← Product Code [FDS](/productcode/FDS) · K180341

# FUJIFILM 600 Series Endoscope EG-600WR v2 (K180341)

_Fujifilm Corporation · FDS · Aug 3, 2018 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K180341

## Device Facts

- **Applicant:** Fujifilm Corporation
- **Product Code:** [FDS](/productcode/FDS.md)
- **Decision Date:** Aug 3, 2018
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The device is intended for the visualization of the upper digestive tract, specifically for the observation, diagnosis, and endoscopic treatment of the esophagus, stomach, and duodenum.

## Device Story

Upper gastrointestinal endoscope; captures images via CMOS sensor at distal tip. Light source transmits light through fiber bundles to illuminate body cavity; CMOS sensor captures visual data. Used in clinical settings by physicians for visualization, diagnosis, and treatment of esophagus, stomach, and duodenum. Connects to compatible video processors and light sources for image display on monitors. Provides real-time visualization to assist clinicians in identifying pathology and performing endoscopic interventions. Benefits include minimally invasive access to upper GI tract for diagnostic and therapeutic procedures.

## Clinical Evidence

Bench testing only. No clinical data presented. Performance testing included electrical safety (ANSI/AAMI ES60601-1, IEC 60601-1-2/6, IEC 60601-2-18), biocompatibility (ISO 10993-1/5/10), endoscopic performance (ISO 8600-1), and reprocessing validation per FDA guidance. Functional testing verified field of view, bending, air/water/suction rates, resolution, and light guide output met pre-defined acceptance criteria.

## Technological Characteristics

Upper GI video endoscope; CMOS image sensor; 2800 light guide fiber bundles; 9.2mm distal diameter; 2.8mm forceps channel; 1100mm working length. Connectivity via video processor/light source. Materials include new epoxy resin. Reprocessing via high-level disinfection/sterilization. Electrical safety per ANSI/AAMI ES60601-1 Ed 3.1.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- Fujifilm 600 Series Endoscopes EC-600WL and EG-600WR ([K132210](/device/K132210.md))

## Reference Devices

- Fujifilm Video Colonoscope Model EC-600WLv2 ([K160196](/device/K160196.md))
- Fujifilm Endoscope Models EC-600HL and EC-600LS ([K162622](/device/K162622.md))
- VP/BL-7000 ([K163675](/device/K163675.md))
- VP-4440FN/XL-4450FN ([K162836](/device/K162836.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

August 3, 2018

FUJIFILM Corporation % Jeffrey Wan Specialist, Regulatory Affairs FUJIFILM Medical Systems U.S.A., Inc. 10 High Point Drive Wayne, NJ 07470

Re: K180341

> Trade/Device Name: FUJIFILM 600 Series Endoscope EG-600WR v2 Regulation Number: 21 CFR§ 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: II Product Code: FDS Dated: July 5, 2018 Received: July 6, 2018

Dear Jeffrey Wan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be avare that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see

https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Glenn B. Bell -S

for

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K180341

Device Name

FUJIFILM 600 Series Endoscope EG-600WR v2

Indications for Use (Describe)

The device is intended for the visualization of the upper digestive tract, specifically for the observation, diagnosis, and endoscopic treatment of the esophagus, stomach, and duodenum.

Type of Use (*Select one or both, as applicable*)

| <span style="font-size:12pt"><b>    Prescription Use (Part 21 CFR 801 Subpart D)   </b></span> | <span style="font-size:12pt"><b>    Over-The-Counter Use (21 CFR 801 Subpart C)   </b></span> |
|------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|
|------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K180341 Page 1 of 5

#### 510(k) SUMMARY

#### FUJIFILM 600 Series Endoscope EG-600WR v2

#### Submitter's Information:

FUJIFILM Corporation 798 Miyanodai Kaisei-Machi Ashigarakami-Gun, Kanagawa, 258-8538, Japan FDA Establishment Registration Number: 3001722928

#### Contact Person:

Jeffrey Wan Specialist, Regulatory Affairs Fujifilm Medical Systems U.S.A., Inc. Telephone: (973) 709-2219 Facsimile: (201) 995-2452 E-Mail: jeffrey.wan@fujifilm.com

