Embozene Microspheres are intended for embolization of arteriovenous malformations and hypervascular tumors, including uterine fibroids (UFE) and hepatoma, and for embolization of prostatic arteries (PAE) for symptomatic benign prostatic hyperplasia (BPH). This device is not intended for neurovascular use.
Device Story
Embozene Color-Advanced Microspheres are spherical, calibrated, biocompatible hydrogel microspheres coated with Polyzene-F polymer; supplied in syringes. Used by interventional radiologists in clinical settings for vascular embolization. Delivered via microcatheter to target vessels to occlude blood flow to hypervascular tumors or prostatic arteries. Color-coded by size (40-1300 µm) for identification. Device is passive; no electronic components or software. Clinical benefit includes symptomatic improvement of BPH (reduced IPSS scores, increased peak flow rates) and tumor/malformation management. Substantial equivalence established via bench testing, biocompatibility, and prospective clinical study of 38 patients.
Clinical Evidence
Prospective single-center study (n=38) evaluated PAE for symptomatic BPH. Primary endpoints: safety and symptomatic improvement (IPSS, Qmax). Results: 100% technical success; 46-47% improvement in IPSS at 6-12 months; QoL improved from 4.4 to 2.3. No serious adverse events or device-related events reported. Preclinical animal studies (porcine kidney) confirmed similar tissue response to predicate.
Technological Characteristics
Spherical, non-resorbable hydrogel microspheres coated with Polyzene-F polymer. Sizes: 40-1300 µm. Passive implant; no energy source. Sterile, supplied in 20 ml syringes. Biocompatible per ISO/ASTM standards (cytotoxicity, sensitization, genotoxicity, hemocompatibility). MR compatible.
Indications for Use
Indicated for embolization of arteriovenous malformations, hypervascular tumors (including uterine fibroids and hepatoma), and prostatic arteries for symptomatic benign prostatic hyperplasia (BPH) in adult males. Not for neurovascular use.
Regulatory Classification
Identification
A prostatic artery embolization device is an intravascular implant intended to occlude the prostatic arteries to prevent blood flow to the targeted area of the prostate, resulting in a reduction of lower urinary tract symptoms related to benign prostatic hyperplasia. This does not include cyanoacrylates and other embolic agents which act by in situ polymerization or precipitation, or embolization devices used in neurovascular applications (see 21 CFR 882.5950).
Special Controls
In combination with the general controls of the FD&C Act. the prostatic artery embolization device is subject to the following special controls:
*Classification.* Class II (special controls). The special controls for this device are:(1) The device must be demonstrated to be biocompatible.
(2) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
(i) Evaluation of suitability for injection through catheters intended for use in embolization; and
(ii) Evaluation of the size distribution of the device.
(3) Performance data must support the sterility and pyrogenicity of the device.
(4) Performance data must support the shelf life of the device by demonstrating continued sterility, package integrity, and device functionality over the identified shelf life.
(5) Clinical data must evaluate post-embolization damage due to non-target embolization under anticipated use conditions.
(6) The labeling must include:
(i) Specific instructions on safe device preparation and use;
(ii) The device shelf life;
(iii) Data regarding urinary retention; and
(iv) Data regarding post-prostatic artery embolization syndrome.
Predicate Devices
Embosphere Microspheres (Embosphere PRO Prostatic Artery Embolization Kit), Merit Medical Systems Inc. (DEN160040)
Reference Devices
Embozene Color-Advanced Microspheres, Boston Scientific Corporation (K141209)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the text "FDA U.S. FOOD & DRUG ADMINISTRATION" on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right is written in a clear, sans-serif font, with "FDA" in a blue box.
April 19, 2018
Boston Scientific Corporation Heidi Shearer Regulatory Affairs Specialist 3 Scimed Place Maple Grove, MN 55311
Re: K180102
Trade/Device Name: Embozene Color-Advanced Microspheres Regulation Number: 21 CFR& 876.5550 Regulation Name: Prostate Artery Embolization Device Regulatory Class: II Product Code: NOY, KRD, NAJ Dated: January 18, 2018 Received: January 19, 2018
Dear Heidi Shearer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Charles Viviano -S
Benjamin R. Fisher, Ph.D. For Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K180102
Device Name Embozene Color-Advanced Microspheres
Indications for Use (Describe)
Embozene Microspheres are intended for embolization of arteriovenous malformations and hypervascular tumors, including uterine fibroids (UFE) and hepatoma, and for embolization of prostatic arteries (PAE) for symptomatic benien prostatic hyperplasia (BPH).
