ACUVUE OASYS (senofilcon A) with Photochromic Additive
K180299 · Johnson & Johnson Vision Care, Inc. · LPL · Apr 10, 2018 · Ophthalmic
Device Facts
Record ID
K180299
Device Name
ACUVUE OASYS (senofilcon A) with Photochromic Additive
Applicant
Johnson & Johnson Vision Care, Inc.
Product Code
LPL · Ophthalmic
Decision Date
Apr 10, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.5925
Device Class
Class 2
Indications for Use
The ACUVUE (senofilcon A) Soft (hydrophilic) Contact Lens with photochromic additive (spherical) is indicated for daily wear for the optical correction of refractive ametropia (myopia and hyperopia) in phakic persons with non-diseased eyes who may have 1.00D or less of astigmatism. The ACUVUE (senofilcon A) Soft (hydrophilic) Contact Lens with photochromic additive TORIC is indicated for daily wear for the optical correction of refractive ametropia (myopia and hyperopia) in phakic or aphakic persons with nondiseased eyes who may have 10.00D or less of astigmatism. The ACUVUE (senofilcon A) Soft (hydrophilic) Contact Lens with photochromic additive MULTIFOCAL is indicated for daily wear for the optical correction of refractive ametropia (myopia and hyperopia) and/or presbyopia in phakic or aphakic persons with non-diseased eyes who may have 0.75D or less of astigmatism. The ACUVUE (senofilcon A) Soft (hydrophilic) Contact Lens with photochromic additive MULTIFOCAL TORIC is indicated for daily wear for the optical correction of refractive ametropia (myopia, and/or astigmatism) and presbyopia in phakic or aphakic persons with non-diseased eyes who may need up to 4.00D of ADD power and may have 10.00D or less of astigmatism. These lenses are also indicated for the attenuation of bright light as they contain a photochromic additive which dynamically absorbs visible light. These lenses contain a UV Blocker to help protect against transmission of harmful UV radiation to the cornea and into the eye. Eye Care Professionals may prescribe the lenses either for daily disposable wear or frequent/planned replacement wear with cleaning, disinfection and scheduled replacement. When prescribed for daily disposable wear, no cleaning or disinfection is required. Lenses should be discarded upon removal. When prescribed for frequent/planned replacement wear, the lenses may be disinfected using a chemical disinfection system only and should be discarded after the recommended wearing period as prescribed by the Eye Care Professional. When the lenses are worn in a frequent/planned replacement modality, they are intended to be worn for up to 2-weeks (14 days).
Device Story
Soft silicone hydrogel contact lens; incorporates photochromic additive (naphthopyran) and UV-absorbing monomer (benzotriazole). Lens dynamically adjusts visible light transmittance (min 23% in activated state) based on UV/HEV exposure; blocks >99% UVB and >90% UVA. Prescribed by eye care professionals for daily or 2-week planned replacement. Patient wears lens to correct refractive errors while simultaneously attenuating bright light. Clinical benefit includes reduced glare/bright light sensitivity without compromising driving performance or visual acuity.
Clinical Evidence
Clinical study (n=24) in Australia; bilateral, randomized, masked, crossover design. Evaluated driving performance and vision in daytime/nighttime conditions on a closed-road circuit. Primary endpoint: overall driving performance score. Secondary endpoints: visual acuity, contrast threshold, road sign recognition, hazard avoidance, pedestrian recognition. Results showed no evidence of driving performance or visual acuity concerns.
Technological Characteristics
Material: 62% senofilcon A, 38% water. Additives: benzotriazole (UV blocker), naphthopyran (photochromic). Oxygen permeability (Dk): 103-122 x 10^-11. Refractive index: 1.42. Sterilization: steam. Form factor: hemispherical/hemitoric shells. Connectivity: N/A.
Indications for Use
Indicated for daily wear optical correction of refractive ametropia (myopia, hyperopia, astigmatism, presbyopia) in phakic or aphakic persons with non-diseased eyes. Also indicated for attenuation of bright light and UV protection.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.