Date Prepared: July 5th, 2018

#### Identification of the Proposed Device:

| Proprietary/Trade Name: | FUJIFILM 600 Series Endoscope EG-600WR v2      |
|-------------------------|------------------------------------------------|
| Common Name:            | Video Endoscope                                |
| Device Class:           | Class II                                       |
| Review Panel:           | Gastroenterology/Urology                       |
| Classification:         | Endoscope and accessories, 21 C.F.R. § 876.150 |
| Product Code:           | FDS                                            |

#### Predicate Device:

Fujifilm 600 Series Endoscopes EC-600WL and EG-600WR, Fujifilm Medical Systems U.S.A, K132210

#### Reference Devices:

- 1. Fujifilm Video Colonoscope Model EC-600WLv2, FUJIFILM Medical Systems U.S.A., K160196
- 2. Fujifilm Endoscope Models EC-600HL and EC-600LS, FUJIFILM Medical Systems U.S.A., K162622

{4}------------------------------------------------

#### Intended Use / Indications for Use:

The device is intended for the visualization of the upper digestive tract, specifically for the observation, diagnosis, and endoscopic treatment of the esophagus, stomach, and duodenum.

#### Device Description:

FUJIFILM 600 Series Endoscope EG-600WR v2 is an upper gastrointestinal endoscope that captures images when used in combination with a video processor and light source. Light travels from the light source, through the glass fiber bundles in the endoscope, and out the tip of the insertion portion to illuminate the body cavity. This provides enough light to the CMOS image sensor to capture an image and display it on the monitor.

### Technoloqical Characteristics:

The proposed device EG-600WR v2 differs from the predicate device in the following minor modifications:

- o Use of a new epoxy resin in the manufacturing process instead of the original epoxy resin. Both the new and original epoxy resins can be used for repair.
- Minor design changes to the optical fiber, distal end cap, and air/water nozzle .
- Addition of VP/BL-7000 (K163675) and VP-4440FN/XL-4450FN (K162836) as compatible . video processors and light sources.
- Expansion of transport and storage conditions
- Use of an improved Air/Water button AW-500 and Suction button SB-500 ●
- Continued compatibility with an improved Water Tank WT-4 ●
- o J Tube JT-500 is changed from a mandatory accessory to an optional accessory.

A comparison of the technological characteristics between the modified and predicate devices is provided in the table below.

