← Product Code [LPL](/productcode/LPL) · K173958

# OxyAqua (olifilcon D) spherical silicone hydrogel soft contact lens, OxyAqua (olifilcon D) multifocal silicone hydrogel soft contact lens, OxyAqua (olifilcon D) toric silicone hydrogel soft contact lens (K173958)

_Visco Vision, Inc. · LPL · Apr 3, 2018 · Ophthalmic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K173958

## Device Facts

- **Applicant:** Visco Vision, Inc.
- **Product Code:** [LPL](/productcode/LPL.md)
- **Decision Date:** Apr 3, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 886.5925
- **Device Class:** Class 2
- **Review Panel:** Ophthalmic

## Indications for Use

The OxyAqua (olifilcon D) Spherical Silicone Hydrogel soft (hydrophilic) contact lenses are daily wear single use soft contact lens indicated for the corrective ametropia (myopia and hyperopia) in phakic or aphakic persons with non-diseased eyes who exhibit refractive astigmatism of 1.00D or less where the astigmatism does not interfere with visual acuity. The OxyAqua (olifilcon D) Toric Silicone Hydrogel Soft (hydrophilic) Contact Lenses are indicated as daily wear for the correction of ametropia (myopia and hyperopia) with astigmatism in aphakic persons with non-diseased eyes and whose powers are from -20.00 to +20.00 diopters and astigmatic corrections are from -0.75 to -2.25 diopters. The OxyAqua (olifilcon D) Multifocal Silicone Hydrogel Soft (hydrophilic) Contact lenses are indicated as daily wear for the correction of ametropia (myopia) and emmetropia with presbyopia in aphakic and non-aphakic persons with non-diseased eyes and whose powers are from -20.00 to +20.00 diopters with add powers from +0.75 to +2.75 diopters. The lenses may be worn by persons who exhibit astigmatism of 1.00 diopters or less where the astigmatism does not interfere with visual acuity. Eye care practitioners may prescribe the lens for daily wear disposable use (single use). The lenses are to be discarded upon removal. Therefore, no cleaning or disinfecting is required.

## Device Story

OxyAqua contact lenses are silicone hydrogel soft lenses (42% olifilcon D, 58% water) designed for daily disposable wear. Lenses incorporate a benzotriazole UV-absorbing monomer and Reactive Blue 19 handling tint. Manufactured via molding process; supplied in sterile buffered saline. Prescribed by eye care practitioners for patients with refractive ametropia, astigmatism, or presbyopia. Lenses are discarded after single use, eliminating need for cleaning or disinfection. Clinical benefit includes vision correction with acceptable ocular biocompatibility and safety profile comparable to predicate devices.

## Clinical Evidence

Three-month prospective clinical study comparing OxyAqua (olifilcon D) to predicate Si-Hy (olifilcon B) in at least 30 subjects. Endpoints included visual acuity (VA), adverse reactions, symptoms, and slit-lamp findings. Results showed mean VA was similar between groups (corrected to 0.1 logMAR or better). No significant slit-lamp findings (Grade 3 or 4) observed; no significant differences between groups.

## Technological Characteristics

Material: 42% olifilcon D, 58% water. UV blocker: benzotriazole monomer. Handling tint: Reactive Blue 19. Oxygen permeability (Dk): 85 (Fatt method). Refractive index: 1.394. Manufacturing: Molded. Standards: ISO 18369. Dimensions: 13.0-15.0 mm diameter, 8.0-9.2 mm base curve, 0.08 mm center thickness (at -3.00D).

## Regulatory Identification

A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.

## Predicate Devices

- Si-Hy (olifilcon B) Silicone Hydrogel Soft Contact Lens ([K160344](/device/K160344.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font below.

