Endocular Viewing Lenses and Silicone Ring

K173944 · Phakos · HJK · Oct 25, 2018 · Ophthalmic

Device Facts

Record IDK173944
Device NameEndocular Viewing Lenses and Silicone Ring
ApplicantPhakos
Product CodeHJK · Ophthalmic
Decision DateOct 25, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 886.1385
Device ClassClass 2

Indications for Use

The endocular viewing lens is a passive device that is used to view the fundus of the eye by placing it on the cornea.

Device Story

Passive diagnostic contact lenses; used to visualize ocular structures (retina, vitreous, lens, iris, anterior chamber) during surgery or consultation; placed directly on patient cornea; stabilizes via optional flexible silicone ring; provides magnified view of fundus; used by ophthalmologists in surgical/clinical settings; no electronic components or software; benefits include improved visualization of intraocular anatomy during procedures.

Clinical Evidence

No clinical data. Bench testing only, including biocompatibility testing per ISO 10993-5 (cytotoxicity) and ISO 10993-10 (irritation and sensitization).

Technological Characteristics

Passive diagnostic contact lenses. Materials: PMMA (polymethyl methacrylate), OKP4 Polyester Optical Plastic, and MED 4035 silicone (for rings). Form factors: Vitrectomy, Gonioprism, Panoramic Wide Field, 3-prism, and Capsulotomy lenses. Non-powered, non-electronic, standalone devices.

Indications for Use

Indicated for visualization of the retina, vitreous humor, lens, iris, anterior chamber, and other ocular structures during surgical procedures or consultation.

