← Product Code [KWQ](/productcode/KWQ) · K173885

# Genesys Spine Binary Lumbar Plate System (K173885)

_Genesys Spine · KWQ · Apr 3, 2018 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K173885

## Device Facts

- **Applicant:** Genesys Spine
- **Product Code:** [KWQ](/productcode/KWQ.md)
- **Decision Date:** Apr 3, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3060
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Genesys Spine Binary Lumbar Plate System is indicated as additional support during fusion via a lateral or anterolateral surgical approach above the great vessels in the treatment of thoracic and thoracolumbar (T1 -L5) spine instability or via the anterior surgical approach, below the bifurcation of the greatment of lumbar and lumbosacral (L1-S1) spine instability as a result of fracture (including dislocation and subluxation), tumor, degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), scoliosis, kyphosis, lordosis, spinal stenos is, or a failed previous spine surgery.

## Device Story

The Genesys Spine Binary Lumbar Plate System is a spinal fixation system consisting of titanium alloy plates and screws. It provides supplemental stabilization during spinal fusion procedures. The device is used by surgeons in an operating room setting to address spinal instability caused by various pathologies, including fractures, tumors, and degenerative conditions. By securing the plates and screws to the vertebral bodies, the system provides mechanical support to the spine, facilitating fusion and potentially reducing pain and improving patient outcomes. The system is available in various sizes to accommodate patient-specific anatomy.

## Clinical Evidence

Bench testing only. Performance was evaluated using static compression bending, static torsion, and dynamic compression bending tests in accordance with ASTM F1717 standards to demonstrate mechanical equivalence.

## Technological Characteristics

Constructed from medical grade Ti-6Al-4V ELI titanium alloy (ASTM F136). System consists of plates and screws in various sizes. Mechanical testing performed per ASTM F1717. No software or electronic components.

## Regulatory Identification

A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.

## Predicate Devices

- Stryker LITe Plate System ([K142699](/device/K142699.md))
- Spinal USA AccuFit Plate System ([K091044](/device/K091044.md))
- Synthes Anterior Tension Band System ([K022791](/device/K022791.md))
- Depuy Aegis Anterior Lumbar Plate ([K052546](/device/K052546.md))
- Synthes TSLP Thoracolumbar Plate ([K020244](/device/K020244.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.

April 3, 2018

Genesys Spine William W. Sowers VP Quality & Engineering 1250 South Capital of Texas Highway, Building 3 Suite 600 Austin, Texas 78746

## Re: K173885

Trade/Device Name: Genesys Spine Binary Lumbar Plate System Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal Intervertebral Body Fixation Orthosis Regulatory Class: Class II Product Code: KWO Dated: March 8, 2018 Received: March 9, 2018

### Dear Mr. Sowers:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);

{1}------------------------------------------------

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

## Indications for Use

510(k) Number (if known)

#### K173885

#### Device Name

Genesys Spine Binary Lumbar Plate System

#### Indications for Use (Describe)

The Genesys Spine Binary Lumbar Plate System is indicated as additional support during fusion via a lateral or anterolateral surgical approach above the great vessels in the treatment of thoracic and thoracolumbar (T1 -L5) spine instability or via the anterior surgical approach, below the bifurcation of the greatment of lumbar and lumbosacral (L1-S1) spine instability as a result of fracture (including dislocation and subluxation), tumor, degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), scoliosis, kyphosis, lordosis, spinal stenos is, or a failed previous spine surgery.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

{3}------------------------------------------------

## 4. 510(K) SUMMARY

| Submitter's Name:                            | Genesys Spine                                         |
|----------------------------------------------|-------------------------------------------------------|
| Submitter's Address:                         | 1250 Capital of Texas Highway South                   |
|                                              | Building Three, Suite 600                             |
|                                              | Austin, Texas 78746                                   |
| Submitter's Telephone:                       | 512-381-7080                                          |
| Submitter's Fax:                             | 800-817-4938                                          |
| Contact Name:                                | William W. Sowers                                     |
| Date Summary was<br>Prepared:                | March 13, 2018                                        |
|                                              |                                                       |
| Trade or Proprietary Name:                   | Genesys Spine Binary Lumbar Plate System              |
| Common or Usual Name:                        | Spinal Fixation System                                |
| Classification:                              | Class II per:                                         |
|                                              | 21 CFR 888.3060 – Spinal intervertebral body fixation |
|                                              | orthosis                                              |
| Product Codes:                               | KWQ                                                   |
| Classification Panel:                        | Orthopedic Devices Panel                              |
| Legally Marketed<br>(unmodified) devices to  | Primary Predicate:                                    |
| Which Substantial<br>Equivalence is Claimed: | Stryker LITe Plate System (K142699)                   |
|                                              | Additional Predicate(s):                              |
|                                              | Spinal USA AccuFit Plate System (K091044),            |
|                                              | Synthes Anterior Tension Band System (K022791),       |
|                                              | Depuy Aegis Anterior Lumbar Plate (K052546),          |
|                                              | Synthes TSLP Thoracolumbar Plate (K020244)            |

## DESCRIPTION OF THE DEVICE SUBJECT TO PREMARKET NOTIFICATION:

The Genesys Spine Binary Lumbar Plate System consists of an assortment of plates and screws. The Binary Lumbar Plate System also contains Class 1 manual surgical instruments and cases that are considered exempt from premarket notification.

Implant components are available in a variety of sizes to suit the individual pathology and anatomical conditions of the patient. The Genesys Spine Binary Lumbar Plate System implants are manufactured from medical grade Ti-6Al-4V ELI titanium alloy per ASTM F136.

{4}------------------------------------------------

#### K173885 Page 2 of 2

## INDICATIONS FOR USE

The Genesys Spine Binary Lumbar Plate System is indicated as additional support during fusion via a lateral or anterolateral surgical approach above the bifurcation of the great vessels in the treatment of thoracic and thoracolumbar (T1 -L5) spine instability or via the anterior surgical approach, below the bifurcation of the great vessels in the treatment of lumbar and lumbosacral (L1-S1) spine instability as a result of fracture (including dislocation and subluxation), tumor, degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), scoliosis, kyphosis, lordosis, spinal stenosis, or a failed previous spine surgery.

## TECHNICAL CHARACTERISTICS

As established in this submission, the Genesys Spine Binary Lumbar Plate System was shown to be substantially equivalent and have equivalent technological characteristics to its predicate devices through comparison in areas including intended use, material composition, principles of operation and design.

## PERFORMANCE DATA

Nonclinical testing was performed to demonstrate that the subject Genesys Spine Binary Lumbar Plate System is substantially equivalent to other predicate devices. The following testing was performed:

- Static Compression Bending per ASTM F1717 ●
- Static Torsion per ASTM F1717
- . Dynamic Compression Bending per ASTM F1717

## CONCLUSION

The overall technology characteristics and mechanical performance data lead to the conclusion that Genesys Spine Binary Lumbar Plate System is substantially equivalent to legally marketed predicate devices for intended use, material composition, principles of operation, and design.

---

**Source:** [https://fda-staging.innolitics.com/device/K173885](https://fda-staging.innolitics.com/device/K173885)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