April 3, 2018

Visco Vision, Inc. % Jennifer Ting Manager Jens Medical Consulting, Ltd. 6F, No. 39 Jixian Rd., Luzhou Dist. New Taipei City, TW 247

Re: K173958

Trade/Device Name: OxyAqua (olifilcon D) spherical silicone hydrogel soft contact lens, OxyAqua (olifilcon D) multifocal silicone hydrogel soft contact lens, OxyAqua (olifilcon D) toric silicone hydrogel soft contact lens Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (Hydrophilic) Contact Lens Regulatory Class: Class II Product Code: LPL, MVN Dated: December 24, 2017 Received: February 5, 2018

Dear Jennifer Ting:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely yours,

# Denise L. Hampton -S

for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known) K173958

#### Device Name

OxyAqua (olifilcon D) Silicone Hydrogel Soft (hydrophilic) Contact Lens

#### Indications for Use (Describe)

The OxyAqua (olifilcon D) Spherical Silicone Hydrogel soft (hydrophilic) contact lenses are daily wear single use soft contact lens indicated for the corrective ametropia (myopia and hyperopia) in phakic or aphakic persons with non-diseased eyes who exhibit refractive astigmatism of 1.00D or less where the astigmatism does not interfere with visual acuity.

The OxyAqua (olifilcon D) Toric Silicone Hydrogel Soft (hydrophilic) Contact Lenses are indicated as daily wear for the correction of ametropia (myopia and hyperopia) with astigmatism in aphakic persons with non-diseased eyes and whose powers are from -20.00 to +20.00 diopters and astigmatic corrections are from -0.75 to -2.25 diopters.

The OxyAqua (olifilcon D) Multifocal Silicone Hydrogel Soft (hydrophilic) Contact lenses are indicated as daily wear for the correction of ametropia (myopia) and emmetropia with presbyopia in aphakic and non-aphakic persons with non-diseased eyes and whose powers are from -20.00 to +20.00 diopters with add powers from +0.75 to +2.75 diopters. The lenses may be worn by persons who exhibit astigmatism of 1.00 diopters or less where the astigmatism does not interfere with visual acuity.

Eye care practitioners may prescribe the lens for daily wear disposable use (single use). The lenses are to be discarded upon removal. Therefore, no cleaning or disinfecting is required.

Type of Use (Select one or both, as applicable)

| <span style="font-family: Arial, sans-serif;"> <span style="font-size: 10pt;"> <span style="color: black;"> <span style="font-weight: bold;">Prescription Use (Part 21 CFR 801 Subpart D)</span> </span> </span> </span> | <span style="font-family: Arial, sans-serif;"> <span style="font-size: 10pt;"> <span style="color: black;"> <span style="font-weight: bold;"></span> </span> </span> </span> |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial, sans-serif;"> <span style="font-size: 10pt;"> <span style="color: black;"> <span style="font-weight: bold;">Over-The-Counter Use (21 CFR 801 Subpart C)</span> </span> </span> </span>  |                                                                                                                                                                              |

CONTINUE ON A SEPARATE PAGE IF NEEDED

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

{3}------------------------------------------------

#### K173958 510 (k) Summary

Preparation Date: March 20, 2018

#### 5.1 Establishment Information:

| Name      | Visco Vision Inc.                                           |
|-----------|-------------------------------------------------------------|
| Address   | No. 1, Xingye St., Guishan Dist., Taoyuan City 33341 Taiwan |
| Contact   | Ted TT Huang                                                |
| Phone No. | 886-3-3490202                                               |
| Fax No.   | 886-3-3596868                                               |

#### 5.2 Contact Person:

| Name     | Evan Huang                    |
|----------|-------------------------------|
| Phone No | 886-3-3490202                 |
| Fax No   | 886-3-3596868                 |
| e-mail:  | evan.huang@viscovision.com.tw |

#### 5.3 Device Identification:

| Proprietary Name    | OxyAqua (olifilcon D) Silicone Hydrogel Soft Contact Lens                                                                                                         |
|---------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name         | Soft (hydrophilic) Contact Lenses                                                                                                                                 |
| Classification Name | Lenses, Soft Contact, Daily Wear,<br>(21 CFR 886.5925, Product Code LPL)<br>Lenses, Soft Contact, Daily Wear (Disposable),<br>(21 CFR 886.5925, Product Code MVN) |

