K180173 · Palo Alto Health Sciences, Inc. · HCC · Aug 23, 2018 · Neurology
Device Facts
Record ID
K180173
Device Name
Freespira
Applicant
Palo Alto Health Sciences, Inc.
Product Code
HCC · Neurology
Decision Date
Aug 23, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5050
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Freespira is intended for use as a relaxation treatment for the reduction of stress by leading the user through guided and monitored breathing exercises. The device is indicated as an adjunctive treatment of symptoms associated with panic disorder (PD) and/or posttraumatic stress disorder (PTSD), to be used under the direction of a healthcare professional, together with other pharmacological and/or non-pharmacological interventions.
Device Story
Freespira is a biofeedback system for stress reduction and adjunctive treatment of PD and PTSD. It consists of a sensor and mobile app. The device measures exhaled carbon dioxide (EtCO2) and respiration rate (RR) to provide real-time feedback on breathing habits. Used at home under healthcare provider oversight; patients follow guided breathing exercises. Data is stored locally on a tablet and uploaded to a cloud server for clinician review. By monitoring EtCO2 and RR, the device helps patients adjust breathing patterns, potentially reducing symptoms associated with PD and PTSD. The system is designed for use alongside other pharmacological or non-pharmacological interventions.
Clinical Evidence
Prospective, single-arm, un-blinded study of 4-week program in PTSD patients. Primary endpoint: CAPS-5 score reduction. Results: 93% responder rate at 2-month follow-up (95% CI 77-99%); 48% remission rate. Mean CAPS-5 score decreased from 52.0 (baseline) to 25.8 (2-months) and 27.0 (6-months). High patient satisfaction reported.
Technological Characteristics
Biofeedback device measuring EtCO2 and respiration rate. Materials evaluated per ISO 10993-1. Electrical safety/EMC compliance: IEC 60601-1, ANSI/AAMI ES60601-1, IEC/EN 60601-1-2, IEC 60601-1-4, and IEC 60601-1-11. Connectivity: Local storage on tablet with cloud upload for clinician review. Power: Battery or AC adapter.
Indications for Use
Indicated for patients with panic disorder (PD) and/or posttraumatic stress disorder (PTSD) as an adjunctive treatment under the direction of a healthcare professional.
Regulatory Classification
Identification
A biofeedback device is an instrument that provides a visual or auditory signal corresponding to the status of one or more of a patient's physiological parameters (e.g., brain alpha wave activity, muscle activity, skin temperature, etc.) so that the patient can control voluntarily these physiological parameters.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter when it is a prescription battery powered device that is indicated for relaxation training and muscle reeducation and prescription use, subject to § 882.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION".
August 23, 2018
Palo Alto Health Sciences, Inc. % Allison Komiyama Principal Consultant AcKnowledge Regulatory Strategies, LLC 2834 Hawthorn St. San Diego, California 92104
Re: K180173
Trade/Device Name: Freespira Regulation Number: 21 CFR 882.5050 Regulation Name: Biofeedback Device Regulatory Class: Class II Product Code: HCC, CCK Dated: July 23, 2018 Received: July 24, 2018
Dear Allison Komiyama:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Pamela D. Scott -S
for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K180173
Device Name Freespira
#### Indications for Use (Describe)
Freespira is intended for use as a relaxation treatment for the reduction of stress by leading the user through guided and monitored breathing exercises. The device is indicated as an adjunctive treatment of symptoms associated with panic disorder (PD) and/or posttraumatic stress disorder (PTSD), to be used under the direction of a healthcare professional, together with other pharmacological and/or non-pharmacological interventions.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|
| <div><span style="font-size:16px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D)</div> | <div><span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)</div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image contains the logo for Palo Alto Health Sciences. The logo consists of a light blue circular design on the left, followed by the text "Palo Alto" in a larger, bolder font. Below "Palo Alto", the words "Health Sciences" are written in a smaller font size. The overall design is clean and professional, suggesting a focus on health and scientific research.
