SERVO-U Ventilator System version 2.1; SERVO-n Ventilator System version 2.1
K180098 · Maquet Critical Care AB · CBK · Mar 19, 2019 · Anesthesiology
Device Facts
Record ID
K180098
Device Name
SERVO-U Ventilator System version 2.1; SERVO-n Ventilator System version 2.1
Applicant
Maquet Critical Care AB
Product Code
CBK · Anesthesiology
Decision Date
Mar 19, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5895
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The SERVO-U ventilator system is: intended for respiratory support, monitoring and treatment of neonatal, pediatric and adult patients to be used only by healthcare providers to be used only in professional healthcare facilities and for transport within these facilities For NAVA and Edi monitoring, it is in addition intended: to provide monitoring of the patient's breathing drive to improve synchrony between the ventilator system and patient when the electrical signal from the brain to the diaphragm is active for use on all patients with no contraindication for insertion/exchange of a nasogastric tube The SERVO-n ventilator system is: intended for respiratory support, monitoring and treatment of neonatal and pediatric patients to be used only by healthcare providers to be used only in professional healthcare facilities and for transport within these facilities For NAVA and Edi monitoring, it is in addition intended: to provide monitoring of the patient's breathing drive to improve synchrony between the ventilator system and patient when the electrical signal from the brain to the diaphragm is active for use on all patients with no contraindication for insertion/exchange of a nasogastric tube
Device Story
SERVO-U/n 2.1 is a continuous ventilator system comprising a patient unit (gas mixing/delivery) and user interface. It delivers controlled or supported breaths via constant flow, constant pressure, or pressure proportional to the patient's electrical activity of the diaphragm (Edi) signal (NAVA mode). Used in professional healthcare facilities and during transport by healthcare providers. Inputs include gas supply, patient pressure/flow sensors, and Edi signal via nasogastric catheter. Outputs include mechanical ventilation, visual/audible alarms, and monitoring data. Features include integrated nebulizer control, optional CO2/Y-sensor monitoring, and battery backup. Clinical benefit includes respiratory support and improved patient-ventilator synchrony. The device assists clinicians in managing ventilation settings and monitoring patient breathing drive.
Clinical Evidence
No clinical investigation performed. Substantial equivalence supported by bench testing, design verification (code review, unit/integration/system tests, regression testing), and adherence to recognized standards (IEC 60601-1, ISO 80601-2-12, etc.). Biocompatibility evaluated per ISO 10993-1.
Technological Characteristics
Continuous ventilator; gas mixing/delivery unit with user interface. Materials: biocompatible per ISO 10993-1. Sensing: pressure, flow, Edi (electrical activity of diaphragm). Connectivity: standalone with battery backup. Software: rule-based ventilation/alarm algorithms. Standards: IEC 60601-1, IEC 60601-1-2, ISO 80601-2-12, ISO 80601-2-55, AAMI ANSI IEC 60601-1-8.
Indications for Use
Indicated for respiratory support, monitoring, and treatment of neonatal, pediatric, and adult patients (SERVO-U) or neonatal and pediatric patients (SERVO-n). Includes NAVA/Edi monitoring for patients without contraindications for nasogastric tube insertion. For use by healthcare providers in professional facilities and during transport.
Regulatory Classification
Identification
A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
March 19, 2019
Maquet Critical Care AB % Mark Dinger Sr. Regulatory Affairs Specialist Maquet Medical System USA 45 Barbour Pond Drive Wayne, New Jersey 07470
Re: K180098
Trade/Device Name: SERVO-U/n Ventilator System 2.1 Regulation Number: 21 CFR 868.5895 Regulation Name: Continuous Ventilator Regulatory Class: Class II Product Code: CBK Dated: February 11, 2019 Received: February 15, 2019
Dear Mark Dinger:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be avare that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# James J. Lee -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K180098
Device Name SERVO-U/n Ventilator System 2.1
| Indications for Use (Describe) | | |
|--------------------------------|--|--|
| | | |
- The SERVO-U ventilator system is:
- · intended for respiratory support, monitoring and treatment of neonatal, pediatric and adult patients
- to be used only by healthcare providers
- · to be used only in professional healthcare facilities and for transport within these facilities
For NAVA and Edi monitoring, it is in addition intended:
- · to provide monitoring of the patient's breathing drive
- · to improve synchrony between the ventilator system and patient when the electrical signal from the brain to the diaphragm is active
- · for use on all patients with no contraindication for insertion/exchange of a nasogastric tube
- The SERVO-n ventilator system is:
- · intended for respiratory support, monitoring and treatment of neonatal and pediatric patients
- to be used only by healthcare providers
- to be used only in professional healthcare facilities and for transport within these facilities
- For NAVA and Edi monitoring, it is in addition intended:
- · to provide monitoring of the patient's breathing drive
- · to improve synchrony between the ventilator system and patient when the electrical signal from the brain to the diaphragm is active
- · for use on all patients with no contraindication for insertion/exchange of a nasogastric tube
Type of Use (Select one or both, as applicable)
| <span> ☑ Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
