# ACCUSYTE 3D Fiducial Marker (K180027)

_Radiation Products, LLC · IYE · Jul 18, 2018 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K180027

## Device Facts

- **Applicant:** Radiation Products, LLC
- **Product Code:** [IYE](/productcode/IYE.md)
- **Decision Date:** Jul 18, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.5050
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

The ACCUSYTE 3D Fiducial Markers are used to mark the location of a tumor cavity after the tumor has been removed. In addition, the Markers are indicated on situations where soft tissue needs to future medical procedure such as subsequent Radiation Therapy.

## Device Story

Implantable fiducial marker consisting of a permanent radiopaque component (Gold or Platinum/Iridium) attached to an absorbable polylactic acid suture; surgically implanted into soft tissue tumor cavity by physician; provides radiographically visible reference point for subsequent radiation therapy targeting; enables accurate localization of tumor site on kV CT, 2D Linac-based kilovoltage/megavoltage X-ray, and cone-beam CT; aids clinicians in radiation treatment planning to limit exposure to adjacent healthy tissue; suture absorbs within one year; device is MR Conditional.

## Clinical Evidence

Bench testing only. Performance data included biocompatibility (ISO-10993), physical testing (USP 30), and residual strength/absorption rate studies for the suture component. Visualization efficacy was validated using a phantom breast model across CT, KV, CBCT, and MV imaging systems. MR safety was established via non-clinical testing for 1.5T and 3T systems, demonstrating a maximum temperature rise of ≤2°C and defined artifact limits. No clinical human study data was provided; the document mentions no complications in a human study context, but this appears to be a reference to general clinical experience rather than a formal clinical trial.

## Technological Characteristics

Materials: Gold or Platinum/Iridium marker, Glycolide-co-Lactide absorbable suture. Dimensions: <0.5 cm marker. Principle: Radiographic visualization of implanted marker. Connectivity: None (passive implant). Sterilization: ETO (SAL 10^-6). MR Safety: MR Conditional (1.5T/3T).

## Regulatory Identification

A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.

## Predicate Devices

- CIVCO Suture-Type Marker ([K071614](/device/K071614.md))

## Reference Devices

- BioZorb Marker ([K143484](/device/K143484.md))
- Foosin WEGO-PGLA Absorbable Suture ([K130735](/device/K130735.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains two logos. The logo on the left is the Department of Health & Human Services-USA logo. The logo on the right is the FDA U.S. Food & Drug Administration logo. The FDA logo has a blue square with the letters FDA in white.

RADIATION PRODUCTS LLC

July 18, 2018

% Mr. Victor Pereira Principal VHP Consulting 7240 NW 63 Terrace PARKLAND FL 33067

Re: K180027

Trade/Device Name: ACCUSYTE 3D Fiducial Marker Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: IYE Dated: June 3, 2018 Received: June 13, 2018

Dear Mr. Pereira:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Michael D'Hara For

Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

#### Indications for Use

K 180027

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known)

Device Name ACCUSYTE 3D Fiducial Marker

Indications for Use (Describe)

The ACCUSYTE 3D Fiducial Markers are used to mark the location of a tumor cavity after the tumor has been removed.

In addition, the Markers are indicated on situations where soft tissue needs to future medical procedure such as subsequent Radiation Therapy.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 5. 510 (k) Summary

This 510 (k) summary information is being submitted in accordance with the requirements of 21 CFR 807.92.

Applicant: Surgical Radiation Products LLC

Date Prepared: April 29, 2018 (amended)

# Contact Person: John D Corbitt MD

Radiation Products LLC 142 JFK Drive Lake Worth FL 33462 Phone: (561) 379-6210 Fax: (561) 439-9902 e-mail: jdcorbitt@aol.com

## Trade Name: ACCUSYTE™ 3D Fiducial Marker

Common Name: Implantable Radiographic Marker Classification Name: Medical charged-particle radiation therapy system Classification Number: 892.5050 Device Classification: Class II Product Code: IYE Predicate Device: CIVCO Suture-Type Marker (K071614) Reference Devices: BioZorb Marker (K143484) Foosin WEGO-PGLA Absorbable Suture (K130735)

# Substantially Equivalent to:

The ACCUSYTE™ 3D Fiducial Marker is substantially equivalent in intended use, principle of operation, and technological characteristics to the CIVCO Suture-Type Marker (K071614), to the BioZorb Marker (K143484), and to the Foosin WEGO-PGLA Absorbable Suture (K130735).