Predicate Devices
ACUVUE OASYS (senofilcon A) Brand Contact Lens (K042275)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.
April 10, 2018
Johnson & Johnson Vision Care, Inc. Victoria Brennand Manager, Regulatory Affairs 7500 Centurion Parkway, Suite 100 Jacksonville, FL 32256
Re: K180299
Trade/Device Name: ACUVUE OASYS (senofilcon A) with Photochromic Additive Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (Hydrophilic) Contact Lens Regulatory Class: Class II Product Code: LPL, MVN Dated: January 31, 2018 Received: February 2, 2018
Dear Victoria Brennand:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Denise L. Hampton -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known)
#### K180299
Device Name
ACUVUE (senofilcon A) Soft (hydrophilic) Contact Lens with photochromic additive
#### Indications for Use (Describe)
The ACUVUE (senofilcon A) Soft (hydrophilic) Contact Lens with photochromic additive (spherical) is indicated for daily wear for the optical correction of refractive ametropia (myopia and hyperopia) in phakic persons with non-diseased eyes who may have 1.00D or less of astigmatism.
The ACUVUE (senofilcon A) Soft (hydrophilic) Contact Lens with photochromic additive TORIC is indicated for daily wear for the optical correction of refractive ametropia (myopia and hyperopia) in phakic or aphakic persons with nondiseased eyes who may have 10.00D or less of astigmatism.
The ACUVUE (senofilcon A) Soft (hydrophilic) Contact Lens with photochromic additive MULTIFOCAL is indicated for daily wear for the optical correction of refractive ametropia (myopia and hyperopia) and/or presbyopia in phakic or aphakic persons with non-diseased eyes who may have 0.75D or less of astigmatism.
The ACUVUE (senofilcon A) Soft (hydrophilic) Contact Lens with photochromic additive MULTIFOCAL TORIC is indicated for daily wear for the optical correction of refractive ametropia (myopia, and/or astigmatism) and presbyopia in phakic or aphakic persons with non-diseased eyes who may need up to 4.00D of ADD power and may have 10.00D or less of astigmatism.
These lenses are also indicated for the attenuation of bright light as they contain a photochromic additive which dynamically absorbs visible light.
These lenses contain a UV Blocker to help protect against transmission of harmful UV radiation to the cornea and into the eye.
Eye Care Professionals may prescribe the lenses either for daily disposable wear or frequent/planned replacement wear with cleaning, disinfection and scheduled replacement. When prescribed for daily disposable wear, no cleaning or disinfection is required. Lenses should be discarded upon removal. When prescribed for frequent/planned replacement wear, the lenses may be disinfected using a chemical disinfection system only and should be discarded after the recommended wearing period as prescribed by the Eye Care Professional. When the lenses are worn in a frequent/planned replacement modality, they are intended to be worn for up to 2-weeks (14 days).
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
#### 510(K) SUMMARY
#### Submitter Information
| Company: | Johnson & Johnson Vision Care, Inc.<br>7500 Centurion Parkway, Suite 100<br>Jacksonville, FL 32256 |
|-----------------|----------------------------------------------------------------------------------------------------|
| Contact Person: | Victoria Brennand |
| Email: | vbrenna2@its.jnj.com |
| Telephone: | (904) 443-3160 |
| FAX: | (904) 443-1424 |
| Date: | January 31, 2018 |
### Identification of the Device
| Common Name: | Soft Contact Lens |
|------------------------|------------------------------------------------------------------------|
| Device/Trade Name: | ACUVUE® (senofilcon A) Soft Contact Lens with<br>Photochromic Additive |
| Classification Name: | Soft (Hydrophilic) Contact Lens, Daily Wear |
| Device Classification: | Class II, 21 CFR 886.5925 (b) (1) |
| Product Code: | LPL, MVN |
## Predicate Device
- . ACUVUE OASYS® (senofilcon A) Brand Contact Lens cleared via K042275
{5}------------------------------------------------
### Description of Device
The subject device is a soft (hydrophilic) contact lens available in a spherical, toric, multifocal and/or multifocal-toric design. The composition of the lens is 62% senofilcon A and 38% water by weight when hydrated and stored in the buffered saline solution with methyl ether cellulose. The lens is supplied sterile (steam) in a foil sealed plastic package. The lenses are hemispherical or hemitoric shells. To date a 2-year shelf-life has been established.
