K173760 · Auris Surgical Robotics, Inc. · EOQ · Mar 22, 2018 · Ear, Nose, Throat
Device Facts
Record ID
K173760
Device Name
Monarch Endoscopy Platform (Monarch Platform)
Applicant
Auris Surgical Robotics, Inc.
Product Code
EOQ · Ear, Nose, Throat
Decision Date
Mar 22, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.4680
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Monarch Endoscopy Platform (Monarch Platform) and its accessories are intended to provide bronchoscopic visualization of and access to patient airways for diagnostic and therapeutic procedures.
Device Story
Monarch Endoscopy Platform is a robotic-assisted system for bronchoscopy. Components: Endoscopy Cart (robotic arms, electronics), Endoscopy Tower (display, computer, controller), and Bronchoscope System (flexible bronchoscope with camera, working channel). Physician operates system via endoscopic controller; inputs transmitted to electromechanical end effectors. Bronchoscope tip articulates in all directions; working channel allows irrigation, aspiration, and delivery of instruments (Aspirating Biopsy Needle, Biopsy Forceps, Cytology Brush). System provides visualization and navigation to target sites in airways. Used in clinical settings by physicians to perform diagnostic/therapeutic procedures. Output is real-time endoscopic video and navigation data on tower monitor. Benefits include enhanced access and visualization for tissue/cell collection in lungs.
Clinical Evidence
No human clinical data. Evidence includes bench testing (electrical safety, EMC, biocompatibility per ISO 10993), software verification/validation, and simulated use testing. Animal and cadaver studies validated user needs and system safety/effectiveness. Comparative study of 'Navigation Yield' and 'Distance to Target' against the superDimension system demonstrated equivalent accuracy.
Technological Characteristics
Robotic-assisted flexible bronchoscope system. Components: Cart, Tower, Bronchoscope. Materials: Polymeric shafts, Nitinol needles, coil pipes. Sensing/Actuation: Electromechanical robotic arms, wire-actuated tip articulation. Connectivity: Networked system components. Sterilization: Ethylene Oxide for bronchoscope/instruments. Software: System control and navigation algorithms. Complies with IEC 60601-1, IEC 60601-2-18, and IEC 60601-1-2.
Indications for Use
Indicated for bronchoscopic visualization and access to patient airways for diagnostic and therapeutic procedures. No specific age, gender, or disease state contraindications provided.
Regulatory Classification
Identification
A bronchoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the bronchoscope and is intended to examine or treat the larynx and tracheobronchial tree. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel or flexible plastic. This generic type of device includes the rigid ventilating bronchoscope, rigid nonventilating bronchoscope, nonrigid bronchoscope, laryngeal-bronchial telescope, flexible foreign body claw, bronchoscope tubing, flexible biopsy forceps, rigid biopsy curette, flexible biopsy brush, rigid biopsy forceps, flexible biopsy curette, and rigid bronchoscope aspirating tube, but excludes the fiberoptic light source and carrier.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is a symbol representing the Department of Health & Human Services - USA, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The FDA logo is a symbol of the agency's mission to protect and promote public health through the regulation and supervision of food safety, tobacco, pharmaceuticals, medical devices, and other related products.
March 22, 2018
Auris Surgical Robotics, Inc. Joy Sacmar Vice President, RA/QA 150 Shoreline Road Redwood City, CA 94065
Re: K173760
Trade/Device Name: Monarch Endoscopy Platform (Monarch Platform) Regulation Number: 21 CFR 874.4680 Regulation Name: Bronchoscope (Flexible or Rigid) and Accessories Regulatory Class: Class II Product Code: EOO Dated: February 21, 2018 Received: February 22, 2018
Dear Joy Sacmar:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Eric A. Mann -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose, and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K173760
Device Name
Monarch Endoscopy Platform (Monarch Platform)
Indications for Use (Describe)
The Monarch Endoscopy Platform (Monarch Platform) and its accessories are intended to provide bronchoscopic visualization of and access to patient airways for diagnostic and therapeutic procedures.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------|----------------------------------------------------------------------------------|
| <div> <span> ☑ Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows a pattern of overlapping ovals that create a symmetrical, butterfly-like shape. The ovals are arranged in a way that they intersect at the center, forming a central void. The color of the ovals is a light shade of yellow, and the lines are thin and delicate. The overall effect is visually appealing and creates a sense of movement and depth.
