← Product Code [LLZ](/productcode/LLZ) · K180006

# AmCAD-UT Detection 2.2 (K180006)

_Amcad Biomed Corporation · LLZ · Aug 31, 2018 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K180006

## Device Facts

- **Applicant:** Amcad Biomed Corporation
- **Product Code:** [LLZ](/productcode/LLZ.md)
- **Decision Date:** Aug 31, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Software as a Medical Device

## Indications for Use

AmCAD-UT® Detection is a Windows-based computer-aided detection (CADe) device intended to assist the medical professionals in analyzing thyroid ultrasound images, acquired from FDA-cleared ultrasound systems, with user-selected regions of interest (ROI). After the initial review of the ultrasound images by the physicians, the device further provides detailed information with quantification of sonographic characteristics of thyroid nodules. The device is intended for use on ultrasound images of discrete thyroid nodules larger than 1cm, for which a biopsy recommendation is required.

## Device Story

AmCAD-UT Detection 2.2 is a Windows-based CADe software application for clinical use by physicians. It processes user-selected regions of interest (ROI) in thyroid ultrasound images acquired from FDA-cleared systems. The device uses statistical pattern recognition and quantification methods to analyze and visualize sonographic characteristics, including hyperechoic foci, echogenicity, texture, margin, orientation, and anechoic areas. The software generates automated reports based on user-annotated findings and preference inputs. Output is viewed on a PC or review station, printed, or saved as a PDF. By providing quantitative data and visualization of nodule characteristics, the device assists physicians in interpreting thyroid ultrasound images, potentially enhancing clinical judgment and decision-making regarding biopsy recommendations.

## Clinical Evidence

Clinical evidence includes a Multiple-Reader-Multiple-Case (MRMC) study and standalone performance testing. The MRMC study validated the device's intended use, demonstrating that physician assistance with AmCAD-UT 2.2 enhances the ability to analyze sonographic characteristics and significantly increases the effectiveness of clinical judgment. Standalone studies confirmed the device's effectiveness in detecting sonographic characteristics (hyperechoic foci, echogenicity, texture, margin, orientation, anechoic areas) across images from various ultrasound system brands.

## Technological Characteristics

Windows-based software application; operates on standard PC or review station. Uses statistical pattern recognition and quantification algorithms. Inputs: DICOM 3.0, Bitmap, JPEG ultrasound images. Outputs: Annotated images, quantitative reports (PDF/print).

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- AmCAD-UT® Detection, Version 2.0 ([K122536](/device/K122536.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. Food & Drug Administration" in blue.

AmCAD BioMed Corporation % Chiu S. Lin President Lin & Associates, LLC 5614 Johnson Avenure BETHESDA MD 20817

August 31st, 2018

Re: K180006

Trade/Device Name: AmCAD-UT Detection 2.2 Regulation Number: 21 CFR 892.2050 Regulation Name: Picture Archiving And Communications System Regulatory Class: Class II Product Code: LLZ Dated: August 1, 2018 Received: August 3, 2018

Dear Chiu Lin:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jeff Balygo

for Robert A. Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K180006

Device Name AmCAD-UT® Detection 2.2

#### Indications for Use (Describe)

AmCAD-UT® Detection is a Windows-based computer-aided detection (CADe) device intended to assist the medical professionals in analyzing thyroid ultrasound images, acquired from FDA-cleared ultrasound systems, with user-selected regions of interest (ROI). After the initial review of the ultrasound images by the physicians, the device further provides detailed information with quantification of sonographic characteristics of thyroid nodules. The device is intended for use on ultrasound images of discrete thyroid nodules larger than 1cm, for which a biopsy recommendation is required.

Type of Use (Select one or both, as applicable)

| <div> <span style="font-size: 10pt;">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> </div> |
|---------------------------------------------------------------------------------------------------|
| <div> <span style="font-size: 10pt;">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div>  |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo and contact information for AmCad BioMed Corporation. The logo features a blue stylized "A" with a "C" inside. The text includes the company name in both Chinese and English, followed by the address: FL.5, NO.167, Fu Hsing N. RD., Taipei 105, Taiwan, ROC. The telephone number is +886-2-27136227 and the fax number is +886-2-25140245.

