K180396 · Miramar Labs, Inc. · NEY · Mar 15, 2018 · General, Plastic Surgery
Device Facts
Record ID
K180396
Device Name
miraDry System
Applicant
Miramar Labs, Inc.
Product Code
NEY · General, Plastic Surgery
Decision Date
Mar 15, 2018
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The miraDry System MD4000 is indicated for use in the treatment of primary axillary hyperhidrosis plus unwanted underarm hair removal, and permanent reduction of underarm hair of all colors for Fitzpatrick skin types I - IV. Permanent hair reduction is defined as long-term, stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime. When used with the treatment of primary axillary hyperhidrosis, the miraDry System may reduce underarm odor.
Device Story
miraDry MD4000 is a microwave-based system for treating axillary hyperhidrosis and removing underarm hair. System comprises a console, handpiece, and disposable bioTip. Console contains microwave generator, vacuum, and cooling systems with touch-screen interface. Handpiece delivers microwave energy to dermal-hypodermal interface; heats tissue to target sweat/odor glands and hair bulbs. Surface cooling protects skin during treatment. Used in clinical settings by trained operators. Output is localized heat absorption. Benefits include reduction of sweat, odor, and hair. Modification involves supplying bioTips as clean/non-sterile instead of sterile; single-patient use remains mandatory.
Clinical Evidence
No clinical data presented. Substantial equivalence supported by bench testing, risk assessment, manufacturing process evaluation (bioburden testing), and user interface/clinician feedback.
Technological Characteristics
Microwave energy source; frequency/power delivery via handpiece. Materials: Polyurethane, Thermoplastic Elastomer (TPE), Polycarbonate. System includes console, handpiece, and disposable bioTip. Cooling system integrated. Software-driven console with touch-screen interface. Single-patient use disposable bioTip. Clean, non-sterile supply.
Indications for Use
Indicated for treatment of primary axillary hyperhidrosis, unwanted underarm hair removal, and permanent reduction of underarm hair of all colors for Fitzpatrick skin types I-IV.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font size below that.
March 15, 2018
Miramar Labs, Inc. Cynthia Kada Sr. Director, QA 2790 Walsh Avenue Santa Clara, California 95051
Re: K180396
Trade/Device Name: miraDry System Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories Regulatory Class: Class II Product Code: NEY, OUB, MWY Dated: February 12, 2018 Received: February 13, 2018
Dear Cynthia Kada:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
## Page 2 - Cynthia Kada
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Jennifer R. Stevenson -S3
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K180396
Device Name miraDry System MD4000
#### Indications for Use (Describe)
The miraDry System MD4000 is indicated for use in the treatment of primary axillary hyperhidrosis plus unwanted underarm hair removal, and permanent reduction of underarm hair of all colors for Fitzpatrick skin types I - IV. Permanent hair reduction is defined as long-term, stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime.
When used with the treatment of primary axillary hyperhidrosis, the miraDry System may reduce underarm odor.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------|---------------------------------------------|
| × Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 2. 510(k) SUMMARY
| Classification: | Class II (special controls) |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------|
| Classification No.: | 21 CFR 878.4400 |
| Classification Name: | Electrosurgical cutting and coagulation device and<br>accessories. |
| Product Code(s): | OUB, NEY, MWY |
| Common Name: | Instrument for Treatment of Hyperhidrosis<br>System, Ablation, Microwave And Accessories<br>System, Microwave, Hair Removal |
| Trade Name | miraDry MD4000 System |
| Predicate Device | miraDry MD 4000 System K160141 |
| Submitter: | Miramar Labs, Inc.<br>2790 Walsh Avenue<br>Santa Clara, CA 95051, USA<br>Tel: 408-940-8700<br>Fax: 408-940-8795 |
| FDA Registration No. | 3008082710 |
| Contact: | Cynthia Kada |
| Date: | February 12, 2018 |
{4}------------------------------------------------
## PURPOSE OF SUBMISSION
The purpose of this Special 510(k) submission is to modify how the treatment bioTip component of the miraDry System is provided to the User. The intended use of the modified device, as described in the labeling, has not changed as a result of this modification.
