 50 Hz/60 Hz or<br>7.2 V/2000 mAh (Ni-MH Battery) |
| Probe connection | Wired | Wired |
| Acoustic Output<br>Power | 3 W | 3 W |
| Nominal Frequency | 2 MHz | 2 MHz |
| Working Frequency | (2.0 ± 10%) MHz | (2.0 ± 10%) MHz |
| <i>Iob</i> | < 20 mW/cm² | < 20 mW/cm² |
| <i>pr</i> | 1 MPa | 1 MPa |
| <i>Ispta</i> | < 100 mW/cm² | < 100 mW/cm² |
| <i>Isata</i> | < 20 mW/cm² | < 20 mW/cm² |
| Wo | 50 mW | 50 mW |
| Mode of operation | Continuous Wave Doppler | Continuous Wave Doppler |
| Effective Radiating Area | (157 $\pm$ 30%) mm² | (245 $\pm$ 15%) mm² |
| FHR Measuring Range | 50 bpm ~ 210 bpm | 50 bpm ~ 240 bpm |
| Accuracy | $\pm$ 2bpm | $\pm$ 2bpm |
| Resolution | 1 bpm | 1 bpm |
{5}------------------------------------------------
#### 7. Non-Clinical Performance Testing
The following performance data were provided in support of the substantial equivalence determination:
# Biocompatibility
The biocompatibility evaluation was conducted in accordance with ISO 10993-1, Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing within a Risk Management Process. Patient contacting materials were subjected to testing that included the following tests:
- i. Cytotoxicity (ISO 10993-5)
- ii. Skin Sensitization (ISO 10993-10)
- iii. Irritation (ISO 10993-10)
# Electrical safety and electromagnetic compatibility (EMC)
Electrical safety and EMC testing were conducted and the device was found to comply with the requirements of the following standards:
- i. IEC 60601-1. Medical Electrical Equipment - Part 1: General requirements for basic safety and essential performance with US deviations per AAMI/ANSI ES 60601-1
- ii. IEC 60601-1-2, Medical Electrical Equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - requirements and tests
- iii. IEC 60601-1-11, General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
# Ultrasound and Acoustic Testing
Bench testing was conducted on the Ultrasonic Doppler and the system was found to comply with the following:
- i. IEC 60601-2-37 Medical electrical equipment - Part 2-37: Particular requirements for the basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment
- Acoustic Output testing per NEMA UD2, Acoustic Output Measurement Standard for Diagnostic ii. Ultrasound Equipment
{6}------------------------------------------------
The acoustic output measurement methodology as recommended in FDA guidance document "Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers," dated September 9, 2008 was followed for Track 1 devices.
# Software Verification and Validation Testing
Software verification and validation testing was conducted and completed with no outstanding anomalies. Software documentation was provided as recommended by FDA guidance document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices," dated May 11, 2005 for a moderate software level of concern.
#### 8. Clinical Performance Testing
Not Applicable
#### 9. Conclusion
Based on the comparison and analysis above, the Ultrasonic Doppler is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.