The subject device is made of a silicone hydrogel material containing an internal wetting agent and UV absorbing monomers. A combination of a benzotriazole UV absorbing monomer and a naphthopyran monomer (photochromic additive) is used to block UV radiation and dynamically absorbs visible light. The transmittance characteristics for these lenses are less than 1.0% in the UVB range of 280 nm to 315 nm and less than 10.0% in the UVA range of 316 nm to 380 nm for the entire power range.
Additionally, the photochromic additive absorbs visible light in the range from 380 nm to 780 nm to a minimum 84% transmittance in the inactivated (closed) state. The activated (open) state dynamically absorbs visible light dependent on the lens thickness and the level of UV and high energy visible (HEV) radiation to a minimum of 23% transmittance.
| Property / Parameter | Subject Device | |
|-----------------------------------------------------------------------------------|----------------------------------------------------------|--|
| Water Content | 38% | |
| Refractive Index | 1.42 | |
| Oxygen Permeability (Fatt method, edge corrected) | 103 x 10-11 (cm2/sec)(mL O2/mL * mm Hg) | |
| Oxygen Permeability<br>(Fatt method, non-edge corrected) | 122 x 10-11 (cm2/sec)(mL O2/mL * mm Hg) | |
| Light Transmittance: Visible - closed form | 84% to 94% | |
| Light Transmittance: Visible - open form<br>calculated from Delta Optical Density | Minimum 23% ª | |
| Light Transmittance: UVA (316 nm to 380 nm) | < 10.0% | |
| Light Transmittance: UVB (280 nm to 315 nm) | < 1.0% | |
| Diameter | 12.0 mm to 15.0 mm | |
| Center Thickness, varies with power | 0.060 — 1.000 mm<br>-3.00D: 0.085 mm<br>+3.00D: 0.168 mm | |
| Base Curve | 7.85 mm to 10.00 mm | |
| Sphere Powers | -20.00D to +20.00D | |
Table 1 details properties and parameters of the subject device. Table 1: Physicochemical Properties and Parameters
{6}------------------------------------------------
| Property / Parameter | Subject Device |
|----------------------|-------------------|
| ADD Powers | +0.25D to +4.00D |
| Axis | 2.5° to 180° |
| Cylinder | -0.25D to -10.00D |
#### Physicochemical Properties and Parameters (Continued) Table 1:
a Luminous transmission from nominal delta optical density range 0.31 to 0.53
#### Indications for Use
The ACUVUE® (senofilcon A) Soft (hydrophilic) Contact Lens with photochromic additive (spherical) is indicated for daily wear for the optical correction of refractive ametropia (myopia and hyperopia) in phakic or aphakic persons with non-diseased eyes who may have 1.00D or less of astigmatism.
The ACUVUE® (senofilcon A) Soft (hydrophilic) Contact Lens with photochromic additive TORIC is indicated for daily wear for the optical correction of refractive ametropia (myopia and hyperopia) in phakic or aphakic persons with non-diseased eyes who may have 10.00D or less of astigmatism.
The ACUVUE® (senofilcon A) Soft (hydrophilic) Contact Lens with photochromic additive MULTIFOCAL is indicated for daily wear for the optical correction of refractive ametropia (myopia and hyperopia) and/or presbyopia in phakic persons with non-diseased eyes who may have 0.75D or less of astigmatism.
The ACUVUE® (senofilcon A) Soft (hydrophilic) Contact Lens with photochromic additive MULTIFOCAL TORIC is indicated for daily wear for the optical correction of refractive ametropia (myopia, hyperopia, and/or astigmatism) and presbyopia in phakic or aphakic persons with non-diseased eyes who may need up to 4.00D of ADD power and may have 10.00D or less of astigmatism.
