s5 Intravascular Ultrasound Imaging and Pressure Systems, s5i Intravascular Ultrasound Imaging and Pressure Systems, CORE Precision Guided Therapy System, CORE Mobile Precision Guided Therapy System
Applicant
Volcano Corporation
Product Code
IYO · Radiology
Decision Date
Apr 11, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1560
Device Class
Class 2
Attributes
Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K173860 · Apr 11, 2018
s5 Intravascular Ultrasound Imaging and Pressure Systems, s5i Intravascular Ultrasound Imaging and Pressure Systems, CORE Precision Guided Therapy System, CORE Mobile Precision Guided Therapy System
Volcano Corporation
Swedish Coronary Angiography and Angioplasty Registry (SCAAR)
The iFR-SWEDEHEART study used registry data to compare iFR-guided revascularization against FFR-guided revascularization, supporting the clinical adoption of a dichotomous 0.89 iFR cut-point.
Patients with coronary artery disease; Sample Size: 2017; Number of Sites: 31 (30 in Sweden, 1 in Iceland)
FFR-guided revascularization
Composite rate of all-cause mortality, non-fatal myocardial infarction, or unplanned revascularization
Indications for Use
The Volcano s5™/s5i/CORE™ Mobile Precision Guided Therapy System is used for the qualitative and quantitative evaluation of vascular morphology in the coronary arteries and vessels of the peripheral vasculature. It is also indicated as an adjunct to conventional angiographic procedures to provide an image of vessel lumen and wall structures. ChromaFlo is indicated for qualitative blood flow information from peripheral and coronary vasculature; flow information can be an adjunct to other methods of estimating blood flow and blood perfusion. VH IVUS is intended to be used in conjunction with imaging catheters during diagnostic ultrasound imaging of the peripheral and coronary vasculature. The Volcano VH IVUS System is intended to semi-automatically visualize boundary features and perform spectral analysis of RF ultrasound signals of vascular features that the user may wish to examine more closely during routine diagnostic ultrasound imaging examinations. The pressure feature is intended for use in all blood vessels, including coronary and peripheral arteries, to measure intravascular blood pressure during diagnostic angiography and/or interventional procedures. Rotational 45MHz feature is intended for the qualitative and quantitative evaluation of vascular morphology in the coronary arteries and vasculature as an adjunct to conventional angiographic procedures to provide an image of the vessel lumen and the wall structures. The Pullback feature of the PIMr withdraws the imaging core within the protective sheath for a maximum of 15 cm. The FFR v2.5 Modality of the s5/s5i/CORE and CORE Mobile Precision Guided Therapy System is indicated in all blood vessels, including coronary and peripheral arteries, to measure intravascular blood pressure during diagnostic angiography and/or interventional procedures. The iFR Modality is intended to be used in conjunction with currently marketed Volcano pressure wires. In the coronary anatomy, the iFR modality has a diagnostic cut-point of 0.89 which represents an ischemic threshold and can reliably guide revascularization decisions during diagnostic catheterization procedure. When used for a pullback assessment, the iFR modality is intended as a visual aid in decision making by indicating the relative location and severity of the stenoses such as, multiple lesions or diffuse disease.
Device Story
System provides intravascular imaging and pressure measurement for coronary and peripheral vasculature. Inputs include ultrasound signals from imaging catheters and pressure data from compatible pressure wires. System processes signals to provide qualitative/quantitative morphology, blood flow (ChromaFlo), and pressure indices (FFR, iFR). Used in cath labs by physicians to guide revascularization. iFR modality uses 0.89 cut-point to identify ischemia, replacing hybrid strategies. Output displayed on monitors to assist in lesion assessment and procedural decision-making. Benefits include reduced need for adenosine-mediated hyperemia and associated patient discomfort.
Clinical Evidence
Supported by ADVISE II, DEFINE-FLAIR, and iFR-SWEDEHEART studies. ADVISE II (n=not specified) showed iFR 0.89 cut-point matches FFR 0.80 (sensitivity 73.0%, specificity 87.8%, AUC 0.90). DEFINE-FLAIR (n=2492) and iFR-SWEDEHEART (n=2017) were prospective randomized trials comparing iFR vs FFR. Primary endpoint (mortality, MI, unplanned revascularization at 12 months) met non-inferiority. iFR showed significant reduction in adenosine-related adverse symptoms (3.0% vs 68.3% in iFR-SWEDEHEART; 3.1% vs 30.8% in DEFINE-FLAIR, p<0.001).
