The DISKOM is intended for use in aspiration of disc material during percutaneous discectomies in the lumbar, thorace and cervical regions of the spine
Device Story
DISKOM is a single-use, battery-powered discectomy probe used by specialized personnel in clinical settings. It functions with an introducer needle (cannula and stylet) positioned via CT or fluoroscopic guidance. Once the cannula is placed, the stylet is removed and the probe is inserted. The probe utilizes a DC motor to drive an internal screw conveyor mechanism, which excises and retrieves nucleus pulposus tissue through the cannula into a transparent collection container. The device aids in spinal decompression procedures. It is provided in various cannula lengths and gauges to accommodate different spinal regions. The healthcare provider uses the device to mechanically remove disc material, potentially relieving pressure on spinal nerves.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing, including functional performance (time of functioning, engine RPM, tissue retrieval quantity, battery shelf life, maximum temperature), radiodetectability, and biocompatibility testing (ISO 10993-5, -10, -11).
Indicated for patients requiring aspiration of disc material during percutaneous discectomies in lumbar, thoracic, and cervical spine regions.
Regulatory Classification
Identification
An arthroscope is an electrically powered endoscope intended to make visible the interior of a joint. The arthroscope and accessories also is intended to perform surgery within a joint.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 26, 2018
Biopsybell s.r.l. % Mr. Maurizio Pantaleoni CEO Isemed s.r.l Via Togliatti 19/X, Imola (BO) 40026, ITALY
Re: K180315
Trade/Device Name: DISKOM Regulation Number: 21 CFR 888.1100 Regulation Name: Arthroscope Regulatory Class: Class II Product Code: HRX Dated: February 1, 2018 Received: February 5, 2018
Dear Mr. Pantaleoni:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely. Jennifer R. Stevenson -S3 For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
510(k) Number (if known) K180315
Device Name
DISKOM
Indications for Use (Describe)
The DISKOM is intended for use in aspiration of disc material during percutaneous discectomies in the lumbar, thorace and cervical regions of the spine
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IE NEEDED!
This section applies only to requirements of the Paperwork Reduction Act of 1995
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduc; or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## Summary of the K180315 DISKOM
This 510(k) Summary is being submitted in accordance with the requirements of the SMDA 1990 and 21 CFR 807.92.
#### General Information 1.
| Submitter : | Biopsybell s.r.l |
|-------------|------------------------------|
| | Via A. Manuzio n.24 |
| | 41037, Mirandola (MO), Italy |
#### Establishment Registration Number: 9617616
| Consultant/ Contact: | Maurizio Pantaleoni |
|----------------------|-----------------------------|
| | Isemed srl |
| | Via Togliatti 19/X |
| | Imola (BO) 40026, ITALY |
| | Mob. +39 348 4435155 |
| | Tel. +39 0542 683803 |
| | Fax +39 0542 698456 |
| | Email: regulatory@isemed.eu |
Summary Prepared Date: 2018-02-01
- 2. Names
Device Name: Classification Name: Product Code: Regulation number: CLASS:
DISKOM ARTHROSCOPE HRX 888.1100 ll
#### 3. Predicate Devices
The DISKOM is substantially equivalent to the following devices legally marketed in US:
| Applicant | Device name | 510(k) Number | Product code |
|-----------|----------------------------------------------------|---------------|--------------|
| GALLINI | Herniatome Percutaneous Discectomy Device | K141557 | HRX |
| STRYKER | Stryker Decompressor Percutaneous Discectomy Probe | K032473 | HRX |
{4}------------------------------------------------
### 4. Device Description
The DISKOM is a single use, active, EO sterilized device, that shall be used by specialized personnel during discectomy procedures.
The DISKOM is a disposable discectomy probe that passes through and works in conjunction with an introducer cannula to remove intervertebral disc nucleus pulposus.
DISKOM is composed by an INTRODUCER NEEDLE and a DISCECTOMY PROBE. The INTRODUCER NEEDLE is made of an aspiration cannula and an introducer stylet. The introducer needle is positioned on the disc (using CT / Fluoroscopic guide), then the stylet is removed and the cannula is used to position the discectomy probe.
The DISCECTOMY PROBE contains a battery source DC motor that causes the internal mechanism to act as a screw conveyor to remove and retrieve the excised debris of the spine through the outer cannula and into the transparent collection container.
The DISKOM is provided in two different models with two different cannula lengths (16cm and 8cm) and two different gauges cannulas (17G and 19G).
#### 5. Indications for Use
The DISKOM is intended for use in aspiration of disc material during percutaneous discectomies in the lumbar, thoracic and cervical regions of the spine.
