K180205 · Convatec Limited · OMP · Oct 19, 2018 · General, Plastic Surgery
Device Facts
Record ID
K180205
Device Name
AVELLE Negative Pressure Wound Therapy System
Applicant
Convatec Limited
Product Code
OMP · General, Plastic Surgery
Decision Date
Oct 19, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4780
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Avelle™ NPWT System is indicated for use on patients that would benefit from a Negative Pressure Wound Therapy (NPWT) device as it may promote wound healing via removal of exudate and infectious materials from low to moderately exuding wound, such as: - · Chronic Wounds e.g. Leg ulcers - · Acute Wounds - · Subacute and dehisced wounds - Traumatic wounds - · Flaps & Grafts - · Surgically closed incision sites Avelle™ NPWT System is suitable for use in a hospital, post-acute and home health environment.
Device Story
System provides negative pressure wound therapy (NPWT) to promote healing; removes exudate and infectious materials. Components: disposable battery-powered pump, absorbent Hydrofiber® dressing, tubing, luer-lock fittings, adhesive fixation strips. Pump maintains nominal -80mmHg negative pressure; draws exudate into dressing layers. Used in hospital, post-acute, or home environments; operated by clinicians or patients. Pump has 30-day lifetime; dressing wear time up to 7 days. Benefits: exudate management, wound healing promotion. Healthcare providers monitor wound status; pump provides automated pressure maintenance.
Clinical Evidence
No clinical data included. Bench testing only: peel testing, bacterial/viral barrier testing, negative pressure distribution, fluid absorption/retention, moisture vapor transmission, pump performance/alarm testing, waterproofness, rucking tests, simulated use, and human factors/usability testing.
Technological Characteristics
Powered suction pump (Class II, OMP). Materials: Hydrofiber® dressing, LDPE release liners. Energy: 3 AAA lithium batteries. Connectivity: None. Sterilization: EtO (dressing), non-sterile (pump). Software: Moderate level of concern, embedded firmware with 30-day usage limit.
Indications for Use
Indicated for patients requiring negative pressure wound therapy for low to moderately exuding wounds, including chronic wounds (e.g., leg ulcers), acute, subacute, dehisced, and traumatic wounds, flaps, grafts, and surgically closed incision sites.
Regulatory Classification
Identification
A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.
Predicate Devices
PICO Single Use Negative Pressure Wound Therapy System (K151436)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 19, 2018
ConvaTec Limited Justin Lovelace Senior Regulatory Affairs Specialist First Avenue Deeside, CH5 2NU Gb
Re: K180205
Trade/Device Name: AVELLE Negative Pressure Wound Therapy System Regulation Number: 21 CFR 878.4780 Regulation Name: Powered Suction Pump Regulatory Class: Class II Product Code: OMP Dated: September 14, 2018 Received: September 17, 2018
Dear Justin Lovelace:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products, and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for
Image /page/1/Picture/5 description: The image shows the name "Cynthia Chang -S" in a large, sans-serif font. The text is black and appears to be centered on a white background. The letters are clear and easy to read, with consistent spacing between them.
Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K180205
Device Name
Avelle™ Negative Pressure Wound Therapy System
Indications for Use (Describe)
The Avelle™ NPWT System is indicated for use on patients that would benefit from a Negative Pressure Wound Therapy (NPWT) device as it may promote wound healing via removal of exudate and infectious materials from low to moderately exuding wound, such as:
- · Chronic Wounds e.g. Leg ulcers
- · Acute Wounds
- · Subacute and dehisced wounds
- Traumatic wounds
- · Flaps & Grafts
- · Surgically closed incision sites
Avelle™ NPWT System is suitable for use in a hospital, post-acute and home health environment.
