← Product Code [DPS](/productcode/DPS) · K173830

# MAC VU360 Resting ECG Analysis System (K173830)

_Ge Medical Systems Information Technologies, Inc. · DPS · Sep 18, 2018 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K173830

## Device Facts

- **Applicant:** Ge Medical Systems Information Technologies, Inc.
- **Product Code:** [DPS](/productcode/DPS.md)
- **Decision Date:** Sep 18, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.2340
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Pediatric

## Indications for Use

The MAC VU360 Resting ECG Analysis System is a non-invasive prescription device. -The device is indicated for use to acquire, analyze, display and print electrocardiograms. -The device is indicated for use to provide interpretation of the data for consideration by a physician. -The device is indicated for use in a clinical setting, by a physician or by trained personnel who are acting on the orders of a licensed physician. It is not intended as a sole means of diagnosis. The interpretations of ECG offered by the device are only significant when used in conjunction with a physician overread as well as consideration of all other relevant patient data. -The device is indicated for use on adult and pediatric (birth through 21 years of age) populations.

## Device Story

Mobile resting electrocardiograph; acquires 3, 6, 12, or 15-lead ECG signals via surface electrodes; processes data using 12SL algorithm to provide interpretive analysis; displays and prints reports. Used in hospitals, clinics, and offices by physicians or trained personnel. Connects to wired LAN or WiFi to interface with cardiology information systems (e.g., GEHC MUSE) for workflow integration; supports PDF reports; displays external web/Citrix content. Output assists physicians in clinical decision-making; requires physician overread. Benefits include standardized ECG analysis and efficient data management.

## Clinical Evidence

No clinical data; bench testing only. Substantial equivalence supported by non-clinical verification and validation, including risk analysis, design reviews, software lifecycle management, module verification, system integration, and safety testing.

## Technological Characteristics

Mobile wheeled trolley form factor; 3, 6, 12, 15-lead ECG acquisition; wired LAN and WiFi connectivity. Employs 12SL algorithm for analysis. No sterilization required. Complies with voluntary standards for electrocardiographs.

## Regulatory Identification

An electrocardiograph is a device used to process the electrical signal transmitted through two or more electrocardiograph electrodes and to produce a visual display of the electrical signal produced by the heart.

## Predicate Devices

- ELI 380 Resting Electrocardiograph ([K142105](/device/K142105.md))

## Reference Devices

- MAC 5500 HD Resting ECG Analysis System ([K073625](/device/K073625.md))
- 12SL (v23) ECG Analysis Program ([K141963](/device/K141963.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.

September 18, 2018

GE Medical Systems Information Technologies, Inc. Marcella Entwisle Regulatory Affairs Leader 9900 West Innovation Drive Wauwatosa. Wisconsin 53226

Re: K173830

Trade/Device Name: MAC VU360 Resting ECG Analysis System Regulation Number: 21 CFR 870.2340 Regulation Name: Electrocardiograph Regulatory Class: Class II Product Code: DPS, DQK, DXH Dated: August 17, 2018 Received: August 20, 2018

Dear Marcella Entwisle:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jessica E. Paulsen -S

for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K173830

Device Name

MAC VU360 Resting ECG Analysis System

#### Indications for Use (Describe)

The MAC VU360 Resting ECG Analysis System is a non-invasive prescription device.

-The device is indicated for use to acquire, analyze, display and print electrocardiograms.

-The device is indicated for use to provide interpretation of the data for consideration by a physician.

-The device is indicated for use in a clinical setting, by a physician or by trained personnel who are acting on the orders of a licensed physician. It is not intended as a sole means of diagnosis.

-The interpretations of ECG offered by the device are only significant when used in conjunction with a physician overread as well as consideration of all other relevant patient data.

-The device is indicated for use on adult and pediatric (birth through 21 years of age) populations.

Type of Use (Select one or both, as applicable)

| <span style="font-size: 10pt;"> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)    </span> |
|---------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-size: 10pt;"> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)    </span>                 |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized font in the center. The circle is surrounded by a white border with a swirling pattern. The logo is simple and recognizable, and it is associated with a well-known and established company.

# 510(k) Summary

## Section 5: 510(k) Summary

In accordance with 21 CFR Part 807.92(a)(1) the following summary of information is provided:

| Date:                  | September 17, 2018                                                                                                                                                       |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter:             | GE Medical Systems Information Technologies, Inc.<br>9900 Innovation Drive<br>Wauwatosa, WI 53226                                                                        |
| Primary Contact:       | Marcella Entwisle<br>Senior Regulatory Affairs Leader<br>Phone: (414) 721-3809<br>Fax: (414) 721-3863<br>email: marcella.entwisle@ge.com                                 |
| Secondary Contact:     | Joel Kent<br>Senior Regulatory Affairs Manager<br>GE Medical Systems Information Technologies, Inc.<br>Phone: (617) 851-0943<br>email: joel.kent@ge.com                  |
| PRODUCT IDENTIFICATION |                                                                                                                                                                          |
| Device Trade Name:     | MAC VU360 Resting ECG Analysis System                                                                                                                                    |
| Common / Usual Name:   | Electrocardiograph                                                                                                                                                       |
| Classification Names   | 21 CFR 870.2340 – Electrocardiograph<br>21 CFR 870.1425 – Programmable Diagnostic Computer<br>21 CFR 870.2920 – Telephone electrocardiograph Transmitter and<br>Receiver |
| Product Codes:         | DPS, DQK, and DXH                                                                                                                                                        |
| Predicate Device(s):   | ELITM 380 Resting Electrocardiograph (K142105)                                                                                                                           |
| Reference Device(s):   | MAC™ 5500 HD Resting ECG Analysis System (K073625)<br>12SLTM (v23) ECG Analysis Program (K141963)                                                                        |

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the General Electric (GE) logo, which is a stylized monogram of the letters 'G' and 'E' in a circular shape. The logo is colored in blue, and the letters are interconnected in a flowing, cursive style. The circular border around the letters has small, stylized water droplet shapes evenly spaced around it.

