K180262 · Sicat GmbH & Co. KG · LLZ · Mar 12, 2018 · Radiology
Device Facts
Record ID
K180262
Device Name
SICAT Endo
Applicant
Sicat GmbH & Co. KG
Product Code
LLZ · Radiology
Decision Date
Mar 12, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
Aiding diagnosis in the oral-maxillofacial region Aiding comparisons of different treatment options Aiding endodontic treatment planning Aiding treatment planning for endodontic surgical guides
Device Story
SICAT Endo is a software-only device for viewing/analyzing medical 3D volume data (CBCT/CT), intraoral images, and 3D optical surface data (optical impressions). Used by dental professionals in clinics to visualize oral-maxillofacial anatomy; identify/mark root canals; define drill canals; and plan endodontic surgical guides. The software provides tools for 3D volume rendering, orthogonal slices, panoramic views, and curved planar reformation (CPR). It allows registration of optical surface data to volume data using user-specified reference points. Output consists of visualized anatomical data and exported planning data for use in CAD or rapid prototyping systems for surgical guide fabrication. By enabling precise 3D visualization and drill path planning, the device assists clinicians in endodontic treatment planning and surgical guide design, potentially improving treatment predictability and quality.
Clinical Evidence
Bench testing only. Verification and validation activities included design reviews, unit tests, code reviews, usability tests, integration tests, and system verification. Special bench testing verified endodontic planning visualization quality and quantitative accuracy of root canal treatment planning. The device passed all verification and validation activities.
Technological Characteristics
Pure software device; runs on Windows 7/8/10 (64-bit). Inputs: DICOM (CBCT/CT), STL, proprietary SSI/SIXD containers. Visualization: 3D volume/surface rendering, orthogonal slices, panoramic views, CPR. Connectivity: Standalone or integrated as SIDEXIS XG/4 module. Software lifecycle: AAMI/ANSI/IEC 62304 compliant. Accuracy: 100 μm length, 1-degree angular measurement.
Indications for Use
Indicated for aiding diagnosis, treatment planning, and comparison of treatment options in the oral-maxillofacial region, specifically for endodontic treatment and endodontic surgical guide planning.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
March 12, 2018
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
SICAT GmbH & Co. KG % Frederik Kunze Quality Management and Regulatory Affairs Brunnenallee 6 Bonn, NRW 53177 GERMANY
Re: K180262
Trade/Device Name: SICAT Endo Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: January 29, 2018 Received: January 30, 2018
Dear Frederik Kunze:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael D. O'Hara For
Robert Ochs, Ph.D Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K180262
Device Name SICAT Endo
Indications for Use (Describe)
- Aiding diagnosis in the oral-maxillofacial region
- Aiding comparisons of different treatment options
- · Aiding endodontic treatment planning
- · Aiding treatment planning for endodontic surgical guides
Type of Use (Select one or both, as applicable)
| | Prescription Use (Part 21 CFR 801 Subpart D) |
|--|----------------------------------------------|
| | Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### Content and format as required by section 21 CFR 807.92
( http://www.fda.gov/ MedicalDevices/ DeviceRequlationandGuidance/ HowtoMarket YourDevice/PremarketSubmissions/PremarketNotification510k/ucm142651.htm)
### 1. SUBMITTER/510(K) HOLDER
SICAT GmbH & Co. KG Brunnenallee 6 53177 Bonn Germany
| Establishment<br>Registration Number: | 3006098230 |
|---------------------------------------|------------------------|
| Telephone: | +49 (228) 854697-82 |
| Facsimile: | +49 (228) 854697-99 |
| Primary Contact: | Mr. Dr. Manfred Breuer |
| Secondary Contact: | Mr. Frederik Kunze |
| Date Prepared: | January 29th, 2018 |
### 2. DEVICE NAME AND DEVICE CLASSIFICATION
| Proprietary Name: | SICAT Endo |
|-------------------------|--------------------------------------------------------------|
| Common/Usual Name: | Radiological Visualization Software for Diagnosis and Dental |
| | Treatment Planning |
| Classification Name: | System, Image Processing, Radiological |
| Regulation Description: | Picture archiving and communications system |
| Product Code: | LLZ |
| Regulation Number: | 21 CFR 892.2050 |
| Classification Class: | Class II Product |
### 3. PREDICATE DEVICES
- SICAT Implant (K103723) (Primary predicate device) .
- 3D Endo™ Software (K171115) .
### 4. DEVICE DESCRIPTION
SICAT Endo is a pure software device.
SICAT Endo is a software tool intended for viewing and analyzing medical information:
{4}------------------------------------------------
- 5 510(k) Summary for SICAT Endo
- medical 3D volume data such as volumetric X-ray data from Cone Beam CT (CBCT) and CT scanners, and
- . intraoral images, and
- . 3D optical surface data like optical impression data from optical scanners, and
SICAT Endo provides tools for analyzing the root canal and to mark it visually. It allows to define a drill canal and ordering of a corresponding surgical quide with a drill sleeve.
