Straumann® Variobase™ for Bridge/Bar Cylindrical are intended to be placed into Straumann dental implants to provide support for prosthetic reconstructions such as bridges and over-dentures
Device Story
Non-engaging dental implant abutments; fabricated from Ti-6Al-7Nb alloy; support bridge or bar reconstructions on two or more dental implants. Used in dental clinics; operated by dental professionals. Workflow involves traditional or CAD/CAM design; patient-specific restorations cemented onto abutments. Basal screw connects abutment to implant. Provides structural support for prosthetic rehabilitations; enables soft tissue healing via temporary restorations; facilitates final functional occlusion post-osseointegration.
Clinical Evidence
Bench testing only. Evidence includes dynamic fatigue testing per ISO 14801, biocompatibility testing per ISO 10993-1, and sterilization validation per ISO 17665-1/2. No clinical data presented.
Technological Characteristics
Material: Ti-6Al-7Nb alloy. Design: Non-engaging Morse taper apical interface; straight wall coronal design with detents. Diameter: 4.5–7.0 mm. Height: 3.5–4.5 mm. Sterilization: Steam autoclave. Workflow: CAD/CAM or traditional wax-up. Connectivity: None.
Indications for Use
Indicated for use as an aid in prosthetic rehabilitations for patients requiring bridge or over-denture restorations on endosseous dental implants. Temporary restorations may be used during the healing phase to maintain, stabilize, and form soft tissue, provided they are placed out of occlusion. Final restorations may be placed into occlusion once the implant is fully osseointegrated.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
Straumann CARES Bone Level Screw Retained Bars (K132844)
Straumann CARES Bone Level Screw Retained Bridges (K132844)
Reference Devices
RN Synocta UCLA Gold Abutment for the Straumann Dental Implant (K041295)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized human figure. To the right of that is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
July 17, 2018
Institut Straumann AG Jennifer Jackson Director, Regulatory Affairs and Quality 60 Minuteman Road Andover, Massachusetts 01810
Re: K173968
Trade/Device Name: Straumann® Variobase™ for Bridge/Bar Cylindrical Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: June 14, 2018 Received: June 18, 2018
Dear Jennifer Jackson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Andrew I. Steen -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known)
K173968
Device Name
Straumann® Variobase™ for Bridge/Bar Cylindrical
Indications for Use (Describe)
Straumann® Variobase™ for Bridge/Bar Cylindrical prosthetic components directly connected to the endosseous dental implants are indicated for use as an aid in prosthetic rehabilitations. The patientspecific prosthetic restoration (bridge or over-denture) can be cemented on the Straumann® Variobase™ for Bridge/Bar Cylindrical prosthetic components. A temporary restoration can be used prior to the insertion of the final components to maintain, stabilize and form the soft tissue during the healing phase. Temporary restorations are indicated to be placed out of occlusion. Straumann® Variobase™ for Bridge/Bar Cylindrical and patient-specific restorations may be placed into occlusion when the implant is fully osseointegrated.
All digitally designed Straumann® Variobase® for Bridge/Bar Cylindrical prosthetic components are intended to be sent to Straumann for manufacture at a validated milling center.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D) Subpart C)
| | Over-The-Counter Use (21 CFR 801
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspectof this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to espond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/14)Equivalent
{3}------------------------------------------------
# 510(k) Summary
| Submitter's Contact Information | |
|---------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | Straumann USA, LLC (on behalf of Institut Straumann AG)<br>60 Minuteman Road<br>Andover, MA 01810<br>Registration No.: 1222315<br>Owner/Operator No.: 9005052 |
| Contact Person: | Jennifer Jackson<br>Director, Regulatory Affairs & Quality |
| Prepared By: | Dr. Gordon Dodds<br>Manager Design Control QM<br>Etkon GmbH |
| Date Prepared: | July 16, 2018 |
| Product Code(s): | NHA (21 CFR 872.3630) |
| Device Class: | II (21 CFR 872.3630) |
| Classification Panel: | Dental |
| Name of the Device | |
| Classification Name: | Endosseous dental implant abutment (21 CFR 872.3630) |
| Proprietary Name: | Straumann® Variobase™ for Bridge/Bar Cylindrical |
| Predicate Device(s): | |
| Predicate Device(s): | K151157 Straumann Variobase for Bridge/Bar K142890 Straumann Variobase Abutments K120822 Straumann CARES Variobase Abutment<br>NNC, RN, WN, NC, RC K132844 Straumann CARES Bone Level Screw<br>Retained Bars, Straumann CARES Bone Level<br>Screw Retained Bridges |
{4}------------------------------------------------
- K041295 RN Synocta UCLA Gold Abutment for . the Straumann Dental Implant
## Device Description
The Straumann® Variobase™ for Bridge/Bar Cylindrical, see Figure 1 and Table 1, are non-engaging (without rotational lock) abutments made from Ti-6Al-7Nb (TAN) that support a bridge or bar reconstruction (framework or full contour) on two or more dental implants. The corresponding basal screw is delivered with the abutment for connecting the abutment to the implant. A dental laboratory technician designs and manufactures the bridge/bar reconstruction via their preferred workflow using traditional or CAD/CAM methods.
