K173975 · Dio Corporation · DZE · Jun 1, 2018 · Dental
Device Facts
Record ID
K173975
Device Name
UF(II) Wide Fixture
Applicant
Dio Corporation
Product Code
DZE · Dental
Decision Date
Jun 1, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3640
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The UF(II) Wide Fixture is indicated for surgical placement in the upper and lower jaw arches, to provide a root form means for single or multiple units' prosthetic attachment to restore a patient's chewing function. The larger (Ø5.9 – Ø6.9mm) implants can be placed with a conventional two stage surgical process with an option for transmucosal healing and are indicated for the molar region with delayed loading.
Device Story
The UF(II) Wide Fixture is a titanium, screw-form, root-form endosseous dental implant. It is surgically placed into the bone of the upper or lower jaw arches by a dental professional to provide a foundation for single or multiple prosthetic units (crowns, bridges, overdentures). The device restores chewing function in patients. It is designed for the molar region and utilizes a conventional two-stage surgical process with an option for transmucosal healing and delayed loading. The implant acts as a mechanical anchor for dental prosthetics.
Clinical Evidence
Bench testing only. Testing included sterilization validation (ISO 11137 series), shelf-life testing (ASTM F1980), fatigue testing (ISO 14801:2016), and mechanical testing (rotational angle, torque, static shear strength, acid residue, and component analysis). Biocompatibility data was leveraged from K122519, supplemented by new cytotoxicity testing.
Technological Characteristics
Material: Titanium Grade 4. Surface: SLA. Dimensions: Widths 5.9/6.4/6.9mm; Lengths 7/8.5/10/11.5/13mm. Sterilization: Gamma Irradiation. Mechanical testing per ISO 14801:2016. Packaging tested for integrity.
Indications for Use
Indicated for patients requiring dental restoration in upper or lower jaw arches; specifically for molar region placement using a two-stage surgical process with delayed loading.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
Predicate Devices
DIO UF HSA Internal Sub-Merged Implant System (K122519)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white. Underneath the blue square, the words "U.S. FOOD & DRUG ADMINISTRATION" are written in blue.
DIO Corporation c/o Peter Kang Business. Manager DIO USA 3470 Wilshire Blvd, #620 Los Angeles, California 90010 June 1, 2018
Re: K173975
Trade/Device Name: UF(II) Wide Fixture Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE Dated: February 23, 2018 Received: March 5, 2018
Dear Peter Kang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
## Mary S. Runner -S
Tina Kiang, Ph.D. for Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for DIO Implant Solution. The logo consists of the letters "DIO" in a bold, sans-serif font. The "DI" is in gray, while the "O" is in pink with a small pink circle above and to the right of it. Below the letters, in a smaller font, are the words "Implant Solution".
# Indication for Use
510(K) Number (if known): K173975
Device Name: UF(II) Wide Fixture
#### Indications for Use:
The UF(II) Wide Fixture is indicated for surgical placement in the upper and lower jaw arches, to provide a root form means for single or multiple units' prosthetic attachment's chewing function. The larger (05.9 - 06.9mm) implants can be placed with a conventional two stage surgical process with an option for transmucosal healing and are indicated for the molar region with delayed loading.
AND/OR Over The-Counter Use Prescription Use X (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
#### Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for DIO Implant Solution. The logo consists of the letters "DIO" in gray, with the "O" stylized as a pink circle with a smaller pink circle above it. Below the letters, the words "Implant Solution" are written in a smaller, gray font.
