Indigo Aspiration System - Penumbra Engine Pump, Indigo Aspiration System - Penumbra Engine Pump Canister
K180105 · Penumbra, Inc. · QEW · Mar 8, 2018 · Cardiovascular
Device Facts
Record ID
K180105
Device Name
Indigo Aspiration System - Penumbra Engine Pump, Indigo Aspiration System - Penumbra Engine Pump Canister
Applicant
Penumbra, Inc.
Product Code
QEW · Cardiovascular
Decision Date
Mar 8, 2018
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.5150
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
INDIGO Aspiration Catheters and Separators As part of the INDIGO Aspiration System, the INDIGO Aspiration Catheters and Separators are indicated for the removal of fresh, soft emboli and thrombi from vessels of the peripheral and venous systems. INDIGO Aspiration Tubing As part of the INDIGO Aspiration System, the INDIGO Sterile Aspiration Tubing is indicated to connect the INDIGO Aspiration Catheters to the Penumbra Aspiration Pump. Penumbra Aspiration Pump The Penumbra Aspiration Pump is indicated as a vacuum source for Penumbra Aspiration Systems.
Device Story
Indigo Aspiration System removes thrombus from peripheral/venous vasculature via continuous aspiration. System components: Aspiration Catheter, Separator, Aspiration Tubing, Penumbra Engine Pump, and Canister. Catheter introduced via guide catheter/sheath to occlusion site; pump provides vacuum pressure; Separator clears catheter lumen if blocked. Engine Pump is electromechanical vacuum source; 1000 mL reservoir Canister with stop-flow filter prevents fluid ingress. Used in OR or interventional labs by physicians. Pump/Canister are non-patient contacting capital equipment. Output is vacuum pressure; clinical benefit is thrombus removal to restore vessel patency.
Clinical Evidence
Bench testing only. No clinical data. Performance verified via electrical safety/EMC testing (IEC 60601-1, IEC 60601-1-2), environmental testing, dimensional inspection, and 500-hour reliability/life testing. All tests met acceptance criteria.
Technological Characteristics
Electromechanical vacuum pump; 100-240 Vac, 50/60 Hz. 1000 mL reservoir canister with stop-flow filter. Compliant with IEC 60601-1, IEC 60601-1-2, and EN ISO 10079-1. Packaging validated per ASTM D4169. Non-sterile capital equipment.
Indications for Use
Indicated for removal of fresh, soft emboli and thrombi from peripheral arterial and venous vessels. Used with Penumbra Aspiration Pump as a vacuum source.
Regulatory Classification
Identification
An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
March 8, 2018
Penumbra, Inc. Mr. Richard Kimura Regulatory Specialist One Penumbra Place Alameda, California 94502
Re: K180105
Trade/Device Name: Indigo Aspiration System - Penumbra Engine Pump and Canister Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy Catheter Regulatory Class: Class II Product Code: DXE Dated: January 15, 2018 Received: January 16, 2018
Dear Mr. Kimura:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
# Page 2 - Mr. Richard Kimura
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Kenneth J. Cavanaugh -S
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K180105
Device Name
Indigo Aspiration System-Penumbra Engine Pump and Canister
Indications for Use (Describe)
INDIGO Aspiration Catheters and Separators
As part of the INDIGO Aspiration System, the INDIGO Aspiration Catheters and Separators are indicated for the removal of fresh, soft emboli and thrombi from vessels of the peripheral and venous systems.
INDIGO Aspiration Tubing
As part of the INDIGO Aspiration System, the INDIGO Sterile Aspiration Tubing is indicated to connect the INDIGO Aspiration Catheters to the Penumbra Aspiration Pump.
Penumbra Aspiration Pump
The Penumbra Aspiration Pump is indicated as a vacuum source for Penumbra Aspiration Systems.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the word "Penumbra" in a red, sans-serif font. To the right of the word is a red circle with a white "P" inside. The "P" is stylized with a horizontal line extending beyond the circle. The logo is simple and modern.
#### 1 510(k) Summary
(as required by 21 CFR 807.92)
Pursuant to Section 12, Part (a)(i)(3A) of the Safe Medical Devices Act of 1990, Penumbra, Inc. is providing the summary of Substantial Equivalence for the Indigo Aspiration System® - Penumbra Engine™ Pump and Canister.