Technological Characteristics
Mobile imaging/pressure system. Components: PIM, custom PCI boards, bedside/control room monitors, pressure wires (SmartWire II, PrimeWire, Verrata). Connectivity: Ethernet (DICOM), USB. Software: FFR v2.5. Electrical safety: Defibrillation-proof isolation via optical isolators and 5KV transformers. Operating range: 10-35°C, 10-95% RH.
Indications for Use
Indicated for patients undergoing diagnostic angiography or interventional procedures in coronary and peripheral vasculature. Used for evaluation of vascular morphology, blood flow, and intravascular pressure measurement. iFR modality indicated for coronary revascularization decision-making using a 0.89 ischemic threshold cut-point.
Regulatory Classification
Identification
An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
Volcano s5/s5i/CORE/CORE Mobile Precision Guided Therapy System (K133323)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
April 11, 2018
Volcano Corporation Christopher McLellan Manager. Regulatory Affairs 9965 Federal Drive Colorado Springs, Colorado 80921
Re: K173860
Trade/Device Name: s5/s5i/CORE/CORE Mobile Precision Guided Therapy System Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic Pulsed Echo Imaging System Regulatory Class: Class II Product Code: IYO Dated: March 9, 2018 Received: March 12, 2018
Dear Christopher McLellan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely, M & Thillel for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K173860
### Device Name
Volcano s5/s5i/CORE/CORE Mobile Precision Guided Therapy System
### Indications for Use (Describe)
The Volcano System is used for the qualitative and quantitative evaluation of vascular morphology in the coronary arteries and vessels of the peripheral vasculature. It is also indicated as an adjunct to conventional angiographic procedures to provide an image of vessel lumen and wall structures.
ChromaFlo is indicated for qualitative blood flow information from peripheral and coronary vasculature; flow information can be an adjunct to other methods of estimating blood flow and blood perfusion.
VH IVUS is intended to be used in conjunction with imaging catheters during diagnostic ultrasound imaging of the peripheral and coronary vasculature. The Volcano VH IVUS System is intended to semi-automatically visualize boundary features and perform spectral analysis of RF ultrasound signals of vascular features that the user may wish to examine more closely during routine diagnostic ultrasound imaging examinations.
The pressure feature is intended for use in all blood vessels, including coronary and peripheral arteries, to measure intravascular blood pressure during diagnostic angiography and/or interventional procedures.
Rotational 45MHz feature is intended for the qualitative evaluation of vascular morphology in the coronary arteries and vasculature as an adjunct to conventional angiographic procedures to provide an image of the vessel lumen and the wall structures. The Pullback feature of the imaging core within the protective sheath for a maximum of 15 cm.
The FFR v2.5 Modality of the s5/s5i/CORE and CORE Mobile Precision Guided Therapy System is indicated in all blood vessels, including coronary and peripheral arteries, to measure intravascular blood pressure angiography and/or interventional procedures.
The iFR Modality is intended to be used in conjunction with currently marketed Volcano pressure wires. In the coronary anatomy, the iFR modality has a diagnostic cut-point of 0.89 which represents an ischemic threshold and can reliably guide revascularization decisions during diagnostic catheterization procedure. When used as for a pullback assessment, the iFR modality is intended as a visual aid in decision making the relative location and severity of the stenoses such as, multiple lesions or diffuse disease.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
### Submitter l.
Volcano Corporation
3721 Valley Center Drive San Diego, CA 92130
Contact Person: Kimberly Simon
Telephone: (858) 720-4113 E-mail: kimberly.simon@philips.com
Date Prepared: December 19, 2017
### II. Device
| Name of Device: | Volcano s5 ™ /s5i/CORE/CORE™ Mobile Precision Guided<br>Therapy System |
|-----------------------|------------------------------------------------------------------------|
| Common or Usual Name: | Ultrasonic pulsed echo imaging system. |
| Classification Name: | System, Imaging, Pulsed Echo, Ultrasonic (21 CFR 892.1560) |
| Regulatory Class: | II |
| Product Code: | IYO |
#### III. Predicate Device
Volcano s5 ™/s5i/CORE/CORE™ Mobile Precision Guided Therapy System (K133323).