This device is not intended for neurovascular use.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No, 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
{3}------------------------------------------------
# 510(k) Summary per 21 CFR §807.92
| Sponsor | Boston Scientific Corporation<br>300 Boston Scientific Way<br>Marlborough, Massachusetts 01752<br>USA |
|---------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Name and<br>Information | Heidi Shearer<br>Three Scimed Place<br>Maple Grove, MN 55311-1566<br>Phone:763-255-0056<br>Fax: 763-494-2222<br>e-mail: Heidi.Shearer@bsci.com |
| Prepared | 12 January 2018 |
| Proprietary Name | EmbozeneTM Color-Advanced Microspheres |
| Common Name | Prostatic Artery Embolization Device<br>Vascular Embolization Device |
| Product Code | NOY<br>KRD, NAJ |
| Classification | Class II, 21 CFR Part 876.5550 (NOY) and 21 CFR<br>Part 870.3300 (KDR, NAJ) |
| Predicate Device | Embosphere® Microspheres (Embosphere PROTM<br>Prostatic Artery Embolization Kit), Merit Medical<br>Systems Inc.<br>DEN160040 (21-Jun-2017) |
| Reference Device | EmbozeneTM Color-Advanced Microspheres, Boston<br>Scientific Corporation<br>K141209 (07-Aug-2014) |
{4}------------------------------------------------
# Device Description
Embozene™ Color-Advanced Microspheres (hereafter referenced as Embozene or Embozene Microspheres) are spherical, tightly calibrated, biocompatible, hydrogel microspheres coated with proprietary polymer (Polyzene®-F). The microspheres are compressible to enable smooth delivery through the indicated delivery catheter and color-coded by size to allow for easy identification.
Microspheres are supplied sterile and packaged in 20 ml syringes with an approximate 7ml fill volume across the range. Embozene microspheres syringes are available in 2 ml microsphere volume for 40 -1300 µm.
# Indications for Use/ Intended Use
Embozene Microspheres are intended for embolization of arteriovenous malformations and hypervascular tumors, including uterine fibroids (UFE) and hepatoma, and for embolization of prostatic arteries (PAE) for symptomatic benign prostatic hyperplasia (BPH).
This device is not intended for neurovascular use.
# Comparison of Technological Characteristics
The Embozene Microspheres are substantially equivalent to the legally marketed Merit Medical Systems Inc. Embosphere® Microspheres, cleared by FDA (DEN160040), in regard to indications for use and similar in technological characteristics. Both devices are spherical, non-resorbable polymer microspheres which are delivered through a microcatheter to embolize the target vessel. There are differences in actual material composition; however, the materials of the Embozene Microspheres are well-known and widely used in the medical environment and are biocompatible for their intended use. The devices are available in similar sizes, Embozene (40-1300 µm) vs Embosphere (40-1200 µm). While Embozene is available in a slightly larger diameter, the size ranqe typically used to treat the prostatic artery are the same between both devices. The differences in materials and size range of these devices do not affect overall safety or effectiveness as demonstrated by clinical data. The Embozene Microspheres are identical to the reference device, Boston Scientific's Embozene Microspheres (K141209), in design, function, device materials, packaging, sterilization method, operating principle, and fundamental technology.
### Summary of Bench Tests
Non-clinical tests performed on the Embozene Microspheres, reference device, provide reasonable assurance that the proposed device has been designed and tested to assure conformance to the requirements for its intended use. A risk assessment was performed and demonstrates that the expanded indication did not require additional bench, biocompatibility, sterilization, or pre-clinical animal testing beyond what was required for the reference device. All existing provided in the previous Embozene Microspheres premarket submissions remains applicable.
