K180375 · Braincool AB · DWJ · May 18, 2018 · Cardiovascular
Device Facts
Record ID
K180375
Device Name
The IQoolTM Warm System
Applicant
Braincool AB
Product Code
DWJ · Cardiovascular
Decision Date
May 18, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.5900
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The IQoolTM Warm System is a temperature regulating system indicated for monitoring and controlling patient temperature.
Device Story
Non-invasive thermal regulating system; monitors and controls patient temperature (32°C to 38.5°C). System comprises ECU 100 refrigeration/control unit, BC COOL (monopropylene glycol coolant), single-use cooling pads (head/neck, torso, thigh), and moisture-absorbing neoprene stabilization insulation. Operated via touchscreen monitor; uses third-party temperature probes for feedback. ECU 100 circulates coolant through pads to remove/add thermal energy. Supports hypothermia (cool/maintain) and normothermia (active rewarming) modes. Includes USB-based BC Stick for configuration/logs/updates. Used in clinical settings (not OR) by healthcare providers. Provides visual temperature graphs and safety alarms, including rapid-warming detection. Benefits patient by maintaining target temperature during treatment.
Clinical Evidence
Bench testing only. Simulated testing demonstrated the device's ability to cool patients to hypothermia, maintain temperature, and rewarm to normothermia, showing performance comparable to the predicate device and published research. No new clinical data provided.
Technological Characteristics
Thermal regulating system; non-invasive surface cooling/heating. Materials: monopropylene glycol coolant, neoprene stabilization insulation. Energy: 115/230 VAC. Dimensions: 105.2 x 50.5 x 61 cm. Connectivity: USB port for configuration/logs. Software-controlled fluid circulation via solenoid valves. Not sterilized. Patient contact materials identical to K162523.
Indications for Use
Indicated for monitoring and controlling patient temperature in patients requiring thermal regulation.
Regulatory Classification
Identification
A thermal regulating system is an external system consisting of a device that is placed in contact with the patient and a temperature controller for the device. The system is used to regulate patient temperature.
Predicate Devices
Arctic Sun® Temperature Management System (K101092)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is a seal with an abstract design, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The text is arranged in three lines, with "FDA" on the first line, "U.S. FOOD & DRUG" on the second line, and "ADMINISTRATION" on the third line. The logo is clean and professional, reflecting the FDA's role in regulating food and drugs in the United States.
May 18, 2018
Brain Cool AB % Adam Harris Associate Director, Regulatory Affairs Target Health Inc 261 Madison Ave New York, New York 10016
Re: K180375
Trade/Device Name: The IQoolTM Warm System Regulation Number: 21 CFR 870.5900 Regulation Name: Thermal regulating system Regulatory Class: Class II Product Code: DWJ Dated: March 28, 2018 Received: March 30, 2018
Dear Adam Harris:
We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good
{1}------------------------------------------------
manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Nicole G. Ibrahim -S
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K180375
Device Name The IQoolTM Warm System
Indications for Use (Describe)
The IQoolTM Warm System is a temperature regulating system indicated for monitoring and controlling patient temperature.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 5. 510(k) Summary
### 510(k) Applicant
Brain Cool AB
Scheelevägen 2, Medicon Village Lund, Sweden SE-223 81
Phone: +46 733 93 70 76
#### Contact Person: Adam Harris, MM, RAC Associate Director, Regulatory Affairs Target Health Inc. 261 Madison Avenue, 24th Floor New York, NY 10016 Telephone: (646) 218-2009
Fax: (212) 682-2105 aharris@targethealth.com
Date of Summary: March 28, 2018
| Device Proprietary<br>Name | The IQool™ Warm System | | |
|--------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------|------|
| Common/Usual Name | Thermal Regulating System | | |
| Classification Names /<br>Numbers and Code | 21 CFR | Classification Name | Code |
| | 870.5900 | Thermal Regulating<br>System | DWJ |
| Regulatory Class | II | | |
| Prescription Status | Prescription Device | | |
| Classification Panel | Cardiovascular | | |
