K180158 · Quanta System Spa · GEX · Feb 14, 2018 · General, Plastic Surgery
Device Facts
Record ID
K180158
Device Name
Surgical Laser fibers
Applicant
Quanta System Spa
Product Code
GEX · General, Plastic Surgery
Decision Date
Feb 14, 2018
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Surgical Laser Fibers are intended to be used in conjunction with cleared surgical lasers equipped with SMA 905 or SMA 906 or compatible connector. Surgical Laser Fibers are indicated for use in general surgical applications for: incision, excision, vaporization, ablation, haemostasis or coagulation of soft tissue in a contact or non-contact mode (with a compatible laser marketed for use in the desired application). Surgical Laser Fibers (surgical fiber optic laser delivery devices) are indicated for use in general surgery, urology, gastroenterology, gynecology, dermatology, vascular surgery, neurosurgery, plastic surgery, ENT/otolaryngology, endovenous occlusion of the greater saphenous vein in the patient with superficial vein reflux and laser assisted lipolysis with an approved compatible laser marketed for use in the desired application. Surgical Laser Fibers are also indicated for use in open or closed endoscopic applications where incision, excision, tissue dissection, excision of external tumors and lesions, complete or partial resection of internal organs, tumors or lesions, tissue vaporization, hemostasis and or coagulation may be indicated with an approved compatible laser marketed for use in the desired application. Surgical Laser Fibers are also intended as an aid for otologic procedures, for use in incision, excision, coagulation and vaporization of soft and fibrous tissue including osseous tissue with an approved compatible laser marketed for use in the desired application. Surgical Laser Fibers are also indicated for use in lithotripsy with an approved compatible laser marketed for use in the desired application. Surgical Laser Fibers are indicated for use with laser devices emitting radiation from 532 nm to 2100 nm, with pulsed and continuous wave (CW) emission mode. Surgical Laser Fibers are indicated, but not limited, for use with Diode laser, Argon, KTP/532, Ho:YAG, Nd:YAG, Tm:YAG pulsed and continuous wave CW laser devices. Surgical Laser Fibers may be used in surgical specialty or procedures for which compatible lasers have received regulatory clearance: for a complete information about applications, contraindications, precautions and warnings when using fiber optic it is necessary to refer to the applicable laser device User Manual.
Device Story
Device is a sterile fiber optic laser delivery system; silica core/clad fiber jacketed with ETFE; 3.0 meters long; SMA 905 proximal connector; various distal tips (flat/ball) and core diameters (200-1000 microns). Used in clinical settings (OR, endoscopy) by physicians to deliver laser radiation (532-2100 nm) from compatible external laser sources to soft or osseous tissue. Operates in contact or non-contact modes. Output is laser energy for tissue interaction; clinical decision-making relies on the physician's selection of laser parameters and fiber configuration based on the specific surgical procedure. Benefits include precise tissue management across multiple surgical specialties.
Clinical Evidence
Bench testing only. Compliance with consensus standards for sterilization (EN 556-1, ISO 11135-1, ISO 11138-2, EN ISO 17665), biocompatibility (ISO 10993-1, -5, -7, -10), packaging (ISO 11607-1, -2), and fiber optic connector interfaces (IEC 61754-22).
Technological Characteristics
Silica core/silica clad fiber; ETFE jacket; 3.0m length; SMA 905 connector; flat or ball distal tips; 200-1000 micron core diameters. Energy source: external laser (532-2100 nm, pulsed/CW). Sterilization: Ethylene oxide or moist heat. Standards: ISO 10993 (biocompatibility), ISO 11135/17665 (sterilization), IEC 61754-22 (connectors).
Indications for Use
Indicated for patients requiring soft tissue incision, excision, vaporization, ablation, hemostasis, or coagulation in general surgery, urology, gastroenterology, gynecology, dermatology, vascular surgery, neurosurgery, plastic surgery, ENT/otolaryngology, otologic procedures, lithotripsy, endovenous occlusion of greater saphenous vein, and laser-assisted lipolysis. Contraindications: refer to the compatible laser device User Manual.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 14, 2018
Quanta System SPA Francesco Dell'Antonio Vice President Regulatory Affairs and QA via acquedotto 109 Samarate (Va), 21017 Italy
Re: K180158
Trade/Device Name: Surgical Laser fibers Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: January 8, 2018 Received: January 19, 2018
Dear Francesco Dell'Antonio:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good
{1}------------------------------------------------
manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Jennifer R. Stevenson -S3
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K180158
#### Device Name
EMPOWER Sterile Surgical Optical Fiber
#### Indications for Use (Describe)
Surgical Laser Fibers are intended to be used in conjunction with cleared surgical lasers equipped with SMA 905 or SMA 906 or compatible connector.