Image /page/4/Picture/4 description: The image shows five dental implant abutments labeled NC, RC, NNC, RN, and WN. The abutments are all silver in color, except for the NC abutment, which has a gold-colored base. The abutments are all different shapes and sizes, with the NC abutment being the smallest and the WN abutment being the largest. The abutments are used to connect dental implants to crowns or other dental prosthetics.
Figure 1 Straumann Variobase for Bridge/Bar Cylindrical
# Table 1 - Listing of devices
| Device Name | Device Number |
|---------------------------------------------------------|---------------------|
| Straumann® NC Variobase™ for Bridge/Bar<br>Cylindrical | 022.0110 / 010.6085 |
| Straumann® RC Variobase™ for Bridge/Bar<br>Cylindrical | 022.0111 / 010.6086 |
| Straumann® RN Variobase™ for Bridge/Bar<br>Cylindrical | 048.378 / 010.6083 |
| Straumann® WN Variobase™ for Bridge/Bar<br>Cylindrical | 048.379 / 010.6084 |
| Straumann® NNC Variobase™ for Bridge/Bar<br>Cylindrical | 048.377 / 010.6082 |
# Intended Use:
{5}------------------------------------------------
Straumann® Variobase™ for Bridge/Bar Cylindrical are intended to be placed into Straumann dental implants to provide support for prosthetic reconstructions such as bridges and over-dentures
# Indications For Use
Straumann® Variobase™ for Bridge/Bar Cylindrical prosthetic components directly connected to the endosseous dental implants are indicated for use as an aid in prosthetic rehabilitations. The patient-specific prosthetic restoration (bridge or overdenture) can be cemented on the Straumann® Variobase™ for Bridge/Bar Cylindrical prosthetic components. A temporary restoration can be used prior to the insertion of the final components to maintain, stabilize and form the soft tissue during the healing phase. Temporary restorations are indicated to be placed out of occlusion. Straumann® Variobase™ for Bridge/Bar Cylindrical and patient-specific restorations may be placed into occlusion when the implant is fully osseointegrated. All digitally designed Straumann® Variobase® for Bridge/Bar Cylindrical prosthetic components are intended to be sent to Straumann for manufacture at a validated milling center.
{6}------------------------------------------------
### Technological Chracteristics
The technological characteristics of the subject devices are compared to the primary predicate and reference devices in the following table.
Table 2: Substantial Equivalence Comparison
{7}------------------------------------------------
| FEATURE | Subject Device<br>Straumann® Variobase™ for Bridge/Bar<br>Cylindrical<br>Subject Submission | Predicate Device<br>Straumann® Variobase™<br>Abutments for Bridge/Bar<br>(K151157) | Reference Predicate<br>Devices<br>Straumann® Variobase™<br>Abutments<br>(K120822, K142890)<br>Straumann® Screws<br>(K132844, K041295) | Equivalence<br>Discussion |
|-----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use | Straumann® Variobase™ for Bridge/Bar<br>Cylindrical prosthetic components directly<br>connected to the endosseous dental implants<br>are indicated for use as an aid in prosthetic<br>rehabilitations. The patient-specific prosthetic<br>restoration (bridge or over-denture) can be<br>cemented on the Straumann® Variobase™ for<br>Bridge/Bar Cylindrical prosthetic components.<br>A temporary restoration can be used prior to<br>the insertion of the final components to<br>maintain, stabilize and form the soft tissue<br>during the healing phase. Temporary<br>restorations are indicated to be placed out of<br>occlusion. Straumann® Variobase™ for<br>Bridge/Bar Cylindrical and patient-specific<br>restorations may be placed into occlusion<br>when the implant is fully osseointegrated.<br>All digitally designed Straumann®<br>Variobase® for Bridge/Bar Cylindrical<br>prosthetic components are intended to be<br>sent to Straumann for manufacture at a<br>validated milling center. | Straumann® Variobase® prosthetic<br>components directly or indirectly<br>connected to the endosseous dental<br>implant are intended for use as an aid<br>in prosthetic rehabilitations. The<br>prosthetic restoration (bridge or<br>overdenture) can be cemented on the<br>Straumann® Variobase® prosthetic<br>components. A temporary restoration<br>can be used prior to the insertion of<br>the final components to maintain,<br>stabilize and form the soft tissue<br>during the healing phase. They may<br>not be placed into occlusion. Final<br>abutments and restorations may be<br>placed into occlusion when the<br>implant is fully osseointegrated. | Not applicable | Identical<br>Section regarding<br>production methods has<br>been added. The same<br>production methods are<br>used for the predicate<br>device. |
