← Product Code [FDS](/productcode/FDS) · K180292

# PENTAX Video Upper G.I. Scope EG34-i10 (K180292)

_Pentax Medical of America, Inc. · FDS · Oct 12, 2018 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K180292

## Device Facts

- **Applicant:** Pentax Medical of America, Inc.
- **Product Code:** [FDS](/productcode/FDS.md)
- **Decision Date:** Oct 12, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

This instrument is intended to be used with a PENTAX video processor (including light source), documentation equipment, monitor, Endotherapy Device such as a Biopsy Forceps, and other ancillary equipment for endoscopy and endoscopic surgery within the upper digestive tract including the esophagus, stomach, and duodenum.

## Device Story

Flexible video endoscope for upper gastrointestinal tract visualization and therapeutic access. Device comprises insertion portion (distal end with objective lens, light guide, instrument channel, air/water/water-jet nozzles), control body (angulation knobs, suction/air/water valves, remote buttons), and PVE connector. Operates by connecting to PENTAX video processor/light source; CCD at distal end captures reflected light from body cavity; video processor converts data to image signal for monitor display. Used in clinical settings by physicians for diagnostic and therapeutic procedures (e.g., biopsy). Provides real-time optical visualization and allows passage of endotherapy devices through instrument channel. Benefits include minimally invasive access to upper GI tract for diagnosis and treatment.

## Clinical Evidence

Bench testing only. Studies included soil accumulation, cleaning efficacy, high-level disinfection (HLD) efficacy, rinsing validation, and sterilization validation. Biocompatibility evaluated per ISO 10993-1. Electrical safety and EMC testing performed per IEC 60601-1, IEC 60601-2-18, and IEC 60601-1-2. Optical performance (light distribution, spectral distribution, luminous flux) and photobiological safety (IEC 62471) confirmed comparable to predicate. Usability testing demonstrated equivalent operation to i10/k10 series scopes.

## Technological Characteristics

Reusable semi-critical flexible endoscope. Components: insertion portion, control body, PVE connector. Sensing: CCD image sensor. Illumination: Light carrying bundle. Connectivity: Proprietary PVE connector to PENTAX video processors. Sterilization: High-level disinfection or sterilization. Electrical safety: IEC 60601-1, IEC 60601-2-18, ANSI/AAMI ES60601-1. EMC: IEC 60601-1-2. Biocompatibility: ISO 10993-1.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- PENTAX Video Upper GI Scopes EG29-i10 ([K131902](/device/K131902.md))

## Reference Devices

- PENTAX Medical EPK-i7010 Video Processor ([K150618](/device/K150618.md))
- PENTAX Medical EPK-i5010 Video Processor ([K143727](/device/K143727.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" and "ADMINISTRATION" in blue text.

October 12, 2018

PENTAX Medical of America, Inc. James W. Monroe Director, Regulatory Affairs 3 Paragon Drive Montvale, NJ 07645-1782

Re: K180292

Trade/Device Name: PENTAX Medical Video Upper GI Scope EG34-i10 Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: II Product Code: FDS Dated: September 14, 2018 Received: September 17, 2018

Dear James W. Monroe:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be avare that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see

https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Image /page/1/Picture/7 description: The image shows the logo for the Food and Drug Administration (FDA). The logo is in a light gray color and features the letters "FDA" in a stylized font. The logo is simple and recognizable, and it is often used to identify products that have been approved by the FDA.

Jeffrey W. Cooper -2 2018.10.12 18:28:33 -04'00'

for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K180292

Device Name

PENTAX Medical Video Upper GI Scope EG34-i10

Indications for Use (Describe)

This instrument is intended to be used with a PENTAX video processor (including light source), documentation equipment, monitor, Endotherapy Device such as a Biopsy Forceps, and other ancillary equipment for endoscopy and endoscopic surgery within the upper digestive tract including the esophagus, stomach, and duodenum.

| Type of Use (Select one or both, as applicable) |                                         |
|-------------------------------------------------|-----------------------------------------|
| Prescription Use (Part 21 CFR 801 Subpart D)    | <span style="font-size: 16px;">☒</span> |
| Over-The-Counter Use (21 CFR 801 Subpart C)     | <span style="font-size: 16px;">☐</span> |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

