ESOTEST MULTI Esophageal Temperature Probe and Temperature Monitoring System
K180047 · Fiab Spa · FLL · Nov 8, 2018 · General Hospital
Device Facts
Record ID
K180047
Device Name
ESOTEST MULTI Esophageal Temperature Probe and Temperature Monitoring System
Applicant
Fiab Spa
Product Code
FLL · General Hospital
Decision Date
Nov 8, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2910
Device Class
Class 2
Indications for Use
The ESOTEST MULTI Temperature Monitoring System is composed of ESOTEST MULTI Monitor and ESOTEST MULTI Probe and is intended for the continuous detection, measurement and visualization (in °C) of esophageal temperature. The intended environments of use are operating rooms and interventional electrophysiology rooms. The ESOTEST MULTI Monitor must be used in conjunction with the ESOTEST MULTI Probe.
Device Story
System comprises ESOTEST MULTI Monitor, ESOTEST MULTI Probe, and patient cable. Probe features 7 French body with 11 French sensors (5 or 7 sensors total) for multi-location esophageal temperature detection. Thermocouple T-type sensors convert temperature to voltage; monitor processes signal to display numeric values and scrolling plots on LCD touch screen. User sets upper/lower alarm thresholds; device triggers visual (flashing red circles) and audible alarms if thresholds exceeded. Used in ORs and electrophysiology rooms by clinicians. Provides real-time temperature visualization to assist in monitoring patient thermal status during procedures.
Clinical Evidence
Bench testing only. Evaluated per ISO 80601-2-56:2009 (accuracy ±0.3°C, precision 0.1°C, 1s response time). Biocompatibility testing per ISO 10993-1 (cytotoxicity, sensitization, intracutaneous reactivity). Electrical safety per IEC 60601-1 and EMC per IEC 60601-1-2. Sterilization validated per ISO 11135-1.
Technological Characteristics
Materials: Polyurethane and Stainless Steel (AISI 304). Sensing: Thermocouple T-type. Form factor: 7Fr probe body, 11Fr sensors. Connectivity: Standalone monitor with external power supply. Software: Moderate level of concern. Sterilization: Ethylene Oxide (EO) gas per ISO 11135-1.
Indications for Use
Indicated for continuous esophageal temperature monitoring in patients in operating rooms and interventional electrophysiology rooms.
Regulatory Classification
Identification
A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.
Special Controls
(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.
Predicate Devices
ESOTEST Esophageal Temperature Probe and Temperature Monitoring System (K123361)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left, and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a sans-serif font.
November 8, 2018
Fiab SpA Francesco Batistini Responsible for regulatory affairs Via Costoli, 4 Vicchio, 50039 It
Re: K180047
Trade/Device Name: ESOTEST MULTI Esophageal Temperature Probe and Temperature Monitoring System Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: Class II Product Code: FLL Dated: August 9, 2018 Received: October 9, 2018
Dear Francesco Batistini:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Alan M.
Stevens -S
Tina Kiang, Ph.D. for Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K180047
Device Name
ESOTEST MULTI Esophageal Temperature Probe and Temperature Monitoring System
Indications for Use (Describe)
The ESOTEST MULTI Temperature Monitoring System is composed of ESOTEST MULTI Monitor and ESOTEST MULTI Probe and is intended for the continuous detection, measurement and visualization (in °C) of esophageal temperature. The intended environments of use are operating rooms and interventional electrophysiology rooms.
The ESOTEST MULTI Monitor must be used in conjunction with the ESOTEST MULTI Probe.
Type of Use (Select one or both, as applicable)
| <span></span> Prescription Use (Part 21 CFR 801 Subpart D) | <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|------------------------------------------------------------|-----------------------------------------------------------|
|------------------------------------------------------------|-----------------------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## K180047 510(k) Summary
| Submitter: | Fiab SpA<br>Via Costoli, 4<br>50039 Vicchio<br>Florence, Italy |
|----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|
| Application Correspondent: | Silvia Calabrò.<br>Title: Official Correspondent<br>Phone: (39) 055 849 79 216<br>Fax: (39) 055 849 79 87<br>E-mail: regulatory@fiab.it |
| Preparation Date: | November 7, 2018 |
| Trade Name: | ESOTEST MULTI Esophageal Temperature Probe<br>and Temperature Monitoring System |
| Common or Usual Name: | Thermometer, Electronic, Clinical |
| Regulation Name: | Clinical Electronic Thermometer. |
| Regulation Number: | 21 CFR 880.2910 |
| Product Code: | FLL |
| Device Class: | Class II |
| Predicate Device: | K123361 – ESOTEST Esophageal Temperature<br>Probe and Temperature Monitoring System |
### Device Indications for Use:
The ESOTEST MULTI Temperature Monitoring System is composed of ESOTEST MULTI Monitor and ESOTEST MULTI Probe and is intended for the continuous detection, measurement and visualization (in °C) of esophageal temperature. The intended environments of use are operating rooms and interventional electrophysiology rooms.