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### K180205 - 510(k) Summary
| Device: | Avelle Negative Pressure Wound Therapy System |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------|
| Applicant: | ConvaTec Limited |
| Contact: | Justin Lovelace<br>Senior Regulatory Affairs Specialist<br>Telephone: 1-336-542-4745<br>Email: Justin.Lovelace@convatec.com |
| Date: | October 18, 2018 |
| Trade Name: | Avelle Negative Pressure Wound Therapy System |
| Classification Name: | Powered Suction Pump (21 CFR 878.4780) |
| Device Class: | Class II |
| Product Code: | OMP |
## Predicate Device Information
| Predicate Device | 510(k) number | Clearance Date |
|-----------------------------------------------------------------|---------------|----------------|
| PICO Single Use<br>Negative Pressure<br>Wound Therapy<br>System | K151436 | 01/28/2016 |
#### Device Description:
The Avelle Negative Pressure Wound Therapy (NPWT) System consists of:
- a disposable battery powered pump ।
- absorbent wound dressing which is connected to the pump via tubing and luer-lock fittings
- adhesive fixation strips
The Avelle System is intended for single-patient use.
The Avelle pump is disposable, portable and battery-powered. The internal software of the pump limits the use to 30 days. The pump requires 3 AAA lithium batteries to operate. Testing shows that the 3 batteries will power the pump for a minimum of 15.6 days based on testing. A total of 6 AAA lithium batteries (2 sets) are provided with the pump which will power the device
{4}------------------------------------------------
for its entire 30-day lifetime. The disposable battery-powered pump unit is attached to the dressing via the supplied tubing and luer-lock fittings.
The Avelle dressing is comprised of gelling fiber (Hydrofiber®) to absorb wound exudate with negative pressure applied indirectly to the dressing via the Avelle pump. The dressing is applied to the wound and secured around its periphery via its adhesive border and adhesive fixation strips are applied to ensure an adequate seal. The dressing's adhesive surface is protected by a split release liner system comprising of 2 or 3 sections of Low Density Polyethylene (LDPE) release liner (depending on dressing size), which are removed prior to dressing application.
The pump is turned on to apply and maintain a nominal negative pressure of -80mmHg to the dressing. The pump will draw excess wound exudate and infectious materials away from the wound bed and into the Hydrofiber® layers as demonstrated in a simulated use test with a moderately exuding flow of 82 mL over 3 days and a second simulated use test with a low exuding flow of 102 mL over 7 days.
The system will deliver up to 30 days' single patient use for the pump component, and up to 7 days' wear time for the dressing. Each dressing may be used for up to 7 days but may require more frequent changes due to clinical situations. After 30 days of use the pump must be discarded and treatment must be stopped.
The pump unit is provided non-sterile. The Absorbing Dressing (with adhesive fixation strips) is provided sterile via ethylene oxide (EtO) sterilization.
#### Indications for Use
The Avelle™ NPWT System is indicated for use on patients that would benefit from a Negative Pressure Wound Therapy (NPWT) device as it may promote wound healing via removal of exudate and infectious materials from low to moderately exuding wound, such as:
- Chronic Wounds e.g. Leg ulcers
- Acute Wounds
- Subacute and dehisced wounds
- Traumatic wounds
- Flaps & Grafts
- Surgically closed incision sites
Avelle™ NPWT System is suitable for use in a hospital, post-acute and home health environment.
### Clinical Data Clinical Data is not included in this 510(k).
{5}------------------------------------------------
### Non-Clinical Performance Data
The AVELLE NPWT is subject to a number of non-clinical performance tests. The AVELLE NPWT meets all the requirements for overall design, sterilization, biocompatibility, and electrical safety results confirming that the design output meets the design inputs and specifications for the device.
The list below summarizes the bench testing undertaken and successfully completed for the AVELLE NPWT system:
- Peel testing of the dressing
- Peel testing of the adhesive fixation strips ●
- Bacterial Barrier testing
- Viral penetration testing
- Distribution of Negative pressure across the dressing
- Absorption assessment of dressing ●
- Fluid Retention testing
- Moisture Vapor Transmission Rate assessment of dressing
- Pump performance and alarm testing
- Assessing Waterproofness
- Rucking Test of the adhesive strips
- Simulated Use Testing
- . Human Factors/Usability Testing
The software documentation in this submission has been assembled according to the recommendations in the FDA document, Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, dated May 11, 2005. The software Level of Concern has been evaluated and determined to be Moderate, and appropriate documentation included, as recommended by the cited FDA guidance.
#### Conclusions
In establishing substantial equivalence to the predicate device, ConvaTec evaluated the indications for use, materials, technology, product specifications, and performance of the device and determined the subject device is substantially equivalent to the predicate. The differences do not raise different questions of safety or effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.