# Titanium TomoFix Medial High Tibia Plate Anatomical (K180213)

_Synthes (USA) Products, LLC · HRS · Oct 19, 2018 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K180213

## Device Facts

- **Applicant:** Synthes (USA) Products, LLC
- **Product Code:** [HRS](/productcode/HRS.md)
- **Decision Date:** Oct 19, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3030
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The TOMOFIX™ Medial High Tibia Plates are indicated for open- and closed-wedge osteotomies, fixation of fractures, and malalignment caused by injury or disease of the medial proximal tibia for the treatment of: Unicompartmental medial or lateral gonarthrosis with malalignment of the proximal tibia Idiopathic or posttraumatic varus or valgus deformity of the proximal tibia

## Device Story

Titanium TomoFix Medial High Tibia Plate Anatomical is a bone fixation plate; precontoured for left/right medial proximal tibia anatomy; used in orthopedic surgery for osteotomies and fracture fixation. Device features seven threaded locking holes for fixed-angle stability and one combination locking/dynamic compression hole. Surgeon uses plate with 5.0 mm Locking Screws, 4.5 mm Cortex Screws, and 5.0 mm Spacers to stabilize bone segments. Provides rigid fixation to support healing of varus/valgus deformities or fractures. Used in clinical/OR settings by orthopedic surgeons. Benefits patient by restoring alignment and stabilizing bone during recovery.

## Clinical Evidence

Bench testing only. Static and dynamic mechanical loading tests were performed to demonstrate that the subject device's mechanical properties are equivalent to the predicate device.

## Technological Characteristics

Material: Commercially pure titanium grade 4 (TiCP4) per ASTM F67, UNS R50700. Design: Anatomically precontoured bone plate with seven threaded locking holes and one combination locking/dynamic compression hole. Compatibility: 5.0 mm Locking Screws, 4.5 mm Cortex Screws, 5.0 mm Spacers. Sterilization: Provided sterile and non-sterile.

## Regulatory Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Titanium TomoFix Medial High Tibia Plate ([K023941](/device/K023941.md))

## Reference Devices

- TomoFix Medial High Tibia Plate, Small ([K100676](/device/K100676.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the text "FDA U.S. FOOD & DRUG ADMINISTRATION" on the right. The symbol on the left is a stylized representation of a human figure, and the text on the right is in blue.

Synthes (USA) Products, LLC Stephan Jepards Sr. Regulatory Affairs Specialist 1301 Goshen Parkway West Chester, Pennsylvania 19380

October 24, 2018

## Re: K180213

Trade/Device Name: Titanium TomoFix Medial High Tibia Plate Anatomical Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: HRS Dated: September 18, 2018 Received: September 19, 2018

Dear Stephan Jepards:

This letter corrects our substantially equivalent letter of October 19, 2018

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Jesse Muir -S 2018.10.24 09:33:50 -04'00'

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

## Indications for Use

510(k) Number (if known)

### K180213

#### Device Name

Titanium TomoFix™ Medial High Tibia Plate Anatomical (line extension to the existing Synthes TomoFix Osteotomy System)

Indications for Use (Describe)

The TOMOFIX™ Medial High Tibia Plates are indicated for open- and closed-wedge osteotomies, fixation of fractures, and malalignment caused by injury or disease of the medial proximal tibia for the treatment of:

- Unicompartmental medial or lateral gonarthrosis with malalignment of the proximal tibia
- Idiopathic or posttraumatic varus or valgus deformity of the proximal tibia

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|                                                 |  |

|X | Prescription Use (Part 21 CFR 801 Subpart D)

__ Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

{3}------------------------------------------------

# 510(k) Summary

Date Prepared: September 17, 2018

| Sponsor:                    | DePuy Synthes<br>Stephan Jepards<br>1301 Goshen Parkway<br>West Chester, PA 19380<br>Office: +41 32 720 41 59                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
|-----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Proprietary<br>Name:        | Titanium TomoFix™ Medial High Tibia Plate Anatomical (to be included within the<br>Synthes TomoFix™ Osteotomy System)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Device<br>common name       | Bone plate                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Classification:             | Classification: Class II<br>Regulation Number: 21 CRF §888.3030<br>Product Code: HRS                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Classification<br>Name      | Single/multiple component metallic bone fixation appliances and accessories.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Predicate<br>Device         | Titanium TomoFix™ Medial High Tibia Plate<br>initially cleared via K023941                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Reference<br>device         | TomoFix™ Medial High Tibia Plate, Small,<br>initially cleared via K100676                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Device<br>Description:      | The subject Titanium TomoFix™ Medial High Tibia Plate Anatomical is anatomically<br>precontoured to fit the medial side of the proximal tibia (also referred to as high tibia)<br>and is provided in a 'left' and 'right' version to accommodate the anatomy of the tibia<br>in the left and right knee, respectively. The plate features seven round, threaded<br>locking holes to provide a locked, fixed-angle construct and one combination locking<br>/ dynamic compression hole to allow temporary compression during the surgery.<br>The subject plate is designed to accept existing, previously cleared 5.0 mm Locking<br>Screws, 4.5 mm Cortex Screws and 5.0 mm Spacers                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Indications for<br>Use:     | The TOMOFIX™ Medial High Tibia Plates are indicated for open- and closed-wedge<br>osteotomies, fixation of fractures, and malalignment caused by injury or disease of the<br>medial proximal tibia for the treatment of:<br>Unicompartmental medial or lateral gonarthrosis with<br>malalignment of the proximal tibia Idiopathic or posttraumatic varus or valgus deformity of the<br>proximal tibia                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Comparison to<br>Predicate  | The subject device and the predicate are identical with respect to indications and<br>intended use.<br>The subject Titanium TomoFix™ Medial High Tibia Plate Anatomical feature<br>equivalent characteristics as the predicate devices. Specifically, identical<br>technological characteristics are as follows:  Plates are compatible with 5.0 mm Locking Screws for angular, stable fixation Plates are compatible with 4.5 mm Cortex Screws to allow temporary compression of the lateral hinge Plates are offered sterile and non-sterile              The differences in design features between the subject plates and the predicate and             reference devices are listed below.  The subject plates are available in a pre-shaped left and right version to                     accommodate the anatomy of the left and right tibia, respectively. The subject plates only feature one LCP combination-hole, where the                     predicate and reference devices each feature three LCP combination-hole in                     total.                                                                                                                                                                   |
| Substantial<br>Equivalence: | The subject Titanium TomoFix™ Medial High Tibia Plate Anatomical has identical<br>indications compared to the predicate device. The features of the subject device are<br>substantially equivalent to the predicate device based on the similarities in intended<br>use and design. Mechanical testing demonstrates substantial equivalence of the<br>subject device to the predicate device.<br><br>In addition, the intended use, packaging, and sterilization of the predicates compared to the subject devices are identical. The subject and predicate devices are both made from commercially pure titanium grade 4 (TiCP4) in accordance with ASTM F67,<br>UNS R50700.<br><br>Mechanical testing demonstrates that the proposed Titanium TomoFix™ Medial High Tibia Plate Anatomical is substantially equivalent in mechanical properties to the predicate device. The subject device was tested under static and dynamic loading conditions to demonstrate mechanical equivalence compared to the predicate device.<br><br>In conclusion, the indications, technological characteristics, and performance testing demonstrate substantial equivalence of the subject Titanium TomoFix™ Medial High Tibia Plate Anatomical. |

{4}------------------------------------------------

---

**Source:** [https://fda-staging.innolitics.com/device/K180213](https://fda-staging.innolitics.com/device/K180213)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).