### Description of the Device subject to Premarket Notification:

The ACCUSYTE™ 3D Fiducial Marker is an implantable marker comprised of an absorbable suture (polylactic acid) component which absorbs completely in less than one year, and a permanent component, a Platinum or Gold Marker attached to the suture by a crimping operation similar to the operation use to attach the Stainless-Steel Needle.

The ACCUSYTE™ 3D Fiducial Marker is provided Sterile for single use and is implantable.

{4}------------------------------------------------

# Indication for Use:

The ACCUSYTE™ 3D Fiducial Markers are used to mark the location of a tumor cavity after the tumor has been removed.

In addition, the Markers are indicated on situations where soft tissue needs to be marked for future medical procedure such as subsequent Radiation Therapy.

# Technical Characteristics:

The ACUSYTE™ 3D Fiducial Marker has similar physical and technical characteristics to the predicative and reference devices, as illustrated in the table below:

| Technical<br>Characteristic                  | Subject Device                                                                                                         | Primary<br>Predicate                                                                               | Reference<br>Device                                                                                     | Reference Device                                                    |
|----------------------------------------------|------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------|
|                                              | ACCUSYTE™ 3D<br>Fiducial Marker                                                                                        | CIVCO Suture-<br>Type Marker<br>(K071614)                                                          | BioZorb Marker<br>(K143484)                                                                             | WEGO-PGLA<br>Absorbable Suture<br>(K130735)                         |
| Overall<br>Technological<br>Characteristic   | Radiographically<br>visible<br>permanent<br>marker on an<br>absorbable<br>suture                                       | Radiographically<br>visible<br>permanent<br>marker on an<br>absorbable<br>suture                   | Radiographicall<br>y visible<br>permanent<br>marker element<br>in a bio-<br>absorbable<br>polymer space | Suture-N/A                                                          |
| Principle of<br>Operation                    | Marker is<br>secured on<br>tumor cavity<br>area for<br>radiographic<br>visualization of<br>tumor site after<br>removal | Marker is<br>positioned into<br>tissue site for<br>radiographic<br>visualization of<br>tissue site | Marker is<br>positioned into<br>tissue site for<br>radiographic<br>visualization of<br>tissue site      | Suture holds<br>marker in place<br>during tissue<br>healing process |
| Absorption<br>time for suture<br>and polymer | Less than 1 year<br>(suture)                                                                                           | Unknown for<br>suture                                                                              | Longer than 1<br>year<br>(Polymer)                                                                      | Suture- Less than 1<br>year                                         |
| Visualization<br>Compatibility               | ikV CT, 2D<br>Linac- based<br>kilovoltage and<br>megavolt-age X-<br>ray imaging<br>systems and kV                      | X-Ray<br>CT (presumed)                                                                             | Mammography,<br>Ultrasound, X-<br>Ray, and CT                                                           | Suture-N/A                                                          |

