← Product Code [DQX](/productcode/DQX) · K173795

# HI-TORQUE TurnTrac Guide Wire Family (K173795)

_Abbott Vascular · DQX · Jan 12, 2018 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K173795

## Device Facts

- **Applicant:** Abbott Vascular
- **Product Code:** [DQX](/productcode/DQX.md)
- **Decision Date:** Jan 12, 2018
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 870.1330
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The HI-TORQUE TurnTrac Guide Wire Family is intended to facilitate the delivery of catheter-based interventional devices during percutaneous transluminal angioplasty (PTA) and percutaneous transluminal coronary angioplasty (PTCA). This guide wire may also be used with compatible stent devices during therapeutic procedures. This device is designed and intended for ONE-TIME USE ONLY. Do not resterilize and /or reuse.

## Device Story

Steerable guide wire used to facilitate delivery of catheter-based interventional devices and stents during PTA/PTCA procedures. Operated by physicians in clinical settings (e.g., cath lab). Device features include varying support levels (Moderate/Extra), tip loads (0.6g/0.9g), hydrophilic coating options, and lengths (190cm/300cm). Includes shaping tool accessory for distal tip customization. Provides mechanical support and navigation for interventional catheters; aids in precise placement of therapeutic devices within vasculature. Single-use only.

## Clinical Evidence

Bench testing only. No clinical data. Testing included tip tensile strength, torsional wire strength, torqueability, integrity (particulate), and friction testing. Biocompatibility testing performed. Results met all acceptance criteria.

## Technological Characteristics

Steerable guide wire; 0.0142" or 0.0145" diameter; stainless steel/nitinol construction (implied by class); hydrophilic coating; single-use; non-powered; manual operation.

## Regulatory Identification

A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.

## Special Controls

*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.

## Predicate Devices

- HI-TORQUE VersaTurn Guide Wire Family ([K141782](/device/K141782.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logos of the U.S. Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. The logos are displayed side-by-side.

January 12, 2018

Abbott Vascular Aniket Khakhadiya Regulatory Affairs Specialist 3200 Lakeside Drive Santa Clara, California 95054

Re: K173795

Trade/Device Name: HI-TOROUE TurnTrac Guide Wire Family Regulation Number: 21 CFR 870.1330 Regulation Name: Catheter guide wire Regulatory Class: Class II Product Code: DQX Dated: December 12, 2017 Received: December 14, 2017

#### Dear Aniket Khakhadiya:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Kenneth J. Cavanaugh -S

for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K173795

Device Name HI-TORQUE® TurnTrac Guide Wire Family

#### Indications for Use (Describe)

The HI-TORQUE TurnTrac Guide Wire Family is intended to facilitate the delivery of catheter-based interventional devices during percutaneous transluminal angioplasty (PTA) and percutaneous transluminal coronary angioplasty (PTCA). This guide wire may also be used with compatible stent devices during therapeutic procedures.

This device is designed and intended for ONE-TIME USE ONLY. Do not resterilize and /or reuse.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size: 2em;">☑</span> |
|----------------------------------------------|----------------------------------------|
| Over-The-Counter Use (21 CFR 801 Subpart C)  | <span style="font-size: 2em;">☐</span> |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## K173795 - 510(K) SUMMARY

The 510(k) Summary is submitted in accordance with 21 CFR §807.92 and the requirements of the Safe Medical Device Act (SMDA) of 1990.

| 1.  | Submitter's Name           | Abbott Vascular                                                           |
|-----|----------------------------|---------------------------------------------------------------------------|
| 2.  | Submitter's Address        | 3200 Lakeside Dr. Santa Clara, CA 95054                                   |
| 3.  | Telephone                  | 408-845-8314                                                              |
| 4.  | Fax                        | 408-845-3689                                                              |
| 5.  | Contact Person             | Aniket Khakhadiya                                                         |
| 6.  | Date Prepared              | December 7, 2017                                                          |
| 7.  | Device Trade Name          | HI-TORQUE® TurnTrac Guide Wire Family                                     |
| 8.  | Device Common Name         | Guide Wire                                                                |
| 9.  | Device Classification Name | Catheter Guide Wire (21 CFR 870.1330; Product Code DQX)                   |
| 10. | Predicate Device Name      | HI-TORQUE VersaTurn Guide Wire Family (K141782 cleared on August 7, 2014) |

### 11. Device Description

The HI-TORQUE (HT) TurnTrac Guide Wire Family includes steerable guide wires offered in several configurations by various support levels, tip loads, hydrophilic coating lengths, tip shape and guide wire lengths.

The HT TurnTrac Guide Wire Family will be available with the features as listed below:

- 2 tip loads: 0.6g and 0.9g ●
- 2 support levels: HT TurnTrac Moderate Support (MS) and HT TurnTrac Extra . Support (ES)
- 2 coating lengths: Fully Coated (HC), Uncoated Tip (pHC) ●
- 1 tip shape: Straight ●
- 2 lengths: 190 cm, 300cm ●

The HI-TORQUE TurnTrac Guide Wire Family has a maximum diameter of 0.0142" or 0.0145" depending on the support level and is compatible with devices designed for use with 0.014" guide wires. The HI-TORQUE TurnTrac Guide Wire Family includes a shaping tool, which is clipped on to the outer packaging coil after the final assembled

{4}------------------------------------------------

product has been inserted into the packaging coil. This accessory is to aid the physician in shaping the distal portion of the guide wire, if desired.

### 12. Indication for Use

The HI-TORQUE TurnTrac Guide Wire Family is intended to facilitate the delivery of catheter-based interventional devices during percutaneous transluminal angioplasty (PTA) and percutaneous transluminal coronary angioplasty (PTCA). This guide wire may also be used with compatible stent devices during therapeutic procedures.

This device is designed and intended for ONE-TIME USE ONLY. Do not resterilize and / or reuse.

### 13. Technological Characteristics

Comparison of the subject device and predicate device demonstrates that the technological characteristics such as product performance, functional specification, operating principle, design and materials (with minor modifications), and indications for use are substantially equivalent to the currently marketed predicate device.

### 14. Performance Data

In vitro bench testing, including tip tensile strength, torsional wire strength, torqueability, integrity (particulate testing), and friction testing were conducted on the modified device. The results from the in vitro bench tests demonstrated that the HI-TORQUE TurnTrac Guide Wire met all acceptance criteria and performed similarly to the predicate device. The results obtained from the biocompatibility testing indicate that the HI-TORQUE TurnTrac Guide Wire Family may be considered substantially equivalent to the predicate device.

### 15. Conclusions

Test results from the in vitro bench testing conducted on the subject device demonstrate that the HI-TORQUE TurnTrac Guide Wire met all acceptance criteria and performed similarly to the predicate device. Therefore, the HI-TORQUE TurnTrac Guide Wire Family may be considered substantially equivalent to the predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K173795](https://fda-staging.innolitics.com/device/K173795)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
