The Habib EndoHPB is a radiofrequency (RF) catheter which provides bipolar energy to perform partial or complete ablation of tissue in the pancreatic and biliary tracts. The Habib EndoHPB is also intended for use to ablate malignant or benign tissue, notably to perform endoscopic biliary drainage or decompression, prior to stent placement or afterwards, to clear occluded stent.
Device Story
8F RF bipolar catheter; introduced via endoscope biopsy channel; connects to external RF generator. Device delivers bipolar RF energy to two ring electrodes at catheter tip; generates heat to ablate surrounding tissue. Used in clinical settings by physicians for endoscopic procedures. Output is thermal ablation of tissue; facilitates biliary drainage, decompression, and stent clearance. Benefits include minimally invasive treatment of pancreatic/biliary obstructions and malignant/benign tissue management.
Clinical Evidence
Clinical studies provided to support expanded indications for use. Bench testing validated 3-year shelf-life and compatibility with new generators. Electrical safety and electromagnetic compatibility testing performed per IEC 60601-1, IEC 60601-1-2, IEC 60601-2-2, and IEC 60601-2-18.
Technological Characteristics
8F bipolar RF catheter; stainless steel electrodes; endoscopic delivery. Energy source: external RF generator. Sterilization: ETO. Shelf-life: 3 years. Compatible with multiple RF generators (e.g., RITA, Olympus, ERBE).
Indications for Use
Indicated for patients requiring partial or complete tissue ablation in the pancreatic and biliary tracts, including ablation of malignant or benign tissue for biliary drainage, decompression, or clearing of occluded stents.
Regulatory Classification
Identification
An endoscopic electrosurgical unit and accessories is a device used to perform electrosurgical procedures through an endoscope. This generic type of device includes the electrosurgical generator, patient plate, electric biopsy forceps, electrode, flexible snare, electrosurgical alarm system, electrosurgical power supply unit, electrical clamp, self-opening rigid snare, flexible suction coagulator electrode, patient return wristlet, contact jelly, adaptor to the cord for transurethral surgical instruments, the electric cord for transurethral surgical instruments, and the transurethral desiccator.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" and the words "U.S. Food & Drug Administration."
March 20, 2018
EMcision Ltd. % Louis-Paul Marin President LOK North America Inc. 2025 Michelin LAVAL, H71 5b7 CANADA
Re: K180165
Trade/Device Name: Habib EndoHPB Regulation Number: 21 CFR 876.4300 Regulation Name: Endoscopic electrosurgical unit and accessories Regulatory Class: Class II Product Code: KNS Dated: January 17, 2018 Received: January 22, 2018
Dear Louis-Paul Marin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Benjamin R. Fisher -S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K180165
Device Name Habib EndoHPB
Indications for Use (Describe)
The Habib EndoHPB is a radiofrequency (RF) catheter which provides bipolar energy to perform partial or complete ablation of tissue in the pancreatic and biliary tracts.
The Habib EndoHPB is also intended for use to ablate malignant or benign tissue, notably to perform endoscopic biliary drainage or decompression, prior to stent placement or afterwards, to clear occluded stent.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------|----------------------------------------------------------------|
| <div> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for "EMcision". The logo is composed of the letters "EMcision" in a bold, sans-serif font. The "E" and "M" are connected, and there is an orange check mark above the "M". The rest of the letters are black.
# 510(k) Summary
| 1. | Type of submission | Traditional |
|----|---------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 2. | Preparation Date | January 17, 2018 |
| 3. | Submitter Address | EMcision Ltd.<br>Department of Surgery<br>Hammersmith Hospital<br>DuCane Road<br>London W12 0HS<br>United Kingdom<br>Phone: 1 (514) 994-9649<br>Contact: François Poulin |
| 4. | Contact Person | Louis-Paul Marin<br>President of LOK North America Inc.<br>Phone: 1 (450) 781-1578 ext. 225<br>Fax: 1 (450) 681-9663<br>Email: marin.lp@lok-corporation.com |
| 5. | Identification of the Device | |
| | Proprietary Name/Trade Name | Habib EndoHPB |
| | Regulation Description | Endoscopic electrosurgical unit and accessories |
| | Device Classification | II |
| | Regulation Number | 876.4300 |
| | Panel | Gastroenterology/Urology |
| | Product Code | KNS |
| 6. | Identification of the Predicate | |
| | Predicate Device Name | Habib EndoHPB |
#### 7. Purpose of this Submission
510(k) Number
The main purpose of this submission is to gain clearance to expand the indication for use of the previously cleared Habib EndoHPB (K083292) by including a treatment indication and update the original 510(k) submission to current configurations in terms of shelf life duration and insignificant changes to the labeling.
K083292
#### 8. Intended use of the Subject Device
The Habib EndoHPB is a radiofrequency (RF) catheter which provides bipolar energy to perform partial or complete ablation of tissue in the pancreatic and biliary tracts.