# 510(k) Summary of Safety and Effectiveness

This 510(k) summary of safety and effectiveness information is submitted as part of the Premarket Notification in compliance with requirements of CFR Part 807, Subpart E and Section 807.92

# 5.1 Identification of Submitter:

| Submitter:            | AmCad BioMed Corporation                                   |
|-----------------------|------------------------------------------------------------|
| Address:              | FL.5-2, NO.167, Fu Hsing N. RD., Taipei 105, Taiwan, R.O.C |
| Phone:                | 886-2-2713-6227                                            |
| Fax:                  | 886-2-2514-0245                                            |
| Contact:              | Jack Yang                                                  |
| Title:                | Vice President                                             |
| Phone:                | 886-2-2713-6227 ext.358                                    |
| Fax:                  | 886-2-2514-0245                                            |
| Email:                | jack.yang@amcad.com.tw                                     |
| Manufacturer:         | AmCad BioMed Corporation                                   |
| US Agent and Contact: | Chiu S. Lin, Ph.D.<br>Lin & Associates, LLC                |
| Address:              | 5614 Johnson Avenue<br>Bethesda, MD 20817                  |
| Phone:                | (O) 301-591-3895                                           |

- Date prepared: December 27, 2017
## 5.2 Identification of Product

E-mail:

| Device Trade Name:           | AmCAD-UT® Detection 2.2                     |
|------------------------------|---------------------------------------------|
| Common and Usual Name:       | Computer-Assisted Detection (CADe) Device   |
| Device Classification Name:  | Picture Archiving and Communications System |
| Regulation Number:           | 21 CFR 892.2050                             |
| Classification Product Code: | LLZ                                         |
| Classification Panel:        | Radiology                                   |

cslin@lin-associates.com

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo and contact information for AmCad BioMed Corporation. The logo features a stylized blue "A" with a "C" inside. The text includes the company name in both Chinese and English, followed by the address: FL.5, NO.167, Fu Hsing N. RD., Taipei 105, Taiwan, ROC. The telephone number is +886-2-27136227 and the fax number is +886-2-25140245.

Classification: Manufacturer:

Class II AmCad BioMed Corporation

# 5.3 Predicate Device

This subject software medical device is substantially equivalent to the device listed below:

| Model:         | AmCAD-UT® Detection, Version 2.0 |
|----------------|----------------------------------|
| Manufacturer:  | AmCad BioMed Corporation         |
| 510(k) Number: | K122536                          |

# 5.4 Device Description

AmCAD-UT® Detection 2.2 is a Windows-based computer-assisted detection (CADe) software application device designed to assist medical professionals in analyzing thyroid ultrasound images of user selected regions of interest (ROI).

After the initial review of thyroid ultrasound images by the physician, he/she can use AmCAD-UT Detection to analyze thyroid images for further interpretation. The physician selects an ROI to define the initial boundary of the ROI. Once the ROI is confirmed, the physician may process the image for detection and quantification of sonographic characteristics (i.e., hyperechoic foci, echogenicity, texture, margin, orientation and anechoic areas) by AmCAD-UT Detection. The device provides more detailed information with quantification and visualization of the sonographic characteristics of thyroid nodule that may assist physician in his/her complete interpretation.

The software application automatically generates reports given the user preference inputs (e.g., the nodule size, nodule location and shape, and the presence or absence of the sonographic characteristics) annotated during the image analysis process. A report form has been designed by AmCad to be consistent with the conventional clinical thyroid report form. An output of the report may be viewed and sent to paper printers or saved on the standalone PC or review station as PDF file.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo and contact information for AmCad BioMed Corporation. The logo features a blue stylized image of a building with a "C" inside a circle on top. The text includes the company name in both Chinese and English, followed by the address: FL.5, NO.167, Fu Hsing N. RD., Taipei 105, Taiwan, ROC. The telephone number is +886-2-27136227 and the fax number is +886-2-25140245.