## DEVICE DESCRIPTION
The miraDry MD4000 System is a microwave device designed to heat tissue located at the dermal- hypodermal interface where the axillary sweat and odor glands and hair bulbs reside using a surface contact applicator. The miraDry MD4000 System consists of:
- MD4000-MC Console: O
- MD4000-HP miraDry Handpiece; and O
- MD4000-BT miraDry bioTip that snaps onto the Handpiece O
The miraDry MD4000 System also includes the following Class I components/accessories:
- MD4000-TS template system, a required component for the miraDry treatment O
- MD4000-PK priming kit; and O
- MD4000-BT-DE demonstration bioTip. O
The MD4000-PK priming kit and the non-sterile "demo" bioTip are required when the system is initially set up at a user facility. Optional accessories include an armrest and disposable ice packs.
The MD4000-MC Console is a software-driven device, which contains circuit boards, a microwave generator, integrated vacuum and cooling systems, and an integrated touch-screen user interface.
The non-invasive miraDry Handpiece is specifically designed to deliver microwave energy to the skin at specified frequency and power levels. The proximal end of the Handpiece has a cable bundle and a console connector that supplies the energy and cooling to the Handpiece. The distal end of the Handpiece has a single patient use disposable, the miraDry bioTip, which contacts the underarm skin of the patient.
### DEVICE MODIFICATION
As described in K160141, the bioTip is disposable and for single patient use, and is provided sterile. The only change to the miraDry MD4000 system cleared under K160141 that is the subject of this Special 510(k) involves supplying the disposable bioTips to the user as clean and non-sterile for single patient use only. The intended use of the modified device, as described in the labeling, has not changed as a result of this modification.
{5}------------------------------------------------
### INDICATIONS FOR USE
The miraDry System MD4000 is indicated for use in the treatment of primary axillary hyperhidrosis plus unwanted underarm hair removal, and permanent reduction of underarm hair of all colors for Fitzpatrick skin types I - IV.
Permanent hair reduction is defined as long-term, stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime.
When used with the treatment of primary axillary hyperhidrosis, the miraDry System may reduce underarm odor.
### SUMMARY OF SUBSTANTIAL EQUIVALENCE
The miraDry MD4000 System described and cleared in 510(k) number K160141 serves as the predicate device for this premarket notification. The miraDry MD4000 System that is the subject of this Special 510(k) has the same technological characteristics and intended use as the device described and cleared in 510(k) number K160141. Furthermore, there have been no changes in design, component materials, chemical composition, or energy source, since FDA's clearance of K160141.
Therefore, the miraDry System MD4000 remains substantially equivalent to the predicate device currently marketed under the Federal Food, Drug and Cosmetic Act except for one characteristic: the condition of use of the bioTips. This difference is rendered inconsequential by the "single patient use" and disposable nature of the bioTips that eliminates the risk of cross contamination between patients. Summarized in the substantial equivalence tables below are the key technological characteristics and indications for use of the miraDry System compared to the predicate device identified in this Special 510(k).