### 5. Indications for Use
Indications for use for SICAT Endo are:
- Aiding diagnosis in the oral-maxillofacial region
- . Aiding comparisons of different treatment options
- . Aiding endodontic treatment planning
- Aiding treatment planning for endodontic surgical quides .
### 6. Device Comparison Table
The following table shows a summary of the intended use, technological characteristics, design and function of SICAT Endo and the predicate devices. The predicate device SICAT Implant is the primary predicate device.
| | SICAT Endo | 3D Endo, Dentsply<br>Sirona<br>(marketed in EU /<br>USA cleared under<br>510(k) K171115) | SICAT<br>Implant,<br>SICAT<br>(marketed in<br>EU / USA<br>cleared under<br>510(k) number<br>K103723) | | | |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|------------------------------------------------------------------------------------------------------|------------|
| | Common / Usual Name | | | | | |
| | Radiological<br>Visualization<br>Software for<br>Diagnosis and<br>Endodontic<br>Treatment Planning | Radiological<br>Visualization Software<br>for Diagnosis and<br>Endodontic Treatment<br>and Re-treatment<br>Planning | Radiological<br>Visualization<br>Software for<br>Diagnosis and<br>Dental Implant<br>Planning | | | |
| Device Description | | | | | | |
| General Description | SICAT Endo is a<br>pure software<br>device.<br>SICAT Endo is a<br>software application<br>for visualization of<br>imaging information<br>of the oral-<br>maxillofacial region.<br>The imaging data | 3D Endo™ is a CBCT<br>based software device<br>designed to improve<br>endodontic treatment<br>planning for more<br>predictability.<br>It is designed to<br>improve treatment<br>quality: to isolate the<br>tooth of interest, to | SICAT Implant<br>V1.2 is a pure<br>software device.<br>SICAT Implant<br>V1.2 is a<br>software<br>application for<br>the visualization<br>of imaging<br>information of | | | |
| SICAT Endo | 3D Endo, Dentsply<br>Sirona<br>(marketed in EU /<br>USA cleared under<br>510(k) K171115) | SICAT<br>Implant,<br>SICAT<br>(marketed in<br>EU / USA<br>cleared under<br>510(k) number<br>K103723) | | | | |
| originates from<br>medical scanners<br>such as CT or CBCT<br>scanners. It is also<br>used as a software<br>system to aid<br>medical<br>professionals with<br>the planning, the<br>evaluation and the<br>comparison of<br>treatment options<br>and the planning of<br>endodontic accesses<br>for an endodontic<br>treatment. The<br>medical<br>professionals'<br>planning data may<br>be exported from<br>SICAT Endo and<br>used for the<br>realization of the<br>planned therapy. | clearly visualize the<br>tooth anatomy in 3D,<br>to identify all canals<br>and to anticipate<br>areas.<br>The use of 3D Endo<br>provides:<br>- Diagnosis and<br>pathology<br>- 3D tooth anatomy<br>- Canal system<br>- 3D canal anatomy<br>- Treatment plan<br>The Wizard allows to:<br>· Diagnose the case;<br>• Isolate the tooth of<br>interest;<br>· Investigate the<br>canal system;<br>• Understand the 3D<br>Canal anatomy;<br>· Create a Treatment<br>plan.<br>The use of 3D Endo<br>allows to:<br>· evaluate 3D working<br>length and cavity<br>access depth<br>• locate canal orifices<br>without opening the<br>tooth<br>• plan for an optimal<br>cavity access and your<br>final instrument<br>· investigate canals'<br>irregularities following<br>their curvatures<br>In addition, the<br>simulated Intra-Root<br>Camera View Visualize | the oral-maxillo-<br>facial region. The<br>imaging data<br>originates from<br>medical scanners<br>such as CT or<br>DVT scanners.<br>SICAT Implant<br>V1.2 is intended<br>for use as<br>planning and<br>simulation<br>software to aid<br>qualified dental<br>professionals in<br>the placement of<br>dental implants<br>and the planning<br>of surgical<br>treatments.<br>SICAT Implant<br>V1.2 allows to<br>name, position,<br>move, rotate,<br>resize and<br>visualize dental<br>implants and<br>other planning<br>objects (i.e.<br>nerve canals)<br>within the<br>visualized 3D<br>volume. Thus,<br>dental<br>professionals like<br>implantologists<br>are enabled to<br>precisely plan<br>the positions,<br>orientations,<br>types and sizes<br>of implants to be<br>placed in the<br>patient's<br>mandible/maxilla | | | | |