Biocompatibility testing provided in the previous Embozene Microspheres premarket notification included: Cytotoxicity, Sensitization, Sub-Acute and Sub-Chronic Systemic Toxicity, Systemic Toxicity- Material Mediated Pyrogenicity, Genotoxicity (Bacterial Mutagenicity, In-vitro Chromosome Aberration, In-Vivo Micronucleous Assay), Implantation, Hemocompatibility (hemolysis, partial thromboplastin time, platelet/leukocyte counts). All tests passed, indicating that the device materials are biocompatible for its intended use.
Performance testing provided in the previous Embozene Microspheres premarket notification included: Microsphere size distribution, visual inspection, pH of transport solution, Osmolality of transport solution, Microsphere suspension, Catheter compatibility, and Elution Test: Determination of leachable substances by UV-Vis, HPLC, and ICP-MS. All tests passed demonstrating the device meets the predetermined performance requirements.
{5}------------------------------------------------
Published literature on four (4) preclinical animal studies assessed the biocompatibility, safety, and efficacy of Embozene Microspheres in comparison with the Embosphere predicate and other embolization devices. The devices were evaluated in a porcine kidney model to determine arterial distribution characteristics, inflammation and recanalization, impact on inflammatory tissue and foreign body reaction, and immunohistochemical inflammatory reactions. The results demonstrate a similar biological tissue response between the Embozene Microsphere predicate device which supports substantial equivalence.
Additionally, for labeling with MR compatibility, the proposed Embozene Microspheres were assessed and met requirements in the Guidance for Industry and Food and Drug Administration Staff Establishing Safety and Compatibility of Passive Implants in the Magnetic Resonance (MR) Environment.
# Clinical Data Summary
Clinical data from a prospective clinical study along with literature on the Embozene Microspheres supports the expanded indication for use and is considered to be substantially equivalent in safety and effectiveness to the predicate Merit Medical's Embosphere Microspheres.
A prospective clinical study at a single center evaluated 38 patients that underwent a prostatic artery embolization (PAE) using Embozene Microspheres for the treatment of symptomatic benign prostatic hyperplasia (BPH). The primary endpoints included safety and the symptomatic improvement of BPH symptoms through follow-up as measured by the International Prostate Symptom Score (IPSS) and the peak flow rate (Qmax). Safety was assessed from the reported adverse events. At baseline, patients underwent a clinical assessment for Qmax, post void residual (PVR), prostate volume (PV) and prostate specific antigen (PSA) levels. All patients also completed the validated symptom and quality of life (QoL) questionnaires. Clinical assessments included the validated IPSS, International Index of Erectile Function (IIEF) and quality of life (QOL) questionnaires.
# Results
Thirty-eight male evaluable subjects have been enrolled in the study with a mean age of 64.6±8.4 years. Bilateral embolization procedures were performed in 94.7% (36/38). Post procedural radiographic evidence demonstrated that the embolization procedures were 100% (38/38) successful without any cases of non-targeted embolization. Table 1 summarizes the baseline characteristics.
| Characteristic | Embozene Prospective Data<br>Mean±SD (n) |
|----------------------|------------------------------------------|
| Age Yrs | 64.6±8.4 (38) |
| IPSS | 21.2±6.7 (37) |
| Quality of Life | 4.4±1.1 (38) |
| Qmax (ml/s) | 10.9±6.9 (30) |
| Prostate Volume (ml) | 79.0±55.7 (38) |
| PSA (ng/ml) | 3.7±3.4 (38) |
# Table 1 Baseline Data
The average 6 month and 12 month IPSS scores demonstrated a symptomatic improvement over baseline at 46% and 47%, respectively. Additionally, 81.8% of the subjects improved at least 3 IPSS points at 12 months (Table 3). The baseline QoL score of 4.4 (mostly dissatisfied) improved to a 2.3 (mostly satisfied) at 3 months which was maintained through 12 months. A prostate volume reduction was also maintained through the follow up with an increase of the peak flow rates (Qmax).