### 5.5 Indications for Use

AmCAD-UT® Detection is a Windows-based computer-aided detection (CADe) device intended to assist the medical professionals in analyzing thyroid ultrasound images, acquired from FDA-cleared ultrasound systems, with user-selected regions of interest (ROI). After the initial review of the ultrasound images by the physicians, the device further provides detailed information with quantification and visualization of sonographic characteristics of thyroid nodules. The device is intended for use on ultrasound images of discrete thyroid nodules larger than 1cm, for which a biopsy recommendation is required.

#### 5.6 Comparison with Predicate Devices

AmCAD-UT® Detection 2.2 is a computer-assisted detection (CADe) device which provides viewing and post-acquisition image processing and analysis of thyroid ultrasound images with user-selected regions of interest (ROI) and automatically generates reports from user inputs annotated during the image analysis process. This software medical device is substantially equivalent to the predicate device listed below:

| Model:         | AmCAD-UT® Detection, Version 2.0 |
|----------------|----------------------------------|
| Manufacturer:  | AmCad BioMed Corporation         |
| 510(k) Number: | K122536                          |

The comparison as described in the following table:

|                                                               | AmCAD-UT® Detection 2.2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | AmCAD-UT® Detection 2.0                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
|---------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer                                                  | AmCad BioMed Corp.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | AmCad BioMed Corp.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| 510(k) Number                                                 | K180006                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | K122536                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Device<br>Common Name                                         | Computer-Assisted Detection<br>(CADe)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Computer-Assisted Detection<br>(CADe)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Regulation<br>Number                                          | 21 CFR 892.2050 - Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | 21 CFR 892.2050 - Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
|                                                               | AmCAD-UT® Detection 2.2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | AmCAD-UT® Detection 2.0                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Regulation Name                                               | Picture archiving and communications system                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Picture archiving and communications system                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Product Code                                                  | LLZ                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | LLZ                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Intended Use                                                  | AmCAD-UT® Detection 2.2 is intended to assist the medical professionals in analyzing thyroid ultrasound images of user-selected regions of interest (ROI). After the initial review of the ultrasound images by the physicians, the device further provides detailed information with quantification and visualization of sonographic characteristics of thyroid nodules.                                                                                                                                                      | AmCAD-UT® Detection 2.0 is intended to assist the medical professionals in analyzing thyroid ultrasound images of user-selected regions of interest (ROI). After the initial review of the ultrasound images by the physicians, the device further provides detailed information with quantification and visualization of sonographic characteristics of thyroid nodules.                                                                                                                                                                                     |
| Indications for Use                                           | AmCAD-UT® Detection 2.2 is a Windows-based computer-aided detection (CADe) device intended to assist the medical professionals in analyzing thyroid ultrasound images, acquired from FDA-cleared ultrasound systems, with user-selected regions of interest (ROI). After the initial review of the ultrasound images by the physicians, the device further provides detailed information with quantification and visualization of sonographic characteristics of thyroid nodules. The device is intended for use on ultrasound | AmCAD-UT® Detection 2.0 is a Windows-based computer-aided detection (CADe) device intended to assist the medical professionals in analyzing thyroid ultrasound images of user-selected regions of interest (ROI). After the initial review of the ultrasound images by the physicians, the device further provides detailed information with quantification and visualization of sonographic characteristics of thyroid nodules. The device is intended for use on Philips HDI5000 images of discrete thyroid nodules larger than 1cm, for which a biopsy has |
|                                                               | AmCAD-UT® Detection 2.2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | AmCAD-UT® Detection 2.0                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
|                                                               | images of discrete thyroid<br>nodules larger than 1cm, for<br>which a biopsy<br>recommendation is required.                                                                                                                                                                                                                                                                                                                                                                                                                    | been recommended. The<br>device performance has been<br>validated on images collected<br>from Philips HDI5000 with a<br>5-12MHz multi-frequency<br>probe.                                                                                                                                                                                                                                                                                                                                                                                                     |
| Functional<br>Capability of<br>Image<br>Processing            | AmCAD-UT® Detection 2.2<br>analyzes the user-selected<br>regions of interest (ROI) of<br>thyroid ultrasound image for<br>the detection and<br>quantification of sonographic<br>characteristics (hyperechoic<br>foci, echogenicity, texture,<br>margin, orientation and<br>anechoic areas). The device<br>further provides detailed<br>information with visualization<br>of sonographic characteristics<br>of thyroid nodules.                                                                                                  | AmCAD-UT® Detection 2.0<br>analyzes the user-selected<br>regions of interest (ROI) of<br>thyroid ultrasound image for<br>the detection and<br>quantification of sonographic<br>characteristics (hyperechoic<br>foci, echo-pattern,<br>echo-texture and anechoic<br>areas). The device further<br>provides detailed information<br>with visualization of<br>sonographic characteristics of<br>thyroid nodules.                                                                                                                                                 |
