K173916 · Finapres Medical Systems B.V. · DXN · Nov 6, 2018 · Cardiovascular
Device Facts
Record ID
K173916
Device Name
Finapres Nova Noninvasive Hemodynamic Monitor
Applicant
Finapres Medical Systems B.V.
Product Code
DXN · Cardiovascular
Decision Date
Nov 6, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1130
Device Class
Class 2
Indications for Use
The Finapres NOVA is intended to be used with patients who have a need for a noninvasive blood pressure and hemodynamic monitor. The noninvasive blood pressure waveform is measured on the subject's finger. The Finapres NOVA provides a noninvasive characterization of the arterial circulation and its beat-to-beat variability in pressure and flow and in various hemodynamic parameters derived from these pressure and flow signals, such as heart rate variability (HRV) and Baroreflex sensitivity (BRS). Cardiac output derived from the flow signal requires a calibration with thermal dilution. The Finapres NOVA has the option to include additional modules to extend its functionality with ECG and SpO2 measurements and blood pressure calibration. When the SpO2 module is present, the Finapres NOVA can additionally monitor the functional oxygen saturation of arterial hemoglobin (SpO2) and the pulse rate. When the ECG module is present, the Finapes NOVA can additionally monitor the ECG parameters of a patient and their pulse rate. Alarms concerning the pulse rate will be available from the monitor. When the blood pressure calibration module is present, the Finapres NOVA can additionally provide an upper arm noninvasive blood pressure measurement to determine the blood pressure value for calibration. The Finapres NOVA is intended to be used for subjects above 18 years of age. The Finapres NOVA is intended for use in hospitals, clinics, and research centers.
Device Story
Device monitors arterial circulation via finger cuff; inputs include photoplethysmography (PPG) signals from finger; optional modules for ECG, SpO2, and upper-arm blood pressure calibration. Operates using Peñáz arterial volume-clamp method and Wesseling Physiocal criteria to maintain finger artery volume constant; pressure required to do so reflects arterial blood pressure. System consists of base station and Nano Core (FEU NC) wrist unit; FEU NC performs pressure generation and signal processing. Used by clinicians in hospitals/clinics. Output includes continuous blood pressure waveforms, heart rate, HRV, BRS, and cardiac output (requires thermal dilution calibration). Data aids hemodynamic assessment and clinical decision-making. Benefits include noninvasive, continuous monitoring of hemodynamic status.
Clinical Evidence
Clinical investigation performed per ISO 81060-2:2013 to validate blood pressure measurement accuracy against invasive reference. Hemodynamic parameters were calculated using both invasive reference and subject device. Results demonstrated compliance with accuracy requirements and substantial equivalence to predicate.
Technological Characteristics
Noninvasive hemodynamic monitor; uses arterial volume-clamp method (Peñáz) and Physiocal criteria. Components: base station, FEU NC wrist unit, finger cuffs. Materials: aluminum/plastic casing, TPU cable. Connectivity: wired interface between FEU NC and base station. Software: C++ firmware (ISO 62304 compliant) for signal processing and pressure control. Power: 100-240 Vac. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-2-27, IEC 80601-2-30.
Indications for Use
Indicated for patients >18 years requiring noninvasive blood pressure and hemodynamic monitoring. Used in hospitals, clinics, and research centers.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
November 6, 2018
Finapres Medical Systems B.V. Iris van Uitert Quality Manager, Finapres Medical Systems V.V. Institutenweg 25 Enschede, 7521 PH NI
### Re: K173916
Trade/Device Name: Finapres NOVA Noninvasive Hemodynamic Monitor Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN, DSB, DQA, DRT Dated: September 24, 2018 Received: September 26, 2018
Dear Iris van Uitert:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Image /page/1/Picture/5 description: The image shows the text "Shawn W. Forrest -A" in a large, sans-serif font. The text is arranged in two lines, with "Shawn W." on the first line and "Forrest -A" on the second line. The background is a light blue color with a subtle pattern of lighter blue rectangles.
sitally signed by Shawn W. Forres HHS OU=FDA OU=Peonle 342 19200300 100 1 1=130040334 e: 2018.11.06 11:21:49 -05'00
for: Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
K173916
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known)
Device Name Finapres NOVA
Indications for Use (Describe)
The Finapres NOVA is intended to be used with patients who have a need for a noninvasive blood pressure and hemodynamic monitor. The noninvasive blood pressure waveform is measured on the subject's finger. The Finapres NOVA provides a noninvasive characterization of the arterial circulation and its beat-to-beat variability in pressure and flow and in various hemodynamic parameters derived from these pressure and flow signals, such as heart rate variability (HRV) and Baroreflex sensitivity (BRS).