| | Subject device | Predicate device (K141557) | Predicate device (K032473) |
|---------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Name | DISKOM | Herniatome Percutaneous<br>Discectomy Device | Stryker Decompressor<br>Percutaneous Discectomy Probe |
| Manufacturer | BIOSPYBELL S.R.L. | GALLINI | STRYKER |
| 510(K) No. | - | K141557 | K032473 |
| Regulation Name | Arthroscope | Arthroscope | Arthroscope |
| Regulation Number | 888.1100 | 888.1100 | 888.1100 |
| Classification | II | II | II |
| Product Code | HRX | HRX | HRX |
| Intended Use | | | |
| Indications for use | The Diskom is intended for use in<br>aspiration of disc material during<br>percutaneous discectomies in the<br>lumbar, thoracic and cervical<br>regions of the spine. | The Herniatome Percutaneous<br>Discectomy Device is intended<br>for use in aspiration of disc<br>material during percutaneous<br>discectomies in the lumbar,<br>thoracic, and cervical regions<br>of the spine. | The Dekompressor<br>percutaneous Discectomy Probe<br>is intended for use in aspiration<br>of disc material during<br>percutaneous discectomies in<br>the lumbar, thoracic and<br>cervical regions of the spine. |
| Technical features | | | |
| Single Use Device | YES | YES | YES |
| Power | DC Motor - 9 Volt Battery | DC Motor - Battery | DC Motor - 9 Volt Battery |
| Maximum use time | 10 minutes | Unknown | 10 minutes |
| Configuration | Single use device with:<br>- introducer needle composed by<br>an external cannula and introducer<br>stylet<br>- Discectomy probe composed by a<br>stylet with screw, a collection<br>container and a grip with internal<br>battery source DC motor<br>- The stylet has 2 screws that enable<br>the collection of disc material | Single use device with:<br>- introducer needle composed<br>by an external cannula and<br>internal stylet<br>- Discectomy probe composed<br>by a stylet with screw, a<br>collection<br>container<br>and<br>internal battery source DC<br>motor | Single use device with:<br>- introducer needle composed<br>by an external cannula and<br>internal stylet<br>- Discectomy probe composed<br>by a stylet with screw, a<br>collection container and internal<br>battery source DC motor<br>- the stylet has 2 screws in the<br>device with length 3" and 6" and<br>3 screws in the one with length<br>9" |
| Dimensions of<br>cannula<br>(length / gauge) | 16 cm / 17 gauge<br>(thoracic / lumbar) | 15 cm / 17 gauge (lumbar) | 6" (15.24 cm) / 17 gauge (also<br>available 13,15,19 gauge) -<br>(thoracic / lumbar)<br><br>9" (22,86 cm) / 17 gauge<br>(lumbar) |
| | 8 cm / 19 gauge (cervical) | 9 cm / 17 gauge (cervical) | 3" (7.62 cm) / 19 gauge<br>(cervical) |
| Aspiration cannula<br>hole position | Lateral hole | Lateral hole | Frontal hole |
| Materials | | | |
| Introducer Stylet | AISI 304 stainless steel | Stainless steel | Unknown |
| Cannula | AISI 304 stainless steel | Stainless steel | Unknown |
| Stylet with screw | Titanium | Stainless steel | Titanium |
| Biocompatibility | | | |
| Citotoxicity | Conforming to ISO 10993 testing<br>(ISO 10993-5) | Conforming | Unknown |
| Sensitization | Conforming to ISO 10993 testing<br>(ISO 10993-10) | Conforming | Unknown |
| Intracutaneous<br>Reactivity | Conforming to ISO 10993 testing<br>(ISO 10993-10) | Conforming | Unknown |
| Acute Systemic<br>toxicity | Conforming to ISO 10993 testing<br>(ISO 10993-11) | Conforming | Unknown |
| Bacterial endotoxins<br>test (LAL test) | Conforming | Unknown | Unknown |
| Pyrogen test | Conforming | Unknown | Unknown |
| Electrical Safety / Electromagnetic Compatibility | | | |
| Electrical Safety /<br>Electromagnetic<br>Compatibility | Conforming to:<br>IEC 60601-1 for electrical safety<br>EN 60601-1-2 for Electromagnetic<br>Compatibility | Conforming to:<br>IEC 60601-1 electrical safety<br>IEC 60601-1-2 for<br>Electromagnetic Compatibility | Conforming to:<br>IEC 60601-1 for electrical safety<br>IEC 60601-1-2 for<br>Electromagnetic Compatibility |
| Sterilization Shelf life | | | |
| Sterilization Method | Ethylene Oxide | Ethylene Oxide | Ethylene Oxide |
| Shelf life | 3 years | 3 years | 3 years |
| Performance test for determination of the SE | | | |
| Performance tests | The following performance tests have been performed on the device, both on newly manufactured and aged products:<br>time of functioning (20 minutes) engine rpm quantity of tissue retrieved during the procedure battery shelf life maximum temperature at the level of the stylet with screw radiodetectability | The following performance tests have been performed on the device:<br>volume flow rate study battery shelf life radiodetectability | Unknown |
#### 6. Substantial equivalence discussion & Summary of Technological Characteristics
{5}------------------------------------------------
# 510(K) SUMMARY
{6}------------------------------------------------
## 510(K) SUMMARY
### 7. Performance Data
#### 7.1 Non clinical tests performed on the subject device
The following comparative tests have been performed for all BIOPSYBELL DISKOM codes:
- 1. Time of functioning (20 minutes). Battery operation test for 20 minutes
- 2. Enqine rpm. Measurement of the number of rpm
- 3. Quantity of tissue retrieved during the procedure.
- 4. Battery shelf life: battery voltage measuring test during operation.
- 5. Maximum Temperature at the level of the stylet with screw measurement
The radiodetectability was another parameter tested. The radiopacity has been tested on subject device and on predicate device. T
Furthermore the following Biocompatibility tests were performed:
- -Citotoxicity
- -Sensitization
- -Intracutaneous Reactivity
- -Acute Systemic toxicity
- -Bacterial endotoxins test (LAL test)
- -Pyrogen test
Based on technological characteristics (intended use, dimensions and features) and performance data (comparative tests and biocompatibility tests) included in the present submission, the BIOPSYBELL DISKOM has been shown to be substantially equivalent to the listed predicate device.
7.2 Clinical tests performed on the subject device No clinical test were performed on the subject device
{7}------------------------------------------------
#### Conclusion 8
Based on technological characteristics (intended use, dimensions and features) and performance data (comparative tests and biocompatibility tests) and comparison discussion included in the present submission, the BIOPSYBELL DISKOM has been shown to be substantially equivalent to the listed predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.