K180217 · Vestagen Protective Technologies, Inc. · QBW · May 18, 2018 · General, Plastic Surgery
Device Facts
Record ID
K180217
Device Name
VESTEX Apparel
Applicant
Vestagen Protective Technologies, Inc.
Product Code
QBW · General, Plastic Surgery
Decision Date
May 18, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4040
Device Class
Class 2
Attributes
Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K180217 · May 18, 2018
VESTEX Apparel
Vestagen Protective Technologies, Inc.
Published clinical literature (Bearman et al., 2012) reporting on a crossover trial in a medical ICU
The study was used to demonstrate the effectiveness of VESTEX apparel in reducing the accumulation of MRSA on healthcare worker clothing in a clinical setting compared to control clothing.
A Crossover Trial of Antimicrobial Scrubs to Reduce Methicillin-Resistant Staphylococcus aureus Burden on Healthcare Worker Apparel (Bearman, GML, et al., 2012); Randomized, blinded, crossover trial; Follow-up/Duration: 12-hour work shift
Healthcare workers in a medical ICU; Number of Sites: 1
Control clothing
Reduction of MRSA accumulation on healthcare worker clothing
Indications for Use
VESTEX® Apparel ("VESTEX"), scrub tops, pants, skirts, and lab coats are apparel that provides fluid repellency and reduces the retention of Methicillin Resistant Staphylococcus aureus (MRSA) on the surface (outer layer) of the apparel. The ability of VESTEX to reduce the retention of MRSA on the fabric surface has not been shown to correlate with the reduction of infections. Clinical studies to evaluate reduction in infection have not been performed for this device. VESTEX Apparel is not intended to replace personal protective equipment.
Device Story
VESTEX Apparel is nonsterile, reusable, fluid-repellent, antimicrobial clothing for healthcare workers; designed for continuous wear to protect against intermittent/unexpected exposure to blood, body fluids, and potentially infectious materials. Fabric features dual-action surface: fluid-repellent barrier to resist bacteria acquisition and antimicrobial agent to disrupt cell membranes/proteins/bio-energetics. Interior side is moisture-wicking. Device does not replace PPE for expected exposure episodes. Healthcare providers wear apparel during clinical duties; output is the physical barrier/antimicrobial surface reducing MRSA burden on clothing. Benefits include reduced pathogen retention on fabric surface, though clinical correlation to infection reduction is not established.
Clinical Evidence
Clinical evidence includes a randomized, blinded, crossover trial (Bearman et al., 2012) in a medical ICU. Study compared VESTEX apparel to control clothing on healthcare workers. Results showed statistically significant reduction in MRSA accumulation on VESTEX apparel at both the beginning and end of 12-hour shifts. Bench testing included biocompatibility (ISO 10993-1:2009), fluid repellency (AATCC22), and splatter testing for MRSA retention.
Indicated for healthcare workers as apparel (scrub tops, pants, skirts, lab coats) to provide fluid repellency and reduce MRSA retention on the outer fabric surface. Not intended to replace personal protective equipment (PPE).
Regulatory Classification
Identification
Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.
Special Controls
*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.
Predicate Devices
Pro-Safe Professional Linens, Inc. Aprons, Scrub Slacks & Tops, Patient Drape (K891212)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 18, 2018
Vestagen Protective Technologies, Inc. Scott Pease Sr. Vice President of Regulatory and Quality 1301 W. Colonial Dr. Orlando, Florida 32804
Re: K180217
Trade/Device Name: VESTEX Apparel ("VESTEX") Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: QBW Dated: April 16, 2018 Received: April 17, 2018
Dear Scott Pease:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Geeta K. Pamidimukkala -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K180217
Device Name VESTEX® Apparel ("VESTEX")
Indications for Use (Describe)
VESTEX® Apparel ("VESTEX"), scrub tops, pants, skirts, and lab coats are apparel that provides fluid repellency and reduces the retention of Methicillin Resistant Staphylococcus aureus (MRSA) on the surface (outer layer) of the apparel.
The ability of VESTEX to reduce the retention of MRSA on the fabric surface has not been shown to correlate with the reduction of infections. Clinical studies to evaluate reduction in infection have not been performed for this device.
VESTEX Apparel is not intended to replace personal protective equipment.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------------|
| Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# VESTEX® Apparel "Vestex"
# Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared
Vestagen Protective Technologies, Inc. 1301 W. Colonial Dr. Orlando. FL 32804
(o) 407.781.2572 (m) 678.689.8010
Contact Person: Scott Pease, Sr. Vice President of Regulatory and Quality (scott.pease@vestagen.com)
Date Prepared: May 3, 2018
| Name of Device | VESTEX® Apparel |
|---------------------------------------------|----------------------------------------------------------------------------------------------|
| Common or Usual Name | Healthcare Practitioner Apparel (Scrub) |
| Classification Name | Suit, Surgical |
| Product Code / Regulation<br>Classification | QBW / 21 CFR § 878.4040 Class II |
| Predicate Devices | K891212 (Aprons, Scrub Slacks & Tops, Patient Drape) -<br>Pro-Safe Professional Linens. Inc. |
#### Indications for Use
VESTEX® Apparel ("VESTEX"), scrub tops, pants, skirts, and lab coats are apparel that provides fluid repellency and reduces the retention of Methicillin Resistant Staphylococcus aureus (MRSA) on the surface (outer layer) of the apparel.
The ability of VESTEX to reduce the retention of MRSA on the fabric surface has not been shown to correlate with the reduction of infections. Clinical studies to evaluate reduction in infection have not been performed for this device.