Regulatory Classification

Identification

A polymethylmethacrylate (PMMA) diagnostic contact lens is a device that is a curved shell of PMMA intended to be applied for a short period of time directly on the globe or cornea of the eye for diagnosis or therapy of intraocular abnormalities.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue. October 25, 2018 PHAKOS % J. D. Webb Official Correspondent The OrthoMedix Group, Inc. 1001 Oakwood Blvd Round Rock, Texas 78681 Re: K173944 Trade/Device Name: Endocular Viewing Lenses and Silicone Ring Regulation Number: 21 CFR 886.1385 Regulation Name: Polymethylmethacrylate (PMMA) Diagnostic Contact Lens Regulatory Class: Class II Product Code: HJK Dated: September 19, 2018 Received: September 25, 2018 Dear J. D. Webb: We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Image /page/1/Picture/5 description: The image shows the name "J. Angelo Green" in a large, bold font. The text is black and appears to be centered on a white background. The name is likely that of a person, and the image may be a title or heading. for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K173944 Device Name ENDOCULAR VIEWING LENSES AND SILICONE RING Indications for Use (Describe) The ENDOCULAR VIEWING LENSES AND SILICONE RING family allows visualizations of the retina, the vireous humor, the lens, the iris, the anterior chamber, as well as other ocular structures during surgical procedures or consultation. | Type of Use (Select one or both, as applicable) | Prescription Use (Part 21 CFR 201.326(a) and D) Over-The-Counter Use (21 CFR 201.326(c)) | |-------------------------------------------------|--------------------------------------------------------------------------------------------| | | Prescription Use (Part 21 CFR 201.326(a) and D) Over-The-Counter Use (21 CFR 201.326(c)) | > Prescription Use (Part 21 CFR 801 Subpart D) | | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## 510(k) Summary: ENDOCULAR VIEWING LENSES AND SILICONE RING In accordance with 21 CFR 807.92 of the Federal Code of Regulations | Date Prepared | October 22, 2018 | |--------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Submitted By | PHAKOS<br>62 Rue Kléber<br>93100 Montreuil<br>FRANCE | | Primary Contact | J.D. Webb<br>1001 Oakwood Blvd<br>Round Rock, TX 78681 512-388-<br>0199 Tele<br>e-mail: jdwebb@orthomedix.net | | Trade Name | Endocular Viewing Lenses and Silicone Ring | | Common Name | Lens, Contact, Polymethylmethacrylate, Diagnostic | | Classification Name | Polymethylmethacrylate (PMMA) diagnostic contact lens | | Class | ll | | Product Code | HJK | | CFR Section | 21 CFR section 886.1385 | | Device Panel | Ophthalmic | | Primary Predicate<br>Device | Disposable Vitrectomy Lens, OCULAR Instruments, Inc. (K012096) | | Secondary Predicate<br>Devices | Disposable Vitrectomy Lens, VOLK Optical, Inc. (K050623/ K151961)<br>Family of Vitrectomy Lenses, SENSOR Medical Technology LLC (K140368/ K142715) | | Device Description | The "ENDOCULAR VIEWING LENSES AND SILICONE RING" family includes devices<br>comprised of an endocular viewing lens and/or of a cornea support that allows for<br>stabilization of this lens on the eye. This range allows one to view the retina, the vitreous<br>humor, the lens, the iris, the angle of the anterior chamber, as well as other ocular<br>structures during surgical procedures or consultation. | | Materials | Conventional lens:<br>PMMA polymethyl methacrylate<br>High Resolution lens:<br>OKP4 Polyester Optical Plastic<br>Flexible removable<br>ring:<br>MED 4035 silicone<br>Silicone ring (alone):<br>MED 4035 silicone | | Intended Use | The endocular viewing lens is a passive device that is used to view the fundus of the eye<br>by placing it on the cornea. | {4}------------------------------------------------ ### K173944 | Substantial<br>Equivalence Claimed<br>to Predicate Devices | The Endocular Viewing Lenses and Silicone Ring is substantially equivalent to the<br>predicate devices in terms of intended use, design, materials used, mechanical safety<br>and performances. | | | | | |--------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------| | Indications for Use | The ENDOCULAR VIEWING LENSES AND SILICONE RING family allows visualizations<br>of the retina, the vitreous humor, the lens, the iris, the angle of the anterior chamber, as<br>well as other ocular structures during surgical procedures or consultation. | | | | | | | Device | PHAKOS | OCULAR | VOLK | SENSOR | | Summary of the<br>Technological<br>Characteristics<br>Compared to<br>Predicate | Intended Use | The endocular<br>viewing lens is<br>a passive<br>device that is<br>used to view<br>the fundus of<br>the eye by<br>placing it on the<br>cornea. | Allow<br>visualization of<br>the ocular<br>fundus,<br>vitreous, and<br>retinal<br>structures<br>during<br>vitrectomy<br>surgery. | For use as<br>diagnostic<br>lenses for eye<br>fundus<br>examinations<br>and use in the<br>therapy of<br>intraocular<br>abnormalities. | Used in the<br>examination of<br>the eye<br>fundus, retina<br>and irido-<br>corneal and<br>vitreous<br>bodies | | | Type of<br>Lenses | - Vitrectomy<br>- Gonioprism<br>- Panoramic<br>Wide Field<br>- 3-prism<br>- Capsulotomy<br>- Silicone Ring | - Vitrectomy<br>- Gonioprism<br>- Panoramic<br>Wide Field<br>- 3-prism<br>- Capsulotomy<br>- Silicone Ring | - Vitrectomy<br>- Gonioprism<br>- Panoramic<br>Wide Field<br>- 3-prism<br>- Capsulotomy | - Vitrectomy<br>- Gonioprism<br>- Panoramic<br>Wide Field<br>- 3-prism<br>- Capsulotomy<br>- Silicone Ring | | | Comparison of<br>Lenses | - All of the systems include the full range of Vitrectomy Lens, with the<br>exception of the SENSOR not having a Wide-Angle Lens. The<br>SENSOR has no HR Vitrectomy Lens<br>- All of the systems include Gonioprism Lens.<br>- All systems include a Panoramic Wide Field Lens. All of the systems<br>include 3-prism Lens.<br>- All systems include a Capsulotomy Lens. | | | | | | Material | - PMMA<br>- OKP 4<br>(acrylic)<br>- silicone | - PMMA<br>- Acrylic<br>- Glass<br>- Quartz | - PMMA<br>- Acrylic<br>- Glass | - PMMA | | Non-clinical Test<br>Summary | The following tests were performed:<br>- ISO 10993-5 In vitro cytotoxicity<br>- ISO 10993-10 Irritation<br>- ISO 10993-10 Sensitization | | | | | | Clinical Test<br>Summary | No clinical studies were performed | | | | | {5}------------------------------------------------ | Conclusions: | PHAKOS considers the Endocular Viewing Lenses and Silicone Ring to be equivalent to | |-----------------------|---------------------------------------------------------------------------------------------| | Non- | the predicate devices listed above. This conclusion is based upon the devices' similarities | | clinical and Clinical | in principles of operation, technology, materials and indications for use |
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...