← Product Code [HRS](/productcode/HRS) · K180239

# Additive Orthopaedics Patient Specific 3D Printed Bone Segments (K180239)

_Additive Orthopaedics, LLC · HRS · May 16, 2018 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K180239

## Device Facts

- **Applicant:** Additive Orthopaedics, LLC
- **Product Code:** [HRS](/productcode/HRS.md)
- **Decision Date:** May 16, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3030
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Additive Orthopaedics 3D Printed Bone Segments are intended to be used for internal bone fixation for bone fractures or osteotomies in the ankle and foot, such as: *Cotton (opening wedge) osteotomies of the medial cuneiform *Evans lengthening osteotomies The Additive Orthopaedics 3D Printed Bone Segments are intended for use with ancillary plating fixation. The Additive Orthopaedics 3D Printed Bone Segments are not intended for use in the spine. It is a patient specific device.

## Device Story

Patient-specific bone segments; additively manufactured from medical-grade titanium alloy (Ti-6AL-4V ELI). Input: patient CT image data. Process: surgeon-approved design based on patient anatomy; 3D printing manufacturing. Output: custom-fit bone wedge for internal fixation. Used in foot/ankle surgery; operated by orthopedic surgeons. Surgeon reviews engineering drawings prior to fabrication to ensure design requirements are met. Benefits: provides patient-specific anatomical fit for bone fixation procedures.

## Clinical Evidence

No clinical data. Substantial equivalence supported by non-clinical bench testing, including morphological characterization, mechanical testing (friction, roughness, durability/abrasion, compressive fatigue), and biocompatibility testing.

## Technological Characteristics

Material: Medical grade titanium alloy (Ti-6AL-4V ELI). Manufacturing: Additive manufacturing (3D printing). Design: Patient-specific geometry derived from CT data. Fixation: Intended for use with ancillary plating. Class II device per 21 CFR 888.3030.

## Regulatory Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Bone Wedge System ([K153207](/device/K153207.md))

## Reference Devices

- Patient Specific Cranial Plate ([K151692](/device/K151692.md))
- TruMatch CMF Titanium 3D Printed Implant System ([K170272](/device/K170272.md))
- Mimics ([K073468](/device/K073468.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the U.S. Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is in blue and includes the words "FDA U.S. FOOD & DRUG ADMINISTRATION".

Additive Orthopaedics, LLC Gregory Kowalczyk President 83 Amelia Circle Little Silver. New Jersev 07739 May 16, 2018

Re: K180239

Trade/Device Name: Additive Orthopaedics Patient Specific 3D Printed Bone Segments Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: HRS, HWC Dated: April 13, 2018 Received: April 16, 2018

Dear Gregory Kowalczyk:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

#### 510(k) Number (if known) K180239

Device Name

Additive Orthopaedics Patient Specific 3D Printed Bone Segments

Indications for Use (Describe)

The Additive Orthopaedics 3D Printed Bone Segments are intended to be used for internal bone fixation for bone fractures or osteotomies in the ankle and foot, such as:

*Cotton (opening wedge) osteotomies of the medial cuneiform *Evans lengthening osteotomies

The Additive Orthopaedics 3D Printed Bone Segments are intended for use with ancillary

plating fixation. The Additive Orthopaedics 3D Printed Bone Segments are not intended for

use in the spine.

It is a patient specific device.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510K Summary (Per 21 CFR 807.92)

| General Company Information:        |                                                                                                                                                                                                                                        | Additive Orthopaedics, LLC.<br>Gregory Kowalczyk<br>President<br>83 Amelia Circle<br>Little Silver, NJ 07739<br>Phone: (732) 882-6633<br>greg@additiveortho.com |  |
|-------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Date Prepared:                      |                                                                                                                                                                                                                                        | April 12, 2018                                                                                                                                                  |  |
| General Device Information:         |                                                                                                                                                                                                                                        |                                                                                                                                                                 |  |
| Proprietary Name:                   | Additive Orthopaedics Patient Specific 3D Printed Bone Segments                                                                                                                                                                        |                                                                                                                                                                 |  |
| System Classification:              | Common Name: Bone Wedge System<br>Regulation Name: Single/multiple component metallic bone<br>fixation appliances and accessories;<br>Product Code: HRS- Class II, HWC- Class II<br>Classification Name and Reference: 21 CFR 888.3030 |                                                                                                                                                                 |  |
| Predicate Devices:                  |                                                                                                                                                                                                                                        |                                                                                                                                                                 |  |
| Company                             | Product Name                                                                                                                                                                                                                           | 510K Number                                                                                                                                                     |  |
| Primary: Additive Orthopaedics, LLC | Bone Wedge System                                                                                                                                                                                                                      | K153207                                                                                                                                                         |  |
| Reference: BioArchitects            | Patient Specific Cranial Plate                                                                                                                                                                                                         | K151692                                                                                                                                                         |  |
| Materialise N.V.                    | TruMatch CMF Titanium 3D<br>Printed Implant System                                                                                                                                                                                     | K170272                                                                                                                                                         |  |
| Materialise N.V.                    | Mimics                                                                                                                                                                                                                                 | K073468                                                                                                                                                         |  |

#### Description

The Additive Orthopaedics Patient Specific 3D Printed Bone Segments is a simple one piece device constructed individually for each patient using CT image data. It is intended to be used for internal bone fixation for bone fractures or osteotomies in the foot and ankle. The

{4}------------------------------------------------

segments are additively manufactured from medical grade titanium alloy (Ti-6AL-4V Eli). It is a patient specific device. The bone segments come in a variety of configurations that depend on the geometry of the application. The surgeon approves the design of the 3D Printed Bone Segments by comparing his/her design requirements to engineering drawings prior to the construction of the implant device.

#### Intended Use (Indications)

The Additive Orthopaedics 3D Printed Bone Segments are intended to be used for internal bone fixation for bone fractures or osteotomies in the ankle and foot, such as:

*Cotton (opening wedge) osteotomies of the medial cuneiform *Evans lengthening osteotomies

The Additive Orthopaedics 3D Printed Bone Segments are intended for use with ancillary plating fixation.

The Additive Orthopaedics 3D Printed Bone Segments are not intended for use in the spine.

It is a patient specific device.

## (a) (6) Technological Characteristics Comparison

The Additive Orthopaedics Patient Specific Bone Segments and the legally marketed predicate device have nearly identical indications, identical materials and identical manufacturing processes. The dimensions and geometry of the subject device, however, are patient-specific. The dimensions and geometry of the Additive Orthopaedics Patient Specific Bone Segments fall within the range of sizes already claimed for the predicate device, are developed in a process equivalent to the reference device, and are verified by the surgeon.

## (b) (1) Substantial Equivalence- Non-Clinical Evidence

Morphological characterization, mechanical testing (including, friction, roughness, durability/abrasion, and compressive fatigue) as well as biocompatibility testing were performed on the predicate device and compared to the Additive Orthopaedics Patient Specific Bone Segments. The results of these comparison activities demonstrate that the Additive Orthopaedics Patient Specific Bone Segments are identical to the predicate device identified.

## (b) (3) Conclusion- Substantial Equivalence

The Additive Orthopaedics Patient Specific Bone Segments possess the same technologic characteristics of the predicate devices. These characteristics include the intended use, basic design, material, size and fundamental technology. The design characteristics of the subject system do not raise any new types of questions of safety and effectiveness. From the evidence submitted in this 510(k), the subject devices can be expected to perform at least as well as the predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K180239](https://fda-staging.innolitics.com/device/K180239)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).