| Predicate Device | K101092 | Arctic Sun® Temperature Management System | |
| Description of Device | The IQool™ Warm System is a non-invasive, thermal<br>regulating system that monitors and controls patient<br>temperature within a range of 32°C to 38.5°C (89.6°F to<br>101.3°F). | | |
| | The IQooL™ Warm System consists of: | | |
| | • ECU 100 – refrigeration and control unit – an integrated<br>control system operated via a touch screen monitor. | | |
| | • BC COOL– a cooling liquid consisting of diluted<br>monopropylene glycol (MPG5). The dilution is made by | | |
| | BrainCool AB to optimally serve the IQool TM Warm<br>System. Five liters of BC COOL are delivered with the<br>system. | | |
| | • Cooling Pads – the single use pads are the only skin<br>contacting component and can be fitted to the head/neck,<br>torso, and thigh. Liquid coolant is circulated from the tank<br>through the pads to control patient temperature. The pads<br>are designed and molded to give a good fit during<br>treatment and are intended for single patient use. | | |
| | • Stabilization insulation – the patented stabilization<br>insulation is made of insulating and moisture-absorbing<br>neoprene which supports the cooling pads and insulates<br>against the ambient environment condensation. The<br>elasticity of the stabilization insulation keeps the cooling<br>pads in place during treatment and ensures maximum<br>contact between the skin and the surface of the cooling<br>pad. The stabilizing insulation is intended for single use<br>only. | | |
| | Accessories: | | |
| | • BC Stick – a USB flash drive used to save system<br>configurations, specifically prepared to communicate with<br>the program of the IQoolTM Warm System. The BC stick<br>does not store or capture any user identifying information.<br>It can also be used to save a log file for system<br>troubleshooting or to update the software. | | |
| | • Filling pitcher – for refilling the tank with BC COOL<br>(coolant). Fill the tank with coolant before or directly after<br>start to avoid damage to the system. | | |
| Indications for Use | The IQoolTM Warm System is a temperature regulating system<br>indicated for monitoring and controlling patient temperature. | | |
{4}------------------------------------------------
{5}------------------------------------------------
| Name | IQool™ Warm System | Arctic Sun® Temperature<br>Management System – Predicate | IQool™ System – Reference |
|----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | BrainCool AB | Medivance, Inc | BrainCool AB |
| 510(K) Number | K180375 | K101092 | K162523 |
| Product Code | DWJ | DWJ | NZE |
| Regulation | 21 CFR 870.5900 | 21 CFR 870.5900 | 21 CFR 870.5900 |
| Indications for<br>Use | The IQool™ Warm System is a<br>temperature regulating system<br>indicated for monitoring and<br>controlling patient temperature. | Thermal regulating system, indicated<br>for monitoring and controlling patient<br>temperature. | Temperature reduction for adult<br>patients where clinically indicated, e.g.<br>in hyperthermic patients. |
| Description | The IQool™ Warm System is a non-<br>invasive, thermal regulating system that<br>monitors and controls patient<br>temperature within a range of 32°C to<br>38.5°C (89.6°F to 101.3°F). | The Arctic Sun Temperature<br>Management System is a<br>thermoregulatory device that monitors<br>and controls patient temperature<br>within a range of 32.0° C to 38.5° C<br>(89.60° F to 101.3° F). | The IQool™ System is a surface<br>cooling device that sustains and<br>monitors patient temperature within a<br>range of 33°C to 37°C.<br>The IQool™ System consists of: |
| | The IQool™ Warm System consists of:<br>• ECU 100 refrigeration and<br>control unit – an integrated<br>control system operated via a<br>touch screen monitor.<br>• BC COOL – a cooling liquid<br>consisting of diluted<br>monopropylene glycol<br>(MPG5). The dilution is made<br>by BrainCool AB to optimally<br>serve the IQool™ Warm<br>System. Five liters of BC | The Arctic Sun System consists of the<br>Arctic Sun Control Module and<br>disposable ArcticGel Pads.<br>A patient temperature probe connected<br>to the Control Module provides patient<br>temperature feedback to an internal<br>control algorithm which automatically<br>increases or decreases the circulating<br>water temperature to achieve a pre-set<br>patient target temperature determined<br>by the clinician.<br>The Arctic Sun pulls temperature- | • ECU 100 – refrigeration and<br>control unit – The ECU 100 is a<br>refrigerator unit with an<br>integrated control system<br>operated via a touch screen<br>monitor.<br>• BC COOL- BC COOL is a<br>cooling liquid consisting of<br>diluted monopropylene glycol<br>(MPG5). The dilution is made<br>by BrainCool AB to optimally<br>serve the IQool™ |
