The ACCUSYTE 3D Fiducial Markers are used to mark the location of a tumor cavity after the tumor has been removed. In addition, the Markers are indicated on situations where soft tissue needs to future medical procedure such as subsequent Radiation Therapy.
Device Story
Implantable fiducial marker consisting of a permanent radiopaque component (Gold or Platinum/Iridium) attached to an absorbable polylactic acid suture; surgically implanted into soft tissue tumor cavity by physician; provides radiographically visible reference point for subsequent radiation therapy targeting; enables accurate localization of tumor site on kV CT, 2D Linac-based kilovoltage/megavoltage X-ray, and cone-beam CT; aids clinicians in radiation treatment planning to limit exposure to adjacent healthy tissue; suture absorbs within one year; device is MR Conditional.
Clinical Evidence
Bench testing only. Performance data included biocompatibility (ISO-10993), physical testing (USP 30), and residual strength/absorption rate studies for the suture component. Visualization efficacy was validated using a phantom breast model across CT, KV, CBCT, and MV imaging systems. MR safety was established via non-clinical testing for 1.5T and 3T systems, demonstrating a maximum temperature rise of ≤2°C and defined artifact limits. No clinical human study data was provided; the document mentions no complications in a human study context, but this appears to be a reference to general clinical experience rather than a formal clinical trial.
Technological Characteristics
Materials: Gold or Platinum/Iridium marker, Glycolide-co-Lactide absorbable suture. Dimensions: <0.5 cm marker. Principle: Radiographic visualization of implanted marker. Connectivity: None (passive implant). Sterilization: ETO (SAL 10^-6). MR Safety: MR Conditional (1.5T/3T).
Indications for Use
Indicated for patients requiring marking of a tumor cavity post-resection or soft tissue marking for future medical procedures, including radiation therapy.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. The logo on the left is the Department of Health & Human Services-USA logo. The logo on the right is the FDA U.S. Food & Drug Administration logo. The FDA logo has a blue square with the letters FDA in white.
RADIATION PRODUCTS LLC
July 18, 2018
% Mr. Victor Pereira Principal VHP Consulting 7240 NW 63 Terrace PARKLAND FL 33067
Re: K180027
Trade/Device Name: ACCUSYTE 3D Fiducial Marker Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: IYE Dated: June 3, 2018 Received: June 13, 2018
Dear Mr. Pereira:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael D'Hara For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
K 180027
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known)
Device Name ACCUSYTE 3D Fiducial Marker
Indications for Use (Describe)
The ACCUSYTE 3D Fiducial Markers are used to mark the location of a tumor cavity after the tumor has been removed.
In addition, the Markers are indicated on situations where soft tissue needs to future medical procedure such as subsequent Radiation Therapy.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 5. 510 (k) Summary
This 510 (k) summary information is being submitted in accordance with the requirements of 21 CFR 807.92.
Applicant: Surgical Radiation Products LLC
Date Prepared: April 29, 2018 (amended)
# Contact Person: John D Corbitt MD
Radiation Products LLC 142 JFK Drive Lake Worth FL 33462 Phone: (561) 379-6210 Fax: (561) 439-9902 e-mail: jdcorbitt@aol.com
## Trade Name: ACCUSYTE™ 3D Fiducial Marker
Common Name: Implantable Radiographic Marker Classification Name: Medical charged-particle radiation therapy system Classification Number: 892.5050 Device Classification: Class II Product Code: IYE Predicate Device: CIVCO Suture-Type Marker (K071614) Reference Devices: BioZorb Marker (K143484) Foosin WEGO-PGLA Absorbable Suture (K130735)
# Substantially Equivalent to:
The ACCUSYTE™ 3D Fiducial Marker is substantially equivalent in intended use, principle of operation, and technological characteristics to the CIVCO Suture-Type Marker (K071614), to the BioZorb Marker (K143484), and to the Foosin WEGO-PGLA Absorbable Suture (K130735).
### Description of the Device subject to Premarket Notification:
The ACCUSYTE™ 3D Fiducial Marker is an implantable marker comprised of an absorbable suture (polylactic acid) component which absorbs completely in less than one year, and a permanent component, a Platinum or Gold Marker attached to the suture by a crimping operation similar to the operation use to attach the Stainless-Steel Needle.
The ACCUSYTE™ 3D Fiducial Marker is provided Sterile for single use and is implantable.
{4}------------------------------------------------
# Indication for Use:
The ACCUSYTE™ 3D Fiducial Markers are used to mark the location of a tumor cavity after the tumor has been removed.
In addition, the Markers are indicated on situations where soft tissue needs to be marked for future medical procedure such as subsequent Radiation Therapy.