AURIS
### 510(k) SUMMARY
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
K173760 510(k) Number: _
#### I. SUBMITTER
Auris Surgical Robotics, Inc. 150 Shoreline Drive Redwood City, CA 94065
| Contact Person: | Joy Sacmar, Vice President RA/QA |
|-----------------|----------------------------------|
| Email: | joy.sacmar@aurisrobotics.com |
| Phone/Fax: | (650) 781-1610 |
| Mobile: | (408) 306-3537 |
| Date Prepared: | December 7, 2017 |
|----------------|-------------------------------------------------------|
| Prepared By: | Elizabeth Osuna, Senior Regulatory Affairs Specialist |
#### II. DEVICE
| Trade/Device Name: | Monarch Endoscopy Platform (Monarch Platform) |
|----------------------|--------------------------------------------------|
| Common Name: | Bronchoscope (flexible or rigid) and accessories |
| Classification Name: | Bronchoscope (flexible or rigid) and accessories |
| Product Code: | EOQ |
| Regulatory Class: | Class II |
| Regulation Number: | 21 CFR 874.4680 |
| Trade/Device Name: | Aspirating Biopsy Needle |
| Common Name: | Bronchoscope (flexible or rigid) and accessories |
| Classification Name: | Bronchoscope (flexible or rigid) and accessories |
| Product Code: | EOQ |
| Regulatory Class: | Class II |
| Regulation Number: | 21 CFR 874.4680 |
{4}------------------------------------------------
| Trade/Device Name: | Biopsy Forceps |
|----------------------|--------------------------------------------------|
| Common Name: | Bronchoscope (flexible or rigid) and accessories |
| Classification Name: | Bronchoscope (flexible or rigid) and accessories |
| Product Code: | EOQ |
| Regulatory Class: | Class II |
| Regulation Number: | 21 CFR 874.4680 |
| Trade/Device Name: | Cytology Brush |
| Common Name: | Bronchoscope (flexible or rigid) and accessories |
| Classification Name: | Bronchoscope (flexible or rigid) and accessories |
| Product Code: | EOQ |
| Regulatory Class: | Class II |
| Regulation Number: | 21 CFR 874.4680 |
#### III. PREDICATE DEVICE(S)
| Trade name: | Karl Storz Bronchoscope |
|----------------------|---------------------------------------------------------------------|
| 510(k) Number: | K071530 |
| Classification Name: | Bronchoscope (flexible or rigid) and accessories<br>21 CFR 874.4680 |
| Product Code: | EOQ |
| Manufacturer: | Karl Storz Endoscopy-America, Inc. |
| Trade name: | PeriFLEX Transbronchial Aspiration Needle |
|----------------------|---------------------------------------------------------------------|
| 510(k) Number: | K162611 |
| Classification Name: | Bronchoscope (flexible or rigid) and accessories<br>21 CFR 874.4680 |
| Product Code: | EOQ |
| Manufacturer: | Spiration, Inc. |
| Trade name: | Olympus FB series biopsy forceps |
|----------------------|---------------------------------------------------------------------|
| 510(k) Number: | K962555 |
| Classification Name: | Bronchoscope (flexible or rigid) and accessories<br>21 CFR 874.4680 |
| Product Code: | EOQ |
| Manufacturer: | Olympus America, Inc. |
{5}------------------------------------------------
| Trade name: | Wang Bronchial Needle Brush |
|----------------------|---------------------------------------------------------------------|
| 510(k) Number: | K944650 |
| Classification Name: | Bronchoscope (flexible or rigid) and accessories<br>21 CFR 874.4680 |
| Product Code: | EOQ |
| Manufacturer: | Mill-Rose Laboratory |
In compliance with FDA Guidance "The 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications [510(k)]' we are including the following as a reference device:
| Trade name: | Auris Robotic Endoscopy System (ARES) |
|----------------|-----------------------------------------------------------------|
| 510(k) Number: | K152319 |
| Manufacturer: | Auris Surgical Robotics, Inc. |
| Trade name: | superDimensionTM Navigation System |
| 510(k) Number: | K151376 |
| Manufacturer: | Covidien LLC. |
| Trade name: | superDimensionTM, SuperTrax Transbronchial Aspirating<br>Needle |
| 510(k) Number: | K860582 |
| Manufacturer: | Covidien LLC. |
| Trade name: | superDimensionTM, SuperTrax Biopsy Forceps |
| 510(k) Number: | K925392 |
| Manufacturer: | Covidien LLC. |
| Trade name: | superDimensionTM Cytology Brush |
| 510(k) Number: | K834402 |
| Manufacturer: | Covidien LLC. |
#### DEVICE DESCRIPTION IV.
The Monarch™ Endoscopy Platform (Monarch Platform) is intended to be used by qualified physicians to provide bronchoscopic visualization of and access to patient airways for diagnostic and therapeutic procedures. The Monarch Platform consists of three major components, (1) Monarch™ Endoscopy Cart, (2) Monarch™ Endoscopy Tower, and (3) Monarch™ Bronchoscope System, and working channel instruments.
{6}------------------------------------------------
The Monarch Cart provides support for the effector arms. It includes two robotic arms and the electronic systems required to power and operate the robotic system.