Cardiac output derived from the flow signal requires a calibration with thermal dilution.
The Finapres NOVA has the option to include additional modules to extend its functionality with ECG and SpO2 measurements and blood pressure calibration.
When the SpO2 module is present, the Finapres NOVA can additionally monitor the functional oxygen saturation of arterial hemoglobin (SpO2) and the pulse rate.
When the ECG module is present, the Finapes NOVA can additionally monitor the ECG parameters of a patient and their pulse rate. Alarms concerning the pulse rate will be available from the monitor.
When the blood pressure calibration module is present, the Finapres NOVA can additionally provide an upper arm noninvasive blood pressure measurement to determine the blood pressure value for calibration.
The Finapres NOVA is intended to be used for subjects above 18 years of age.
The Finapres NOVA is intended for use in hospitals, clinics, and research centers.
| Type of Use (Select one or both, as applicable) | | |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
| <table border="0"><tr><td><span style="font-family: DejaVu Sans, sans-serif">❌</span> Prescription Use (Part 21 CFR 801 Subpart D)</td><td><span style="font-family: DejaVu Sans, sans-serif">□</span> Over The Counter Use (21 CFR 801 Subpart G)</td></tr></table> | <span style="font-family: DejaVu Sans, sans-serif">❌</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: DejaVu Sans, sans-serif">□</span> Over The Counter Use (21 CFR 801 Subpart G) |
| <span style="font-family: DejaVu Sans, sans-serif">❌</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: DejaVu Sans, sans-serif">□</span> Over The Counter Use (21 CFR 801 Subpart G) | |
> Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
# Change of front-end unit of Finapres Nova to front-end unit Nano Core (FEU NC)
This 510(k) summary is being submitted in accordance with the requirements of SMDA and 21CFR § 807.92.
# 1. Submitter's information
| Submitter: | Finapres Medical Systems B.V.<br>Institutenweg 25<br>7521 PH Enschede<br>The Netherlands<br>Phone number: +31 88 115 2700<br>Fax Number: +31 20 609 0677<br>Operator Number: 9051428<br>Registration number: 3003803088 |
|-----------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact person: | Iris van Uitert, PhD<br>Quality Manager<br>Phone number: +31 88 115 2700<br>E-mail:iris.van.uitert@finapres.com |
Date of preparation: November 1, 2018
# 2. Device information Finapres NOVA
| Trade name: | Finapres NOVA Noninvasive Hemodynamic Monitor |
|---------------------------------------------------------|---------------------------------------------------------------------------------------------|
| Common name(s) | Noninvasive Blood Pressure Monitor<br>Hemodynamic Monitor<br>Electrocardiograph<br>Oximeter |
| Classification name | See Table 1 |
| Device classification | Class II |
| 510(k) number under which it was<br>previously cleared. | K141460 and K160967 |
| | | Table 1. Classification name Finapres NOVA | |
|--|--|--------------------------------------------|--|
| | | | |
| Classification name | 21 CFR Section | Product Code |
|-------------------------------------------------------|----------------|--------------|
| Noninvasive Blood Pressure Monitor Measurement System | 870.1130 | DXN |
| Plethysmograph, Impedance | 870.2770 | DSB |
| Oximeter | 870.2700 | DQA |
| Cardiac monitor | 870.2300 | DRT |
{4}------------------------------------------------
# 3. Predicate Device for the proposed change
The Finapres NOVA remains substantially equivalent in design (methodology) and indications for use to the primary predicate shown in Table 2
The additional AT software module (not present in K141460) is substantially equivalent in design (methodology) to the software algorithms used on the secondary predicate shown in Table 2
The indications for use of the Finapres NOVA is substantially equivalent to the devices shown in Table 2 that has previously been cleared.