# Technical Characteristics:
The ACUSYTE™ 3D Fiducial Marker has similar physical and technical characteristics to the predicative and reference devices, as illustrated in the table below:
| Technical<br>Characteristic | Subject Device | Primary<br>Predicate | Reference<br>Device | Reference Device |
|----------------------------------------------|------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------|
| | ACCUSYTE™ 3D<br>Fiducial Marker | CIVCO Suture-<br>Type Marker<br>(K071614) | BioZorb Marker<br>(K143484) | WEGO-PGLA<br>Absorbable Suture<br>(K130735) |
| Overall<br>Technological<br>Characteristic | Radiographically<br>visible<br>permanent<br>marker on an<br>absorbable<br>suture | Radiographically<br>visible<br>permanent<br>marker on an<br>absorbable<br>suture | Radiographicall<br>y visible<br>permanent<br>marker element<br>in a bio-<br>absorbable<br>polymer space | Suture-N/A |
| Principle of<br>Operation | Marker is<br>secured on<br>tumor cavity<br>area for<br>radiographic<br>visualization of<br>tumor site after<br>removal | Marker is<br>positioned into<br>tissue site for<br>radiographic<br>visualization of<br>tissue site | Marker is<br>positioned into<br>tissue site for<br>radiographic<br>visualization of<br>tissue site | Suture holds<br>marker in place<br>during tissue<br>healing process |
| Absorption<br>time for suture<br>and polymer | Less than 1 year<br>(suture) | Unknown for<br>suture | Longer than 1<br>year<br>(Polymer) | Suture- Less than 1<br>year |
| Visualization<br>Compatibility | ikV CT, 2D<br>Linac- based<br>kilovoltage and<br>megavolt-age X-<br>ray imaging<br>systems and kV | X-Ray<br>CT (presumed) | Mammography,<br>Ultrasound, X-<br>Ray, and CT | Suture-N/A |
{5}------------------------------------------------
| | cone-beam CT. | | | |
|---------------------------------------|--------------------------------------------------------------------------------------------------------|-----------------------------------------|--------------------------------------------------------------|-------------------------------------------------------------------|
| Materials of<br>Construction | Gold or<br>Platinum/Iridium Marker<br>component,<br>absorbable<br>suture<br>(Glycolide-co-<br>Lactide) | Gold Marker and<br>absorbable<br>suture | Titanium<br>Markers and<br>absorbable<br>polymer<br>(spacer) | Suture-Glycolide-<br>co Lactide with<br>Stainless Steel<br>Needle |
| Overall Device<br>Length | <0.5 cm<br>(Marker) | <5cm (Marker) | 2-5 cm (Marker) | Suture-45 cm |
| Typical<br>Anatomic<br>Treatment Site | Soft tissue<br>tumor cavity site<br>including breast | Breast | Soft tissue<br>including breast | Suture- soft tissue |
| -Method of<br>Marker<br>Deployment | Surgically<br>implanted | Surgically<br>implanted | Manual, open<br>surgical | Suture- surgical<br>placement |
| Marker<br>Stability | Sutured in place | Sutured in place | Tissue retention | Suture retains<br>marker in place<br>through healing<br>process |
| How Provided | Sterile for single<br>use | Sterile single use | Sterile single<br>use | Suture-Sterile<br>Single Use |
| Sterilization<br>Method | ETO<br>validated cycle<br>with SAL of 10-6 | Unknown | Radiation | Suture- ETO<br>validated cycle<br>with SAL of 10-6 |
# Performance Data:
The Absorbable Suture component on the candidate device was tested as part of its 510 (k) submission (K130735) with the following tests:
- Bacterial Endotoxin (USP Pharmacopeia <85>) - Pyrogenicity
- Biocompatibility (ISO-10993) - suture component - Implant
- Physical Testing - USP 30 <861, <871, and <881 - suture component
- Residual Strength and Absorption Rate studies as outlined in FDA's Class II Special Controls Guidance Document: Surgical Sutures.
## Performance data for the ACCUSYTE™ 3D Fiducial Marker device included:
- . The Gold and Platinum/Iridium markers were bench tested for clarity of visualization using a phantom breasts model across computed tomography (CT), kilovoltage (KV), cone-beam CT (CBCT), and megavoltage (MV) linear accelerator imaging
- Radiographic Imaging 6-8 weeks after implantation
- Mammography imaging after 1-year post implantation
{6}------------------------------------------------
#### MRI Safety:
The ACCUSYTE™ 3 D Fiducial Markers are considered "MR Conditional"
Non-clinical testing has demonstrated the ACCUSYTE™3D Fiducial Marker device is MR Conditional. A patient with this device can be safely scanned in an MR system meeting the following conditions:
- Static magnetic field of 1.5-Tesla or 3-Tesla
- Maximum spatial gradient magnetic field of 4,000-gauss/cm (40-T/m)
- Maximum MR system reported, whole body averaged specific absorption rate (SAR) of 2- W/kg in the Normal Operating Mode
- Under the scan conditions defined above, the Accusyte™ 3D Fiducial marker is expected to produce a maximum temperature rise of ≤2℃ after 15-minutes of continuous scanning at 3 Tesla
- The image artifact caused by the Accusyte™ 3D Fiducial is not expected to extend more than 10-mm from this device when imaged with a gradient echo pulse sequence and a 3- Tesla MRI system.
The Accusyte™ 3D Fiducial and may be safely scanned with MRI under the conditions listed above. Scanning under different conditions may result in severe patient injury.
The ACCUSYTE™ 3D Fiducial Marker performance is equivalent to that of the predicate and reference devices in that it provides clear visualization of the tumor site, the candidate device secures the markers on soft tissue of the tumor cavity for a more accurate localization of the site, aiding in targeting the area for subsequent Radiation therapy and limiting exposure to adjacent tissue.
There were no complications reported from the candidate markers device during the human study, such as bleeding, pain, inflammation, or tissue reactions. Therefore; demonstrating that the candidate device is safe and effective for its intended use.
### Basis for Determination of Substantial Equivalence:
Upon reviewing the safety and efficacy information provided in this submission and comparing intended use, principle of operation, and overall technological characteristics, the ACCUSYTE™ 3D Fiducial Marker is determined by Radiation Products LLC to be substantially equivalent to the existing legally marketed predicate device CIVCO Suture Type Marker.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.