{6}------------------------------------------------
| Table 2 Mean Results for Baseline, 3 Months, 6 Months and 12 Months | | | | | |
|---------------------------------------------------------------------|---------------------------|----------------|----------------|----------------|--|
| | Embozene Prospective Data | | | | |
| Characteristic | Baseline | 3 months | 6 months | 12 months | |
| | Mean±SD(n) | Mean±SD (n) | Mean±SD (n) | Mean±SD (n) | |
| IPSS | 21.2±6.7 (37) | 10.8±7.7 (36) | 11.5±8.1 (37) | 11.3±6.3 (23) | |
| Quality of Life | 4.4±1.1 (38) | 2.3±1.2 (33) | 2.1±1.3 (36) | 2.3±1.4 (23) | |
| Qmax (ml/s) | 10.9±6.9 (30) | 17.6±16.8 (12) | 15.4±7.7 (23) | 18.5±10.2 (10) | |
| Prostate Volume (ml) | 79.0±55.7 (38) | 69.3±35.1 (7) | 65.1±41.4 (33) | 56.8±24.3 (5) | |
# Table 3 Proportion of Patients with at least 1 and 3 point IPSS Improvement
| | Embozene Prospective Data | | |
|----------------------------------|---------------------------------|---------------------------------|---------------------------------|
| IPSS Points | 3 months<br>% (n/N)<br>95% CI | 6 months<br>% (n/N)<br>95% CI | 12 months<br>% (n/N)<br>95% CI |
| At least 1 point<br>improvement | 85.7% (30/35)<br>[69.7%, 95.2%] | 88.9% (32/36)<br>[73.9%, 96.9%] | 95.5% (21/22)<br>[77.2%, 99.9%] |
| At least 3 points<br>improvement | 82.9% (29/35)<br>[66.4%, 93.4%] | 80.6% (29/36)<br>[64.0%, 91.8%] | 81.8% (18/22)<br>[59.7%, 94.8%] |
The adverse events are presented in Table 4 by system organ class (SOC) and Preferred Term (PT). There were no serious adverse events or device related events reported. The investigative site did not report any post-prostatic artery embolization syndrome events.
{7}------------------------------------------------
| System/Organ Class | Preferred Term | Total<br>Events<br>(Subjects) | Rate of<br>Subjects<br>with Event |
|-------------------------------------------------------|----------------------|-------------------------------|-----------------------------------|
| Total | Total | 22 (15) | 39.5% |
| Renal and urinary<br>disorders | Total | 12 (9) | 23.7% |
| | Hematuria | 3 (3) | 7.9% |
| | Dysuria | 2 (2) | 5.3% |
| | Micturition urgency | 2 (2) | 5.3% |
| | Urine flow decreased | 2 (2) | 5.3% |
| | Pollakiuria | 1 (1) | 2.6% |
| | Urinary incontinence | 1 (1) | 2.6% |
| | Urinary retention | 1 (1) | 2.6% |
| Infections and<br>infestations | Total | 3 (3) | 7.9% |
| | Cystitis | 2 (2) | 5.3% |
| | Penile infection | 1 (1) | 2.6% |
| Nervous system<br>disorders | Total | 2 (2) | 5.3% |
| | Paraesthesia | 2 (2) | 5.3% |
| Reproductive system and<br>breast disorders | Total | 2 (2) | 5.3% |
| | Penis disorder | 1 (1) | 2.6% |
| | Spermatic cord pain | 1 (1) | 2.6% |
| Blood and lymphatic<br>system disorders | Total | 1 (1) | 2.6% |
| | Leukocytosis | 1 (1) | 2.6% |
| Musculoskeletal and<br>connective tissue<br>disorders | Total | 1 (1) | 2.6% |
| | Muscle tightness | 1 (1) | 2.6% |
| Psychiatric disorders | Total | 1 (1) | 2.6% |
| | Sleep disorder | 1 (1) | 2.6% |
### Conclusion
Based on the indications for use, technological characteristics, safety and performance testing, and clinical evidence, the Embozene Color-Advanced Microspheres have been shown to support safety and effectiveness and are considered to be substantially equivalent to Embosphere Microspheres when indicated for embolization of prostatic arteries for symptomatic benign hyperplasia (BPH),
The Embozene Microspheres demonstrates compliance with the Special Controls under 21 CFR 876.5550 for the expanded indication for use and 21 CFR 870.3300.
In addition, the MR Compatability meets the requirements of the Guidance Document Establishing Safety and Compatibility of Passive Implants in the Magnetic Resonance (MR) Environment.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.