|-----------------------------------------------------------------------------|----------------------------------------------------------------------------|
|-----------------------------------------------------------------------------|----------------------------------------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) SUMMARY
# 510(k) SUMMARY as required by section 21 CFR 807.92
# Submitter Name & Address
Maquet Critical Care AB Röntgenvägen 2 SE-171 54 Solna, Sweden Phone: (011) 46 10 335 7300
| Contact Persons for this submission: | Jerker Åberg<br>Regulatory Affairs Manager<br>Phone: direct: (011) 46 10 335 7300<br>Email: jerker.aberg@getinge.com |
|--------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Application Correspondent: | Mark Dinger<br>Sr. Regulatory Affairs Specialist<br>Maquet Medical Systems USA<br>45 Barbour Pond Drive<br>Wayne, NJ 07470<br>Email: mark.dinger@getinge.com<br>Phone: 973-709-7691<br>Fax: 973-909-9954 |
| 510(k) Number | K180098 |
Date prepared: March 18, 2019
| Trade Name: | Model: | Model no: |
|---------------------------------|---------|-----------|
| SERVO-U/n Ventilator System 2.1 | SERVO-U | 66 94 800 |
| | SERVO-n | 66 88 600 |
# Device Classification
| Common Name | Classification<br>Number | Class | Regulation Number |
|--------------------------------------|--------------------------|-------|-------------------|
| Ventilator, continuous, facility use | CBK | II | 21 CFR 868.5895 |
# Predicate Device Identification
| Legally marketed devices to which equivalence is being claimed | 510(k) # |
|----------------------------------------------------------------|----------|
| SERVO-U/n Ventilator System version 1.1 | K151814 |
{4}------------------------------------------------
### Reference Device Identification
| Legally marketed devices to which equivalence is being claimed | 510(k) # |
|----------------------------------------------------------------|----------|
| AIRVO 2 HUMIDIFIER, MYAIRVO 2 HUMIDIFIER | K131895 |
## Device Description
The SERVO-U/n 2.1 is available in two models, SERVO-U and SERVO-n. The SERVO-U/n 2.1 consists of a Patient Unit where gases are mixed and administered, and a User Interface where the settings are made and ventilation is monitored.
The SERVO-U/n 2.1 is based on the cleared predicate device SERVO-U/n 1.1 (K151814), with some improvements. The ventilation modes in the SERVO-U/n 2.1 are similar as in the predicate device, even though the standard configurations of available modes and optional modes differ between the devices, i.e. SERVO-U, SERVO-n and the cleared predicate device SERVO-U/n 1.1.
The ventilator delivers controlled or supported breaths to the patient, with constant flow, constant pressure or pressure proportional to the Edi signal (the electrical activity of the diaphragm) of the patient, using a set oxygen concentration.
SERVO-U/n contains a dedicated controller circuit for the Aerogen Pro and Solo nebulizers (included as standard).
Accessories for CO2 monitoring and flow and pressure measurements at the Y piece (Y sensor) are integrated as options.
The SERVO-U/n Ventilator System will produce visual and audible alarms if any parameter varies beyond pre-set or default limits and produce alarm recordings.
The system contains provisions for battery modules to supply the system in the case of mains power failure or during intra-hospital transport.
#### System parts:
The SERVO-U/n Ventilator System consists of the following parts:
- · User interface, where user interactions are performed.
- · Patient unit with all connections to the patient, to power and gases.
- · Mobile cart, on wheels, for using the ventilator on either the left or the right side of the patient.
{5}------------------------------------------------
## Indications for Use The SERVO-U ventilator system is:
- intended for respiratory support, monitoring and treatment of neonatal, pediatric and adult patients
- to be used only by healthcare providers ●
- to be used only in professional healthcare facilities and for transport within these . facilities
For NAVA and Edi monitoring, it is in addition intended:
- to provide monitoring of the patient's breathing drive ●
- to improve synchrony between the ventilator system and patient when the . electrical signal from the brain to the diaphragm is active
- for use on all patients with no contraindication for insertion/exchange of a nasogastric tube
# The SERVO-n ventilator system is:
- intended for respiratory support, monitoring and treatment of neonatal and . pediatric patients
- to be used only by healthcare providers ●
- to be used only in professional healthcare facilities and for transport within these . facilities
For NAVA and Edi monitoring, it is in addition intended:
- to provide monitoring of the patient's breathing drive ●
- to improve synchrony between the ventilator system and patient when the . electrical signal from the brain to the diaphragm is active
- . for use on all patients with no contraindication for insertion/exchange of a nasogastric tube
{6}------------------------------------------------
The second part of the indications for use is related to the NAVA (Neurally Adjusted Ventilatory Assist) ventilation mode which is a supported mode for SERVO-U/n that uses the Edi signal (the electrical activity of the diaphragm) as an addition to the flow/pressure trigger to synchronize the patient efforts with the onset and cycle off of supported breaths. NAVA is available in invasive and non-invasive modes.
#### Intended use of the Device
Intended use is identical to indications for use.
#### CHANGES AND SUBSTANTIAL EQUIVALENCE DISCUSSION 1.1
#### Comparison of Intended Use
The intended use/indication for use for the proposed device, SERVO-U/n 2.1 Ventilator System, is identical to the predicate device SERVO-U/n 1.1 (K151814).