Table 2. Primary and secondary predicate devices for the Finapres Nova with front-end unit Nano Core
| Device name | Manufacturer | 510(k) |
|--------------------------|--------------------------|---------|
| Primary: Finapres Nova | Finapres Medical Systems | K141460 |
| Secondary: Finapres Nova | Finapres Medical Systems | K160967 |
### 4. Description of the Finapres with FEU NC
The Finapres NOVA with FEU NC is based on existing Finapres Medical Systems B.V. product: the Finapres Nova. The changes between the Finapres NOVA with FEU NC and the previous systems are described below.
The current hardware that generates pressure in the Finapres NOVA contains components which are end-of-life. Therefore, they will need to be replaced. It was decided not to redesign these PCBs, but to redesign the whole pressure control hardware for reasons mentioned below.
In the new NOVA with FEU NC, the wrist unit FEU of the current NOVA system is replaced by the wrist unit FEU Nano Core. The FEU Nano Core wrist unit has the capability of generating pressure in the wrist unit, whereas with the FEU wrist unit the pressure was generated in the NOVA base station. This means that the pump, printed circuit boards (PCBs) and FEU have been replaced by a new power supply PCB and the FEU NC. By moving the pressure generation and control to the Nano Core wrist unit, there is no need for a pressure cable from the base station to the wrist unit anymore, which increases the mobility of the system and patient.
# 5. Indications for use
The Finapres NOVA is intended to be used with patients who have a need for a noninvasive blood pressure and hemodynamic monitor. The noninvasive blood pressure waveform is measured on the subject's finger. The Finapres NOVA provides a noninvasive characterization of the arterial circulation and its beat-to-beat variability in pressure and flow and in various hemodynamic parameters derived from these pressure and flow signals, such as heart rate variability (HRV) and Baroreflex sensitivity (BRS).
Cardiac output derived from the flow signal requires a calibration with thermal dilution.
The Finapres NOVA has the option to include additional modules to extend its functionality with ECG and SpO2 measurements and blood pressure calibration.
When the SpO2 module is present, the Finapres NOVA can additionally monitor the functional oxygen saturation of arterial hemoglobin (SpO2) and the pulse rate.
{5}------------------------------------------------
When the ECG module is present, the Finapres NOVA can additionally monitor the ECG parameters of a patient and their pulse rate. Alarms concerning the pulse rate will be available from the monitor.
Using the blood pressure calibration module, the Finapres NOVA can additionally provide an upper arm non-invasive blood pressure measurement to determine the blood pressure value for calibration.
The Finapres NOVA is intended to be used for subjects above 18 years of age.
The Finapres NOVA is intended for use in hospitals, clinics and research centers.
{6}------------------------------------------------
### 6. Comparison of technical characteristics to predicate devices K141460 and K160967
The modified Finapres NOVA with FEU NC has the finapres Nova from submission K141460 and K160967:
- Both devices use the same operating principles and use the same algorithms, see Table 3 below. .
In summary, the Finapres NOVA with FEU NC described in this submission is, in our opinion, substantially equivalent to the predicate device Finapres NOVA.