{5}------------------------------------------------

|                                       | cone-beam CT.                                                                                          |                                         |                                                              |                                                                   |
|---------------------------------------|--------------------------------------------------------------------------------------------------------|-----------------------------------------|--------------------------------------------------------------|-------------------------------------------------------------------|
| Materials of<br>Construction          | Gold or<br>Platinum/Iridium Marker<br>component,<br>absorbable<br>suture<br>(Glycolide-co-<br>Lactide) | Gold Marker and<br>absorbable<br>suture | Titanium<br>Markers and<br>absorbable<br>polymer<br>(spacer) | Suture-Glycolide-<br>co Lactide with<br>Stainless Steel<br>Needle |
| Overall Device<br>Length              | <0.5 cm<br>(Marker)                                                                                    | <5cm (Marker)                           | 2-5 cm (Marker)                                              | Suture-45 cm                                                      |
| Typical<br>Anatomic<br>Treatment Site | Soft tissue<br>tumor cavity site<br>including breast                                                   | Breast                                  | Soft tissue<br>including breast                              | Suture- soft tissue                                               |
| -Method of<br>Marker<br>Deployment    | Surgically<br>implanted                                                                                | Surgically<br>implanted                 | Manual, open<br>surgical                                     | Suture- surgical<br>placement                                     |
| Marker<br>Stability                   | Sutured in place                                                                                       | Sutured in place                        | Tissue retention                                             | Suture retains<br>marker in place<br>through healing<br>process   |
| How Provided                          | Sterile for single<br>use                                                                              | Sterile single use                      | Sterile single<br>use                                        | Suture-Sterile<br>Single Use                                      |
| Sterilization<br>Method               | ETO<br>validated cycle<br>with SAL of 10-6                                                             | Unknown                                 | Radiation                                                    | Suture- ETO<br>validated cycle<br>with SAL of 10-6                |

# Performance Data:

The Absorbable Suture component on the candidate device was tested as part of its 510 (k) submission (K130735) with the following tests:

- Bacterial Endotoxin (USP Pharmacopeia <85>) - Pyrogenicity
- Biocompatibility (ISO-10993) - suture component - Implant
- Physical Testing - USP 30 <861, <871, and <881 - suture component
- Residual Strength and Absorption Rate studies as outlined in FDA's Class II Special Controls Guidance Document: Surgical Sutures.

## Performance data for the ACCUSYTE™ 3D Fiducial Marker device included:

- . The Gold and Platinum/Iridium markers were bench tested for clarity of visualization using a phantom breasts model across computed tomography (CT), kilovoltage (KV), cone-beam CT (CBCT), and megavoltage (MV) linear accelerator imaging
- Radiographic Imaging 6-8 weeks after implantation
- Mammography imaging after 1-year post implantation

{6}------------------------------------------------

#### MRI Safety:

The ACCUSYTE™ 3 D Fiducial Markers are considered "MR Conditional"

Non-clinical testing has demonstrated the ACCUSYTE™3D Fiducial Marker device is MR Conditional. A patient with this device can be safely scanned in an MR system meeting the following conditions:

- Static magnetic field of 1.5-Tesla or 3-Tesla
- Maximum spatial gradient magnetic field of 4,000-gauss/cm (40-T/m)
- Maximum MR system reported, whole body averaged specific absorption rate (SAR) of 2- W/kg in the Normal Operating Mode
- Under the scan conditions defined above, the Accusyte™ 3D Fiducial marker is expected to produce a maximum temperature rise of ≤2℃ after 15-minutes of continuous scanning at 3 Tesla
- The image artifact caused by the Accusyte™ 3D Fiducial is not expected to extend more than 10-mm from this device when imaged with a gradient echo pulse sequence and a 3- Tesla MRI system.

The Accusyte™ 3D Fiducial and may be safely scanned with MRI under the conditions listed above. Scanning under different conditions may result in severe patient injury.

The ACCUSYTE™ 3D Fiducial Marker performance is equivalent to that of the predicate and reference devices in that it provides clear visualization of the tumor site, the candidate device secures the markers on soft tissue of the tumor cavity for a more accurate localization of the site, aiding in targeting the area for subsequent Radiation therapy and limiting exposure to adjacent tissue.

There were no complications reported from the candidate markers device during the human study, such as bleeding, pain, inflammation, or tissue reactions. Therefore; demonstrating that the candidate device is safe and effective for its intended use.

### Basis for Determination of Substantial Equivalence:

Upon reviewing the safety and efficacy information provided in this submission and comparing intended use, principle of operation, and overall technological characteristics, the ACCUSYTE™ 3D Fiducial Marker is determined by Radiation Products LLC to be substantially equivalent to the existing legally marketed predicate device CIVCO Suture Type Marker.

---

**Source:** [https://fda-staging.innolitics.com/device/K180027](https://fda-staging.innolitics.com/device/K180027)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).