Classification II

#### 5.4 Legally Marketed Equivalent Device:

| Predicate Device Name | Si-Hy (olifilcon B) Silicone Hydrogel Soft Contact Lens |
|-----------------------|---------------------------------------------------------|
| Manufacturer          | Visco Vision Inc.                                       |
| 510(k) Number         | K160344                                                 |
| Product Code          | MVN                                                     |

#### 5.5 Device Description

The OxyAqua (olifilcon D) Silicone Hydrogel Soft Contact Lens is made of silicone hydrogel material with UV blocker available as spherical lens, toric lens and multifocal lens. The composition of the lens is 42% olifilcon D and 58 % water. A light blue color tinted with "reactive Blue19" is for handling visibility purpose. A benzotriazole UV

{4}------------------------------------------------

absorbing monomer is used to block UV radiation. The transmittance characteristics are 2.4% (< 5%) in the UVB range of 280-315nm and 13.7% (< 50%) in the UVA range of 316-380nm. It is supplied in a sterile state packaged in a buffered saline solution.

#### 5.6 Indication for Use:

The OxyAqua (olifilcon D) Spherical Silicone Hydrogel soft contact lenses are daily wear single use soft contact lens indicated for the correction of refractive ametropia (myopia and hyperopia) in phakic or aphakic persons with non-diseased eyes who exhibit refractive astigmatism of 1.00D or less where the astigmatism does not interfere with visual acuity.

The OxyAqua (olifilcon D) Toric Silicone Hydrogel Soft Contact Lenses are indicated as daily wear for the correction of ametropia (myopia and hyperopia) with astigmatism in aphakic and non-aphakic persons with non-diseased eyes and whose powers are from -20.00 to +20.00 diopters and astigmatic corrections are from -0.75 to -2.25 diopters.

The OxyAqua (olifilcon D) Multifocal Silicone Hydrogel Soft Contact lenses are indicated as daily wear for the correction of ametropia (myopia and hyperopia) and emmetropia with presbyopia in aphakic and non-aphakic persons with non-diseased eyes and whose powers are from -20.00 to +20.00 diopters with add powers from +0.75 to +2.75 diopters. The lenses may be worn by persons who exhibit astigmatism of 1.00 diopters or less where the astigmatism does not interfere with visual acuity.

Eye care practitioners may prescribe the lens for daily wear disposable use (single use). The lenses are to be discarded upon removal. Therefore, no cleaning or disinfecting is required.

#### 5.7 Technological characteristic

OxyAqua SPHERICAL Soft Contact Lenses characteristics:

- Diameter Range: 13.0 to 15.0 mm .
- Base Curve:8.0 to 9.2 mm .
- Center Thickness : 0.08 mm for -3.00D (varies with power) .
- . Power:+20.00 to -20.00 D

#### OxyAqua TORIC Soft Contact Lenses characteristics:

- Diameter Range: 13.0 to 15.0 mm .

{5}------------------------------------------------

- Base Curve:8.0 to 9.2 mm •
- Center Thickness:0.08 mm for -3.00D (varies with power) •
- Power:+20.00 to -20.00 D •
- Cylinder: -0.75D ~ -2.25D .
- Axis: 10° to 180° (in 10° increments) •

OxyAqua MULTIFOCAL Soft Contact Lenses characteristics:

- Diameter Range:13.0 to 15.0 mm
- . Base Curve:8.0 to 9.2 mm
- Center Thickness:0.08 mm for -3.00D (varies with power) .
- Power:+20.00 to -20.00 D .
- Additional Powers: +0.75D ~ +1.25D (LOW) .
	- +1.50D ~ +2.00D (MID)
	- +2.25D ~ +2.75D (HIGH)