| Name | IQool™ Warm System | Arctic Sun® Temperature<br>Management System – Predicate | IQool™ System- Reference |
| | COOL are delivered with the<br>system.<br>Cooling Pads – the single use<br>pads are the only skin<br>contacting component and can<br>be fitted to the head/neck,<br>torso, and thigh. Liquid coolant<br>is circulated from the tank<br>through the pads to control<br>patient temperature. The pads<br>are designed and molded to give<br>a good fit during treatment and<br>are intended for single patient<br>use. Stabilization insulation – The<br>patented stabilization<br>insulation is made of<br>insulating and moisture-<br>absorbing neoprene which<br>supports the cooling pads and<br>insulates against the ambient<br>environment condensation.<br>The elasticity of the<br>stabilization insulation keeps<br>the cooling pads in place<br>during treatment and ensures<br>maximum contact between the<br>skin and the surface of the<br>cooling pad. The stabilizing<br>insulation is intended for<br>single use only. | controlled water ranging between 4° C<br>and 42 ° C (39.2° F and 107.6° F)<br>through the ArcticGel Pads, resulting<br>in heat exchange between the water<br>and the patient. | System.<br>Five liters of BC COOL are<br>delivered with the system.<br>Cooling Pads – the single use<br>cooling pads are the only skin<br>contacting component and can<br>be fitted to the head/neck,<br>torso, and thigh. Liquid<br>coolant is circulated from the<br>tank through the pads to cool<br>patients. The cooling pads are<br>designed and molded to give a<br>good fit during treatment and<br>are intended for single patient<br>use. Stabilization insulation – The<br>patented stabilization insulation<br>is made of insulating and<br>moisture-absorbent neoprene.<br>The stabilization insulation<br>supports the cooling pads and<br>insulates against the ambient<br>environment condensation. The<br>elasticity of the stabilization<br>insulation keeps the Cooling<br>Pads in place during treatment<br>and ensures maximum cooling<br>between the skin and the<br>surface of the Cooling Pad. The |
| Name | IQool™ Warm System | Arctic Sun® Temperature<br>Management System – Predicate | IQool™ System – Reference |
| | Accessories:<br>BC Stick – a USB flash drive<br>used to save system<br>configurations, specifically<br>prepared to communicate with<br>the program of the IQool™<br>Warm System. The BC Stick<br>does not store or capture any<br>user identifying information.<br>It can also be used to save a<br>log file for system<br>troubleshooting or to update<br>the software. Filling pitcher – for refilling<br>the tank with coolant. Fill the<br>tank with coolant before or<br>directly after start to avoid<br>damage to the system. The ECU 100 pushes temperature-<br>controlled BC COOL ranging<br>between 4°C and 40°C through the<br>Cooling Pads at approximately 1.2<br>liter per minute per pad. This results<br>in heat exchange between the BC<br>COOL and the patient. Patient<br>temperature is monitored by one or | | stabilizing insulation is<br>intended for single use only.<br>Accessories:<br>BC STICK — The BC STICK<br>is a USB flash drive used to<br>save system configurations,<br>specifically prepared to<br>communicate with the program<br>of the IQool™ System. The BC<br>stick does not store or capture<br>any user identifying<br>information. It can also be used<br>to save a log file for system<br>troubleshooting or to update the<br>software. Filling pitcher - The filling<br>pitcher is for refilling the tank<br>with BC COOL (coolant). Fill<br>the tank with coolant before or<br>directly after start to avoid<br>damage to the system. The ECU 100 pushes temperature-<br>controlled BC COOL ranging between<br>4°C and 30°C through the Cooling<br>Pads at approximately 1.2 liter per<br>minute per pad. This results in heat |
| | two commercially available third-<br>party temperature probes. The<br>IQool™ Warm System maintains a<br>controlled patient temperature during | | exchange between the BC COOL and<br>the patient. Patient temperature is<br>monitored by one or two commercially<br>available third-party temperature |
| Name | IQool™ Warm System | Arctic Sun® Temperature<br>Management System – Predicate | IQool™ System- Reference |
| | | | |
