The Cadwell Apollo System is indicated for prescription use to acquire, record, transmit, and display physiological and environmental data for electroencephalographic (EEG) and polysomnographic (PSG) ambulatory and/or clinical studies of patients of all ages.
Device Story
Cadwell Apollo System acquires, records, transmits, and displays physiological/environmental data for EEG/PSG studies; used in home and clinical environments. System components: recorder, 32/64-channel amplifiers, patient event button, microphone, cables; optional photic stimulator, video camera, cart. Operates via Arc acquisition software; supports wired (USB/Ethernet) or wireless connectivity to PC. Healthcare providers use Arc Sentinel software at central nurse stations for monitoring multiple records. Device provides high-fidelity signal acquisition (1 MHz sampling, >110 dB CMRR) to assist clinicians in diagnostic EEG/PSG interpretation. Benefits include extended recording duration (up to 96 hours) and improved signal quality for patient monitoring.
Clinical Evidence
Bench testing only. No clinical data provided. Performance verified against IEC 60601-2-26 (EEG safety/performance), IEC 60601-1 (general safety), IEC 60601-1-11 (home healthcare), and IEC 60601-1-2 (EMC). Software verified and validated per IEC 62304 and FDA guidance.
Technological Characteristics
System includes recorder, 32/64-channel amplifiers, photic stimulator. Connectivity: USB, Ethernet, wireless. Power: 3.8V 3880 mAh Li-ion battery or USB. Sampling: 1 MHz. CMRR: >110 dB. Software: Arc acquisition software (Moderate level of concern). Standards: IEC 60601-1, IEC 60601-1-11, IEC 60601-1-2, IEC 60601-2-26, IEC 62304, ISO 15004-2.
Indications for Use
Indicated for prescription use in patients of all ages requiring EEG or PSG ambulatory or clinical studies to acquire, record, transmit, and display physiological and environmental data.
Regulatory Classification
Identification
An electroencephalograph is a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text. The word "Administration" is on the second line.
July 17, 2018
Cadwell Industries, Inc. Alison Hull EEG Product Manager 909 North Kellogg Street Kennewick, Washington 99336
Re: K180269
Trade/Device Name: Cadwell Apollo System Regulation Number: 21 CFR 882.1400 Regulation Name: Electroencephalograph Regulatory Class: Class II Product Code: GWQ, GWE, GWL, OLT, OLV, OMC Dated: April 17, 2018 Received: April 18, 2018
Dear Alison Hull:
We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Jay R. Gupta -S
For Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K180269
Device Name Cadwell Apollo System
Indications for Use (Describe)
The Cadwell Apollo System is indicated for prescription use to acquire, record, transmit, and display physiological and environmental data for electroencephic (EEG) and polysomnographic (PSG) ambulatory and/or clinical studies of patients of all ages.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## K180269 510(k) Summary
| Submitter's Name<br>and Address | Cadwell Industries, Inc.<br>909 North Kellogg Street<br>Kennewick, WA 99336 |
|---------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person | Ms. Alison Hull<br>Phone: +1 (800) 245-3001, Extension 210<br>Email: alisonh@cadwell.com |
| Date Summary<br>Prepared | July 6, 2018 |
| Trade Name | Cadwell Apollo System |
| Common Name | Electroencephalograph (EEG) |
| Classification Name<br>& Product Code | <b>PRIMARY</b><br>Class II<br>21 CFR 882.1400 GWQ Full-montage Standard<br>Electroencephalograph<br><br><b>SUBSEQUENT</b><br>21 CFR 882.1400 OLT Non-normalizing Quantitative<br>Electroencephalograph Software<br>21 CFR 882.1400 OLV Standard Polysomnograph (PSG) with<br>Electroencephalograph<br>21 CFR 882.1400 OMC Reduced-Montage Standard<br>Electroencephalograph<br>21 CFR 882.1890 GWE Evoked Response Photic Stimulator<br>21 CFR 882.1835 GWL Physiological Signal Amplifier |
| Predicate Device | K133355 Cadwell Flex EEG/PSG<br>21 CFR 882.1400; GWQ<br><br>Subsequent<br>Product Codes GWE, GWL OLT, OLV, OMC |
{4}------------------------------------------------
| Device Description | The Cadwell Apollo System (Apollo) is used to acquire, record, transmit, and display physiological and environmental data for ambulatory electroencephalograph (EEG) and polysomnography (PSG) and/or clinical studies of patients of all ages.<br><br>The Apollo system consists of a Recorder, a 64-channel and/or 32-channel Amplifier, a patient event button, a microphone, and various connecting cables. Optional accessories include the Arc photic stimulator, a video recording camera, and a cart for the equipment. Previously FDA cleared accessories used with the Apollo system consist of various electrodes, leads, and cables.<br><br>The Apollo system utilizes Cadwell Arc acquisition software (previously cleared in K133355) with support for Apollo hardware using single or combinations of amplifiers, and a photic stimulator.<br><br>Additional channels can be added with multiple amplifiers. Apollo is intended for use in both home healthcare and professional healthcare environments. | | |
|--------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|----------------------------------------------|
| Indications for Use | The Cadwell Apollo System is indicated for prescription use to acquire, record, transmit, and display physiological and environmental data for electroencephalographic (EEG) and polysomnographic (PSG) ambulatory and/or clinical studies of patients of all ages. | | |