VESTEX Apparel is not intended to replace personal protective equipment.
### Device Description / Technological Characteristics
VESTEX® Apparel is nonsterile, reusable, fluid repellant apparel containing an antimicrobial agent that is designed for continuous wear to provide protection to healthcare workers where intermittent or unexpected exposure to microorganisms from blood, body fluids and other potentially infectious material (OPIM) can occur.
VESTEX® does not replace PPE worn for specific use and during episodes of expected blood, body fluid and other potentially infectious material (OPIM) exposure.
VESTEX® has a dual mechanism of action that is fluid repellant and an antimicrobial agent. VESTEX® has a proprietary application method used to covalently bond a fluid repellent and an antimicrobial agent to the outer surface of the fabric.
{4}------------------------------------------------
The primary mechanism of action is a fluid repellant fabric surface which acts to resist bacteria acquisition on the fabric. As the fluids and pathogens are repelled, the numbers of microorganisms that remain on the fabric are significantly reduced.
The secondary mechanism of action is an antimicrobial agent. It has one of several possible modes of action, including disruption of the cell membrane, denaturation of cell surface and transmembrane proteins, and inactivation of bio-energetic systems to prevent the microorganisms from being retained on the fabric. This secondary mechanism of action reduces Methicillin Resistant Staphylococcus aureus (MRSA) on the surface (outer layer) of the apparel.
### Performance Data
Biocompatibility and functional bench testing performed by Vestagen Protective Technologies, Inc. demonstrates VESTEX® Apparel substantial equivalence, in terms of the safety and effectiveness, to the referenced predicate device. In vitro bench testing included an assessment of all design input requirements and confirmation that the output of the design process met all design input requirements was completed, including those relating to appropriate standards and guidance's, as follows:
#### Biocompatibility
- ISO 10993-1:2009 Biological Evaluation of Medical Devices- Part I: ● Evaluation and Testing
- Skin Sensitization O
- Irritation O
- Cytotoxicity O
### Functional
- ASTM D1424-09:2013 Tearing Strength of Fabrics by Falling-Pendulum ● (Elemdorf-Type) Apparatus
- . Fabric Clinical Environment / Reuse
- Pre-Conditioning o
- O Fluid Repellency: Spray Test
- Splatter Testing (Reduction of MRSA Retention) O
- Cleaning Validation O
Additionally, in a randomized, blinded, crossover trial within a medical ICU, VESTEX® apparel demonstrated its effectiveness in reducing accumulation of Methicillin Resistant Staphylococcus aureus (MRSA) on health care workers (HCW) clothing 1 compared to control clothing. The MRSA reduction as statically significant both at the beginning and end of a 12 hour work shift.
- 1) Bearman, GML, et al. A Crossover Trial of Antimicrobial Scrubs to Reduce Methicillin-Resistant Staphylococcus aureus Burden on Healthcare Worker Apparel. Infection Control and Hospital Epidemiology. Vol. 33, No. 3 (March 2012), pp. 268-275.
{5}------------------------------------------------
| Feature | Subject Device (K180217) | Predicate Device<br>(K891212) |
|-------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------|
| Intended Use /<br>Indications | VESTEX® Apparel (“VESTEX”),<br>scrub tops, pants, skirts, and lab coats<br>are apparel that provides fluid<br>repellency and reduces the retention of<br>Methicillin Resistant Staphylococcus<br>aureus (MRSA) on the surface (outer<br>layer) of the apparel.<br>The ability of VESTEX to reduce the<br>retention of MRSA on the fabric<br>surface has not been shown to<br>correlate with the reduction of<br>infections. Clinical studies to evaluate<br>reduction in infection have not been<br>performed for this device. | Protective apparel for<br>dentists to prevent the<br>migration of bacteria and<br>liquids. |
| | VESTEX Apparel is not intended to<br>replace personal protective equipment. | |
| Materials of<br>Construction | 79% Polyester, 18% Rayon, 3%<br>Spandex | 65% polyester, 35% cotton<br>back with Barrier Supreme®<br>50% cotton 50% polyester<br>front. |
| Design Features | One layer of fabric constructed as an<br>active barrier. Exterior has a dual<br>mechanism of action to include a fluid<br>barrier and antimicrobial agent.<br>Interior has moisture wicking on the<br>skin touching side of the fabric. | One layer of fabric Scrub<br>Top and Slacks constructed<br>with a fluid resistant fabric in<br>front of the garment. |
| Performance<br>Testing | Biocompatibility:<br>Yes - ISO 10993-1 | Biocompatibility:<br>Not available |
| | Fluid Repellency (@ 0 & 50 washes):<br>Yes - Simulated Worst Case Study<br>(Variation of AATCC22) | Fluid Repellency (@ 0 & 50<br>washes):<br>Yes - AATCC22 |
| | Durability & Tear Strength:<br>Yes - ASTM D1424 | Durability & Tear Strength:<br>Yes - ASTM D2261 |
| | Splatter Challenge (Reduction of<br>MRSA Retention @ 0 & 50 Washes):<br>Yes - Simulated Worst Case Study | Reduction of MSRA<br>Retention:<br>No |
| | Clinical Trial<br>Yes - Bearman, GML, et al. A<br>Crossover Trial of Antimicrobial<br>Scrubs to Reduce Methicillin-<br>Resistant Staphylococcus aureus<br>Burden on Healthcare Worker Apparel | Clinical Trial<br>No |
# Summary of Substantial Equivalence
Based upon the device description, technical characteristics and test data provided within this submission, VESTEX® Apparel is substantially equivalent to the referenced predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.