| Parameter | Finapres NOVA K141460 | Finapres NOVA K160967 | Finapres NOVA with FEU NC | Different? |
|-------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>use | The Finapres NOVA is intended to be<br>used with patients who have a need<br>for a noninvasive blood pressure and<br>hemodynamic monitor. The<br>noninvasive blood pressure waveform<br>is measured on the subject's finger.<br>The Finapres NOVA provides a<br>noninvasive characterization of the<br>arterial circulation and its beat-to-beat<br>variability in pressure and flow and in<br>various hemodynamic parameters<br>derived from these pressure and flow<br>signals.<br>Cardiac output derived from the flow<br>signal requires a calibration with<br>thermal dilution.<br><br>The Finapres NOVA has the option to<br>include additional modules to extend<br>its functionality with ECG and SpO2<br>measurements and blood pressure<br>calibration. | The Finapres NOVA is intended to be<br>used with patients who have a need<br>for a noninvasive blood pressure and<br>hemodynamic monitor.<br>The noninvasive blood pressure<br>waveform is measured on the<br>subject's finger. The Finapres NOVA<br>provides a noninvasive<br>characterization of the arterial<br>circulation and its beat-to-beat<br>variability in pressure and flow and in<br>various hemodynamic parameters<br>derived from these pressure and flow<br>signals, such as heart rate variability<br>(HRV) and Baroreflex sensitivity<br>(BRS).<br>Cardiac output derived from the flow<br>signal requires a calibration with<br>thermal dilution.<br><br>The Finapres NOVA has the option to<br>include additional modules to extend<br>its functionality with ECG and SpO2<br>measurements and blood pressure<br>calibration. | The Finapres NOVA with FEU NC is<br>intended to be used with patients who<br>have a need for a noninvasive blood<br>pressure and hemodynamic monitor.<br>The noninvasive blood pressure<br>waveform is measured on the<br>subject's finger. The Finapres NOVA<br>with FEU NC provides a noninvasive<br>characterization of the arterial<br>circulation and its beat-to-beat<br>variability in pressure and flow and in<br>various hemodynamic parameters<br>derived from these pressure and flow<br>signals, such as heart rate variability<br>(HRV) and Baroreflex sensitivity<br>(BRS).<br>Cardiac output derived from the flow<br>signal requires a calibration with<br>thermal dilution.<br><br>The Finapres NOVA with FEU NC has<br>the option to include five additional<br>modules to extend its functionality<br>with ECG, SpO2, blood pressure<br>calibration and data transfer to and<br>from the device. | Yes, see note 1 below<br>this table. |
| Parameter | Finapres NOVA K141460 | Finapres NOVA K160967 | Finapres NOVA with FEU NC | Different? |
| | When the SpO2 module is present, the<br>Finapres NOVA can additionally<br>monitor the functional oxygen<br>saturation of arterial hemoglobin<br>(SpO2) and the pulse rate. | When the SpO2 module is present, the<br>Finapres NOVA can additionally<br>monitor the functional oxygen<br>saturation of arterial hemoglobin<br>(SpO2) and the pulse rate. | When the SpO2 module is present, the<br>Finapres NOVA with FEU NC can<br>additionally monitor the functional<br>oxygen saturation of arterial<br>hemoglobin (SpO2) and the pulse<br>rate. | |
| | When the ECG module is present, the<br>Finapres NOVA can additionally<br>monitor the ECG parameters of a<br>patient and their pulse rate. Alarms<br>concerning the pulse rate will be<br>available from the monitor. | When the ECG module is present, the<br>Finapres NOVA can additionally<br>monitor the ECG parameters of a<br>patient and their pulse rate. Alarms<br>concerning the pulse rate will be<br>available from the monitor. | When the ECG module is present, the<br>Finapres NOVA with FEU NC can<br>additionally monitor the ECG<br>parameters of a patient and their<br>pulse rate. Alarms concerning the<br>pulse rate will be available from the<br>monitor. | |
| | When the blood pressure calibration<br>module is present, the Finapres NOVA<br>can additionally provide an upper arm<br>non-invasive blood pressure<br>measurement to determine the blood<br>pressure value for calibration. | When the blood pressure calibration<br>module is present, the Finapres NOVA<br>can additionally provide an upper arm<br>non-invasive blood pressure<br>measurement to determine the blood<br>pressure value for calibration. | When the blood pressure calibration<br>module is present, the Finapres NOVA<br>with FEU NC can additionally provide<br>an upper arm non-invasive blood<br>pressure measurement to determine<br>the blood pressure value for<br>calibration. | |
| | The Finapres NOVA is intended to be<br>used for subjects above 18 years of<br>age. | The Finapres NOVA is intended to be<br>used for subjects above 18 years of<br>age. | The Finapres NOVA with FEU NC is<br>intended to be used for subjects above<br>18 years of age. | |