Surgical Laser Fibers are indicated for use in general surgical applications for: incision, vaporization, ablation, haemostasis or coagulation of soft tissue in a contact mode (with a compatible laser marketed for use in the desired application).
Surgical Laser Fibers (surgical fiber optic laser delivery devices) are in general surgery, urology, gastroenterology, gynecology, dermatology, vascular surgery, plastic surgery, ENT/otolaryngology, endovenous occlusion of the greater saphenous vein in the patient with superficial vein reflux and laser assisted lipolysis with an approved compatible laser marketed for use in the desired application.
Surgical Laser Fibers are also indicated for use in open or closed endoscopic applications where incision, tissue dissection, excision of external tumors and lesions, complete or partial resection of internal organs, tumors or lesions, tissue vaporization, hemostasis and or coagulation may be indicated with an approved compatible laser marketed for use in the desired application.
Surgical Laser Fibers are also intended as an aid for otologic procedures, for use in incision, coagulation and vaporization of soft and fibrous tissue including osseous tissue with an approved compatible laser marketed for use in the desired application.
Surgical Laser Fibers are also indicated for use in lithotripsy with an approved compatible laser marketed for use in the desired application.
Surgical Laser Fibers are indicated for use with laser devices emitting radiation from 532 nm to 2100 mm, with pulsed and continuous wave (CW) emission mode.
Surgical Laser Fibers are indicated, but not limited, for use with Diode laser, Argon, KTP/532, Ho:YAG, Tm:YAG pulsed and continuous wave CW laser devices.
Surgical Laser Fibers may be used in surgical specialty or procedures for which compatible lasers have received regulatory clearance: for a complete informations, contraindications, precautions and warnings when using fiber optic it is necessary to refer to the applicable laser device User Manual.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 5. 510(K) SUMMARY
| Applicant /<br>Manufacturer<br>Name and Address: | Quanta System SPA<br>Via Acquedotto, 109<br>Samarate (VA)<br>Italy 21017 |
|--------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Contact Person: | Francesco Dell'Antonio<br>Vice President Regulatory Affairs and QA<br>Quanta System SPA<br>Email: francesco.dellantonio@quantasystem.com<br>Phone: +39-0331-376797<br>Fax: +39-0331-367815 |
| Date Prepared: | January 8th 2018 |
| Trade Name: | EMPOWER Sterile Surgical Optical Fiber |
| Common Name: | Sterile Surgical Optical Fiber |
| Classification: | Class II |
| Classification Name: | Laser surgical instrument for use in general and plastic surgery<br>and in dermatology. |
| Regulation Number: | 21 CFR 878.4810 |
| Product Code: | GEX |
| Basis for Submission: | change to labeling information |
| Primary predicate device | Surgical Laser fibers (K160513), Quanta System SpA |
The modified device Surgical Laser fibers is derived from the legally marketed (unmodified) device Surgical Laser fibers (K160513).
#### Performance Standards:
There are no mandatory performance standards for this device.
# Description of the device:
The Description of the modified device and the unmodified device is exactly the same, as follows. The Surgical Laser Fibers are a Fiber Optic Laser Delivery System meaning a device intended for the delivery of laser radiation to soft tissue.
{5}------------------------------------------------
The Surgical Laser Fibers are surgical fiber optic laser delivery devices based on a silica core/silica clad fiber jacketed with ethylene tetrafluoroethylene (ETFE). The devices are 3.0 meters (9.8 ft) long and are terminated with a laser specific SMA 905 connector plus a strain relief on the proximal end. Different distal tip configurations (flat and ball shaped) and several core diameter sizes (from 200 to 1000 microns) are offered.
# Intended use
Surgical Laser Fibers are intended to be used in conjunction with cleared surgical lasers equipped with SMA 905 or SMA 906 or compatible connector.