K173975
# 510(k) Summary
This 510(k) Summary is being submitted in accordance with requirement of 21 CFR part 807.92
## Submitter:
Jiae Park DIO Corporatioin. 66 Centum seo-ro, Haeundae-gu, Busan, Republic of Korea Phone +82-51-745-7836 Fax +82-51-745-7781
## Contact / US agent:
Peter Kang DIO USA. 3470 Wilshire Blvd. #620 Los Angeles, CA Phone 213-365-2875 Fax 213-365-1595
## Device Information:
Trade Name: UF(II) Wide Fixture Classification Name: Endosseous Dental Implant Classification: Class II Product Code: DZE Regulation : 21 CFR 872.3640 Date Prepared: 05/31/2018
## General Description
UF(II) Wide Fixture is especially designed for use in dental implant surgery. It is intended to be surgically placed in the bone of the upper or lower jaw arches. It consists of titanium, screw-form, root-form endosseous dental implant.
## Indication For Use
The UF(II) Wide Fixture is indicated for surgical placement in the upper and lower jaw arches, to provide a root form means for single or multiple units' prosthetic attachment to restore a patient's chewing function. The larger (Ø5.9 – Ø6.9mm) implants can be placed with a conventional two stage surgical process with an option for transmucosal healing and are indicated for the molar region with delayed loading.
## Predicate devices
Primary Predicate Device : DIO UF HSA Internal Sub-Merged Implant System(K122519) Reference Device : Rescue Dental Internal Implant System (K063216)
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for DIO Implant Solution. The logo consists of the letters "DIO" in a bold, sans-serif font. The "DI" is in gray, while the "O" is in pink and has a small pink circle above and to the right of it. Below the letters, in a smaller font, are the words "Implant Solution".
## Substantial Equivalence Comparison
The UF(II) Wide Fixture is substantially equivalent in design, function, material and indications for use to Predicate device of K122519, K063216.
The UF(II) Wide Fixture has the same material, surface treatment and similar design to the Primary Predicate device as K122519, the similar with reference devices as K063216.
The Indications for Use listed for the Subject and Primary predicate devices (K122519) do not raise new questions of substantial equivalence because smaller diameter implants are not part of this submission and all the subject devices retain the same intended use. The Indications for Use were adequately modified to reflect this difference.
Any differences in technology characteristics are accompanied by information that demonstrated the device is substantially equivalent as the predicate and do not raise different questions of safety and effectiveness than the predicate.
| | Subject Device | Primary<br>Predicate Device | Reference device |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Applicant | DIO Corporation | DIO Corporation | Megagen Implant Co., Ltd. |
| Trade Name | UF(II) Wide Fixture | DIO UF HSA Internal Sub-<br>Merged Implant System | RESCUE INTERNAL<br>IMPLANT SYSTEM |
| Classification<br>Name | Endosseous Dental Implant<br>(872.3640) | Endosseous Dental Implant<br>(872.3640) | Endosseous Dental Implant<br>(872.3640) |
| 510(k) No. | Not yet assigned | K122519 | K063216 |
| Product Code | DZE | DZE | DZE |
| Class | II | II | II |
| Material | Titanium Gr4 | Titanium Gr4 | Titanium Gr4 |
| Surface | SLA | SLA | SLA |
| Width(Φ) | 5.9/6.4/6.9 | 6.0/6.5/7.0 | 5.0/6.0/6.5/7.0/8.0 |
| Length | 7/8.5/10/11.5/13 | 7/8.5/10.5/12.5/14.5 | 7/8.5/10/11.5/13 |
| Image | Image: Implant | Image: Implant | Image: Implant |
| Indications<br>For Use | The UF(II) Wide Fixture is<br>indicated for surgical<br>placement in the upper and<br>lower jaw arches, to<br>provide a root form means<br>for single or multiple units'<br>prosthetic attachment to<br>restore a patient's chewing<br>function. The larger (Ø5.9<br>– Ø6.9mm) implants can<br>be placed with a<br>conventional two stage<br>surgical process with an<br>option for transmucosal | The DIO UF HSA Internal<br>Sub-Merged Implant System<br>is indicated for surgical<br>placement in the upper and<br>lower jaw arches, to provide<br>a root form means for single<br>or multiple units' prosthetic<br>attachment to restore a<br>patient's chewing function.<br>The smaller (Ø3.8 ~ Ø5.5)<br>implants can be placed with<br>a conventional two stage<br>surgical process with an<br>option for transmucosal | The Rescue Internal Implant<br>System is intended to be<br>surgically placed in the<br>maxillary or mandibular<br>molar areas for the purpose<br>providing prosthetic support<br>for dental restorations<br>(Crown, bridges, and<br>overdentures) in partially of<br>fully edentulous individuals.<br>These implants are intended<br>to be used where smaller<br>implants have failed. |