| | The Finapres NOVA is intended for use<br>in a professional medical environment. | The Finapres NOVA is intended for use<br>in hospitals, clinics and research<br>centers. | The Finapres NOVA is intended for use<br>in hospitals, clinics and research<br>centers, see note 1 below this table. | |
| Measurement<br>method /<br>Algorithm | The Finapres NOVA uses the algorithm<br>arterial volume-clamp method of J.<br>Peñáz and the Physiocal criteria (Servo<br>SelfAdjust) of K.H. Wesseling for the<br>continuous measurement of blood<br>pressure | The Finapres NOVA uses the algorithm<br>arterial volume-clamp method of J.<br>Peñáz and the Physiocal criteria (Servo<br>SelfAdjust) of K.H. Wesseling for the<br>continuous measurement of blood<br>pressure | The Finapres NOVA with FEU NC uses<br>the algorithm arterial volume-clamp<br>method of J. Peñáz and the Physiocal<br>criteria (Servo SelfAdjust) of K.H.<br>Wesseling for the continuous<br>measurement of blood pressure | No |
| Parameter | Finapres NOVA K141460 | Finapres NOVA K160967 | Finapres NOVA with FEU NC | Different? |
| | The Finapres NOVA uses the<br>NOVAScope/Modelflow software<br>technology to analyze the trend<br>hemodynamic parameters on the<br>device and main module firmware to<br>control the measurement | The Finapres NOVA uses the<br>NOVAScope/Modelflow software<br>technology to analyze the trend<br>hemodynamic parameters on the<br>device and main module firmware to<br>control the measurement | The Finapres NOVA with FEU NC uses<br>the NOVAScope/Modelflow software<br>technology to analyze the trend<br>hemodynamic parameters on the<br>device. The firmware of the FEU NC<br>controls the measurement | Yes, see note 2 below<br>this table |
| Maximum<br>measurement<br>duration | Maximum 24 h continuous<br>measurement | Maximum 24 h continuous<br>measurement | Maximum 24 h continuous<br>measurement | No |
| How supplied | Stationary Base station to which<br>subject is connected and to generate<br>pressure | Stationary Base station to which<br>subject is connected and to generate<br>pressure | Stationary Base station to which<br>subject is connected (no pressure<br>generation) | Yes, different materials<br>and design, see note 3<br>below this table |
| | Front-end unit to interface the signals<br>from the electronics and air supply in<br>the main module to the finger cuffs | Front-end unit to interface the signals<br>from the electronics and air supply in<br>the main module to the finger cuffs | Front-end unit to generate air pressure<br>signals to and interpret electronic<br>signals from the finger cuffs. Receives<br>data and power from base station,<br>sends blood pressure data to base<br>station | Yes, different material<br>of the FEU cable and<br>design, see note 3<br>below this table |
| | Finger cuffs for continuous blood<br>pressure measurement | Finger cuffs for continuous blood<br>pressure measurement | Finger cuffs for continuous blood<br>pressure measurement | no |
| | Hydrostatic height correction unit<br>compensates for vertical movement of<br>the hand relative to heart level<br>(optional). | Hydrostatic height correction unit<br>compensates for vertical movement of<br>the hand relative to heart level<br>(optional). | Hydrostatic height correction unit<br>compensates for vertical movement of<br>the hand relative to heart level<br>(optional). | no |
| | Analog input/output module (optional) | Analog input/output module (optional) | Analog input/output module (optional) | no |
| | Upper arm calibration module with<br>arm cuff for pressure calibration<br>(optional) | Upper arm calibration module with<br>arm cuff for pressure calibration<br>(optional) | Upper arm calibration module with<br>arm cuff for pressure calibration | no |
| | ECG module with 3 or 5 lead ECG<br>cable (optional) | ECG module with 3 or 5 lead ECG<br>cable (optional) | ECG module with 3 or 5 lead ECG<br>cable (optional) | no |
| | SpO2 module with SpO2 finger clip<br>(optional) | SpO2 module with SpO2 finger clip<br>(optional) | SpO2 module with SpO2 finger clip<br>(optional) | no |
| Parameter | Finapres NOVA K141460 | Finapres NOVA K160967 | Finapres NOVA with FEU NC | Different? |
| | No separate software package for data<br>downloading. Finapres wave form data<br>can be analyzed by the NOVAScope<br>software on the Finapres NOVA. | No separate software package for data<br>downloading. Finapres wave form data<br>can be analyzed by the NOVAScope<br>software on the Finapres NOVA. | No separate software package for data<br>downloading. Finapres wave form data<br>can be analyzed by the NOVAScope<br>software on the Finapres NOVA. | no |
| Required<br>accessories |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.