The ESOTEST MULTI Monitor must be used in conjunction with the ESOTEST MULTI Probe.
### Device Description:
The ESOTEST MULTI Temperature Monitoring System is composed of the following device components:
- Monitor ESOTEST MULTI MONITOR
- Esophageal probe ESOTEST MULTI PROBE
{4}------------------------------------------------
## - Patient cable ESOTEST MULTI PATIENT CABLE
The LCD display of ESOTEST MULTI monitor shows the temperatures detected by the thermocouple sensors located in the probe, which is measuring the temperature of the patient's esophagus. Temperatures are displayed as numeric values and as scrolling plots. The user is able to set the value of an upper alarm threshold and that of a lower alarm threshold. If any measured temperature exceeds the upper alarm threshold or falls below the lower alarm threshold, ESOTEST MULTI monitor immediately emits visual and audible alarm signals.
The ESOTEST MULTI esophageal probe has a 7 French body equipped with 11 French sensors and is intended for the detection of esophageal temperature at multiple locations. The probe is connected to the monitor by ESOTEST MULTI patient cable.
| Element of<br>comparison | Subject Device: K180047FIAB<br>- ESOTEST MULTI<br>Esophageal Temperature<br>Probe and Temperature<br>Monitoring System | Primary Predicate: K123361 -<br>ESOTEST Esophageal<br>Temperature Probe and<br>Temperature Monitoring<br>System | Comparison<br>(Same/Different) |
|-----------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Thermometer type | Esophageal | Esophageal | Same |
| Indications for use | The ESOTEST MULTI<br>Temperature Monitoring System<br>is composed of ESOTEST<br>MULTI Monitor and ESOTEST<br>MULTI Probe and is intended<br>for the continuous detection,<br>measurement and visualization<br>(in °C) of esophageal<br>temperature. The intended<br>environments of use are<br>operating rooms and<br>interventional electrophysiology<br>rooms. | ESOTEST Probe is intended for<br>continuous esophageal<br>temperature monitoring.<br>ESOTEST Monitor is intended<br>for display continuous<br>temperature measurement (°C)<br>from 3 sensors temperature<br>probe. | Same<br>(both devices are<br>intended for the same<br>use as temperature<br>monitoring systems, but<br>the indications for use<br>of ESOTEST MULTI<br>are more detailed)) |
| Components | Temperature probe, Patient<br>cable, Monitor. The package of<br>the monitor includes an external<br>power supply and an<br>equipotential cable | Temperature probe, Interconnect<br>cable, Monitor. The package of<br>the monitor includes an<br>equipotential cable | Different<br>(the main difference is<br>that ESOTEST MULTI<br>is powered by an<br>external power supply) |
| Temperature<br>measurement | 0 - 75 | 25 - 45 | Different |
| Element of<br>comparison | Subject Device: K180047FIAB<br>- ESOTEST MULTI<br>Esophageal Temperature<br>Probe and Temperature<br>Monitoring System | Primary Predicate: K123361 -<br>ESOTEST Esophageal<br>Temperature Probe and<br>Temperature Monitoring<br>System | Comparison<br>(Same/Different) |
| range [°C] | | | |
| Number of<br>temperature<br>sensors | 5 or 7 | 3 | Different |