## Secondary Predicate Device
FFR v2.5 (K170133)
Reference Submissions for the Volcano s5 ™/s5i/CORE/CORE™ Mobile Precision Guided Therapy System:
- K133641
- . K140291
- K150441
- K153369
{5}------------------------------------------------
#### IV. Device Description
The Volcano s5 TM/s5i/CORETM Mobile Precision Guided Therapy System is a mobile imaging and pressure management system as previously described in K133323. The subject device incorporates the Volcano iFR® Modality cleared in K133323.
Pressure measurement is captured through the use of currently marketed pressure wires compatible with the currently marketed s5/s5i/CORE/CORE Mobile imaging and pressure measurement system.
### V. Indications for Use
The Volcano s5™/s5i/CORE™ Mobile Precision Guided Therapy System is used for the qualitative and quantitative evaluation of vascular morphology in the coronary arteries and vessels of the peripheral vasculature. It is also indicated as an adjunct to conventional angiographic procedures to provide an image of vessel lumen and wall structures.
ChromaFlo is indicated for qualitative blood flow information from peripheral and coronary vasculature; flow information can be an adjunct to other methods of estimating blood flow and blood perfusion.
VH IVUS is intended to be used in conjunction with imaging catheters during diagnostic ultrasound imaging of the peripheral and coronary vasculature. The Volcano VH IVUS System is intended to semi-automatically visualize boundary features and perform spectral analysis of RF ultrasound signals of vascular features that the user may wish to examine more closely during routine diagnostic ultrasound imaging examinations.
The pressure feature is intended for use in all blood vessels, including coronary and peripheral arteries, to measure intravascular blood pressure during diagnostic angiography and/or interventional procedures.
Rotational 45MHz feature is intended for the qualitative and quantitative evaluation of vascular morphology in the coronary arteries and vasculature as an adjunct to conventional angiographic procedures to provide an image of the vessel lumen and the wall structures. The Pullback feature of the PIMr withdraws the imaging core within the protective sheath for a maximum of 15 cm.
The FFR v2.5 Modality of the s5/s5i/CORE and CORE Mobile Precision Guided Therapy System is indicated in all blood vessels, including coronary and peripheral arteries, to measure intravascular blood pressure during diagnostic angiography and/or interventional procedures.
The iFR Modality is intended to be used in conjunction with currently marketed Volcano pressure wires. In the coronary anatomy, the iFR modality has a diagnostic cut-point of 0.89 which represents an ischemic threshold and can reliably guide revascularization decisions during diagnostic
{6}------------------------------------------------
catheterization procedure. When used for a pullback assessment, the iFR modality is intended as a visual aid in decision making by indicating the relative location and severity of the stenoses such as, multiple lesions or diffuse disease.
#### VI. Comparison with the Predicate Device
The predicate device is the currently marketed Volcano s5/s5i/CORE/CORE Mobile Precision Guided Therapy System iFR Modality (cleared under K133323 with the same trade name). This submission is for a change in the indications for use of the currently marketed device. The indications for use for the iFR Modality of the subject device has been revised to reflect the currently recognized, and clinically supported physiological, dichotomous 0.89 intravascular pressure index from a hybrid approach of 0.75 – 0.80 to provide information to help inform decisions on whether to perform or defer percutaneous coronary intervention (PCl). In the coronary anatomy, the iFR Modality has a diagnostic cut-point of 0.89 which represents an ischemic threshold and can reliably guide revascularization during diagnostic catheterization procedure. When used for a pullback assessment, the iFR modality is intended as a visual aid in decision making by indicating the relative location and severity of the stenoses such as, multiple lesions or diffuse disease.
The Volcano FFR and iFR Modality Operator's Manual is being revised to include a dichotomous cutpoint for Instantaneous Wave-free Ratio (iFR) guidance in coronary revascularization decision-making to replace the hybrid approach described in the Volcano FFR and iFR Modality Operator's Manual of the currently marketed device. There are no changes to the device materials or intended use.
The Volcano s5/s5i/CORE Mobile Intravascular Imaging and Pressure System and the predicate device share the same intended use. These devices are software measure blood pressure in the coronary and peripheral vasculature.
The Volcano s5™/s5i/CORE™ Mobile Precision Guided Therapy System and the predicate device share the same general operating principal. Both devices use the same pressure wires and are both compatible with only the Volcano imaging and pressure systems.