| | SICAT Endo | 3D Endo, Dentsply<br>Sirona<br>(marketed in EU /<br>USA cleared under<br>510(k) K171115) | SICAT<br>Implant,<br>SICAT<br>(marketed in<br>EU / USA<br>cleared under<br>510(k) number<br>K103723) | | | |
| | | throughout the canal<br>the theoretical cutting<br>envelope of the<br>Dentsply Sirona<br>Endodontics<br>instruments. | together with the<br>related surgical<br>procedures.<br>The dental<br>professionals'<br>planning data<br>may be exported<br>from SICAT<br>Implant V1.2 and<br>used as input<br>data for CAD or<br>Rapid<br>Prototyping<br>Systems. | | | |
| Indications for Use | | | | | | |
| Indications for<br>Use | Aiding diagnosis in<br>the oral-maxillofacial<br>region<br>Aiding comparisons<br>of different<br>treatment options<br>Aiding endodontic<br>treatment planning<br>Aiding treatment<br>planning for<br>endodontic surgical<br>guides | 3D Endo™ Software is<br>intended to aid in the<br>visualization,<br>diagnosis and<br>planning of<br>endodontic treatment<br>and re-treatment<br>cases utilizing DICOM<br>images. | SICAT Implant is<br>a software<br>application for<br>the visualization<br>of imaging<br>information of<br>the oral-<br>maxillofacial<br>region. The<br>imaging data<br>originates from<br>medical scanners<br>such as CT or<br>DVT scanners.<br>SICAT Implant is<br>intended for use<br>as planning and<br>simulation<br>software to aid<br>qualified dental<br>professionals in<br>the placement of<br>dental implants<br>and the planning<br>of surgical<br>treatments. The<br>dental<br>professionals | | | |
| | SICAT Endo | 3D Endo, Dentsply<br>Sirona<br>(marketed in EU /<br>USA cleared under<br>510(k) K171115) | SICAT<br>Implant,<br>SICAT<br>(marketed in<br>EU / USA<br>cleared under<br>510(k) number<br>K103723) | | | |
| | | | planning data<br>may be exported<br>from SICAT<br>Implant and<br>used as input<br>data for CAD or<br>Rapid<br>Prototyping<br>Systems. | | | |
| Contraindications | None | Do not use 3D Endo™<br>Software for implants<br>or any other dental<br>procedure outside<br>endodontics.<br>There are no known<br>serious adverse<br>reactions or potential<br>safety hazards. | None | | | |
| | Medical Data Viewing | | | | | |
| Data types<br>visualized | 3D volume data, 3D<br>surface data (optical<br>impressions), 2D<br>radiographs | 2D and 3D visualized<br>CBCT data | 3D volume data,<br>3D surface data<br>(optical<br>impressions,<br>implant objects) | | | |
| Imaging data<br>visualization<br>region | Oral-maxillofacial<br>region | Oral-maxillofacial<br>region | Oral-maxillofacial<br>region | | | |
| | Viewing Modes typically used for dental procedures | | | | | |
| *Orthogonal<br>slices | Axial, coronal,<br>sagittal within the<br>Endoline wizard | Axial | Axial, coronal,<br>sagittal | | | |
| *Panoramic view | Panoramic view (ray<br>sum) based on a<br>panoramic curve.<br>Panoramic curve can<br>be manually<br>adjusted. | No | Panoramic view<br>(ray sum) based<br>on a panoramic<br>curve.<br>Panoramic curve<br>can be manually<br>adjusted. | | | |
| Curved Planar<br>Reformation | Curved planar<br>reformation based | Yes | None | | | |
| | SICAT Endo | 3D Endo, Dentsply<br>Sirona<br>(marketed in EU /<br>USA cleared under<br>510(k) K171115) | SICAT<br>Implant,<br>SICAT<br>(marketed in<br>EU / USA<br>cleared under<br>510(k) number<br>K103723) | | | |
| (CPR) also called<br>Multi Planar<br>Reformation<br>(MPR) that is<br>less precise | on a spline<br>positioned within the<br>volume. By<br>positioning the<br>spline the view is<br>adjusted.<br>No equidistant scale. | | | | | |
| *Panoramic slice | Movable "Panoramic<br>Slicing Window"<br>embedded in<br>panoramic view | No | Movable<br>"Panoramic<br>Slicing Window"<br>embedded in<br>panoramic view | | | |
| *TSA and LSA | Transversal slice<br>(TSA) and<br>longitudinal slice<br>(LSA), both with<br>respect to<br>panoramic curve | Transversal slice | Transversal slice<br>(TSA) and<br>longitudinal slice<br>(LSA), both with<br>respect to<br>panoramic curve | | | |
| *3D volume<br>rendering | Yes…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.