| Abutment<br>Material | Titanium-Aluminum-Niobium alloy (Ti-6Al-7Nb) | Titanium-Aluminum-Niobium alloy (Ti-<br>6Al-7Nb) | Not applicable | Identical |
| FEATURE | Subject Device<br>Straumann® Variobase™ for Bridge/Bar<br>Cylindrical<br>Subject Submission | Predicate Device<br>Straumann® Variobase™<br>Abutments for Bridge/Bar<br>(K151157) | Reference Predicate<br>Devices<br>Straumann® Variobase™<br>Abutments<br>(K120822, K142890)<br>Straumann® Screws<br>(K132844, K041295) | Equivalence<br>Discussion |
| Abutment<br>Diameter | 4.5 – 7.0 mm | 4.2 – 7.0 mm | Not applicable | Equivalent<br>The subject device NC<br>interface is 0.3 mm wider<br>and gives more support |
| Abutment<br>Height | 3.5 – 4.5 mm | 3.5 – 4.5 mm | Not applicable | Identical |
| Abutment<br>Apical<br>Design | Non-engaging (no rotational lock) Morse taper | Non-engaging (no rotational lock)<br>Morse taper | Not applicable | Identical |
| Abutment<br>Coronal<br>Design | Straight wall with detents | 10° or15° conical taper | K120822, K142890 -<br>Straight wall with detents | Equivalent<br>The coronal design is<br>identical to the reference<br>predicate devices and has<br>the equivalent function as<br>the predicate devices. |
| Basal Screw<br>Design | NC/RC Conical Head Screw Bone Level<br>RN/WN Conical Head Screw Tissue Level | | K132844, K041295 -<br>NC/RC Conical Head Screw<br>Bone Level<br>RN/WN Conical Head<br>Screw Tissue Level | Identical |
| FEATURE | Subject Device<br>Straumann® Variobase™ for Bridge/Bar<br>Cylindrical<br>Subject Submission | Predicate Device<br>Straumann® Variobase™<br>Abutments for Bridge/Bar<br>(K151157) | Reference Predicate<br>Devices<br>Straumann® Variobase™<br>Abutments<br>(K120822, K142890)<br>Straumann® Screws<br>(K132844, K041295) | Equivalence<br>Discussion |
| Restoration<br>Types<br>Supported | Patient-specific prosthetic bridge or over-<br>denture | Patient-specific prosthetic bridge or<br>over-denture | Not applicable | Identical |
| Restoration<br>Material | Materials cleared by the FDA under 21 CFR<br>872.6660 or exempt materials as described<br>under 21 CFR 872.3060 (Noble metal alloys)<br>and 21 CFR 872.3710 (Base metal alloys) | Materials cleared by the FDA under<br>21 CFR 872.6660 or exempt<br>materials as described under 21 CFR<br>872.3060 (Noble metal alloys) and 21<br>CFR 872.3710 (Base metal alloys) | Not applicable | Identical |
| Design<br>Workflow | Wax-up or CAD | Wax-up or CAD | Not applicable | Identical |
| Manufacturing<br>Workflow | Traditional casting or Straumann Milling Center | Traditional casting, pressing or<br>Straumann Milling Center | Not applicable | Identical |
| Mechanical<br>Stability | Dynamic fatigue test<br>Pass | Dynamic fatigue test<br>Pass | Not applicable | Identical |
| Sterilization | Steam autoclave | Steam autoclave | Not applicable | Identical |
| FEATURE | Subject Device<br>Straumann® Variobase™ for Bridge/Bar<br>Cylindrical<br>Subject Submission | Predicate Device<br>Straumann® Variobase™<br>Abutments for Bridge/Bar<br>(K151157) | Reference Predicate<br>Devices<br>Straumann® Variobase™<br>Abutments<br>(K120822, K142890)<br>Straumann® Screws<br>(K132844, K041295) | Equivalence<br>Discussion |
| Mode of Action | Screw-retained or cement retained | Screw-retained or cement retained | Not applicable | Identical |
| Reusable | No | No | Not applicable | Identical |
{8}------------------------------------------------
{9}------------------------------------------------
{10}------------------------------------------------
{11}------------------------------------------------
### Performance Data
Per Guidance for Industry and FDA Staff - Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Abutments dated May 12, 2004, the substantial equivalence of the subject device(s) are satisfactorily addressed via bench studies.
The following testing has been conducted or substantial equivalence to tested predicate devices has been demonstrated:
- Dynamic fatigue testing conforming to FDA guidance and ISO 14801. ●
- . Software validation conforming to the requirements of IEC 62304.
- . Sterilization validation conforming to ISO 17665-1 and ISO/TS 17665-2 (K151157 Straumann Variobase for Bridge/Bar).
- Biocompatibility testing conforming to ISO-10993-1 (K120822 Straumann CARES . Variobase Abutment NNC, RN, WN, NC, RC)
### Conclusions
Based upon our assessment of the design and applicable performance data, the subject devices have been determined to be substantially equivalent to the identified predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.