# 510(k) Summary K180173
## DATE PREPARED
August 23, 2018
### MANUFACTURER AND 510(k) OWNER
Palo Alto Health Sciences, Inc. 12020 113th Ave NE, Building C, Suite 215 Kirkland, WA 98034, U.S.A. Telephone: (925) 786-8409 Fax: (844) 394-2533 Official Contact: Simon Thomas, COO
## REPRESENTATIVE/CONSULTANT
Allison C. Komiyama, Ph.D., R.A.C. Aarthi Karkal, M.S. AcKnowledge Regulatory Strategies, LLC Telephone: Email: akomiyama@acknowledge-rs.com
#### PROPRIETARY NAME OF SUBJECT DEVICE
Freespira®
## COMMON NAME / DEVICE CLASSIFICATION
Device, Biofeedback (21 CFR 882.5050, HCC, Class II) Analyzer, Gas, Carbon-Dioxide, Gaseous-Phase (21 CFR 868.1400, CCK, Class II)
#### PREMARKET REVIEW
ODE/DNPMD/Neurodiagnostic and Neurosurgical Devices Branch (NDNB) ODE/DAGRID/Anesthesiology Devices Branch (ANDB)
#### INDICATIONS FOR USE
Freespira® is intended for use as a relaxation treatment for the reduction of stress by leading the user through guided and monitored breathing exercises. The device is indicated as an adjunctive treatment of symptoms associated with panic disorder (PD) and/or posttraumatic stress disorder (PTSD), to be used under the direction of a healthcare professional, together with other pharmacological and/or non-pharmacological interventions.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the logo for Palo Alto Health Sciences. The logo features the text "Palo Alto" in a bold, sans-serif font, with the words "Health Sciences" appearing in a smaller font size beneath "Palo Alto". To the left of the text is a circular graphic element in a light blue color. The overall design is clean and professional, suggesting a focus on health and scientific research.
## DEVICE DESCRIPTION
Freespira® is intended for use as a relaxation treatment for the reduction of stress by leading the user through guided and monitored breathing exercises. The device is indicated as an adjunctive treatment of symptoms associated with panic disorder (PD) and/or posttraumatic stress disorder (PTSD), to be used under the direction of a healthcare professional, together with other pharmacological and/or nonpharmacological interventions. Freespira is authorized and overseen by a licensed healthcare provider. Patients are trained to use the Freespira Sensor and the Freespira Mobile App to measure and display their exhaled carbon dioxide (EtCO₂) level and respiration rate (RR) and how different breathing habits affect EtCO2 levels.
## PREDICATE DEVICE IDENTIFICATION
Freespira is substantially equivalent to the following predicate:
| 510(k) Number | Predicate Device Name / Manufacturer | Primary Predicate |
|---------------|--------------------------------------|-------------------|
| K131586 | Canary Breathing™ System | Yes |
## SUMMARY OF NON-CLINICAL TESTING
No FDA performance standards have been established for Freespira. Reports on the following tests were provided to demonstrate safety based on current industry standards:
Biocompatibility: Patient contacting material was evaluated for biocompatibility in compliance to ISO 10993-1:2009(R)2013 Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process.
Software Verification: The software development and testing was executed according to FDA's Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices.
Electromagnetic Compatibility and Electrical Safety: The subject device was tested in compliance to methods outlined in IEC 60601-1:1988 +A1:1991 +A2:1995 - Medical Electrical Equipment Part 1: General Requirements for Safety; ANSI/AAMI ES60601-1:2005 Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance; IEC/EN 60601-1-2:2007 - Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility -Requirements and tests; IEC 60601-1-4:1996 + A1:1999 - Medical electrical equipment - Part 1-4: General requirements for safety - Collateral standard: Programmable electrical medical systems; and IEC 60601-1-11:2010 - Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment.
{5}------------------------------------------------
### 510(k) Summary
Image /page/5/Picture/1 description: The image contains the logo for Palo Alto Health Sciences. The logo features a stylized, light blue circular design on the left. To the right of the circular design, the words "Palo Alto" are written in a bold, dark font on the top line. Below "Palo Alto", the words "Health Sciences" are written in a smaller, lighter font.
## SUMMARY OF CLINICAL TESTING
Palo Alto Health Sciences Inc. performed a prospective, single arm, un-blinded investigation of Freespira in the indicated population. Patients with a primary diagnosis of PTSD (military and non-military, combat and non-combat) were enrolled into a four-week Freespira program. At enrollment, and then at specified intervals, patients underwent assessment using validated questionnaires (PHQ-9, SF-36, PDSS, CGC-S and CAPS-5 30-day score). In addition, every time a patient used Freespira, their EtCO2 and RR data were stored locally (on the tablet) and securely uploaded to a cloud server for clinician review.