#### 510(k) Summary

#### SUBMITTER .

PENTAX of America, Inc., HOYA Corporation PENTAX Division 3 Paragon Drive Montvale, New Jersey 07645-1782 Phone: 800-431-5880 Ext 2129 Fax: 201-571-2340

Contact: James Monroe Date Prepared: January 31, 2018

#### II. DEVICE

The purpose of this 510(k) is to obtain clearance of the PENTAX Medical Video Upper GI Scope EG34-i10.

| Device Name          | PENTAX Medical Video Upper GI Scope EG34-i10 |
|----------------------|----------------------------------------------|
| Common Name          | Gastroscope and Accessories, Flexible/Rigid  |
| Classification Name  | Endoscope and accessories                    |
| Regulation No.       | 876.1500                                     |
| Device Class         | 2                                            |
| Product Code         | FDS                                          |
| Classification Panel | Gastroenterology/Urology                     |

Table 5.1: Regulatory Classification of PENTAX Medical Video Upper GI Scope EG34-i10.

#### III. PREDICATE DEVICES

The predicate device for this submission, PENTAX Video Upper GI Scopes EG29-i10 (K131902), is virtually identical to the PENTAX Medical Video Upper GI Scope EG34-i10. The subject and predicate devices differ slightly with regard to dimensional specifications and materials.

#### IV. DEVICE DESCRIPTION

The PENTAX Medical Video Upper GI Scope EG34-i10 is intended to provide optical visualization of (via a video monitor), and therapeutic access to, the upper gastrointestinal tract. The PENTAX Medical Video Upper GI Scope EG34-i10 is compatible with the PENTAX Medical EPK-i7010 Video Processor (K150618) and PENTAX Medical EPK-i5010 Video Processor (K143727).

The PENTAX Medical Video Upper GI Scope EG34-i10 is composed of three main components: an Insertion Portion, Control Body and PVE Connector. The Insertion is inserted into the body cavity of patient. The Insertion Portion includes the Distal End and Bending Section. The

{4}------------------------------------------------

Distal End of the Insertion Portion includes the Objective Lens, Light Guide, Instrument Channel, Air/Water Nozzle, and Water Jet Nozzle.

The Air/Water Feeding System is used to deliver the air and water to the Objective Lens from the Air/Water Nozzle. The Air/Water Feeding Valve is attached to the Air/Water Cylinder. When the hole at the top of Air/Water Feeding Valve is covered, the air is delivered. When the Air/Water Feeding Valve is pushed, the water is delivered. A Water Jet System is used to stream forward the sterile water from the Water Jet Nozzle.

The Control Body is held by the user's hand. The Control Body includes the Angulation Control Knob, Angulation Lock Knob/Lever, Air/Water Cylinder, Suction Cylinder, Remote Control Button, and Instrument Channel Inlet. The Bending Section is bent by the Angulation Control Knob to operate the endoscope angulation Lock Knob/Lever is used to adjust the rotation torque of the Angulation Control Knob. The Remote Button is used to operate the function of video processor and external device from the control body, as necessary.

The Inlet Seal is attached to the Instrument Channel Inlet. Endotherapy Device such as Biopsy Forceps may be inserted from the Instrument Channel Inlet into the body cavity through the instrument channel.

The PVE Connector is connected to the Video Processor via an Electrical Contacts. The Light Guide of the Distal End is used to illuminate the body cavity by light that is carried through the Light Carrying Bundle. The Light Carrying Bundle guides the light from Light Guide Plug that is connected to the Light Source inside the video processor. The CCD built into the Distal End receives reflected light (image data) from the body cavity, and sends the image data to the Video Processor through the video cable. The image data is converted into the image signal by the Video Processor, and the image inside the body cavity is displayed on the Monitor.

The PENTAX Medical Video Upper Gl Scope EG34-i10 is a reusable semi-critical device. Since it is packaged non-sterile, it must be high-level disinfected or sterilized before initial use. Prior to each subsequent procedure, it must be subjected to an appropriate cleaning and either highlevel disinfection or sterilization processes.

## V. INDICATIONS FOR USE

This instrument is intended to be used with a PENTAX video processor (including light source), documentation equipment, monitor, Endotherapy Device such as a Biopsy Forceps, and other ancillary equipment for endoscopy and endoscopic surgery within the upper digestive tract including the esophagus, stomach, and duodenum.