Volcano s5™/s5i/CORE/CORE™ Mobile Precision Guided Therapy System and the predicate device share the same technological features.
The only difference between the Volcano s5/s5i/CORE/CORE Mobile Intravascular Imaging and Pressure System with iFR "Modality and the predicate device is the indications for use. The indications for use for the iFR Modality of the subject device has been changed to measure intravascular blood pressure using the iFR Modality diagnostic cut-point of 0.89 which represents an ischemic threshold and can reliably guide revascularization decisions during diagnostic catheterization procedure. When used for a pullback assessment, the iFR modality is intended as a visual aid in decision making by indicating the relative location and severity of the stenoses such as, multiple lesions or diffuse disease.
The similarities between the subject Volcano s5/s5i/CORE/CORE Mobile Intravascular Imaging and Pressure System and the predicate device are illustrated in the table below.
{7}------------------------------------------------
| Attribute/Feature | Volcano s5/s5i/CORE/CORE Mobile Precision<br>Guided Therapy System iFR Modality – K133323<br>(Predicate Device) | Volcano s5/s5i/CORE/CORE<br>Mobile Precision Guided<br>Therapy System<br>(Subject Device) |
|------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | The iFR Modality of the s5™/s5i/CORE/CORE™<br>Mobile Series Precision Guided Therapy System is<br>indicated in all blood vessels, including coronary<br>and peripheral arteries, to measure intravascular<br>blood pressure during diagnostic angiography<br>and/or interventional procedures. The iFR Modality<br>is intended to be used in conjunction with<br>currently marketed Volcano pressure wires. | The Volcano<br>s5™/s5i/CORE/CORE™ Mobile<br>Precision Guided Therapy<br>System is used for the<br>qualitative and quantitative<br>evaluation of vascular<br>morphology in the coronary<br>arteries and vessels of the<br>peripheral vasculature. It is<br>also indicated as an adjunct to<br>conventional angiographic<br>procedures to provide an<br>image of vessel lumen and wall<br>structures.<br>ChromaFlo is indicated for<br>qualitative blood flow<br>information from peripheral<br>and coronary vasculature; flow<br>information can be an adjunct<br>to other methods of estimating<br>blood flow and blood<br>perfusion.<br>VH IVUS is intended to be used<br>in conjunction with imaging<br>catheters during diagnostic<br>ultrasound imaging of the<br>peripheral and coronary<br>vasculature. The Volcano VH<br>IVUS System is intended to<br>semi-automatically visualize<br>boundary features and perform<br>spectral analysis of RF<br>ultrasound signals of vascular<br>features that the user may wish |
| Attribute/Feature | Volcano s5/s5i/CORE/CORE Mobile Precision<br>Guided Therapy System iFR Modality – K133323<br>(Predicate Device) | Volcano s5/s5i/CORE/CORE<br>Mobile Precision Guided<br>Therapy System<br>(Subject Device) |
| | | routine diagnostic ultrasound<br>imaging examinations.<br>The pressure feature is<br>intended for use in all blood<br>vessels, including coronary and<br>peripheral arteries, to measure<br>intravascular blood pressure<br>during diagnostic angiography<br>and/or interventional<br>procedures.<br>Rotational 45MHz feature is<br>intended for the qualitative<br>and quantitative evaluation of<br>vascular morphology in the<br>coronary arteries and<br>vasculature as an adjunct to<br>conventional angiographic<br>procedures to provide an<br>image of the vessel lumen and<br>the wall structures. The<br>Pullback feature of the PIMr<br>withdraws the imaging core<br>within the protective sheath for<br>a maximum of 15 cm.<br>The FFR v2.5 Modality of the<br>s5/s5i/CORE and CORE Mobile<br>Precision Guided Therapy<br>System is indicated in all blood<br>vessels, including coronary and<br>peripheral arteries, to measure<br>intravascular blood pressure<br>during diagnostic angiography<br>and/or interventional<br>procedures. |
| Attribute/Feature | Volcano s5/s5i/CORE/CORE Mobile Precision<br>Guided Therapy System iFR Modality – K133323<br>(Predicate Device) | Volcano s5/s5i/CORE/CORE<br>Mobile Precision Guided<br>Therapy System<br>(Subject Device) |