| Temperature<br>sensor type | Thermocouple T type(accuracy<br>±0.3°C) | Thermocouple T type (accuracy<br>±0.3°C) | Same |
| Measurement<br>presentation/User<br>Interface | LCD monitor<br>Touch screen monitor | LED display<br>Up and down buttons | Different |
| Alarm<br>temperature range<br>[°C] | 37-41 upper threshold (Tmax)<br>12-24 lower threshold (Tmin) | 36-41 upper threshold | Different |
| Alarm signal | Visual (flashing red circles on<br>the LCD display)<br>Audible (intermittent sound) | Visual (flashing red LEDs)<br>Audible (intermittent sound) | Same |
| Power<br>requirements | 100 - 240 Vac | 100-120/230 Vac | Different |
| Monitor<br>classification | I, CF, defib protected | I, CF, defib protected | Same |
| Dimensions [cm] | Monitor: 34(width) x25(height)<br>x6.5 (depth)<br>Patient cable (length): 290<br>Temperature probe (length and<br>diameter): 83; 7Fr body, 11Fr<br>sensors | Monitor: 36.7x24.9x11.0<br>Interconnect cable (length): 250<br>Temperature probe (length and<br>diameter): 95; 7Fr body, 11Fr<br>sensors | Different |
| Introduction | Esophageal (nose/throat) | Esophageal (nose/throat) | Same |
| Signal processing<br>and display | Actual temperature is a function<br>of the thermocouple voltage.<br>Temperature displayed in 0.1°C<br>increments. | Actual temperature is a function<br>of the thermocouple voltage.<br>Temperature displayed in 0.1°C<br>increments. | Different |
| Element of<br>comparison | Subject Device: K180047FIAB<br>- ESOTEST MULTI<br>Esophageal Temperature<br>Probe and Temperature<br>Monitoring System | Primary Predicate: K123361 -<br>ESOTEST Esophageal<br>Temperature Probe and<br>Temperature Monitoring<br>System | Comparison<br>(Same/Different) |
| | 1 input (single probe) available<br>5/7 sensors per probe<br>measurements and user-selected<br>alarm limits are displayed on<br>LCD monitor. | 1 input (single probe) available<br>3 sensors per probe<br>measurements and user-selected<br>alarm limit are displayed on<br>LED displays. | |
| Materials (patient<br>contacting) | Polyurethane and Stainless Steel<br>(SST) AISI 304 | Polyurethane and Stainless Steel<br>(SST) AISI 304 | Same |
| Ambient<br>temperature | +10°C to +40°C<br>Non condensed relative<br>humidity:<br>30% to 75% | +10°C to +40°C<br>Non condensed relative<br>humidity:<br>30% to 75% | Same |
| Accuracy | 0.3°C within rated output range<br>(ISO 80601-2-56:2009<br>requirements for clinical<br>thermometers) | 0.3°C within rated output range<br>(ISO 80601-2-56:2009<br>requirements for clinical<br>thermometers ) | Same |
| Precision and<br>repeatability [°C] | 0.1 | 0.1 | Same |
| Response time(1) | Both heating and cooling<br>response time are approximately<br>1 second | Both heating and cooling<br>response time are approximately<br>1 second | Same |
## Technological Characteristics and Substantial Equivalence:
{5}------------------------------------------------
{6}------------------------------------------------
## Substantial Equivalence Discussion
The ESOTEST MULTI Temperature Monitoring System is intended for the same use as the predicate device. The differences between the ESOTEST MULTI Temperature Monitoring System and the predicate device are listed and discussed in the table below:.