{4}------------------------------------------------
The Habib EndoHPB is also intended for use to ablate malignant or benign tissue, notably to perform endoscopic biliary drainage or decompression, prior to stent placement or afterwards, to clear occluded stent.
#### Device Description 9.
The Habib EndoHPB is an 8F RF bipolar catheter that consists of a catheter with 2 ring electrodes, introduced via an endoscope's biopsy channel. It has an attached cable which connects the device to an RF Generator. RF is applied to the two electrodes at the top of the catheter so that heat is applied to tissue surrounding the catheter.
#### 10. Performance Data
Compared to the predicate device, there has been no change or modification in design, material, energy source or manufacturing process in the subject device. The only modifications that have required performance testing are the following:
- New claimed shelf-life; and -
- -Compatibility with new generators.
The results obtained support the claimed 3-year shelf-life and the compatibility with new generators.
#### 11. Clinical Data
The subject device indications for use is significantly different when compared to the predicate device. Therefore, multiple clinical studies are being submitted to demonstrate that the Habib EndoHPB is as safe and effective as the predicate. All the clinical studies provided within this submission have been performed using the subject device and clearly support the new intended use.
#### 12. Electromagnetic and Electrical Testing
Electromagnetic compatibility and electrical safety testing have been performed on the Habib EndoHPB to validate that it meets the requirements of the latest Recognized Consensus Standards:
- -IEC 60601-1 Medical Electrical Equipment - Part 1: General requirements for basic safety and essential performance
- -IEC 60601-1-2 Medical Electrical Equipment -- Part 1-2: General Requirements for Basic Safety and Essential Performance - Collateral Standard: Electromagnetic Disturbances -Requirements and Tests
- -IEC 60601-2-2 Medical Electrical Equipment - Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories
- -IEC 60601-2-18 Medical Electrical Equipment – Part 2-18: Particular Requirements for the Basic Safety and Essential Performance of Endoscopic Equipment
{5}------------------------------------------------
#### 13. Substantial Equivalence Determination
The overall technological comparisons between the subject device and the predicate are summarized in the following table:
| Item | Predicate Device | Subject Device |
|---------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Habib EndoHPB<br>(K083292) | Habib EndoHPB |
| | Substantial Equivalence | |
| Classification | II | II |
| Code of Federal<br>Regulation | 876.4300 | 876.4300 |
| Prescription Medical<br>Devices | Yes | Yes |
| Indications for Use | The Habib EndoHPB is<br>intended to be used in the<br>coagulation of tissue during<br>Endoscopic surgical procedures<br>in the Gastro-intestinal Tract. | The Habib EndoHPB is a<br>radiofrequency (RF) catheter<br>which provides bipolar energy to<br>perform partial or complete<br>ablation of tissue in the<br>pancreatic and biliary tracts.<br><br>The Habib EndoHPB is also<br>intended for use to ablate<br>malignant or benign tissue,<br>notably to perform endoscopic<br>biliary drainage or<br>decompression, prior to stent<br>placement or afterwards, to clear<br>occluded stent. |
| Anatomical sites | Gastro-intestinal Tract | Pancreatic and Biliary Tracts |
| Operating Mode | Bipolar RF energy | Bipolar RF energy |
| Active Electrode<br>Material | Stainless Steel | Stainless Steel |
| Catheter Size | 7.5F | 8F |
| Number of Active<br>Electrodes | 2 | 2 |
| Electrode Shape | Central electrode assembly and<br>outer ring electrode | Central electrode assembly and<br>outer ring electrode |
| Delivery Mode | Endoscopic | Endoscopic |
| Single Use | Yes | Yes |
| Sterilization | ETO | ETO |
| Packaging | Coil inside a Tyvek pouch<br>mounted in a carton | Coil inside a Tyvek pouch<br>mounted in a carton |
| Shelf-Life | 6 months | 3 years |
| Compatible<br>Generators | RITA Medical Systems 1500<br>RITA Medical Systems 1500X<br>Radionics Cosman Coagulator<br>CC-1 | RITA Medical Systems 1500<br>RITA Medical Systems 1500X<br>Olympus ESG-100<br>Genii GI 4000<br>ERBE VIO 200 |
{6}------------------------------------------------
| ERBATOM ICC 200, 300, 350 |
|---------------------------|
|---------------------------|
The evidence collected in the table above clearly demonstrate that, except for insignificant differences, the subject device, the Habib EndoHPB, is identical to the previously cleared Habib EndoHPB (K083292) in terms of fundamental technology, design attributes, materials, features, functions, and performance.
The differences between the subject and the predicate device are as follows:
- -Expanded indications for use;
- Extended shelf-life; and -
- Compatibility with more generators. -
Performance data provided within this submission support the substantial equivalence of the subject device to the predicate. Additionally, clinical and performance data provided demonstrated the Habib EndoHPB to be safe and effective for the expanded indication for use, extended shelf-life, and compatibility with additional generators.
#### 14. Conclusion
The information provided by EMcision in this submission demonstrates that the subject device Habib EndoHPB is substantially equivalent to the currently cleared predicate device (K083292) and raises no new questions of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.