|                                        | Proposed Device                 | Predicate Device                                                                                                                                                                                   |                                                                                                                                                                                       |
|----------------------------------------|---------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device name                            | EG-600WR v2                     | EG-600WR                                                                                                                                                                                           |                                                                                                                                                                                       |
| Common name                            | Same as K132210                 | Endoscope and accessories                                                                                                                                                                          |                                                                                                                                                                                       |
| Manufacturer                           | Same as K132210                 | FUJIFILM Corporation                                                                                                                                                                               |                                                                                                                                                                                       |
| 510(k) number                          | Same as K132210                 | K132210                                                                                                                                                                                            |                                                                                                                                                                                       |
| Intended<br>Use/Indications<br>for Use | Same as K132210                 | The device is intended for the visualization of the<br>upper digestive tract, specifically for the observation,<br>diagnosis, and endoscopic treatment of the<br>esophagus, stomach, and duodenum. |                                                                                                                                                                                       |
| Appearance                             |                                 | Same as K132210                                                                                                                                                                                    | Image: Endoscope                                                                                                                                                                      |
| Viewing direction                      |                                 | Same as K132210                                                                                                                                                                                    | Forward / 0 degrees                                                                                                                                                                   |
| Observation<br>range                   |                                 | Same as K132210                                                                                                                                                                                    | 2-100mm                                                                                                                                                                               |
| Field of view                          |                                 | Same as K132210                                                                                                                                                                                    | 140 degrees                                                                                                                                                                           |
| F# of the objective<br>lens            |                                 | Same as K132210                                                                                                                                                                                    | 9.1                                                                                                                                                                                   |
| Resolution                             |                                 | Same as K132210                                                                                                                                                                                    | At 5mm of working distance: 0.07mm of line pair on<br>the square wave chart is readable.<br>At 100mm of working distance: 0.9mm of line pair on<br>the square wave chart is readable. |
| Distortion<br>characteristics          |                                 | Same as K132210                                                                                                                                                                                    | Orthogonal Projection                                                                                                                                                                 |
| Magnification of<br>lens(es)           |                                 | Same as K132210                                                                                                                                                                                    | 0.48-0.01                                                                                                                                                                             |
| Focal length                           |                                 | Same as K132210                                                                                                                                                                                    | 1.2mm                                                                                                                                                                                 |
| Image sensors                          |                                 | Same as K132210                                                                                                                                                                                    | CMOS                                                                                                                                                                                  |
| # of Light Guide<br>Fiber Bundles      |                                 | Same as K132210                                                                                                                                                                                    | 2800                                                                                                                                                                                  |
| Distal end<br>diameter                 |                                 | Same as K132210                                                                                                                                                                                    | 9.2mm                                                                                                                                                                                 |
| Flexible portion<br>diameter           |                                 | Same as K132210                                                                                                                                                                                    | 9.3mm                                                                                                                                                                                 |
| Maximum<br>insertion diameter          |                                 | Same as K132210                                                                                                                                                                                    | 10.7mm                                                                                                                                                                                |
| Bending                                | Up                              | Same as K132210                                                                                                                                                                                    | 210 degrees                                                                                                                                                                           |
| capability                             | Down                            | Same as K132210                                                                                                                                                                                    | 90 degrees                                                                                                                                                                            |
|                                        | Left                            | Same as K132210                                                                                                                                                                                    | 100 degrees                                                                                                                                                                           |
|                                        | Right                           | Same as K132210                                                                                                                                                                                    | 100 degrees                                                                                                                                                                           |
| Forceps channel<br>diameter            |                                 | Same as K132210                                                                                                                                                                                    | 2.8mm                                                                                                                                                                                 |
| Working length                         |                                 | Same as K132210                                                                                                                                                                                    | 1100mm                                                                                                                                                                                |
| Total length                           |                                 | Same as K132210                                                                                                                                                                                    | 1400mm                                                                                                                                                                                |
| WJ Function                            |                                 | Same as K132210                                                                                                                                                                                    | ?                                                                                                                                                                                     |
| Location of WJ<br>inlet                |                                 | Same as K132210                                                                                                                                                                                    | On the light guide connector                                                                                                                                                          |
| Video Processor                        |                                 | Light source: BL-7000<br>Processor: VP-7000                                                                                                                                                        | EPX-4440HD<br>Light source: XL-4450<br>Processor: VP-4440HD                                                                                                                           |
|                                        |                                 | EPX-4440HD<br>Light source: XL-4450<br>Processor: VP-4440HD                                                                                                                                        | EPX-4440FN<br>Light source: XL-4450FN                                                                                                                                                 |
|                                        | Processor: VP-4440FN            |                                                                                                                                                                                                    | 500 Series                                                                                                                                                                            |
| Video/Light guide connector            | Same as K132210                 |                                                                                                                                                                                                    |                                                                                                                                                                                       |
| Peripherals                            | Same as K132210                 |                                                                                                                                                                                                    | Water Tank WT-2                                                                                                                                                                       |
|                                        |                                 |                                                                                                                                                                                                    | Water Tank WT-4                                                                                                                                                                       |
|                                        |                                 |                                                                                                                                                                                                    | Monitor                                                                                                                                                                               |
|                                        |                                 |                                                                                                                                                                                                    | Printer                                                                                                                                                                               |
|                                        |                                 |                                                                                                                                                                                                    | Electrosurgical Instruments                                                                                                                                                           |
|                                        |                                 |                                                                                                                                                                                                    | Foot Switch                                                                                                                                                                           |
|                                        |                                 |                                                                                                                                                                                                    | Cart                                                                                                                                                                                  |
| Accessories                            | Cleaning Brush (WB11002FW2)     | Cleaning Brush (WB11002FW2)                                                                                                                                                                        |                                                                                                                                                                                       |
|                                        | Cleaning Brush (WB4321FW2)      | Cleaning Brush (WB4321FW2)                                                                                                                                                                         |                                                                                                                                                                                       |
|                                        | Cleaning Adapter Kit (CA-510/A) | Cleaning Adapter Kit (CA-510/A)                                                                                                                                                                    |                                                                                                                                                                                       |
|                                        | Forceps Valve (FOV-DV7)         | Forceps Valve (FOV-DV7)                                                                                                                                                                            |                                                                                                                                                                                       |
|                                        | Ventilation Adapter (AD-7)      | Ventilation Adapter (AD-7)                                                                                                                                                                         |                                                                                                                                                                                       |
|                                        | Air/Water button (AW-500)       | J Tube (JT-500)                                                                                                                                                                                    |                                                                                                                                                                                       |
|                                        | Suction button (SB-500)         | Air/Water button (AW-500)                                                                                                                                                                          |                                                                                                                                                                                       |
|                                        | Water Jet Inlet cap             | Suction button (SB-500)                                                                                                                                                                            |                                                                                                                                                                                       |
|                                        |                                 | Water Jet Inlet cap                                                                                                                                                                                |                                                                                                                                                                                       |
| Optional Items                         | Air leak tester LT-7F           | Air leak tester LT-7F                                                                                                                                                                              |                                                                                                                                                                                       |
|                                        | J Tube (JT-500)                 |                                                                                                                                                                                                    |                                                                                                                                                                                       |
| Electrical Safety Compliance           | ANSI AAMI ES60601-1 Edition 3.1 | ANSI AAMI ES60601-1 Edition 3.0                                                                                                                                                                    |                                                                                                                                                                                       |