| | | The iFR Modality is intended to<br>be used in conjunction with<br>currently marketed Volcano<br>pressure wires. In the coronary<br>anatomy, the iFR modality has<br>a diagnostic cut-point of 0.89,<br>which represents an ischemic<br>threshold and can reliably<br>guide revascularization<br>decisions during diagnostic<br>catheterization procedure.<br>When used for a pullback<br>assessment, the iFR modality is<br>intended as a visual aid in<br>decision making by indicating<br>the relative location and<br>severity of the stenoses such<br>as, multiple lesions or diffuse<br>disease. |
| Pressure Wires | SmartWire II, PrimeWire, PrimeWire Prestige,<br>PrimeWire Prestige Plus, Verrata, Verrata PLUS | SAME |
| Patient Isolation | Contained in the Patient Interface Module (PIM):<br>Electronic and Electrical Defibrillation proof<br>Isolation through 8mm bare fiberglass PCA<br>creepage with optical isolators and specially<br>wound transformers with 5KV insulation. | SAME |
| Environmental<br>Operating<br>Temperature,<br>Humidity and<br>Pressure Range | Operating Temperature: +10°C to +35°C Operating / Storage Humidity: 10% ≤ RH ≤ 95% Non-condensing Operating Pressure Range: 70-106kPa (526.3 – 797 mmHG) Storage Pressure Range: 50-106kPa (376 – 797 mmHg) | SAME |
| Display | Control room | SAME |
| Attribute/Feature | Volcano s5/s5i/CORE/CORE Mobile Precision<br>Guided Therapy System iFR Modality – K133323<br>(Predicate Device) | Volcano s5/s5i/CORE/CORE<br>Mobile Precision Guided<br>Therapy System<br>(Subject Device) |
| | 19" Non-medical grade powered through an<br>isolation transformer<br>Exam room | |
| | Use an existing Monitor in "Monitor Bank" Add a 19: non-medical grade to the x-ray boom (requires isolated power) 15" or 17" DC Powered Medical Grade Monitor mounted on the patient exam table 19" Monitor mounted on an additional boom (requires isolated power) | |
| User Controls | Control Console<br>Custom control panel that can be placed in control room or mounted bedside Touch Pad Custom touch pad that is mounted bedside Joy Stick Custom joystick that can be mounted bedside | SAME |
| Signal Processing | Base Band (identical custom PCI boards used) | SAME |
| FFR | Through identical custom PCI board | SAME |
| iFR® | Through identical custom PCI board (same custom PCI board as used for FFR) | SAME |
| Inputs | 100-240 VAC 50/60Hz Power Catheter/PIM connection ECG. Aortic high-level (from hemodynamic system) Ethernet (for DICOM Worklist) USB (for control devices) PS2 (for control devices) Touchpad | SAME |
| Attribute/Feature | Volcano s5/s5i/CORE/CORE Mobile Precision<br>Guided Therapy System iFR Modality – K133323<br>(Predicate Device) | Volcano s5/s5i/CORE/CORE<br>Mobile Precision Guided<br>Therapy System<br>(Subject Device) |
| | - Joystick | |
| Outputs | - Monitor display: 1280 x 1024 resolution<br>- 19" Monitor<br>- 19" Boom Monitor<br>- 17" Bedside Monitor<br>- Printer | SAME |
| Patient Interface<br>Module (PIM) | Compatible with Digital IVUS, Rotational IVUS, FFR<br>Pimmette | SAME |
| Connection to PIMs | Via direct connection to CPU or Connection Box | SAME |
| Accessories and<br>Peripherals | Bedside Touchpad Controller, Joystick Controller,<br>Connection Box, Control Console | SAME |
| Imaging<br>Specifications | Imaging boards and software | SAME |
| FFR Software<br>(includes iFR<br>Modality) | FFR Software Version v2.3 | FFR Software Version v2.5<br>(cleared in K170133) |
{8}------------------------------------------------
{9}------------------------------------------------
{10}------------------------------------------------
{11}------------------------------------------------
### VII. Performance Data
The subject device is the same as the currently marketed predicate device. There are no changes to the pressure measurement wires for collection of data using the iFR" Modality. Performance data establishing the electrical safety and electromagnetic compatibility (EMC) of the device was previously reviewed in K133641 and K140291.