#### 1.1 Sponsor/Applicant Name and Address
Penumbra, Inc. One Penumbra Place Alameda, CA 94502 USA
#### 1.2 Sponsor Contact Information
Richard Kimura Regulatory Affairs Specialist Phone: (510) 995-2034 FAX: (510) 217-6414 Email: rkimura@penumbrainc.com
#### 1.3 Date of Preparation of 510(k) Summary
January 12, 2018
#### 1.4 Device Trade or Proprietary Name
Indigo Aspiration System® - Penumbra Engine™ Pump and Canister
#### 1.5 Device Classification
| Regulatory Class: | II |
|-----------------------|-----------------------|
| Classification Panel: | Cardiovascular |
| Classification Name: | Catheter, Embolectomy |
| Regulation Number: | 21 CFR §870.5150 |
| Product Code: | DXE |
#### 1.6 Predicate and Reference Devices
| 510(k) Number / Clearance Date | Name of Device | Name of Manufacturer |
|------------------------------------|--------------------------|---------------------------------------------------------------|
| Predicate Device | | |
| K161523 cleared on July 1, 2016 | Indigo Aspiration System | Penumbra, Inc.<br>One Penumbra Place<br>Alameda, CA 94502 USA |
| Reference Device | | |
| K122756 cleared on October 2, 2012 | Penumbra Pump MAX | Penumbra, Inc.<br>One Penumbra Place<br>Alameda, CA 94502 USA |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the word "Penumbra" in red font. To the right of the word is a red circle with a white "P" inside. The "P" is stylized with a horizontal line through the middle. The logo is simple and modern.
#### Predicate Comparison 1.7
| | Predicate Device | Subject Device |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------|
| Trade Name | Indigo System (with Penumbra Pump MAX<br>and Canister) | Indigo System (with Penumbra Engine<br>Pump and Canister) |
| 510(k) No. | K161523 | To Be Determined |
| Classification | Class II, DXE | SAME |
| Indication for Use | INDIGO Aspiration Catheters and<br>Separators<br>As part of the INDIGO Aspiration System,<br>the INDIGO Aspiration Catheters and<br>Separators are indicated for the removal of<br>fresh, soft emboli and thrombi from vessels<br>of the peripheral arterial and venous<br>systems.<br><br>INDIGO Aspiration Tubing<br>As part of the INDIGO Aspiration System,<br>the INDIGO Sterile Aspiration Tubing is<br>indicated to connect the INDIGO Aspiration<br>Catheters to the Penumbra Aspiration<br>Pump.<br><br>Penumbra Aspiration Pump<br>The Penumbra Aspiration Pump is indicated<br>as a vacuum source for Penumbra<br>Aspiration Systems. | SAME |
| Aspiration Pump | | |
| Trade Name | Penumbra Pump MAX and Canister | Penumbra Engine Pump and Canister |
| IEC 60601-1<br>Compliance | Yes | SAME |
| IEC 60601-1-2<br>Compliance | Yes | SAME |
| Voltage | 100-115 Vac/230 Vac | 100-240 Vac |
| Frequency | 50 Hz/60 Hz | SAME |
| Sterilization | Non sterile | SAME |
| Shelf Life | N/A | SAME |
| The Aspiration Catheters, Separators, and Aspiration Tubings are unchanged and remain identical to<br>those of the currently cleared Indigo System (K161523). | | |
#### 1.8 Device Description
The Indigo Aspiration System ("Indigo System") is intended for the removal of fresh, soft emboli and thrombi from vessels of the peripheral and venous systems using the Indigo Aspiration Catheter, Indigo Separator, Indigo Aspiration Tubing, and Penumbra
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the Penumbra company logo. The logo consists of the word "Penumbra" in a bold, red font. To the right of the word is a red circle with a white "P" inside. The "P" is stylized with a horizontal line extending from the top of the "P" to the edge of the circle.
Aspiration Pump. The Indigo System was most recently cleared under K161523.
The Indigo System is designed to remove thrombus from the vasculature using continuous aspiration. The Aspiration Catheter targets aspiration from the Aspiration Pump directly to the thrombus. The Separator may be used to clear the lumen of the Aspiration Catheter should it become blocked with thrombus. The use of the Separator may not be necessary when using an Aspiration Catheter with an I.D. of 0.054in or larger. The Aspiration Catheter is introduced through a guide catheter or long femoral sheath and into the site of the primary occlusion. The Aspiration Catheter is used with the Aspiration Pump to aspirate thrombus from an occluded vessel. As needed, a Separator may be deployed from the Aspiration Catheter to assist with thrombus removal. The Separator is advanced and retracted through the Aspiration Catheter at the proximal margin of the primary occlusion to facilitate clearing of the thrombus from the Aspiration Catheter tip. For the aspiration source, the Aspiration Catheter is used in conjunction with the Aspiration Pump, which is connected using the Aspiration Tubing and Canister. The Aspiration Catheter is provided with a steam shaping mandrel and rotating hemostasis valve, and a peelable sheath. The Separator is provided with an introducer and torque device. The Aspiration Catheters and Separators are visible under fluoroscopy.
