K173955 · Halyard Health, Inc. · KNT · Jan 26, 2018 · Gastroenterology, Urology
Device Facts
Record ID
K173955
Device Name
Mic-Key* SF Over-the-Wire Stoma Measuring Device
Applicant
Halyard Health, Inc.
Product Code
KNT · Gastroenterology, Urology
Decision Date
Jan 26, 2018
Decision
SESK
Submission Type
Special
Regulation
21 CFR 876.5980
Device Class
Class 2
Indications for Use
MIC-KEY* SF Over-the-Wire Stoma Measuring Device is indicated for measuring the length of a stoma prior to placement of a low profile feeding tube.
Device Story
Device is a 10 Fr tubular shaft with graduated scale, inflation valve, and retaining balloon; used to measure stoma length prior to low-profile feeding tube placement. Operated by clinicians in hospital environments. Input is physical stoma dimensions; device is inserted over a guidewire, balloon is inflated, and length is measured via graduated scale. Output is a physical measurement of stoma length, which informs the selection of the appropriate feeding tube size. Benefits include accurate sizing to ensure proper fit and function of the subsequent feeding tube.
Clinical Evidence
Bench testing only. Verification tests included tensile strength (EN 1615:2000), balloon concentricity and diameter (ASTM F 2528-06), radiopacity (ASTM F640-7), sterile barrier integrity (ASTM F 2096-11), and system leak tests. Biocompatibility testing performed per ISO 10993-1, -5, -10, and -11. All tests met acceptance criteria.
Technological Characteristics
Polyurethane tubing; 10 Fr shaft; EtO sterilized; single-use. Features include graduated scale, inflation valve, and retaining balloon. Complies with EN 1615:2000 (tensile strength), ASTM F 2528-06 (balloon/tube dimensions), ASTM F640-7 (radiopacity), and ASTM F 2096-11 (sterile barrier). Biocompatibility per ISO 10993 series.
Indications for Use
Indicated for measuring stoma length prior to placement of a low profile feeding tube in patients requiring such placement.
Regulatory Classification
Identification
A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).
Special Controls
*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.
Predicate Devices
MIC-KEY* SF Over the Wire Stoma Measuring Device (K122653)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 26, 2018
Halyard Health, Inc. David M. Lee, JD Associate Director, Regulatory Affairs 5405 Windward Parkway Alpharetta, GA 30004
Re: K173955
> Trade/Device Name: MIC-KEY* SF Over-the-Wire Stoma Measuring Device Regulation Number: 21 CFR§ 876.5980 Regulation Name: Gastrointestinal Tube and Accessories Regulatory Class: II Product Code: KNT Dated: December 26, 2017 Received: December 28, 2017
Dear David M. Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit/tray. The general controls of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Benjamin R. Fisher -S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K173955
Device Name
MIC-KEY* SF Over-the-Wire Stoma Measuring Device
Indications for Use (Describe)
MIC-KEY* SF Over-the-Wire Stoma Measuring Device is indicated for measuring the length of a stoma prior to placement of a low profile feeding tube.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------|
| <div style="display:flex; align-items:center;"><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</div> | <div style="display:flex; align-items:center;"><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(K) SUMMARY- K173955
#### Date: January 20, 2018
#### Applicant's Name, Address, Telephone, Contact Person
Halvard Health. Inc. 5405 Windward Parkway Alpharetta, Georgia 30004
Contact Name: David M. Lee, JD Associate Director, Requlatory Affairs Halvard Health 470 448 5182 (Phone) David.lee@hyh.com
#### 1. CLASSIFICATION, COMMON OR USUAL NAME, DEVICE NAME
| Classification: | Class II, per 21 CFR 876.5980 |
|----------------------|--------------------------------------------------|
| Classification Name: | Gastrointestinal tube and accessories |
| Common/Usual Name: | Stoma Measuring Device |
| Product Code: | KNT |
| Trade Name: | MIC-KEY* SF Over-the-Wire Stoma Measuring Device |
#### 2. PREDICATE DEVICE
MIC-KEY* SF Over the Wire Stoma Measuring Device cleared under the MIC-KEY*LOW PROFILE GASTROSTOMY TUBE AND ACCESSORIES-K122653.