| Reading<br>Paradigm                                           | AmCAD-UT® Detection 2.2 is<br>to provide quantification and<br>visualization of sonographic<br>characteristics after<br>physicians' initial review of the<br>images.                                                                                                                                                                                                                                                                                                                                                           | For use as "second reader"<br>meaning that the function of<br>AmCAD-UT® Detection 2.0 is<br>to provide quantification and<br>visualization of sonographic<br>characteristics after<br>physicians' initial review of the<br>images.                                                                                                                                                                                                                                                                                                                            |
| Output<br>Generated by<br>the CAD Device                      | The image can be annotated<br>with the detected sonographic<br>characteristics and be<br>recorded by the device. The<br>software also automatically<br>generates reports given the<br>user preference inputs in the<br>analysis process.                                                                                                                                                                                                                                                                                       | The image can be annotated<br>with the detected sonographic<br>characteristics and be<br>recorded by the device. The<br>software also automatically<br>generates reports given the<br>user preference inputs in the<br>analysis process.                                                                                                                                                                                                                                                                                                                      |
| Type of Film to<br>be Processed<br>by the CAD<br>Device       | Digital ultrasound image                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Digital ultrasound image                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
|                                                               | AmCAD-UT® Detection 2.2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | AmCAD-UT® Detection 2.0                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Software Design                                               | Based on Statistical Pattern<br>Recognition and Quantification<br>method                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Based on Statistical Pattern<br>Recognition and Quantification<br>method                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Ground Truth<br>Establishment                                 | The ground truth to be<br>established for performance<br>studies of the device includes<br>the ROI, the presence of each<br>sonographic characteristic,<br>and the surgical pathology<br>examination result.                                                                                                                                                                                                                                                                                                                   | The ground truth to be<br>established for performance<br>studies of the device includes<br>the ROI, the presence of each<br>sonographic characteristic, and<br>the surgical pathology<br>examination result.                                                                                                                                                                                                                                                                                                                                                  |
| Platform                                                      | Window-based                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Window-based                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Operating<br>System                                           | Standard PC or review station                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Standard PC or review station                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Clinical<br>Application                                       | Thyroid cancers                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Thyroid cancers                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Image Type                                                    | Ultrasound Image                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Ultrasound Image                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Image Format                                                  | DICOM3.0, Bitmap, JPEG                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | DICOM3.0, Bitmap, JPEG                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| ROI<br>Quantification                                         | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Automatically<br>Generating<br>Report                         | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Report Storage                                                | Paper printers, Local disk                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Paper printers, Local disk                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Performance<br>Testing Data to<br>Support SE<br>Determination | Results from standalone<br>performance testing and<br>clinical performance testing<br>(MBMC study)                                                                                                                                                                                                                                                                                                                                                                                                                             | Standalone performance<br>assessment and clinical MRMC<br>study.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image contains the logo for AmCad BioMed Corporation. The logo consists of a blue letter 'A' with a circle in the upper right corner, followed by the company name in both Chinese and English. The Chinese characters are "安克生醫股份有限公司", and the English name is "AmCad BioMed Corporation".