These lenses are also indicated for the attenuation of bright light as they contain a photochromic additive which dynamically absorbs visible light.
These lenses contain a UV Blocker to help protect against transmission of harmful UV radiation to the cornea and into the eye.
Eye Care Professionals may prescribe the lenses either for daily disposable wear or frequent/planned replacement wear with cleaning, disinfection and scheduled replacement. When prescribed for daily disposable wear, no cleaning or disinfection is required. Lenses should be discarded upon removal. When prescribed for frequent/planned replacement wear, the lenses may be disinfected using a chemical disinfection system only and should be discarded after the recommended wearing period as prescribed by the Eye Care Professional. When the lenses are worn in a frequent/planned replacement modality, they are intended to be worn for up to 2-weeks (14 days).
{7}------------------------------------------------
#### Technological Characteristics
The technological characteristics of the subject device are compared to the characteristics of the predicate device in Table 2 and Table 3.
| Table 2: | Material & Physicochemical Comparison | | |
|----------|---------------------------------------|--|--|
|----------|---------------------------------------|--|--|
| Property | Predicate Device | Subject Device |
|-------------------------------|---------------------------------------------------------|---------------------------------------------------------|
| Material | senofilcon A | senofilcon A |
| ISO Classification Groupa | Group 5C<br>(Silicone hydrogel: low water<br>sub group) | Group 5C<br>(Silicone hydrogel: low water<br>sub group) |
| UV Blocker | Yes | Yes |
| Photochromic Additive | No | Yes |
| Water Content, % | 38 | 38 |
| Refractive Index | 1.42 | 1.42 |
| Oxygen Permeability (Dk)b | 103c<br>122d | 103c<br>122d |
| Specific Gravity (calculated) | 0.98-1.12 | 0.98-1.12 |
4 As referenced in ISO 18369-1:2017
b Dk units = x 10-11 (cm2/sec)(mL O2/mL * mm Hg)
ぐ 35℃ Fatt method, edge corrected
d 35°C Fatt method, non-edge corrected
{8}------------------------------------------------
| Indication | Predicate Device | Subject Device |
|--------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Spherical | For the optical correction of refractive<br>ametropia (myopia and hyperopia) in phakic<br>or aphakic persons with non-diseased eyes<br>who have 1.00D or less of astigmatism. | Same as predicate |
| Toric | For the optical correction of visual acuity in<br>phakic or aphakic persons with non-<br>diseased eyes that are hyperopic or myopic<br>and may have 10.00D or less of<br>astigmatism. | Same as predicate |
| Multifocal | For the optical correction of distance and<br>near vision in presbyopic, phakic or aphakic<br>persons with non-diseased eyes who may<br>have 0.75D or less of astigmatism. | Same as predicate |
| Multifocal-Toric | For the optical correction of distance and<br>near vision in presbyopic phakic or aphakic<br>persons with non-diseased eyes who may<br>have 10.00D of astigmatism or less. | Same as predicate |
| Attenuation of bright<br>light | N/A | These lenses are also<br>indicated for the<br>attenuation of bright light<br>as they contain a<br>photochromic additive<br>which dynamically<br>absorbs visible light. |
| UV Statement | Contain a UV Blocker to help protect<br>against transmission of harmful UV<br>radiation to the cornea and into the eye. | Same as predicate |
Table 3a: Indication Comparison
| Table 3b: | Wear/ Replacement Schedule Comparison | |
|-----------|---------------------------------------|--|
|-----------|---------------------------------------|--|
| Wear/<br>Replacement<br>Schedule | Predicate Device | Subject Device |
|----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|
| | Eye Care Professionals may prescribe the lenses either for single-use disposable wear or frequent/planned replacement wear with cleaning, disinfection and scheduled replacement. When prescribed for frequent/planned replacement wear, the lenses may be cleaned and disinfected using a chemical disinfection system only. | Same as predicate |
{9}------------------------------------------------
#### Non-clinical Performance Data
A series of in-vitro and in-vivo tests were performed to assess the properties and safety and effectiveness of the contact lens following the 1994 FDA Guidance Document for Daily Wear Contact Lenses. All biocompatibility tests were conducted in accordance with the GLP regulation (21 CFR Part 58). All other testing was conducted according to valid scientific protocols.