| Characteristics<br>miraDry<br>System | Predicate Device<br>miraDry MD4000<br>K160141 | Modified Device<br>miraDry MD4000 | Comparison |
|--------------------------------------|-----------------------------------------------|-----------------------------------|------------|
| Device Class | II | II | Same |
| Energy Type | Microwave | Microwave | Same |
| Mode of Action | Generation of localized<br>heat | Generation of localized<br>heat | Same |
| Product Code | NEY, OUB, MWY | NEY, OUB, MWY | Same |
| | miraDry System Substantial Equivalence Comparison Table | | |
|--|---------------------------------------------------------|--|--|
| | | | |
{6}------------------------------------------------
| Characteristics<br>miraDry<br>System | Predicate Device<br>miraDry MD4000<br>K160141 | Modified Device<br>miraDry MD4000 | Comparison |
|--------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------|
| Indications for<br>Use | The miraDry System<br>MD4000 is indicated for<br>use in the treatment of<br>primary axillary<br>hyperhidrosis plus<br>unwanted underarm hair<br>removal, and permanent<br>reduction of underarm<br>hair of all colors for<br>Fitzpatrick skin types I -<br>IV. | The miraDry System<br>MD4000 is indicated for<br>use in the treatment of<br>primary axillary<br>hyperhidrosis plus<br>unwanted underarm hair<br>removal, and permanent<br>reduction of underarm<br>hair of all colors for<br>Fitzpatrick skin types I -<br>IV. | Same |
| | Permanent hair reduction<br>is defined as long-term,<br>stable reduction in the<br>number of hairs regrowing<br>when measured at 6, 9 and<br>12 months after the<br>completion of a treatment<br>regime. | Permanent hair reduction<br>is defined as long-term,<br>stable reduction in the<br>number of hairs regrowing<br>when measured at 6, 9 and<br>12 months after the<br>completion of a treatment<br>regime. | Same |
| | When used with the<br>treatment of primary<br>axillary hyperhidrosis, the<br>miraDry System may<br>reduce underarm odor. | When used with the<br>treatment of primary<br>axillary hyperhidrosis, the<br>miraDry System may<br>reduce underarm odor. | Same |
| Function | Heat absorption by tissue<br>located at the dermal-<br>hypodermal interface<br>where the axillary sweat<br>(wetness and odor) glands<br>and hair bulbs reside | Heat absorption by tissue<br>located at the dermal-<br>hypodermal interface<br>where the axillary sweat<br>(wetness and odor) glands<br>and hair bulbs reside | Same |
| Overall System<br>structure | Microwave<br>source/amplifier, coolant<br>supply system, operator<br>interface, and microwave,<br>electrical and coolant<br>lines that connect to the<br>applicator | Microwave<br>source/amplifier, coolant<br>supply system, operator<br>interface, and microwave,<br>electrical and coolant<br>lines that connect to the<br>applicator | Same |
| Key<br>Components | Console, Handpiece,<br>disposable bioTip | Console, Handpiece,<br>disposable bioTip | Same |
| Console Control<br>Mechanism | Electronic user interface | Electronic user interface | Same |
| Coolant usage | Delivers cooling to the<br>skin surface | Delivers cooling to the<br>skin surface | Same |
| Characteristics<br>of bioTip | Predicate Device<br>miraDry MD4000<br>K160141 | Modified Device<br>miraDry MD4000 | Comparison |
| Materials | Polyurethane,<br>Thermoplastic<br>Elastomer (TPE),<br>Polycarbonate | Polyurethane,<br>Thermoplastic<br>Elastomer (TPE),<br>Polycarbonate | Same |
| Biocompatible | Yes | Yes | Same |
| Packaging<br>Configuration | Thermo-formed (PETG)<br>plastic tray with Tyvek<br>lid | Thermo-formed (PETG)<br>plastic tray with Tyvek<br>lid | Same |
| Treatment<br>Condition | Sterile | Clean, non-sterile | Change to how it<br>is provided |
| Condition of Use | Single patient use,<br>disposable | Single patient use,<br>disposable | Same |
{7}------------------------------------------------
#### miraDry System bioTip Component Substantial Equivalence Comparison Table
#### SUMMARY OF VERIFICATION AND VALIDATION ACTIVITIES
The Special 510(k) for this device modification to the cleared miraDry System (K160141) utilized Miramar Labs established quality system and design control requirements in accordance with the Quality System Regulation (21 CFR 820).
Miramar Labs Inc. declares conformance to design controls in making this change and utilized the following risk based assessments:
- 1. Risk assessment including Hazard Analysis
- 2. Complaints incidence and review
- 3. Manufacturing environment and process evaluation including bioburden testing and monitoring trends
- 4. Instructions for Use
- 5. User interface and clinician user feedback
Risk control measures were focused on manufacturing environment, user interface evaluation, and postmarket surveillance and user feedback. No new risks were identified as a result of this change.
#### CONCLUSION
Based on all information in this Special 510(k), the miraDry MD4000 System and accessories is substantially equivalent to the identified predicate device currently marketed under the Federal Food, Drug and Cosmetic Act. The safety and effectiveness of the device modification are reasonably assured with no new or increased risks justifying 510(k) clearance.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.