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for AmCad BioMed Corporation. The logo consists of a blue stylized letter "A" with a circle inside, followed by the company name in both Chinese and English. The Chinese characters are placed above the English text, which reads "AmCad BioMed Corporation."

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for AmCad BioMed Corporation. The logo consists of a blue letter "A" with a circle in the upper right corner, followed by the company name in both Chinese and English. The Chinese characters are above the English text, and both are in black font. The logo appears to be for a biomedical company.

NO.167. Fu Hsing N. RD. Tel: +886-2-27136227

AmCAD-UT® Detection 2.2 is substantially equivalent to AmCAD-UT® Detection 2.0 that provides display and post-acquisition image analysis of ultrasound images assisting the physician in analyzing the ultrasound images of thyroid nodules. The standalone and clinical reader performance assessment results of AmCAD-UT® Detection 2.2 are substantially equivalent to the standalone performance assessment and clinical MRMC reader study of AmCAD-UT® Detection 2.0. Minor technological characteristics differences do not raise any new questions of safety and effectiveness.

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo and contact information for AmCad BioMed Corporation. The logo features a blue stylized "A" with a "C" inside. Below the logo is the company name in both Chinese and English. The address is listed as FL.5, NO.167, Fu Hsing N. RD., Taipei 105, Taiwan, ROC, and the telephone number is +886-2-27136227, while the fax number is +886-2-25140245.

Thus, AmCAD-UT® Detection 2.2 is substantially equivalent to the predicate device as a Computer-Assisted Detection (CADe) device intended to assist the physicians in clinical practice.

# 5.7 Performance Standards

No applicable FDA performance standards have been issued under the authority of Section 514.

## 5.8 Software

Software development for the AmCAD-UT® Detection 2.2 follows documented processes for software design, verification and validation testing. A risk assessment has been completed to identify potential design hazards that could cause an error or injury based on the use of the quantification results. Appropriate steps have been taken to control all identified risks for this type of image viewing and quantification device.

## 5.9 Summary of Performance Data to Support Substantial Equivalence

AmCad BioMed Corporation has conducted standalone and clinical reader performance studies to validate and assess the performance of the AmCAD-UT® Detection 2.2 for its intended use. The standalone studies evaluated the performance of the quantified sonographic characteristics (hyperechoic foci, echogenicity, texture, margin, orientation and anechoic areas) on images acquired from ultrasound systems of different brands and showed that the device is effective in detecting the sonographic characteristics of thyroid nodules acquired from various ultrasound systems.

The intended use of the AmCAD-UT® Detection 2.2 was validated in a clinical Multiple-Reader-Multiple-Case (MRMC) study. The results of the MRMC study demonstrated that the physician reading thyroid nodule sonography images with the assistance of AmCAD-UT® Detection 2.2 can enhance their ability in analyzing the sonographic characteristics and has led to a significant increase in effectiveness of making clinical judgment.

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the logo and contact information for AmCad BioMed Corporation. The logo features a blue stylized letter "C" inside a square. Below the logo is the company name in both Chinese and English. The contact information includes the address: FL.5, NO.167, Fu Hsing N. RD., Taipei 105, Taiwan, ROC, as well as the telephone number: +886-2-27136227 and fax number: +886-2-25140245.

#### 5.10 Conclusions

AmCAD-UT® Detection 2.2, being a computer assisted detection (CADe) software device, has the same intended use as the predicate device. In addition to generalizing the use of the device for ultrasound images acquired from various FDA-cleared ultrasound systems, two quantified and visualized sonographic characteristics, i.e. margin distinctness and tumor orientation, are added. On the user interface of the proposed device, gauge meters expressing the quantified values of sonographic characteristics are also added for convenience of the medical professionals' reading. The performance data demonstrates that the proposed device performs as safely and effectively as the predicate devices. Therefore, AmCAD-UT® Detection 2.2 is substantially equivalent to the predicate devices as the CADe device intended to assist the medical professionals in analyzing ultrasound images of thyroid nodules in clinical practice.

---

**Source:** [https://fda-staging.innolitics.com/device/K180006](https://fda-staging.innolitics.com/device/K180006)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).