K180175 · View Point Medical, Inc. · GDW · Dec 7, 2018 · General, Plastic Surgery
Device Facts
Record ID
K180175
Device Name
SignalMark Lung Biopsy Site Marker
Applicant
View Point Medical, Inc.
Product Code
GDW · General, Plastic Surgery
Decision Date
Dec 7, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4750
Device Class
Class 2
Indications for Use
The SignalMark Lung Biopsy Site Marker is intended to provide accuracy in marking a biopsy location for visualization during surgical resection.
Device Story
Device consists of applicator and implantable marker; used by physicians in surgical settings (hospitals/clinics) to percutaneously place marker in lung tissue biopsy site. Applicator (plastic/stainless steel) pushes marker into tissue. Marker (porcine gelatin-based hydrogel with methylene blue-colored silicon dioxide microspheres) provides visual aid for surgeons to locate biopsy site during subsequent resection. Output is visual marker in tissue; assists surgeon in identifying biopsy site; facilitates accurate tissue/tumor resection; benefits patient by ensuring precise surgical removal of biopsied area.
Clinical Evidence
No human clinical data. Evidence includes bench testing (visual inspection, deployment, dimensional, stroke length, compression, tensile, hydration, ultrasound visualization, wipe test, packaging) and animal studies (rodent biodistribution, porcine safety, efficacy, and biologic response). Biocompatibility testing performed per ISO 10993-1.
Technological Characteristics
Applicator: plastic and stainless steel. Marker: USP-grade porcine gelatin-based hydrogel with methylene blue-colored silicon dioxide microspheres. Manual percutaneous delivery. Biocompatibility per ISO 10993-1.
Indications for Use
Indicated for adults undergoing percutaneous lung biopsies to mark biopsy locations for visualization during surgical resection.
Regulatory Classification
Identification
An implantable staple is a staple-like device intended to connect internal tissues to aid healing. It is not absorbable.
NovaBone MacroPor-Si+ – Bioactive Synthetic Bone Graft (K110925)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services USA logo, which features a stylized caduceus. To the right is the FDA U.S. Food & Drug Administration logo, with the FDA acronym in a blue square and the full name in blue text.
December 7, 2018
ViewPoint Medical, Inc. Thomas Kane Vice President, Regulatory Affairs & Quality Assurance 1235 Puerta del Sol #600 San Clemente, California 92673
Re: K180175
Trade/Device Name: SignalMark Lung Biopsy Site Marker Regulation Number: 21 CFR 878.4750 Regulation Name: Implantable Staple Regulatory Class: Class II Product Code: GDW Dated: November 2, 2018 Received: November 6, 2018
Dear Thomas Kane:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Cynthia Chang -S
for
Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
# K180175
Device Name
# SignalMark Lung Biopsy Site Marker
Indications for Use (Describe)
The SignalMark Lung Biopsy Site Marker is intended to provide accuracy in marking a biopsy location for visualization during surgical resection.
Type of Use (Select one or both, as applicable)
즈 Prescription Use (Part 21 CFR 801 Subpart D)
_Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) Summary
Image /page/3/Picture/1 description: The image shows the logo for View Point Medical. The words "View Point" are in a large, blue, sans-serif font. Below that is a logo consisting of a stylized letter "L" with a yellow circle in the upper left corner. To the right of the logo is the word "MEDICAL" in a smaller, blue, sans-serif font.
# 510(k) Summary K180175
#### DATE PREPARED
December 6, 2018
#### MANUFACTURER AND 510(k) OWNER
View Point Medical Inc. 5621 Palmer Way, Suite F Carlsbad, CA 92010, U.S.A. Telephone: (760) 707-7901 Official Contact: Thomas Kane, Vice President of Regulatory Affairs and Quality Assurance
#### REPRESENTATIVE/CONSULTANT
Allison C. Komiyama, Ph.D., R.A.C. AcKnowledge Regulatory Strategies, LLC Telephone: +1 (619) 208-7888 Email: akomiyama@acknowledge-rs.com
#### PROPRIETARY NAME OF SUBJECT DEVICE
SignalMark Lung Biopsy Site Marker
### COMMON NAME
Biopsy site marker
### DEVICE CLASSIFICATION
Staple, Implantable (Product code GDW, Class II) 21 CFR 878.4750 Implantable staple
#### INDICATIONS FOR USE
The SignalMark Lung Biopsy Site Marker is intended to provide accuracy in marking a biopsy location for visualization during surgical resection.