| Difference | Impact on safety and effectiveness |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ESOTEST MULTI makes use of an external power supply (included in the package of ESOTEST | The adoption of a medical grade external power supply and of a custom locking connector does not introduce new |
| Difference | Impact on safety and effectiveness |
| MULTI monitor) | questions of safety and effectiveness. It has been verified by<br>testing the power supply system together with the device<br>according to the requirements of the applicable Electrical<br>Safety standards, |
| ESOTEST MULTI has a larger measurement range | The larger measurement range does not introduce new<br>questions of safety and effectiveness. It has been verified<br>though performance accuracy testing. |
| ESOTEST MULTI probe has 5 to 7 sensors instead<br>of 3 as the probe of the predicate device | The greater number of sensors allows the device to monitor<br>a longer portion of esophagus temperature and does not<br>introduce new questions of safety and effectiveness. The<br>probes were verified through performance testing. |
| ESOTEST MULTI makes use of a touch screen<br>user interface instead of employing LED displays<br>and physical buttons as the predicate device | The adoption of an LCD display with adequate size and<br>brightness, along with the capacitive touch screen. This<br>does not introduce new questions of safety and<br>effectiveness. |
| ESOTEST MULTI has a lower alarm threshold in<br>addition to the upper alarm threshold of the<br>predicate device | The presence of a second alarm threshold allows the user to<br>employ the alarm both during heating and cooling<br>treatments. This does not introduce new questions of safety<br>and effectiveness. The alarm thresholds were validated<br>through performance testing and the software was<br>verified/validated. |
| ESOTEST MULTI can tolerate a slightly larger<br>mains supply voltage range | This does not introduce new questions of safety and<br>effectiveness. It has been verified by testing the device in<br>the mains supply rated voltage range, according to<br>requirements of applicable Electrical Safety standards |
| The main unit of ESOTEST MULTI is smaller and<br>lighter and the interconnection cables are longer<br>than the corresponding components of the predicate<br>device | The fact that the main unit is smaller and lighter does not<br>introduce new questions of safety and effectiveness. |
| ESOTEST MULTI measures and displays the<br>temperature detected by 5 to 7 sensors instead of 3<br>as the predicate device | The LCD display of the greater number of sensors does not<br>introduce new questions of safety and effectiveness. |
l Response time is defined as the mean value of the measured time intervals associated to temperature increase (tr) or to temperature decrease (tf) necessary to cover the 63.3% of the total temperature excursion.
{7}------------------------------------------------
# Non-Clinical Testing Summary:
Non-clinical tests were conducted to verify and validate that the proposed device meets requirement specifications and is substantially equivalent (SE) to the ESOTEST Esophageal Temperature Probe and Temperature Monitoring System. The device was subjected to the following non-clinical testing to demonstrate that it performs as intended:
{8}------------------------------------------------
- ISO 80601-2-56:2009 Medical electrical equipment. Part 2-56: Particular requirements for . basic safety and essential performance of clinical thermometers for body temperature measurement. Performance parameters that were evaluated are listed in the following table:
| Parameter | ESOTEST requirement |
|--------------------------------------------|------------------------------------------------------------------|
| Accuracy of the device | ±0.3°C<br>in the rated range +10 to +45°C |
| Precision in condition of<br>repeatability | 0.1°C |
| Response Time | 1 s<br>temperature step +27 to +43 °C |
| Alarm Threshold | Auditory and visual alarm is triggered<br>when outside threshold |
- Software documentation including verification & validation was provided in accordance with . FDA Guidance: Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices for a software with a moderate level of concern.
- . Verification/validation was completed for Electrical Safety according to the standard IEC 60601-1:2005 - Medical electrical equipment - Part 1: General requirements for basic safety and essential performance, and for EMC according to the standard IEC 60601-1-2:2014 - Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance -Collateral Standard: Electromagnetic disturbances - Requirements and tests
- The process for sterilizing ESOTEST MULTI probe is an EO gas cycle in sterilization chamber. . The process is validated according to the standard ISO 11135-1:2007 - Sterilization of health care products - Ethylene oxide - Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices. The product is released after each sterilization cycle only after having verified the sterility of the biological indicators (BIs) inserted in the load according to the validation protocol; BIs are tested according to the standard ISO 11737-2:2009 - Sterilization of medical devices - Microbiological methods - Part 2: Tests of sterility performed in the definition, validation and maintenance of a sterilization process.
- The patient contact device, ESOTEST MULTI probe, has been tested according to the ● requirements of the standard ISO 10993-1:2009 - Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process. Based on the contact categorization as "mucosal membrane" for a "limited exposure" duration time, as requested by ISO 10993-1 the following tests were performed: in vitro cytotoxicity test according to ISO 10993- 5:2009; sensitization test and intracutaneous reactivity test according to ISO 10993-10:2010.
- A risk analysis was conducted in accordance with ISO 14971: 2007 Medical devices -. Application of risk management to medical devices.
In all testing, the pre-determined acceptance criteria were met.
{9}------------------------------------------------
### Substantial Equivalence Conclusion
Differences between the intended use and technological characteristics of the subject device compared to the predicate do not raise different questions of safety and effectiveness. The performance of the device is supported by non-clinical testing and risk management activities. The ESOTEST MULTI Esophageal Temperature Probe and Temperature Monitoring System is Substantially Equivalent (SE) to the ESOTEST Esophageal Temperature Probe and Temperature Monitoring System, cleared under K123361.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.