The Penumbra Engine Pump and Canister
The Penumbra Engine Pump ("Engine Pump") is an electromechanical device designed to create vacuum pressure for use with the Indigo System. The Engine Pump maintains vacuum pressure for 3 hours of continuous use and has an operating life of ≥ 500 hours. The pump can be used in environments with 100 - 240 Vac and 50/60 Hz. The Engine Pump is intended for use in operating rooms or interventional catheterizations laboratories. The Engine Pump and Canister do not come into contact with the patient. The Engine Pump Canister is a 1000 mL minimum volume reservoir which is an accessory to the Engine Pump. The Canister has a lid with a stop-flow filter to prevent excess fluid from entering the pump. The patient port on the lid is sized to accept the Suction Connector on the Indigo Aspiration Tubing.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the Penumbra company logo. The logo consists of the word "Penumbra" in red font, followed by a red circle with a white "P" inside. The "P" is stylized with a horizontal line extending beyond the circle.
#### 1.9 Indications for Use
# INDIGO Aspiration Catheters and Separators
As part of the INDIGO Aspiration System, the INDIGO Aspiration Catheters and Separators are indicated for the removal of fresh, soft emboli and thrombi from vessels of the peripheral arterial and venous systems.
### INDIGO Aspiration Tubing
As part of the INDIGO Aspiration System, the INDIGO Sterile Aspiration Tubing is indicated to connect the INDIGO Aspiration Catheters to the Penumbra Aspiration Pump.
### Penumbra Aspiration Pump
The Penumbra Aspiration Pump is indicated as a vacuum source for Penumbra Aspiration Systems.
### 1.10 Summary of Non-Clinical Data
As required under Section 12, Part (a)(i)(3A) of the Safe Medical Devices Act of 1990, a summary of any information regarding substantial equivalence of the device follows.
Included in this section are summary descriptions of the testing (or rationale for not testing if not applicable to the Engine Pump and Canister) which substantiates the performance of the subject Indigo System with Engine Pump and Canister as well as its substantial equivalence to the predicate device:
- Biocompatibility .
- Design Verification (Bench-Top Testing and Electrical Safety/EMC Testing) .
- Shelf Life ●
- Sterilization .
- Packaging
The subject Indigo System with Engine Pump and Canister met all established requirements.
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for Penumbra, Inc., a global healthcare company focused on innovative therapies. The logo features the word "Penumbra" in a red, sans-serif font. To the right of the name is a circular graphic, also in red, with a stylized "P" shape formed by white lines within the circle.
### 1.10.1 Biocompatibility
The subject and predicate Indigo System sterile devices are identical. There are no changes to the previously provided biocompatibility data of the Indigo System materials sterile devices, which were reviewed and cleared under K161523. No additional biocompatibility testing is required or was performed for the Indigo System sterile devices.
The Indigo System Engine Pump is a non-sterile reusable piece of capital equipment. The pump does not contact the patient, nor is it introduced into the sterile field. As such, biocompatibility testing is not required and was not performed for the pump. The Indigo System Engine Pump Canister also does not contact the patient, nor is it introduced into the sterile field. Furthermore, blood or body fluids collected in the canister are not reintroduced to the patient. As such, biocompatibility testing is not required and was not performed for the Canister.
### 1.10.2 Design Verification (Bench-top Testing and Electrical Safety/EMC Testing)
The subject and predicate Indigo System sterile devices are identical. There are no changes to the design specifications and performance characteristics of the Indigo System sterile devices. Therefore, all previous bench-top testing data which were reviewed and cleared under K161523 continue to support the subject Indigo System sterile devices. No additional bench-top testing is required or was performed for the Indigo System sterile devices.
Bench-top testing was conducted to evaluate the physical and mechanical properties of the subject Indigo System Engine Pump and Canister. All bench-top studies were conducted using good scientific practices and statistical sampling methods as required by the Penumbra Design Control procedures. Performance testing was based on the design specifications, risk analysis, performance standards, and guidance documents.
The Indigo System Engine Pump and Canister also underwent electrical safety and EMC testing in accordance with the requirements of IEC 60601-1 and IEC 60601-1-2. Testing was performed by Intertek, a nationally recognized test laboratory. The Engine Pump and Canister passed all tests and met all acceptance criteria.