Surgical Laser Fibers are indicated for use in general surgical applications for: incision, excision, vaporization, ablation, haemostasis or coagulation of soft tissue in a contact or non-contact mode (with a compatible laser marketed for use in the desired application).
Surgical Laser Fibers (surgical fiber optic laser delivery devices) are indicated for use in general surgery, urology, gastroenterology, gynecology, dermatology, vascular surgery, neurosurgery, plastic surgery, ENT/otolaryngology, endovenous occlusion of the greater saphenous vein in the patient with superficial vein reflux and laser assisted lipolysis with an approved compatible laser marketed for use in the desired application.
Surgical Laser Fibers are also indicated for use in open or closed endoscopic applications where incision, excision, tissue dissection, excision of external tumors and lesions, complete or partial resection of internal organs, tumors or lesions, tissue vaporization, hemostasis and or coagulation may be indicated with an approved compatible laser marketed for use in the desired application.
Surgical Laser Fibers are also intended as an aid for otologic procedures, for use in incision, excision, coagulation and vaporization of soft and fibrous tissue including osseous tissue with an approved compatible laser marketed for use in the desired application.
Surgical Laser Fibers are also indicated for use in lithotripsy with an approved compatible laser marketed for use in the desired application.
Surgical Laser Fibers are indicated for use with laser devices emitting radiation from 532 nm to 2100 nm, with pulsed and continuous wave (CW) emission mode.
Surgical Laser Fibers are indicated, but not limited, for use with Diode laser, Argon, KTP/532, Ho:YAG, Nd:YAG, Tm:YAG pulsed and continuous wave CW laser devices.
Surgical Laser Fibers may be used in surgical specialty or procedures for which compatible lasers have received regulatory clearance: for a complete information about applications, contraindications, precautions and warnings when using fiber optic it is necessary to refer to the applicable laser device User Manual.
{6}------------------------------------------------
The wording differs from the one of the predicate device (K160513) limitedly to the removal of the restriction of use with Quanta System laser devices.
#### Performance data:
Same as the unmodified device, the Surgical Laser fibers devices comply with the following recognized consensus standards:
EN 556-1:2001/AC:2006 Sterilization of medical devices -Requirements for medical devices to be designated "STERILE" – Part 1: requirements for terminally sterilized medical devices.
ISO 15223-1:2012Medical devices. Symbols to be used with medical device labels, labelling and information to be supplied. General requirements
EN 1041:2008 Information supplied by the manufacturer of medical devices.
ISO 10993-1:2009 Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process.
ISO 10993-5:2009 Biological evaluation of medical devices – Part 5: Tests for in vitro cytotoxicity.
ISO 10993-7:2008 Biological evaluation of medical devices - Part 7: Ethylene oxide sterilization residuals.
ISO 10993-10:2010 Biological evaluation of medical devices – Part 10: Test for irritation and delayed-type hypersensitivity.
ISO 11135-1:2007 Sterilization of health-care products - Ethylene oxide - Part 1: Requirements for development, validation and routine control of a sterilization processes for medical devices.
ISO 11138-2:2006 Sterilization of health-care products – Biological indicators – Part 2: biological indicators for moist heat sterilization processes .
ISO 11607-1:2006 Packaging for terminally sterilized medical devices – Part 1: requirements for materials, sterile barrier systems and packaging systems.
ISO 11607-2:2006 Packaging for terminally sterilized medical devices – Part 2: validation requirements for forming, sealing and assembly processes.
ISO 11737-1:2006 Sterilization of medical devices – Microbiological methods – Part 1: Determination of a population of microorgamisms on product.
ISO 14971:2007 Medical devices - Application of risk management to medical devices.
IEC 61754-22:2005 Fibre optics connector interfaces – Part 22: type SMA connector family.
EN ISO 17665:2006 Sterilization of health care products -- Moist heat -- Part 1: Requirements for the development, validation and routine control of a sterilization process for medical devices.
{7}------------------------------------------------
# Biocompatibility:
The materials of the subject and unmodified device have no differences that affect biocompatibility.
# Comparison with predicate device:
The subject device and its predicate (K160513) have the same intended use and the same technological characteristics.
Any minor differences to the labeling do not present any new types of safety or effectiveness questions.
# Summary
Surgical Laser Fibers are substantially equivalent to the legally marketed predicate device (K160513).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.