Source of illumination for curing photo-activated dental restorative materials and adhesives.
Device Story
Portable dental LED curing light system; provides illumination for polymerization of photo-activated dental materials. System includes handpiece, charging stand, barrier sleeves, light shields, and power adapter. Handpiece features aluminum construction, rechargeable battery, and interchangeable LED light heads (395-480nm or 440-480nm). Operator uses LCD display and two buttons to select timer and curing modes (plasma or pulse). Device used in dental clinics by dental professionals. Light shield and barrier sleeves provide eye protection and infection control. High-power LED technology delivers curing light to restorative materials; facilitates efficient polymerization. Benefits include improved ease of use via modular swivel head design and effective curing across various dental material photoinitiator ranges.
Clinical Evidence
Bench testing only. Performance evaluated per FDA guidance 'Dental Curing Lights – Premarket Notification [510(k)] Submissions'. Testing included optical irradiance, spectral irradiance, curing time, curing modes, heat management, electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), and ANSI/ADA 48-2 compliance. Software and hardware verification/validation and risk analysis were performed.
Technological Characteristics
Portable LED curing light; aluminum handpiece body; rechargeable lithium-ion battery. Interchangeable LED heads; peak wavelengths 395-480nm or 440-480nm. Microprocessor-controlled user interface with LCD and two buttons. Compliance: IEC 60601-1, IEC 60601-1-2, ANSI/ADA 48-2. Non-sterile; uses barrier sleeves for infection control.
Indications for Use
Indicated for use as a source of illumination for curing photo-activated dental restorative materials and adhesives in dental patients.
Regulatory Classification
Identification
An ultraviolet activator for polymerization is a device that produces ultraviolet radiation intended to polymerize (set) resinous dental pit and fissure sealants or restorative materials by transmission of light through a rod.
{0}------------------------------------------------
# 510(k) Summary for FUSION Plus Curing Lights
#### 1. APPLICANT
K180065
Dentli
DentLight Inc. 1825 Summit Ave., Suite 210 Plano, TX 75074
Phone: 972-889-8857 Fax: 972-346-6550 Contact Person: Richard Liu Date Prepared: April 19, 2019
#### 2. DEVICE NAME
Proprietary Name: FUSION Plus Curing Light Common/Usual Name: Dental Curing Light Classification Name: Ultraviolet activator for polymerization (872.6070) Product Code: EBZ
#### 3. PRIMARY PREDICATE DEVICE
Valo (K083647), manufactured by Ultradent Products, Inc
#### REFERENCE DEVICE
FUSION (K052593), manufactured by DentLight Inc.
#### 4. INTENDED USE / INDICATIONS FOR USE
Source of illumination for curing photo-activated dental restorative materials and adhesives.
# 5. DEVICE DESCRIPTION
Fusion Plus Curing Light is a portable dental LED curing light system. It provides a source of illumination for polymerization of photo-activated dental materials. The principle of operation is the same as the predicate devices: illumination for polymerization or curing of photo-activated dental materials. It uses our patented Fusion optical technology in combination with high-power LED to deliver high efficiency and high-power curing light.
The system consists of a curing handpiece, a charging stand, 100 barrier sleeves, light shields, a power adapter, and instructions for use. The handpiece is constructed of aluminum and consists of a main control, a light head, and a rechargeable battery assembly. The main control houses the control circuitry with microprocessor and user interface consisting of LCD display and two control buttons. A mode button selects operation timer and two curing modes: plasma or pulse mode. A ON/OFF button activates and deactivates the light. The handpiece is non-sterile utilizing the barrier sleeves for infection control. Similar to the predicate devices, the light shield and curing cap absorb emitted light
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for "DentLight". The word "Dent" is in gray, while "Light" is in blue. There is a light blue glow above the "i" in "Light".
and perform the function of eye protection. The charging stand allows for the placement of the handpiece in a holder with electrical connection to the power adapter as the source for charging the rechargeable battery. FUSION Plus has an interchangeable LED light head. Depending on the light head connected with the main control or added later, the kit can be configured in three options:
| Trade Name | LED Peak Wavelength (nm) |
|-------------------------------------------------------------------------------------|--------------------------|
| FUSION Plus | 395-480 |
| FUSION Plus Standard | 440-480 |
| The instructions for use detail the use of each head in curing of dental materials. | |
# 6. DEVICE CHARACTERISTICS AND SUBSTANDTIAL EQUIVALENCE
FUSION Plus is substantially equivalent to the predicate device Valo (K083647) manufactured by Ultradent Products Inc. and FUSION (K052593) designed and manufactured by DentLight Inc. A comparison table with the predicate and reference devices lists key features and characteristics of the subject and predicate devices. It shows the product to be substantially equivalent to the predicate devices.