## VI. COMPARISON OF TECHNOLOGY CHARACTERISTICS WITH THE PREDICATE

The subject device is virtually identical to the predicate device. The subject device has the same fundamental technology and operating principles of the predicate device, including the same intended use and design technological characteristics, such as Insertion Portion, Control Body and optical characteristics. The minor differences between the subject and predicate devices

{5}------------------------------------------------

exist with regard to dimensional specifications and materials. These modifications do not impact the intended use and do not raise different questions of safety and effectiveness.

#### VII. PERFORMANCE DATA

#### a. Reprocessing Validation

The PENTAX Medical PENTAX Medical Video Upper GI Scope EG34-i10 is reusable semi-critical devices. Since they are packaged non-sterile, they must be high-level disinfected or sterilized before initial use. Prior to each subsequent procedure, they must be subjected to an appropriate cleaning and either high-level disinfection or sterilization processes.

The following testing has been performed for the PENTAX Medical Video Upper GI Scope EG34i10:

- Soil Accumulation Study
- Cleaning Efficacy Study
- High Level Disinfection (HLD) Efficacy Study
- Rinsing Validation
- . Sterilization Validation

All acceptance criteria were satisfied.

#### b. Biocompatibility

The biocompatibility evaluation of the patient contacting materials of the PENTAX Medical Video Upper GI Scope EG34-i10 was conducted in accordance with "Guidance for Industry and Food and Drug Administration Staff: Use of International Standard ISO 10993-1, 'Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process' (June 16, 2016)" with passing results.

#### Electrical Safety and Electromagnetic Compatibility (EMC) ﻥ

The PENTAX Medical Video Upper GI Scope EG34-i10 complies with the following standards for electrical safety and EMC.

## Electrical Safety

- IEC 60601-1: 2005 + CORR. 1:2006 + CORR. 2:2007 + AM1:2012
- IEC 60601-2-18:2009
- ANSI/AAMI ES60601-1:2005 + A2:2010 + A1:2012

## Electromagnetic Capability

- IEC 60601-1-2:2007/EN 60601-1-2:2007
## d. Software Verification and Validation

The subject devices utilize the same software as the predicate device.

#### e. Performance Testing

## Optical Testing

{6}------------------------------------------------

PENTAX Medical completed optical testing of the PENTAX Medical Video Upper GI Scope EG34i10 compared to the predicate, PENTAX Medical Video Upper Gl Scope EG29-i10. The scopes were evaluated for Light Distribution, Spectral Distribution, Total Luminous Flux, and Photobiological Safety. Testing was conducted using both the PENTAX Medical EPK-i5010 and EPK-i7010 Video Processors. The results of the testing indicate that the light distributions, intensity and wavelength, and total luminous flux of the PENTAX Medical Video Upper GI Scope EG34-i10 were comparable to the PENTAX Medical Video Upper GI Scope EG29-i10.

The photobiological safety of the lamps and lamp systems was evaluated in accordance with IEC 62471, Photobiological safety of lamps and lamp systems. Testing revealed that IEC 62471 evaluation values of all subject and predicate device combinations are risk group 2 or lower, and the safety of the subject device combinations are considered to be equivalent to the predicate device combinations.

#### Usability Testing

A comparison of the usability of the NTAX Medical Video Upper GI Scope EG34-i10 to PENTAX Medical i10 and k10 series of scopes was conducted. The i10 series of scopes include the predicate device, EG29-i10. Usability testing of various operating conditions was performed which demonstrated that there is no difference in all operations and their usability is equivalent.

#### VIII. CONCLUSION

The subject device is virtually identical to the predicate device with only minor changes to materials and dimensional specifications. The intended use of the subject and predicate devices are identical. Any differences in technological characteristics between the subject and predicate devices do not raise different questions of safety and effectiveness.

The data submitted support the performance and reprocessing for the PENTAX Medical Video Upper GI Scope EG34-i10 and demonstrate that the scope performs as intended and can be reprocessed in a safe and effective manner. The data provided in this 510(k) Premarket Notification support the equivalence of the subject and predicate devices.

---

**Source:** [https://fda-staging.innolitics.com/device/K180292](https://fda-staging.innolitics.com/device/K180292)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