| | healing and are indicated | healing or they can be | |
| | for the molar region with | placed in a single stage | |
| | delayed loading. | surgical process for | |
| | | immediate loading when | |
| | | good primary stability is | |
| | | achieved with appropriate | |
| | | occlusal loading. The larger | |
| | | ( $ Ø6.0 ~ Ø7.0 $ ) implants can | |
| | | be placed with a | |
| | | conventional two stage | |
| | | surgical process with an | |
| | | option for transmucosal | |
| | | healing and are | |
| | | indicated for the molar | |
| | | region with delayed loading. | |
| Sterilization | Gamma Irradiation | Gamma Irradiation | Gamma Irradiation |
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for DIO Implant Solution. The logo consists of the letters "DIO" in a bold, sans-serif font, with the "D" and "I" in gray and the "O" in pink. Above the "O" is a small pink circle. Below the letters is the text "Implant Solution" in a smaller, gray, sans-serif font.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for DIO Implant Solution. The logo consists of the letters "DIO" in a bold, sans-serif font. The letters "DI" are in gray, while the "O" is in pink with a small pink circle above and to the right of it. Below the letters, in a smaller font, are the words "Implant Solution" in gray.
## Non-Clinical Test Data
The result of non-clinical testing demonstrate that the results have met the standards, and the subject device is substantially equivalent to the predicate device. This testing included:
#### Sterilization Validation and Shelf Life Testing
Sterilization validating testing has been performed in accordance with ISO 11137-1, ISO 11137-2, ISO 11137-3, ISO 11737-1, and ISO 11737-2 for gamma sterilization. Test results have demonstrated that the SAL of 10-6 was achieved and all testing requirements were met.
For the subject devices provided sterile, accelerated aging shelf life testing was conducted according to ASTM F1980 and real time testing is being conducted to support accelerated aging results.
#### Fatigue test
The fatigue test was performed on the subject device in accordance with ISO 14801:2016 Dentistry-Implants-Dynamic loading test for endosseous dental implants. The worst case scenario was chosen based on the FDA guidance "Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Implant Abutments".
#### Biocompatibility
The Biocompatibility Test are leveraged from previous submission(K122519) such as Cytotoxicity test, Sensitization test, Irritation or Intracutaneous reactivity test, Systemic toxicity test, Sub-chronic toxicity test, Genotoxocity test and Implantation test. Also, new cytotoxicity test conducted for subject device has the same material, surface treatment, and manufacturing process as the Primary predicate device (K122519).
#### Mechanical Testing
The following tests were conducted to demonstrate substantial equivalence to the predicate device: Appearance Test confirm external appearance of subject device (smooth without crack or damage) DimensionTest - confirm same fundamental technology through dimensional analysis RotationalAngle Test - confirm that the subject device will not significantly rotate Maximum Torque for dental implant screwTest - confirm equivalent rotational torque Removal Torque for dental implant screwTest - confirm equivalent removal torque Static Shear Strength Test accordingto ISO 14801 .- confirm equivalent static shear strength Packaging Test - confirm that the packaging is adequate and equivalent Acid Residue Test - confirm that F-, Cl-, NO2-, NO3-, SO42-, Br- and PO43- should be less than 1ppm. Component analysis - confirm that no foreign materials are observed on the surface.
## Conclusions
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807, and based on the information provided in this premarket notification, DIO Corporation concludes that the UF(II) Wide Fixture is substantially equivalent to predicate devices as described herein.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.