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the word "Penumbra" in red font. To the right of the word is a red circle with a white outline of the letter "P" inside. The logo is simple and modern, with a focus on the company name and a stylized initial.
| Attribute | Sample<br>Size | Specification | Acceptance Criteria | Results |
|------------------------------------------------------------|----------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------|-----------|
| Pump - Electrical<br>Safety | N=1 | The Pump conforms to IEC 60601-1<br>and IEC 60601-1-2 requirements<br>including international worldwide<br>variants (CB Scheme). The Pump is<br>compliant with EN ISO 10079-1. | 100% Pass | 100% Pass |
| Pump - Environmental<br>Testing | N=5 | Pump performance specifications<br>under environmental conditions | 100% Pass | 100% Pass |
| Pump - Dimensional<br>Inspection | N=5 | Dimensional specifications per<br>Product Specification | 100% Pass | 100% Pass |
| Pump - Inspection of<br>Design Features | N=5 | Design specifications per Product<br>Specification | 100% Pass | 100% Pass |
| Pump - Performance | N=5 | Performance specifications per<br>Product Specification | 100% Pass | 100% Pass |
| Pump - Performance at<br>variable voltage and<br>frequency | N=5 | Performance specifications per<br>Product Specification | 100% Pass | 100% Pass |
| Pump and Canister -<br>500 hour Use Testing | N=5 | Performance specifications after 500<br>hour use | 100% Pass | 100% Pass |
| Pump Canister -<br>Inspection of Design<br>Features | N=5 | Design specifications per Product<br>Specification | 100% Pass | 100% Pass |
| Pump Canister -<br>Performance | N=5 | Performance specifications per<br>Product Specification | 100% Pass | 100% Pass |
| Pump and Canister -<br>Simulated Use | N=5 | Pump and Canister use<br>specifications | 100% Pass | 100% Pass |
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image shows the Penumbra company logo. The word "Penumbra" is written in a red, sans-serif font. To the right of the word is a red circle with a white "P" inside. The logo is simple and modern.
### 1.10.3 Shelf Life
The subject and predicate Indigo System sterile devices are identical. There are no changes to the previously provided shelf life data for the sterile devices which were reviewed and cleared under K161523. No additional shelf life testing is required or was performed for the Indigo System sterile devices.
The proposed Indigo System Engine Pump is a reusable piece of capital equipment that is provided non-sterile. Therefore, shelf life testing is not applicable to the Engine Pump and shelf life testing was not performed. The Engine Pump is established for 500 hours of use based on completed life (reliability) testing. The Engine Pump Canister does not have an established shelf life.
### 1.10.4 Sterilization
The subject and predicate Indigo System sterile devices are identical. There are no changes to the previously provided sterilization data of the devices, which were reviewed and cleared under K161523. No additional sterilization testing is required or was performed for these devices.
Sterilization testing is not applicable to the proposed Indigo System Engine Pump and Canister. Both are supplied non-sterile and are not intended to be sterilized.
### 1.10.5 Packaging
The packaging materials and process of the subject and predicate Indigo System sterile devices are identical. There are no changes to the previously provided packaging material listing or the packaging process for these devices, which were reviewed and cleared under K161523. No additional packaging testing is required or was performed for the Indigo System sterile devices.
The packaging materials for the proposed Indigo System Engine Pump and Canister are similar to those used for the predicate Indigo System Penumbra Pump MAX and Canister. The Engine Pump and Canister are packaged to ensure that damage does not occur during shipping. The Engine Pump is packaged in a protective, corrugated, cardboard 275 B/C Flute Double Wall RSC shipping container with custom shaped foam inserts. Pumps used for Design Verification and Packaging Validation testing first underwent transportation conditioning per ASTM D4169, Distribution Cycle 3,
{10}------------------------------------------------
Image /page/10/Picture/1 description: The image shows the logo for Penumbra, Inc., a global healthcare company focused on innovative therapies. The logo features the word "Penumbra" in a bold, red font. To the right of the name is a stylized "P" inside a red circle, with a white line bisecting the "P" and extending beyond the circle's edge.
Assurance Level 3 to ensure packaging integrity. The Canister is packaged individual in 200 pound E-Flute container product boxes. The Canister box is then packaged in protective, corrugated, cardboard 275 B/C Flute Double Wall RSC shipping containers (8 canister boxes per shipper).
# 1.11 Summary of Substantial Equivalence
The subject Indigo System disposable devices are unchanged and remain identical to those of the predicate Indigo System with regard to indications, function, design, materials, biocompatibility, packaging, and sterilization. The subject Indigo System Engine Pump and Canister are substantially equivalent to the predicate device with regard to intended use, operating principle, design concept, materials, and packaging processes.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.