| Comparison of<br>Technological<br>Characteristics to<br>Predicate Device | Characteristic | K133355<br>Predicate Device | K180269<br>Subject Device |
| | Intended Patient<br>Population | All ages | Same |
| | System<br>Configuration | Computer based equipment with<br>dedicated hardware<br>peripherals/components | Same |
| | Recording<br>Modality | Attended and unattended | Same |
| | Intended<br>Environment of<br>Use | Hospital or home | Same |
| | Recorder to<br>Personal<br>Computer (PC)<br>Connectivity/<br>Networking | Wired via an Ethernet cable | Wired via a USB cable |
| | Recorder to<br>Amplifier<br>Connectivity | Wired via a cable | Same |
| | Amplifiers<br>Available | 32 channel amplifier | 32 channel amplifier<br>64 channel amplifier |
| | Sampling Rage | 3200 Hz | 1 MHz |
| | Other Inputs | 8 active/ reference, 2 other | Up to 10 active/reference, 2 other |
| | Impedance<br>Check | Yes | Yes |
| | Number of<br>Amplifiers that<br>can Connect to | One (1) | Two (2) |
| Continued: Comparison of Technological Characteristics | | | |
| Characteristic | K133355<br>Predicate Device | K180269<br>Subject Device | |
| Power Source | Rechargeable Batteries (D Alkaline)<br>or USB powered | Rechargeable Batteries (3.8V 3880<br>mAh Lithium Ion Battery) or USB<br>powered | |
| Acquisition<br>Software | Arc | Same | |
| Arc Sentinel<br>Software, a<br>Central Nurse<br>Station | No | Yes; for viewing and monitoring<br>multiple data records | |
| Remote<br>Monitoring | Yes | Same | |
| Photic Flash<br>Rate | 1 to 60 Hz | Same | |
| Photic Interface | USB | Same | |
| Microphone<br>Input Ability | Yes | Yes | |
| Wireless<br>Communication<br>Capability | No | Yes; wireless<br>communication capability<br>between Recorder and PC<br>with Arc acquisition<br>software. | |
| CMMR | > 92 dB | Improved; > 110 dB | |
| Recording<br>Duration | > 48 hours per 2 D batteries | Up to 96 hours on battery;<br>Unlimited on battery | |
| Storage Rate | 250 Hz | 16 kHz | |
| Other Inputs | Patient Event Button<br>Patient Microphone<br>Q-Video Mobile 2 (in-home<br>recordings) | Patient Event Button<br>Patient Microphone<br>Q-Video Mobile 3 (in-home<br>recordings) | |
| Cameras | IP and USB Cameras | IP and USB Cameras | |
{5}------------------------------------------------
{6}------------------------------------------------
| Summary of Nonclinical Testing to Demonstrate Substantial Equivalence | |
|-----------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Software | The Arc acquisition software corresponds to a MODERATE level of<br>concern. Software was designed and developed per a robust software<br>development process, and was verified and validated. Software<br>information is provided in accordance with internal requirements, the<br>following guidance documents, and IEC software standard: |
| | • FDA guidance: The content of premarket submissions for software<br>contained in medical devices, issued May 11, 2005. |
| | • FDA guidance: Off-the-shelf software use in medical devices, issued<br>September 09, 1999. |
| | • FDA guidance: General principles of software validation; Final<br>guidance for industry and FDA staff, issued January 02, 2011. |
| | • FDA guidance: Content of premarket submissions for management of<br>cybersecurity in medical devices, October 02, 2014. |
| | • IEC 62304: 2006, Medical device software - Software life cycle<br>processes |
| | Test results indicate that the Arc acquisition software conforms to<br>predetermined specifications and the applicable software guidance<br>documents. |
| Electrical Safety | The Apollo System was tested for performance in accordance with the<br>following standard: |
| | • IEC 60601-1: 2005, Medical electrical equipment – Part 1: General<br>requirements for basic safety and essential performance. |
| | • IEC 60601-1-11: 2010, Medical electrical equipment – Part 1-11:<br>General requirements for basic safety and essential performance –<br>Collateral Standard: Requirements for medical electrical equipment<br>and medical electrical systems used in the home healthcare<br>environment |
| | • ISO 15004-2: 2007, Ophthalmic Instruments – Fundamental<br>requirements and test methods – Part 2: Light hazard protection |
| | Test results indicate that the Apollo System conforms to the above<br>standards. |
| Electromagnetic Compatibility | The Apollo was tested for performance in accordance with the following<br>standard: |
| | • IEC 60601-1-2: 2007, Medical electrical equipment – Part 1-2:<br>General requirements for basic safety and essential performance –<br>Collateral standard: Electromagnetic disturbances – Requirements<br>and tests. |
| | Test results demonstrate that the Apollo conforms to the above standard. |
| Performance<br>Testing – Bench | The Apollo was tested for performance in accordance with internal<br>requirements and the following standards:<br>IEC 60601-2-26: 2012, Medical electrical equipment – Part 2-26:<br>Particular requirements for the basic safety and essential<br>performance of electroencephalographs IEC 60601-1-6: 2010, Medical electrical equipment - Part 1-6:<br>General requirements for basic safety and essential performance -<br>Collateral standard: Usability IEC 62366: 2007, Medical devices – Application of usability<br>engineering to medical devices. Test results indicate that the Apollo conforms to its predetermined<br>specifications and the above standards. |
| Conclusion | The results of the aforementioned performance data demonstrate that the<br>Cadwell Apollo System is substantially equivalent to the Cadwell Flex<br>EEG/PSG. |
T
{7}------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.