The primary endpoint of this study was: quantitative improvements using the Clinician Administered PTSD Scale (CAPS-5). Using this scale, "Response" was defined as a reduction of 6 or more points. "Remission" was defined as Response plus no longer meeting clinical symptom criteria and having a severity score < 25.
The primary effectiveness hypothesis was that from baseline to the 2-month follow-up, at least 50% of study participants will experience a response to the treatment (defined as a 6-point reduction in CAPS-5 score from baseline).
Results demonstrate that this hypothesis was met: at 2-month post-treatment follow-up, the proportion of responders was 93% [with 95% Cl between 77-99%]. In addition, at 2-months the rate of remission was 48% and the mean CAPS score decreased from 52.0 at baseline to 25.8.
Post-treatment (week 5) the proportion of responders was 88% [with an unadjusted 95% Cl between 71-96%] and the rate of remission was 25%. Mean CAPS score decreased from 52.0 at baseline to 32.0.
At the 6-months post-treatment follow-up, the proportion of responders was 91% [with an unadjusted 95% Cl between 71-99%] and the rate of remission was 50%. The mean CAPS-5 score decreased from 52.0 at baseline to 27.0 showing stability of improvement.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image contains the logo for Palo Alto Health Sciences. The logo features the text "Palo Alto" in a bold, sans-serif font, with the words "Health Sciences" appearing in a smaller font size beneath "Palo Alto". To the left of the text is a stylized, abstract graphic element composed of curved lines, possibly representing a wave or a stylized letter 'P'. The overall design is clean and professional, suggesting a focus on health and scientific research.
## 510(k) Summary
A consistently large proportion of subjects achieved a decrease in their CAPS-5 score as the threshold definition of response increased in magnitude (from 11% to 40% of the change from baseline).
| | >=6 pts<br>decrease<br>from<br>baseline<br>[=11%]* | >=10 pts<br>decrease<br>from<br>baseline<br>[=19%]* | >=13 pts<br>decrease<br>from<br>baseline<br>[=25%]* | >=16 pts<br>decrease<br>from<br>baseline<br>[=30%]* | >=21 pts<br>decrease<br>from<br>baseline<br>[=40%]* |
|----------|----------------------------------------------------|-----------------------------------------------------|-----------------------------------------------------|-----------------------------------------------------|-----------------------------------------------------|
| Post-tx | 88% | 81% | 71% | 59% | 38% |
| 2-months | 93% | 83% | 84% | 69% | 59% |
| 6-months | 90% | 86% | 86% | 73% | 59% |
#### Response Rates with increasing Thresholds of Response Definition
*% change from mean baseline score of 52.0
The following adverse events were reported as possibly related to the treatment: chest pain, cold symptoms, and headache.
Patient satisfaction with Freespira was high post-treatment (84%), at 2-months (74%), and 6-months (90%) post-treatment. These percentages were based on the number of patients who gave a score of 3 or 4 on a 0-4 scale (where "0" meant "not satisfied" and "4″ meant "very satisfied") at each time point.
The clinical data demonstrate that Freespira provides a safe and effective, adjunct at-home treatment for the symptoms associated with PTSD that can be easily and widely disseminated. Freespira is completed in 4-weeks at-home without unpleasant side-effects. This likely contributed to the relatively low dropout rate (20%), high protocol adherence (77%), and high satisfaction scores (84%).
#### EQUIVALENČE TO PREDIČATE DEVIČE
Freespira is substantially equivalent to the predicate device based on the information summarized here:
The subject device has the same design and dimensions, and uses identical materials as the device cleared in K131586. The subject device has the same technological characteristics to the device cleared in K131586 except that the device can now also be used while it is plugged into the AC power adaptor. The fundamental difference between the two devices is that the subject device has expanded indications that include its use as an adjunct treatment for symptoms associated with PTSD. The subject device has undergone bench and clinical testing to ensure the device is as safe and effective as the predicate.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image contains the logo for Palo Alto Health Sciences. The logo features the text "Palo Alto" in a bold, sans-serif font, with the words "Health Sciences" appearing in a smaller font size beneath "Palo Alto". To the left of the text is a circular graphic element in a light blue color.
### 510(k) Summary
## CONCLUSION
Based on the testing performed, namely a prospective clinical study, it can be concluded that the subject device does not raise new issues of safety or efficacy compared to the predicate device. The similar technological characteristics and performance characteristics for the proposed Freespira are assessed to be substantially equivalent to the predicate device. The device is considered safe and effective for its intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.