Device Description: The MAC VU360 Resting ECG Analysis System is a mobile electrocardiograph designed to acquire, analyze, display, and record ECG signals from surface ECG electrodes. The MAC VU360 can capture 3, 6, 12, or 15 lead electrocardiograms, provide interpretive analysis, and print reports. The MAC VU360 is indicated for use on adult and pediatric (birth through 21 years of age) populations. The device can connect to a network, either through a wired LAN connection or via wireless WiFi access points. Once on the network, the device can optionally interface with the cardiology information systems such as the GEHC MUSE® system to participate in a complete electrocardiology workflow. The device can also optionally display data from external systems which provide web or Citrix content. The device provides state-of-the-art information technology security features and a contemporary user interface. Mobility is provided via a wheeled trolley. The MAC VU360 is intended to be a high-performance, Intended Use / multichannel resting electrocardiograph. As a resting electrocardiograph, the MAC VU360 simultaneously acquires data from each lead. Once the data is acquired, it can be analyzed, reviewed, stored, printed or transmitted. It is a device primarily intended for use in hospitals, but may be used in medical clinics and offices of any size. The MAC VU360 is intended to be used under the direct supervision of a licensed healthcare practitioner, by trained operators in a hospital or medical professional's facility. Indications for Use / The MAC VU360 Resting ECG Analysis System is a non-invasive prescription device. -The device is indicated for use to acquire, analyze, display and print electrocardiograms. -The device is indicated for use to provide interpretation of the data for consideration by a physician. -The device is indicated for use in a clinical setting, by a physician or by trained personnel who are acting on the orders of a licensed physician. It is not intended as a sole means of diagnosis. The interpretations of ECG offered by the device are only significant when used in conjunction with a physician overread as well as consideration of all other relevant patient data. -The device is indicated for use on adult and pediatric (birth through 21 years of age) populations.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" intertwined within a blue circle. There are three white water droplet shapes surrounding the circle. 

| Contraindications:     | The MAC VU360 Resting ECG Analysis System is not intended for<br>use in the following manner:<br>-During patient transport<br>-With high-frequency surgical units<br>-As an intra-cardiac application<br>-As a sole means of diagnosis<br>-As a vital signs physiological monitor                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Technology:            | The MAC VU360 employs the same fundamental scientific<br>technology, basic design, construction, materials, energy source,<br>control mechanism, operating principles as the predicate device ELI<br>380 and the MAC 550 HD in acquiring, analyzing, recording,<br>displaying and printing ECG data for both adult and pediatric<br>populations. The MAC VU360 utilizes the 12SL Algorithm as does<br>the MAC 5500 HD.<br><br>The basic system prints 6 or 12 leads of ECG and provides optional<br>transmission and reception of ECG data to and from a central ECG<br>cardiovascular information system.<br><br>The MAC VU360 is similar to the ELI 380 Resting ECG Analysis<br>System, K142105, in the technology of downloading orders and<br>patient demographics from a central ECG cardiovascular information<br>system (e.g. MUSE) as well as supporting ECG reports in PDF. Both<br>are able to use WiFi communication.<br><br>For more on the predicate device comparison refer to Section 12 of<br>this submission. |
| Performance Standards: | The MAC VU360 Resting ECG Analysis System complies with the<br>voluntary standards as detailed in Section 9 of this submission.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Sterilization:         | The MAC VU360 Resting ECG Analysis System does not require<br>sterilization.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a circle. The circle has small, wave-like elements around its perimeter, giving it a dynamic appearance. The logo is presented in a blue color scheme.

| Determination of<br>Substantial Equivalence: | Summary of Non-Clinical Tests:                                                                                                                                                                                                                                           |
|----------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                              | The MAC VU360 Resting ECG Analysis System and its applications<br>comply with voluntary standards. The following quality assurance<br>measures were applied to the development of the system:                                                                            |
|                                              | ● Risk Analysis<br>● Required Reviews<br>● Design Reviews<br>● Software Development Lifecycle<br>● Testing on unit level (Module verification)<br>● Integration testing (System verification)<br>● Performance testing (Verification)<br>● Safety testing (Verification) |
| Summary of Clinical<br>Tests:                | The subject of this premarket submission, MAC VU360 Resting<br>ECG Analysis System, did not require clinical studies to support<br>substantial equivalence.                                                                                                              |
| Conclusion:                                  | GE Healthcare considers the MAC VU360 Resting ECG Analysis<br>System to be as safe, and as effective, and performance is<br>substantially equivalent to the predicate device, ELI 380 Resting<br>Electrocardiograph.                                                     |

---

**Source:** [https://fda-staging.innolitics.com/device/K173830](https://fda-staging.innolitics.com/device/K173830)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