### DEVICE DESCRIPTION
The SignalMark Lung Biopsy Site Marker is a medical device used by a physician to percutaneously place a small implantable hydrogel marker in lung tissue biopsy to "mark" the location of the biopsy site. It is intended to be used on adults undergoing percutaneous lung biopsies, in surgical settings, such as hospitals or medical clinics with operating suites. The SignalMark Lung Biopsy Site Marker consists of two components:
{4}------------------------------------------------
510(k) Summary
Image /page/4/Picture/1 description: The image shows the logo for View Point Medical. The words "View Point" are stacked on top of the word "MEDICAL". The words are in a sans-serif font and are blue. To the left of the word "MEDICAL" is a stylized image of a circle with lines coming out of it.
- Applicator: Component made of plastic and stainless steel that pushes the marker into ● the tissue.
- Marker: Component made of USP-grade porcine gelatin-based hydrogel with methylene blue-colored silicon dioxide microspheres. The marker aids in the visualization of tissue allowing surgeons to readily locate the biopsy site for subsequent tissue or tumor resection.
### PREDICATE DEVICE IDENTIFICATION
SignalMark Lung Biopsy Site Marker is substantially equivalent to the following predicate:
| 510(k) Number | Predicate Device Name / Manufacturer | Primary Predicate |
|---------------|-----------------------------------------------------------|--------------------|
| K041331 | Lung Biopsy Site Marker / Biopsy Sciences, LLC | ✓ |
| K071937 | BonAlive granules and plates / Vivoxid, Ltd. | (Reference Device) |
| K110925 | NovaBone MacroPor-Si+ – Bioactive Synthetic Bone<br>Graft | (Reference Device) |
### SUMMARY OF NON-CLINICAL TESTING
No FDA performance standards have been established for SignalMark Lung Biopsy Site Marker.
Patient-contacting material was subjected to biocompatibility testing in compliance to ISO 10993-1 Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process.
The following performance testing was completed in order to demonstrate equivalence to the predicate device:
- . Performance testing (Bench)
- Visual Inspection of the Applicator о
- Applicator Deployment Test O
- Applicator Dimensional Inspection O
- Applicator Stroke Length Test O
- Applicator Compression Test O
- Applicator Tensile Test O
- Visual Inspection of the Marker Pad O
- Marker Pad Diameter O
- Marker Pad Length O
- Marker Pad Hydration O
- Marker Pad Ultrasound Visual Test O
- Wipe test with 70% IPA O
- Packaged Contents Verification O
- Performance testing (Animal) ●
- Biodistribution in rodents o
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for ViewPoint Medical. The words "View Point" are in a large, blue, sans-serif font. Below that, the word "MEDICAL" is in a smaller, blue, sans-serif font. To the left of the word "MEDICAL" is a blue graphic with a yellow circle.
# 510(k) Summary
- Safety and efficacy in porcine o
- Biologic response in porcine O
# EQUIVALENCE TO PREDICATE DEVICE
The SignalMark Lung Biopsy Site Marker is substantially equivalent to the predicate device based on the information summarized here:
The subject device uses similar or identical materials as the devices cleared in K041331, K071937, and K110925. The subject device has the same intended use, intended population, and use environments as the predicate device cleared in K041331. The subject device has similar technological characteristics to the device cleared in K041331. Unlike the predicate device, the SignalMark Lung Biopsy Site Marker uses silicon dioxide microspheres and USP grade porcine gelatin-based hydrogel in the marker. These technological characteristics have undergone bench and animal testing to ensure the device is substantially equivalent to the predicate.
# CONCLUSION
Based on the testing performed, including biocompatibility and non-clinical performance testing (bench and animal), it can be concluded that the subject device does not raise new issues of safety or effectiveness compared to the predicate device. The similar technological characteristics and performance characteristics for the proposed SignalMark Lung Biopsy Site Marker device are assessed to be substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.