### VIII. Clinical Data
Clinical evidence suggests the adoption of a dichotomous cut-point for Instantaneous Wave-free Ratio (iFR) guidance in coronary revascularization decision-making rather than the use of a hybrid approach as was previously documented in the Volcano FFR and iFR Modality Operator's Manual.
The clinical data submitted to support the proposed modifications is documented in the ADVISE II, DEFINE-FLAIR and iFR SWEDEHEART studies. The ADVISE II study demonstrated that Traditional 510(k) for Volcano s5/s5i/CORE/CORE Mobile Precision Guided Therapy System
{12}------------------------------------------------
an iFR cut-point of 0.89 matches best with an FFR ischemic cut-point of 0.80 with a specificity of 87.8% and sensitivity of 73.0% (C statistic: 0.90 (95% confidence interval (CI): 0.88 to 0.92, p < 0.001]).1
The use of this dichotomous cut-point for iFR-guided revascularization was compared to FFR-guided revascularization in two large prospective randomized controlled trials. 3 DEFINE-FLAIR (Functional Lesion Assessment of Intermediate Stenosis to Guide Revascularization) enrolled 2492 patients with coronary artery disease in a 1:1 ratio to undergo revascularization using either FFR or iFR-guidance. The second study, iFR-SWEDEHEART (Instantaneous Wave-free Ratio versus Fractional Flow Reserve in Patients with Stable Angina Pectoris or Acute Coronary Syndrome) was similar; it enrolled 2017 patients with coronary artery disease included in the Swedish Coronary Angiography and Angioplasty Registry (SCAAR). The registry contains data on patients from all 30 coronary intervention centers in Sweden plus a single site in Iceland.
Both studies had a primary end-point that was the composite rate of all-cause mortality, non-fatal myocardial infarction or unplanned revascularization within 12 months after the index procedure. The non-inferiority margin was 3.4 in DEFINE-FLAIR, and 3.2 in iFR-SWEDEHEART. Both studies met their respective primary endpoint by demonstrating non-inferiority at 12 months:
| DEFINE-FLAIR: | 6.8% (iFR) vs 7.0% (FFR); D = -0.2%; 95% CI, -2.3 to 1.8; P<0.001 |
|-----------------|-------------------------------------------------------------------|
| iFR-SWEDEHEART: | 6.7% (iFR) vs 6.1% (FFR); D = 0.7; 95% CI, -1.5-2.8, P=0.007 |
Both studies also demonstrated a significant reduction in adverse effects associated with the adenosine-mediated hyperemia required for FFR-guidance. 68.3% of patients in the FFR arm of iFR-SWEDEHEART reported chest discomfort, while only 3.0% of patients in the iFR arm reported any chest discomfort (P<0.001). Similarly, DEFINE-FLAIR patients in the FFR group reported adverse procedural signs or symptoms in 30.8% of cases compared to 3.1% in the iFR group (P<0.001).
The iFR guided procedures demonstrated a significant reduction in acute adverse signs or symptoms associated with the adenosine-mediated hyperemia required for FFR. The use of iFR in stable ischemic heart disease is now considered appropriate by the major American cardiology societies. A hybrid iFR-FFR strategy is thus outdated.
<sup>4</sup> Escaned J, Echavarria-Pinto M, Garcia HM, et. al. Prospective Assessment of the Diagnostic Accuracy of Instantaneous Wave-Free Ratio to Assess Coronary Stenosis Results of ADVISE II International, Multicenter Study (ADenosine Vasodilator Independent Stenosis Evaluation II). JACC Cardiovasc Interv. 2015 May;8(6):824-33.
<sup>2</sup> Davies JE, Sen S, Dehbi H-M, et. al. Use of the Instantaneous Wave-free Ratio or Fractional Flow Reserve in PCI. N Engl J Med. 2017 May 11;376(19):1824-1834.
<sup>3</sup> Götberg M, Christiansen EH, Gudmundsdottir IJ, et al. Instantaneous Wave-free Ratio versus Fractional Flow Reserve to Guide PCI. N Engl J Med. 2017 May 11;376(19):1813-1823.
{13}------------------------------------------------
### IX. Conclusions
Based upon the information submitted in this premarket notification 510(k), the Volcano s5™/s5i/CORE/CORE™ Mobile Precision Guided Therapy System is substantially equivalent to the currently marketed predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.