| Parameter | | FUSION Plus | Valo (Primary<br>Predicate Device) | FUSION<br>(Reference Device) |
|----------------------------------|------------|-----------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | | Source of illumination<br>for curing photo-<br>activated dental<br>restorative materials<br>and adhesives | Source of illumination<br>for curing photo-<br>activated dental<br>restorative materials<br>and adhesives | Photopolymerization of dental resins,<br>restorative composite materials, and<br>orthodontic brackets, bonding, and<br>sealing that are photo-polymerized in<br>the 440-480nm wavelength range |
| 510(k) Number | | K180065 | K083647 | K052593 |
| Product Code | | EBZ | EBZ | EBZ |
| Regulation Number | | 21 CFR 872.6070 | 21 CFR 872.6070 | 21 CFR 872.6070 |
| Device Class | | II | II | II |
| Common Name of<br>Classification | | Ultraviolet activator<br>for polymerization | Ultraviolet activator<br>for polymerization | Ultraviolet activator<br>for polymerization |
| Light Source | | High-power LED | High-power LED | High-power LED |
| Power Supply | | Input: 100-240V -<br>50/60Hz; Output: 5V<br>1-2A | Input: 100-<br>240VAC@50-60Hz;<br>Output: 12V 5A | Input: 100-240V - 50/60Hz; Output: 5V<br>1-2A |
| Peak Wavelength Range | | 395-480<br>(Plus head)<br>440-480 nm<br>(Plus Standard) | 395-480 nm | 440-480 nm |
| Curing Optical | Plasma | 4000 | 4500 | 2700 |
| Intensity | Pulse/high | 2000 | 1400 | 1350 |
| mW/cm² | Standard | NA | 1000 | NA |
| User Control Interface | | 2 Buttons | 2 Buttons | 2 Buttons |
| Weight (grams) | | 112 | 75 grams | 111 |
| Battery | | Lithium ion | NA | Lithium ion |
| Structure | | Ergonomic wand | Ergonomic wand | Ergonomic wand |
| Materials | | Aluminum body,<br>plastic barrier control<br>and light shield | Aluminum body,<br>plastic barrier control<br>and light shield | Aluminum body, plastic barrier control<br>and light shield |
| Compliance Standards | | IEC 60601-1; IEC<br>60601-1-2 | IEC 60601-1; IEC<br>60601-1-2 | IEC 60601-1; IEC 60601-1-2 |
{2}------------------------------------------------
Similarities between the FUSION Plus Curing Light and Valo:
- · The FUSION Plus has the same indications for use as the primary predicate device Valo.
- · The FUSION Plus has substantially equivalent curing modes as the primary predicate device Valo.
- · The FUSION Plus has substantially equivalent performance characteristics as the primary predicate device Valo.
- · The FUSION Plus is substantially equivalent to the primary predicate device Valo in standards compliance.
### Differences between the FUSION Plus Curing Light and Valo:
The primary predicate Valo has a uni-body construction with a fixed aluminum body. The FUSION Plus has a modular aluminum body construction with a swivel interchangeable light head. Plus Head and Plus Standard Head. The Plus Head has a wavelength range of 395- 480nm that is the same as the primary predicate Valo and wider than the reference device FUSION. These wavelengths cover VLC materials with photoinitiator sensitive to all photoactivated dental materials (395- 480mm). The difference in construction improves the operator's ease of use and does not raise any substantial equivalence concerns as it has the same wavelengths of the primary predicate Valo.
## Similarities between the FUSION Plus Curing Light and the reference device FUSION:
- · The FUSION Plus and the reference device FUSION has the same construction. FUSION Plus Standard Head has the same center wavelengths of 440 nm - 480 nm as the reference device FUSION. The wavelengths cover VLC materials with photoinitiators sensitive to photoactivated dental materials in the 440 nm - 480 nm (CQ and PPD).
- · The FUSION Plus and the reference device FUSION use the same optoelctronic design, FUSION optical beam technology, and construction materials except the Plus Head LED.
Both FUSION Plus head and FUSION Plus Standard head are powered and controlled by the same main control under the same condition with the only difference being the choice of LED and its placement. These differences do not raise concerns about safety and effectiveness as the performances are similar to the primary predicate device and the reference device.
# 7. NON-CLINICAL PERFORMANCE TESTING
The following non-clinical tests were conducted to evaluate the functionality, performance, safety, and substantial equivalence of the FUSION Plus Curing Light to the predicate device as suggested by the FDA's guidance for industry and FDA staff titled "Dental Curing Lights – Premarket Notification [510(k)] Submissions" issued on March 27, 2006:
- · device performance
- o optical irradiance
- o spectral irradiance
- o curing time
- o curing modes and safety controls
- o heat management
- · software and hardware verification and validation,
- · risks to health and traceability analysis
- · electrical safety and electromagnetic compatibility
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for DentLight. The word "Dent" is in gray, and the word "Light" is in blue. There is a small, light blue circle above the "i" in "Light."
FUSION Plus complies with the electrical and optical safety requirements as the predicate devices: IEC 60601-1, IEC 60601-1-2, and ANSI/ADA 48-2.
## 8. CONCLUSION
Based on similarities in indications for use and technology together with results from performance testing, we believe that FUSION Plus Curing Light is substantially equivalent to the predicate devices.
{4}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known) K180062
Device Name FUSION Plus Curing Light
Indications for Use (Describe) Source of illumination for curing photo-activated dental restorative materials and adhesives
Trade Name FUSION Plus FUSION Plus Standard
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
April 19, 2019
DentLight Inc. Richard Liu President 1825 Summit Ave. Ste 210 Plano, Texas 75074
Re: K180065
Trade/Device Name: FUSION Plus Curing Light Regulation Number: 21 CFR 872.6070 Regulation Name: Ultraviolet Activator For Polymerization Regulatory Class: Class II Product Code: EBZ Dated: March 19, 2019 Received: March 20, 2019
Dear Richard Liu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809 : medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
#### Enclosure
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.