The Monarch Tower is the primary user (i.e. physician) procedural display interface. It contains a monitor for user viewing and computers running the system software. The tower provides connectivity for the bronchoscope camera and lighting, as well as the fluidics system. In addition, the tower includes an endoscopic controller that allows the user to control the system during a procedure.
The user controls the system with an endoscopic controller which transmits user inputs through the electromechanical system to the end effectors. The Monarch Bronchoscope System comprising of a bronchoscope and sheath is attached at the end effector of a robotic arm with multiple degrees of freedom. The flexible bronchoscope has a working channel and a camera at the tip. The bronchoscope has an articulated tip that can bend in all directions. The working channel of the bronchoscope is used for irrigation, aspiration and to deliver the working channel instruments.
The single-use manually controlled working channel instruments compatible with the Monarch Platform include the Aspirating Biopsy Needle, Biopsy Forceps, and Cytology Brush. All three have the same technological characteristics as the respective predicates and reference devices, that is, they are intended to operate in the same manner for collection of tissue or cells in the patient's lungs through a flexible endoscope or other working channel.
The Aspirating Biopsy Needle is comprised of a polymeric shaft with a needle joined to the distal end. A Luer fitting at the proximal end of the shaft allows the user to provide suction to the needle via the supplied syringe. A handle mechanism interfaces with the shaft to provide extension and retraction of the needle, and an outer jacket provides protection when the needle is not in use.
The Cytology Brush is comprised of an outer sheath and an inner catheter assembly. The inner catheter assembly consists of a thumb ring at the proximal end and a shaft to connect to the distal end which terminates in a brush. When the catheter is inserted into a working channel, the distal brush would be in a retracted position inside the outer sheath. When the catheter is in position, the brush can be extended into the tissue to obtain samples by advancing the proximal thumb ring.
The Biopsy Forceps instrument is comprised of a coil pipe with a mechanical jaw assembly joined to the distal end. The jaws are opened by sliding the outer handle component distally relative to the inner, longer handle component. The jaws are closed by squeezing the two components together.
#### V. INTENDED USE/INDICATIONS FOR USE
The Monarch Endoscopy Platform (Monarch Platform) and its accessories are intended to provide bronchoscopic visualization of and access to patient airways for diagnostic and therapeutic procedures.
{7}------------------------------------------------
#### VI. COMPARISON OF TECHNOLOGICAL CHARACTERSTICS WITH THE PREDICATE AND REFERENCE DEVICE
Overall, the Monarch Platform, Karl Storz Bronchoscope and ARES are based on the following similar basic technological elements:
- Device contains a flexible bronchoscope to visualize and gain access to the A tracheobronchial tree.
- A Device is intended for procedures performed by physician users via access through the bronchoscope working channel.
- A Device requires continuous direct control by physician user to move the bronchoscope.
- A Device moves the distal tip by pulling wires.
- A Device enables use of working channel instruments to perform diagnostic and therapeutic procedures via working channel.
- A Monarch Bronchoscope System is sterile.
The Karl Storz bronchoscope lacks the electro-mechanical control mechanism and navigation included with the Monarch Platform. For this reason, we are using the cleared ARES and SuperDimension Navigation System as reference devices. The cleared reference devices have been selected to address issues of safety and effectiveness for the differing technologies.
The Aspirating Biopsy Needle and the PeriFLEX Transbronchial Aspiration Needle have similar technological characteristics. Both devices operate in the same manner to obtain a tissue biopsy using a bronchoscope. The primary difference between the subject and predicate devices is the handle design. Other differences between the subject and predicate include the working outer diameter, needle protrusion length, needle tip material, needle gauge, and the working length. There are no differences that raise different questions of safety and effectiveness relative to the predicate.
The Biopsy Forceps and the Olympus FB series biopsy forceps have the same intended use with minor technological differences, namely the jaw diameter and the working length; and do not raise new or different questions of safety and effectiveness relative to the predicate.
The Cytology Brush and Wang Bronchial Needle Brush have the same intended use and minor technological characteristic differences, namely brush length, brush diameter and working length; and do not raise new or different questions of safety and effectiveness relative to the predicate.
#### VII. PERFORMANCE DATA
The Monarch Platform has been subjected to and successfully tested for function, performance, and safety per FDA-recognized standards. In addition, the Monarch Platform has been evaluated for electrical safety and electromagnetic compatibility, biocompatibility and toxicity testing of patient-contacting materials. The following performance data was provided in support of the substantial equivalence determination.
{8}------------------------------------------------
### Biocompatibility testing
The biocompatibility evaluation for the direct and indirect body contacting components of the Monarch Platform - specifically the Monarch Bronchoscope System , working channel instruments, and accessories - was conducted in accordance with the FDA Blue Book Memorandum #95-1 "Use of International Standard ISO-10993, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing," May 1, 1995, and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA.