#### Comparison of Technology Characteristics
The SERVO-U/n 2.1 is based on the cleared predicate device SERVO-U/n 1.1 (K151814), with some improvements noted in the description below. Software algorithms for ventilation and alarms are re-used. Furthermore, the ventilation modes in the SERVO-U/n 2.1 are similar as in the cleared predicate device, even though the standard configurations of available modes and optional modes differ between the devices.
The following changes have been made to SERVO-U/n 2.1 compared to the cleared predicate device SERVO-U/n 1.1 (K151814):
- Software and the Graphical User Interface (GUI) have been updated to include ● the lock/unlock functionality.
- High Flow Therapy (HFT) has been added as a new optional breathing therapy.
- SERVO Compass has been added as a visual clinical support tool to facilitate . protective ventilation.
- Implementation of Volume Control improvements including decelerating flow ● and de-activation of flow adaptation.
- . Implementation of NAVA improvements by removing NAVA(PS) as backup and some minor additional changes.
- Addition of configuration SERVO-U Adult. This adds a possibility to market ● SERVO-U with a SW configuration with fewer ventilation modes compared to predicate device.
- . Changes in Alarm sound to comply with standard AAMI ANSI IEC 60601-1-8:2006 + A1:2012. Verification to comply with A1:2012 of IEC 60601-1:2005 and A1:2012 of AAMI ANSI IEC 60601-1-8:2006 , updated test reports and CB certifications available.
{7}------------------------------------------------
- . Changes of circuit board for expiratory flow measurement, including HW and SW updates, due to LTB. Renaming from PC2004 to PC2024. Functionality is unchanged.
- Material changes in the filter in the patent gas pathway. Addition of alternative pressure sensors in gas pathway.
- Cleaning instructions have been updated with additional cleaning agents. ●
- Minor usability improvements for alarm handling
- The battery runtime has been increased in the IFU from 45 min to 1 hour. ●
- . Minor improvements of the PRVC mode.
- Improvements of internal software process interaction and supervision. ●
# Similarities and differences
The subject device is similar in all respects, except from what is stated above, to the predicate device.
# Non-clinical Testing and Performance
Support for the substantial equivalence of the SERVO-U/n 2.1 to the cleared predicate device SERVO-U/n 1.1 (K151814) is provided as a result of design verification and bench testing.
The design verification activities consist of:
- Code review and static code analysis
- Unit tests
- Integration tests ●
- System tests (including safety related functions from risk analysis) .
- Free User Testing (FUT) .
- . Regression testing
- Verification of applicable product standards .
The following product standards are included in the verification:
- ANSI/AAMI ES60601-1:2005/(R)2012 and A1:2012. Recognition Number 19-4. ●
- . IEC 60601-1-2 Edition 3: 2007-03, Recognition Number 19-1.
- IEC 60601-1-2:2014 RFID Immunity \ AIM Standard 7351731 Rev 1 ●
- AAMI ANSI IEC 60601-1-8:2006 & A1:2012, Recognition Number 5-92. ●
- ISO 80601-2-12:2011, Recognition Number 1-98. ●
- ISO 80601-2-55:2011, Recognition Number 1-96. .
- . EN13544-1:2007. this standard is not recognized by FDA.
Biocompatibility evaluation of the SERVO-U/n 2.1 is in accordance with AAMI / ANSI / ISO 10993-1:2009, recognition number 2-156 included the extent of recognition. Biocompatibility testing (i.e. cytotoxicity, sensitization, genotoxicity and implantation) including extractables and leachables and evaluation/risk assessment has been performed
{8}------------------------------------------------
regarding exposure to volatile organic compounds, particulate matter, colorants and leachable (in case of condensate).
The connector to High-Pressure gas is in accordance with CGA V-5:2008, Recognition Number 1-81.
Design validation has been performed in order to ensure that the product meets both its intended use and user needs, including usability.
Bench testing of waveforms have been performed to verify substantial equivalence to the predicate device.
Design verification and validation have demonstrated that the SERVO-U/n 2.1 performs within its specifications and within the limits of the applied product performance standards.
## Clinical Investigation
No clinical investigation has been performed since it has been concluded based on literature data, state of the art knowledge and applicable product standards that SERVO-U/n 2.1 has no new clinical aspects or risks which are not already discussed and evaluated in the 510(k) submission for the cleared predicate device SERVO-U/n 1.1 (K151814).
#### 1.1.1 Conclusion for substantial equivalence
SERVO-U/n 2.1 and the predicate device SERVO-U/n 1.1 (K151814) have identical intended use. There are no new type questions of safety and effectiveness for SERVO-U/n 2.1 as compared to the predicate device SERVO-U/n 1.1 (K151814). MAQUET has conducted risk analysis and performed the necessary verification and validation activities to demonstrate that the design outputs meet the design input requirements and the appropriate product standards. MAQUET concludes that the performance data for SERVO-U/n 2.1 shows that it is substantially equivalent to the cleared predicate device SERVO-U/n 1.1 (K151814).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.