#### 3. INDICATIONS FOR USE
MIC-KEY* SF Over-the-Wire Stoma Measuring Device is indicated for measuring the length of a stoma prior to placement of a low profile feeding tube.
#### 4. DESCRIPTION OF DEVICE
MIC-KEY* SF Over-the-Wire Stoma Measuring Device is designed for measuring the length of a stoma prior to placement of a low profile feeding tube. The Stoma Measuring Device comprises a tubular shaft 10 Fr with graduated scale, an inflation valve and retaining balloon. The MIC-KEY* SF Over the Wire Stoma Measuring Device is made of polyurethane tubing, EtO sterilized and for single use. Like the predicate device, it is intended to be use in hospital environment.
#### 5. TECHNOLOGICAL CHARACTERISTICS
The MIC-KEY* SF Over-the-Wire Stoma Measuring Device is substantially equivalent to the predicate device. The following changes from the predicate device were made:
- . the printed device labeling (printed marking was added to the polyurethane molded head indicating the balloon inflation volume)
- minor changes to the geometry of the molded polyurethane head ●
- minor change to the geometry of the molded tethered cap ●
- the ink used for device marking was changed o
There are no other significant differences in the design, materials, performance, and technological characteristics from the predicate device.
{4}------------------------------------------------
### 6. SUMMARY OF NONCLINICAL TESTS
Performance testing on the device under submission was conducted to demonstrate the modified device continued to meet performance specification. Results of design verification and validation activities did not raise any new issues of safety or effectiveness. Non-clinical verification was conducted through the risk management process according to ISO 14971:2012. The following verification tests were conducted:
| Test | Acceptance Criterion | Result |
|--------------------------------------|-------------------------------------------------------------------------------------|--------|
| Tensile Strength, Head/Tube | ≥ 15 N per EN 1615:2000 | Pass |
| Balloon Concentricity | < 2:1 per ASTM F 2528-06 | Pass |
| Deflated Balloon Outer Diameter | Pass through 4 Fr size larger ring gauge<br>per ASTM F 2528-06 (without lubricant) | Pass |
| Deflated Balloon Outer Diameter | Pass through 2 Fr size larger ring gauge<br>per ASTM F 2528-06 (with lubricant) | Pass |
| Tube Outer Diameter | Pass through 1 Fr size larger ring gauge<br>per ASTM F 2528-06 | Pass |
| Balloon Valve Luer Tip Compatibility | Able to inflate balloon with a Luer tip<br>syringe | Pass |
| Printed Ink Durability | Printing remains legible following cleaning | Pass |
| Device Removal Force | ≤ 0.48 lbf | Pass |
| Balloon System Leak Test | No leakage when the balloon is inflated<br>for five minutes | Pass |
| Collar Movement | The collar can slide along the tubing for<br>the entire length of the printed ruler | Pass |
| System Leak Test, Cap/Head | No leakage when pressurized to 0.5 psi | Pass |
| Guidewire Insertion | Device can slide along a 0.038" diameter<br>guidewire | Pass |
| Device Radiopacity | ASTM F640-7 | Pass |
| Sterile Barrier | No bubbles per ASTM F 2096-11 | Pass |
The following biocompatibility tests were also conducted, and all met acceptance criteria:
ISO 10993-1 Fourth edition 2009-10-15 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
AAMI ANSI ISO 10993-5, 2009: Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity
ISO 10993-10:2010: Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization
ISO 10993-11:2006/(R2010): Biological evaluation of medical devices - Part 11: Tests for systemic toxicity
#### 7. OVERALL PERFORMANCE CONCLUSIONS
The above nonclinical studies demonstrate MIC-KEY* SF Over-the-Wire Stoma Measuring Device is substantial equivalence to the predicate MIC-KEY* SF Over-the-Wire Stoma Measuring Device (K122653) in intended use, design, materials, performance, and biocompatibility attributes. There are no new questions of safety and effectiveness as compared to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.