The Ethylene Oxide sterilized devices were validated as biocompatible with the appropriate selection of tests indicated for Externally Communicating Devices, in contact with Tissue/Bone/Dentin, with limited contact duration (less than 24 hours). Tests included Cytotoxicity, Sensitization, Irritation, Acute Systemic Toxicity and Material-Mediated Pyrogenicity.
The non-sterilized devices have been validated as biocompatible with the appropriate suite of biocompatibility tests indicated for Surface Devices, in contact with the Skin, with limited contact duration (less than 24 hours). Tests included Cytotoxicity, Sensitization, and Irritation.
### Electrical safety and electromagnetic compatibility (EMC)
Electrical safety and EMC testing were conducted on the Monarch Platform, consisting of the Monarch Tower, Monarch Cart, and Monarch Bronchoscope System. The system complies with the IEC 60601-1, and IEC 60601-2-18 standards for safety and the IEC 60601-1-2 standard for EMC.
### Software Verification and Validation Testing
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" issued May 11, 2005. Results of verification and validation testing confirm that the Monarch Platform conforms to design specifications and meets the needs of the intended users.
### Monarch Platform Performance Testing
- Monarch Cart
- Monarch Tower
- Robotic Arms and Instrument Device Manipulators ●
- Endoscopic Controller
- Vision
- Navigation
- Fluidics
- System Algorithms
- Bronchoscope and Sheath ●
- Working Channel Instruments and Accessories
- Simulated Use Testing ●
{9}------------------------------------------------
## Aspirating Biopsy Needle Performance Testing
- Puncture Force
- Insertabilty into Bronchoscope Working Channel Bronchoscope Deflection ●
- Positive/Negative Pressure and Leak Test ●
- Inadvertent Needle Deployment
- Minimum Needle Deployment
- Handle Assembly Strength and Durability
- Handle Actuation Torque and Force
- Nitinol Needle Buckling and Shear
- Catheter to Handle Joint Strength
- Nitinol Needle to Catheter Joint Strength
- Device Radiopacity ●
- Device Durability
- Gold Marker Adhesion
- Needle Targeting Accuracy
- Dimensional Testing ●
- Simulated Use
- Sterilization Validation ●
- Packaging and Shelf Life .
# Cytology Brush Performance Testing
- Insertabilty into Bronchoscope Working Channel ●
- Bronchoscope Deflection ●
- Inadvertent Brush Deployment
- Minimum Brush Deployment
- Ability to Cut with Scissors
- Atraumatic Ball Retention Strength
- Handle Assembly Strength and Durability
- Catheter to Handle Joint Strength
- Brush Head to Catheter Joint Strength
- Device Radiopacity
- Device Durability ●
- Dimensional Testing
- Simulated Use
- Sterilization Validation ●
- Packaging and Shelf Life
# Biopsy Forceps Performance Testing
- Insertabilty into Bronchoscope Working Channel ●
- Bronchoscope Deflection
- Handle Assembly Strength and Durability
- Catheter to Handle Joint Strength
- Distal Assembly Strength ●
- Jaw Opening Angle
{10}------------------------------------------------
- Device Radiopacity ●
- Device Durability ●
- Dimensional Testing ●
- . Simulated Use
- Sterilization Validation .
- Packaging and Shelf Life .
### Navigational Accuracy Comparative Study
A comparative study of the 'Navigation Yield' and 'Distance to Target' metrics of the Medtronic/Covidien superDimension™ System (superDimension™) against the Monarch Platform was completed. The Monarch Platform Navigation System and superDimension™ System have substantially equivalent segmental, sub-segmental accuracy and on target accuracy, and thus equivalent Navigation Yield.
### Animal and Cadaver Testing
Auris performed animal and cadaver testing to evaluate the Monarch Platform under simulated use conditions to validate the user needs, including the safety and effectiveness of the system as per its intended clinical use. These studies demonstrated that the Monarch Platform design meets the intended user requirements and facilitates safe and effective use.
### Usability/Human Factors Testing
Auris performed usability and human factors testing of the Monarch Platform. Simulated testing was performed in accordance with finalized guidance: "Applying Human Factors and Usability Engineering to Medical Devices" issued February 3, 2016. This testing assessed the Monarch Platform for safety and effective use by representative users during a simulated use bronchoscopy procedure after training on using the Monarch Platform. The testing demonstrated that Monarch Platform design meets the intended user requirements and facilitates safe and effective user interactions with little chance of committing dangerous user errors.
#### VIII. CONCLUSION
Based on the indications for use, technological characteristics, and performance testing, Auris has demonstrated that the Monarch Platform and Auris working channel instruments are as safe and effective as the predicate and reference devices for the stated intended use to collect tissue samples, and that the Monarch Platform and Auris' working channel instruments are, as a result, substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.