Zircon Blanks are used for milling using digital stl-files to produce Crowns, Bridges and monolithic works. We propose the K2 Zircon Blanks for anterior and posterior restorations. All K2 Zircon Blanks are indicated to be colored with Dip & Brush K2 CAD CAM Zircon Liquids. K2 Zircon Blanks should be not used by the general public or over-the-counter.
Device Story
Yeti K2 Zircon Blanks are ceramic discs composed of zirconium oxide (ZrO2 + HfO2 + Y2O3, Al2O3, and other oxides). Used in dental clinics/laboratories by dental professionals; blanks are machined via CAD/CAM or copy milling systems using digital STL files to produce crowns, bridges, and monolithic dental restorations. Blanks may be colored using Dip & Brush K2 CAD CAM Zircon Liquids prior to sintering. Final restorations provide structural support for dental prosthetics. Benefits include patient-specific, durable, and aesthetic dental restorations.
Clinical Evidence
No clinical data. Biocompatibility testing performed per ISO 10993-1, including cytotoxicity, haemolysis, acute systemic toxicity, irritation, sensitization, pyrogenicity, chromosomal aberration, and long-term implantation. Bench testing conducted on raw material and product batches for chemical composition and physical properties.
Technological Characteristics
Ceramic discs (zirconium oxide). Composition: ZrO2 + HfO2 + Y2O3 (>99.5%), Al2O3, and other oxides. Dimensions: 95 or 98.5 mm diameter, 10-25 mm height. Sintering parameters: 1450°C for 2 hours or 1500°C for 1 hour. Single-use, non-sterile. Mechanical properties: bulk density, grain size, flexural bending strength, fracture toughness, Vickers hardness, thermal coefficient, radioactivity, and solubility.
Indications for Use
Indicated for dental patients requiring anterior or posterior restorations, including single crowns and bridges (up to 2 intermediate elements for white classic/translucent, 1 for extreme translucent). Not for general public or OTC use.
Regulatory Classification
Identification
Porcelain powder for clinical use is a device consisting of a mixture of kaolin, felspar, quartz, or other substances intended for use in the production of artificial teeth in fixed or removable dentures, of jacket crowns, facings, and veneers. The device is used in prosthetic dentistry by heating the powder mixture to a high temperature in an oven to produce a hard prosthesis with a glass-like finish.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Yeti Dentalprodukte Gmbh Thomas Biskupski General Manager Industriestrasse 3a Engen, 78234 GERMANY November 15, 2018
Re: K173971
Trade/Device Name: K2 Zirkon Blank White Classic; K2 Zirkon Blank Translucent; K2 Blank Extreme Translucent Regulation Number: 21 CFR 872.6660 Regulation Name: Porcelain Powder For Clinical Use Regulatory Class: Class II Product Code: EIH Dated: August 20, 2018 Received: August 23, 2018
Dear Thomas Biskupski:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Digitally signed by Mary S. Runner -S3 Mary S. Runner -S3 13:23:04 -05'00'
forTina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K173971
Device Name
K2 Zircon Blank white classic, K2 Zircon Blank translucent, K2 Zircon Blank extreme translucent
#### Indications for Use (Describe)
Zircon Blanks are used for milling using digital stl-files to produce Crowns, Bridges and monolithic works. We propose the K2 Zircon Blanks for anterior and posterior restorations.
Table 1: Indication of use and maximum intermediate elements.
| | Single crown | | Bridges | |
|----------------------------------------|--------------|-----------|----------|-----------|
| | Anterior | Posterior | Anterior | Posterior |
| K2 Zircon Blank<br>white classic | X | X | 2 | 2 |
| K2 Zircon Blank<br>translucent | X | X | 2 | 2 |
| K2 Zircon Blank<br>extreme translucent | X | X | 1 | - |
All K2 Zircon Blanks are indicated to be colored with Dip & Brush K2 CAD CAM Zircon Liquids. K2 Zircon Blanks should be not used by the general public or over-the-counter.
Type of Use (Select one or both, as applicable)
× Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW .*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Yeti Dental. The logo features a mountain range above the word "YETI" in a bold, shadowed font. Below "YETI" is the word "DENTAL" in a smaller, bold font.
# 510(k) Summary K173971
| Submitter | Yeti Dentalprodukte GmbH<br>Industriestrasse 3<br>78234 Engen - Germany |
|----------------------------------------|-------------------------------------------------------------------------------------------------------------------------------|
| Contact Person | Thomas Biskupski |
| Date Prepared | August 15th, 2018 |
| K Number | K173971 |
| Device | |
| Name of Device | K2 CAD CAM Zircon Blank white classic,<br>K2 CAD CAM Zircon Blank translucent,<br>K2 CAD CAM Zircon Blank extreme translucent |
| Common Name | Powder, Porcelain |
| Classification Name | Porcelain powder for clinical use |
| Regulatory Class | II |
| Product Code | EIH |
| Predicate Device | |
| K2 Zirconia Blank white classic: | Z-CAD HD (K072569) |
| Reference Devices | |
| K2 Zirconia Blank translucent: | Ceramill® zi (K063511) |
| K2 Zirconia Blank extreme translucent: | Zolid FX (K152383) |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows a logo for "YETI DENTAL". The logo features a stylized mountain range above a blue banner. The word "YETI" is written in large, bold, black letters on the blue banner, and the word "DENTAL" is written in smaller, bold, black letters below the banner.