{5}------------------------------------------------

{6}------------------------------------------------

Substantial equivalence was determined based on the performance testing described below.

#### Performance Data:

Electrical safety testing was conducted on the proposed device EG-600WR v2 using the following consensus standards: ANSI/AAMI ES60601-1:2005+AMD1:2012 CSV, IEC 60601-1-2:2007, IEC 60601-1-6:2013, and IEC 60601-2-18:2009.

The proposed device EG-600WR v2 was adopted into biocompatibility testing conducted on the reference device EC-600WL v2 using the following consensus standards: ISO 10993-1:2009, ISO 10993-5:2009, and ISO 10993-10:2010.

Endoscopic performance testing was conducted on the proposed device EG-600WR v2 according to the consensus standard ISO 8600-1:2015.

Cleaning, high-level disinfection, and sterilization testing was conducted on the proposed device EG-600WR v2 in accordance with the FDA guidance "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling" (March 17, 2015).

The proposed device EG-600WR v2 was adopted into storage and transportation validation testing conducted on the reference devices EC-600HL/LS.

{7}------------------------------------------------

Fujifilm conducted the following performance testing on the proposed device EG-600WR v2 to ensure that the modified device performs equivalently to the predicate device:

- . Field of view
- Bending capability
- Air supply rate
- Water supply rate
- Suction rate
- o Working length
- Forceps channel diameter
- Viewing direction ●
- o Resolution
- LG output

In all cases, the device met the pre-defined acceptance criteria for the test.

#### Substantial Equivalence:

The company's EG-600WR v2 has the same intended use as the previously cleared predicate EG-600WR (K132210). In addition, the proposed device EG-600WR v2 has similar indications for use, technological characteristics, and principles of operation as its predicate. The minor differences between the proposed device and its predicate device do not raise new or additional questions of safety or effectiveness of the proposed device. Thus, the proposed device EG-600WR v2 is substantially equivalent to its predicate device.

### Conclusions:

The modified EG-600WR v2 is substantially equivalent to the predicate EG-600WR and conforms to applicable medical device safety and performance standards.

---

**Source:** [https://fda-staging.innolitics.com/device/K180341](https://fda-staging.innolitics.com/device/K180341)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