Non-Clinical testing performed includes:
- Physicochemical Properties ●
- Refractive Index
- Oxygen Permeability
- Specific Gravity
- Modulus
- -Tensile Strength
- Elongation
- Dynamic Contact Angle —
- । Leachables
- Light Transmittance l
- | Photochromic Performance
- Biocompatibility .
- Bacterial Reverse Mutation Study (according to ISO 10993-3:2003) —
- In Vitro Chromosomal Aberration Study in Chinese Hamster Ovary Cell | (according to ISO 10993-3:2003)
- । Cytotoxicity Study Using the ISO Direct Contact Method (according to ISO 10993-5:2009). Conducted before and after photochromic transitioning of the lens
- । Cytotoxicity Study Using the Colony Assay (Extraction Method) (according to ISO 10993-5:2009). Conducted before and after photochromic transitioning of the lens
- 22-Day Rabbit Contact Lens Study (according to ISO 9394:2012) -
- ISO Ocular Irritation Study in Rabbits (according to ISO 10993-10:2010) —
- ISO Acute Systemic Toxicity Study in Mice (according to ISO 10993-11:2006)
- । Guinea Pig Maximization Sensitization Study (according to ISO 10993-10:2010)
- Solution Compatibility (according to ISO 11981:2009) and Preservative Uptake ● and Release (according to ISO 11986:2010)
{10}------------------------------------------------
- Clear Care® Plus Cleaning and Disinfecting Solution ।
- RevitaLens® Blink® Multi-Purpose Disinfecting Solution ।
- OPTI-FREE® PureMoist® Multi-Purpose Disinfecting Solution ।
- Biotrue® Multi-Purpose Solution ।
The results of the non-clinical testing on the subject device demonstrate that:
- . the lens material and extracts are non-toxic and non-irritating, and
- lens physical and material properties are consistent with currently marketed lenses with the exception of photochromic performance (Delta OD)
### Clinical Performance Data
A clinical study was conducted in Australia to evaluate the effect of the subject device on vision and driving performance in both daytime and nighttime lighting under real world driving conditions. This was achieved through field-based driving studies on a closed-road driving circuit at night and during the day. Ouantitative methods were used to assess vision and driving performance under a range of challenging conditions. Appropriate masking, order of testing, randomization and control conditions were used.
The study evaluated 24 subjects in a bilateral, non-dispensing, randomized, partially masked (subject), crossover study. Primary endpoint was overall driving performance score. Secondary endpoint(s) included low luminance high contrast distance visual acuity; distance visual acuity (~1 lux); low luminance (~1 lux) contrast threshold; road sign recognition (percentage); percentage of hazards avoidance and pedestrian recognition distance.
The results of the clinical testing demonstrate there was no evidence of driving performance or visual acuity concerns while wearing the subject device during daytime and nighttime driving.
## Conclusions Drawn from the Non-clinical and Clinical Tests
Substantial Equivalence: Information presented in this Premarket Notification establishes that the subject device is as safe and effective as the predicate device when used for the proposed indication, in accordance with the labeled directions for use. All potential risks associated with the proposed modification, the addition of a photochromic additive, intended for the attenuation of bright light, have been thoroughly addressed. No new questions of safety and efficacy compared to the predicate device have been identified.
Risk and Benefits: The risks of the subject device are the same as those normally attributed to the wearing of soft (hydrophilic) contact lenses on a daily wear basis. The benefits to the patient are the same as those for other soft (hydrophilic) contact lenses with the additional benefit of the attenuation of bright light.
## Other Information
Not applicable.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.