| | the entire treatment period. Any<br>deviations from the set temperature<br>are automatically re-adjusted by the<br>system. The treatment settings for<br>temperature and time can be changed<br>through the touchscreen monitor.<br>Alarms and notifications are<br>activated if any errors are detected.<br>Temperature graphs for each<br>treatment are shown on the<br>touchscreen display for visual<br>monitoring. | | probes. The IQool™ System maintains<br>a controlled patient temperature during<br>the entire treatment period. Any<br>deviations from the default temperature<br>are automatically adjusted by the<br>system. The default treatment settings<br>for temperature and time can be<br>changed through the touchscreen<br>monitor. Alarms and notifications are<br>activated if any errors are detected.<br>Temperature graphs for each treatment<br>are shown on the touchscreen display<br>for visual monitoring. |
| Technological<br>Characteristics | Achieves thermal regulation (cooling<br>and rewarming) by circulating liquid<br>coolant through patient contacting<br>cooling pads. Temperature is<br>controlled by the settings on the<br>device. | Achieves cooling and heating through<br>patient contacting gel pads<br>through which water is circulated.<br>Temperature is controlled by the<br>settings on the device. | Achieves cooling with patient<br>contacting Cooling Pads through which<br>coolant is circulated. Cooling is<br>controlled by the settings on the<br>device. Consistent temperature<br>reduction is maintained and controlled<br>by the system pushing more or less<br>coolant through the pads as needed<br>achieve or maintain desired cooling. |
| Therapy Modes | Hypothermia: cool, maintain<br>Normothermia: actively rewarm | Hypothermia: cool, maintain<br>Normothermia: actively rewarm | Hypothermia: cool, maintain<br>Normothermia: monitor and control<br>rewarming |
| Cooling<br>Mechanism | Employs a single tank which employs<br>three solenoid valves (ON/OFF) to<br>push coolant to the pads intermittently<br>to achieve and maintain desired<br>temperature automatically | Employs two separate tanks, one for<br>cooling and another for heating. To<br>obtain a certain temperature liquid<br>from the two tanks is mixed passes<br>continuously into the cooling pads on | Employs a single tank which employs<br>three solenoid valves (ON/OFF) to<br>push coolant to the pads intermittently<br>to achieve and maintain desired<br>temperature automatically. |
| Name | IQool™ Warm System | Arctic Sun® Temperature Management System – Predicate | IQool™ System- Reference |
| | | the patient. | |
| Heating Capability | Yes | Yes | No |
| Heating Capacity | 2500 BTU/hr 750 Watts | 2500 BTU/hr / 750 Watts | N/A |
| Cooling Rates | 1.38 to 1.61 °C/hour | 1.2 to 2.0 per hour °C/hour | 1.38 to 1.61 °C/hour |
## Table 1: Substantial Equivalence Comparison Table
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
## Other Specifications
| Cooling<br>medium | Diluted monopropylene glycol | Water | Diluted monopropylene glycol |
|---------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|
| Reservoir<br>Capacity | 4.0 liters | 3.5 liters | 4.0 liters |
| Water Flow<br>Rate | 1.5 – 6 liters per minute | 5 liters per minute | 1.5 – 6 liters per minute |
| Patient Probe<br>Type | YSI 400 Series compatible (rectal) | YSI 400 Series compatible | YSI 400 Series compatible (rectal) |
| Patient<br>Temperature<br>Inputs | Patient Temp 1: control, monitor, alarm<br>Patient Temp 2: monitor, alarm | Patient Temp 1: control, monitor, alarm<br>Patient Temp 2: monitor, alarm | Patient Temp 1: control, monitor, alarm<br>Patient Temp 2: monitor, alarm |
| Patient<br>Temperature<br>Measurement | ±0.4°C (10°C to 32°C)<br>±0.2°C (32°C to 38°C)<br>±0.4°C (38°C to 44°C) | ±0.4°C (10°C to 32°C)<br>±0.2°C (32°C to 38°C)<br>±0.4°C (38°C to 44°C) | ±0.4°C (10°C to 32°C)<br>±0.2°C (32°C to 38°C)<br>±0.4°C (38°C to 44°C) |
| Accuracy | Includes ±0.1°C external probe | Includes ±0.1°C external probe | Includes ±0.1°C external probe |
| Patient<br>Temperature<br>Range | 32°C to 38.5°C<br>89.6°F to 101.3°F<br>0.1 °C/°F increments | 32°C to 38.5°C<br>89.6°F to 101.3°F<br>0.1 °C/°F increments | 33°C to 37°C<br>91.4°F to 98.6°F<br>0.1 °C/°F increments |
| Water/Fluid<br>Temperature | -50°C to 99°C / -58°F to 210.2°F<br>0.1 °C/°F increments | -50°C to 99°C / -58°F to 210.2°F<br>0.1 °C/°F increments | -50°C to 99°C / -58°F to 210.2°F<br>0.1 °C/°F increments |
| Display Range | 4°C to 40°C / 39.2°F to 105.8°F | 4°C to 42°C / 39.2…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.