### Device Description
The Yeti K2 Zircon Blanks are ceramic blanks, which are indicated for the production of dentalprosthetic restorations. The blanks named above can be machined with CAM, CAD/CAM or copy milling systems commonly used in dental medical technology and for milling using digital STL files to produce Crowns, Bridges and monolithic works made from zirconium oxide.
The Yeti Dip & Brush K2 liquids are ready for use water-based products. These liquids are specifically formulated for the colorization of K2 Zircon Blanks from Yeti Dentalprodukte GmbH.
## Indications for Use
Zircon Blanks are used for milling using digital STL files to produce Crowns, Bridges and monolithic works. We propose the K2 Zircon Blanks for anterior and posterior restorations.
| | Single crown | | Bridges | |
|----------------------------------------|--------------|-----------|----------|-----------|
| | Anterior | Posterior | Anterior | Posterior |
| K2 Zircon Blank<br>white classic | x | x | 2 | 2 |
| K2 Zircon Blank<br>translucent | x | x | 2 | 2 |
| K2 Zircon Blank<br>extreme translucent | x | x | 1 | - |
Table 1: Indication of use and maximum intermediate elements
All K2 Zircon Blanks are indicated to be colored with Dip & Brush K2 CAD CAM Zircon Liquids. K2 Zircon Blanks should be not used by the General public or over-the-counter.
### Comparison of Technological Characteristics with the predicate device
All compared characteristics: Chemical Composition and specifications are the same for the subject and the predicate devices. There are no significant deviations.
The material consists of ZrO2 + HfO2 + Y2O3, Y2O3, Al2O3 and other Oxides in the subject and the predicate devices.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image is a logo for Yeti Dental. The logo features a mountain range above the word "YETI" in a bold, sans-serif font, with a blue background behind the word. Below "YETI" is the word "DENTAL" in a smaller, sans-serif font.
The shapes and dimensions are also comparable for the K2 Zircon Blank white classic, translucent and extreme translucent, with they're predicate devices.
The defined specifications for the K2 Zircon Blank white classic, translucent and extreme translucent: Bulk density, Grain Size, Flexural Bending Strength, Fracture Toughness, Hardness (Vicker's Hardness), Coefficient of Thermal, Radioactivity (raw material) and Solubility (acetic acid) have no significant deviations to the predicate devices.
Except for the K2 Zircon Blank white classic, there was no comparable value found for the specifications translucency and Thermal conductivity.
### Performance Data
### Biocompatibility testing:
The biocompatibility evaluation for the K2 Zircon Blanks was conducted to several standards for cytotoxicity, haemolysis, acute systematic toxicity, irritation, sensitization, pyrogen, chromosomal abbreviation and long-term implantation. All tests were performed in accordance to ISO 10993-1.
### Bench test for each raw material or product batch:
The chemical composition and several properties are tested for each raw material batch or product batch.
### Shelf Life:
Shelf-life for the K2 Zircon Blanks was set at 10 years by Yeti.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows a logo for Yeti Dental. The logo features a mountain range above the word "YETI" in bold, shadowed letters. Below "YETI" is the word "DENTAL" in a smaller, bold font.
Technological Characteristics:
| | New Device | Predicate Device |
|--------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------|
| | K2 Zircon Blank white classic | Z-CAD HD |
| 510(k) number | K173971 | K072569 |
| 510(k)<br>submitter/ holder | YETI DENTALPRODUKTE | METOXIT |
| Product Code | EIH | EIH |
| Applications | CAD / CAM fabrication | CAD / CAM<br>Fabrication |
| Material<br>ZrO2 + HfO2 + Y2O3 | > 99.5 % | > 99.5 % |
| Shape | Disks | Disks |
| Dimensions | Various<br>(Diameter: 95 or 98.5 mm<br>Height: 10 to 25 mm) | Various<br>(Diameter: 95 or 98.5 mm<br>Height: 10 to 25 mm) |
| Sterility | No | No |
| Single Use | Yes | Yes |
| Sintering Parameters | 1450°C - 2 hours<br>1500°C - 1 hour | 1450°C - 2 hours<br>1500°C - 1 hour |
| Indications for use | Zircon Blanks are used for milling<br>using digital STL-files to produce<br>Crowns, Bridges and monolithic<br>works. We propose the K2 Zircon<br>Blanks for anterior and posterior<br>restorations. | Metoxit blanks are indicated for<br>use as a substructure for<br>porcelain fused ceramic fixed<br>dental restorations. |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Yeti Dental. The logo features a mountain range above the word "YETI" in a bold, shadowed font, set against a blue background. Below "YETI" is the word "DENTAL" in a smaller, sans-serif font.
## Conclusions
Based on all information shown above and all performed testing the K2 Zircon Blank white classic, K2 Zircon Blank translucent and K2 Zircon Blank extreme translucent from Yeti Dentalprodukte GmbH are substantially equivalent to the predicate and reference devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.