#### 5.8 Comparison table:

The characteristic comparison to predicate device is summarized in the following table.

| Similarities and differences      |                                                                 |                                                            |
|-----------------------------------|-----------------------------------------------------------------|------------------------------------------------------------|
| Item                              | Device                                                          | Predicate (K160344)                                        |
| Product Name                      | OxyAqua (olifilcon D)<br>Silicone Hydrogel soft<br>contact lens | Si-Hy (olifilcon B) Silicone<br>Hydrogel soft contact lens |
| Manufacturer                      | VISCO VISION Inc.                                               | VISCO VISION Inc.                                          |
| Intended Use                      | Myopia, Hyperopia                                               | Myopia, Hyperopia,<br>astigmatism, Presbyopia              |
| Lens Design                       | Spherical, toric, or<br>multifocal                              | Spherical, toric, or<br>multifocal                         |
| Replacement Schedule              | Daily Disposable (Single<br>use)                                | Daily Disposable (Single<br>use)                           |
| Chemical composition              | Olifilcon D                                                     | Olifilcon B                                                |
| Classification                    | Group II<br>(High water, nonionic)                              | Group I (Nonionic, Low<br>water)                           |
| Water Content                     | 58 %                                                            | 47 % (<50%),                                               |
| Oxygen Permeability<br>(DK, 35°C) | 85<br>(Fatt method)                                             | 120<br>(Fatt method)                                       |

{6}------------------------------------------------

| Base Curve Range (mm) | 8.2~9.0                                             | 8.0~9.2                                          |
|-----------------------|-----------------------------------------------------|--------------------------------------------------|
| Diameter (mm)         | 13.8~14.4                                           | 13.0~15.0                                        |
| Center Thickness      | Varies with design and power<br>(0.08 mm at -3.00D) | Varies with design and power (0.08 mm at -3.00D) |
| Powers                | -20.00D to +20.00D in 0.25 steps                    | -20.00D to +20.00D                               |
| Refractive Index      | 1.394                                               | 1.410                                            |
| Light Transmittance   | 94%                                                 | 94%                                              |
| Blue handling tint    | Reactive Blue19                                     | Reactive Blue19                                  |
| Method of Manufacture | Molded                                              | Molded                                           |

#### 5.9 Nonclinical Tests Performed

- 5.9.1 Physiochemical studies were conducted according to ISO 18369 First edition 2006-08-15, Ophthalmic optics - Contact lenses (Ophthalmic). The physical, optical and chemical properties of the lens are within established specifications for the lenses.
- 5.9.2 Toxicology studies report shows that the lenses are non-toxic and biocompatibility result is acceptable in ocular environment.

#### 5.10 Clinical Studies

A three-month clinical study was conducted to demonstrate the safety and effectiveness of the OxyAqua (olifilcon D) silicone hydrogel soft contact lens by comparison with Si-Hy (olifilcon B) silicone hydrogel soft contact lens when worn on a daily wear basis. At least 30 evaluable subjects were participated. Parameters measured include visual acuities, adverse reactions, symptom, problem and complaints, slit lamp findings, as well as lens wearing time. It was found that the mean VA for both lenses was similar at each visit. The visual acuity could be corrected to 0.1 (log MAR) or better. No significant slit lamp findings (i.e. Grade 3 or Grade 4) were noted. No significant differences in slit-lamp findings were found between groups.

The clinical study provided data to establish substantial equivalence with the predicate, control lens, Si-Hy (olifilcon B) silicone hydrogel soft contact lens, Visco Vision Inc. (K160344), with respect to the safety and effectiveness.

{7}------------------------------------------------

## 5.11 Conclusion

Comparison to the predicate device for chemical composition, physical and optical properties, it shows that "OxyAqua silicone hydrogel soft contact lens" is as safe, as effective and performs as well as the predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K173958](https://fda-staging.innolitics.com/device/K173958)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
