VACUETTE EVOPROTECT Safety Blood Collection / Infusion Set
K173757 · Greiner Bio-One Na, Inc. · FMI · Apr 19, 2018 · General Hospital
Device Facts
Record ID
K173757
Device Name
VACUETTE EVOPROTECT Safety Blood Collection / Infusion Set
Applicant
Greiner Bio-One Na, Inc.
Product Code
FMI · General Hospital
Decision Date
Apr 19, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5570
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The VACUETTE EVOPROTECT Safety Blood Collection/Infusion Set is indicated for venous blood collection and/or the short-term infusion of intravenous fluids. It is to be used by appropriately trained healthcare professionals in accordance with the instructions.
Device Story
Single-use, sterile, winged needle device for venous blood collection or short-term IV fluid infusion. Features flexible tubing with Luer connector and semi-automatic safety mechanism. Operated by trained healthcare professionals in clinical settings. Safety mechanism activated manually by user; spring-driven motion covers needle immediately post-venipuncture to prevent accidental needlestick injuries. Luer port allows connection to standard accessories like holders or adapters. Device provides physical protection for clinicians; facilitates safe blood draw and fluid delivery.
Clinical Evidence
Bench testing only. Performance verified against ISO 7864, ISO 23908, ISO 9626, ISO 8536-4, and ASTM standards. Biocompatibility confirmed per ISO 10993 series. Sterility validated per ISO 11137-1. All tests passed.
Technological Characteristics
Winged needle (AISI 304 stainless steel) with flexible PVC tubing. Components: PP needle cap, PC protector, PMMA hub, ABS stopper/connector, TPE wing. Semi-passive safety mechanism uses AISI 301 stainless steel spring. Sterilized via irradiation (SAL 10^-6). Complies with ISO 7864, 23908, 9626, 8536-4, 11137-1, 10993, and ASTM F2096/F1929.
Indications for Use
Indicated for venous blood collection and/or short-term intravenous fluid infusion. For use by trained healthcare professionals.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
Predicate Devices
VACUETTE SAFETY BLOOD COLLECTION SET & SAFETY INFUSION SET (K121908)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name, "U.S. Food & Drug Administration," written in a sans-serif font.
April 19, 2018
Greiner Bio-One NA Inc Manfred Abel Quality Systems and Regulatory Affairs 4238 Capital Drive Monroe, North Carolina 28110
Re: K173757
Trade/Device Name: VACUETTE EVOPROTECT Safety Blood Collection / Infusion Set Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: Class II Product Code: FMI Dated: March 12, 2018 Received: March 13, 2018
Dear Manfred Abel:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Geeta K. Pamidimukkala -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K173757
Device Name
VACUETTE EVOPROTECT Safety Blood Collection/Infusion Set
Indications for Use (Describe)
The VACUETTE EVOPROTECT Safety Blood Collection/Infusion Set is indicated for venous blood collection and/or the short-term infusion of intravenous fluids. It is to be used by appropriately trained healthcare professionals in accordance with the instructions.
Type of Use (Select one or both, as applicable)
| <span></span> <label><input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> |
|-------------------------------------------------------------------------------------------------------------------|
| <span></span> <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### SECTION 5. PREMARKET NOTIFICATION 510(K) SUMMARY: K173757
#### 1. SUBMITTER Applicant Name: Greiner Bio-One GmbH. Bad Haller Strasse 32 4550 Kremsmuenster, Austria
- Manfred Abel, M.S., MBA Contact person: Greiner Bio-One NA Inc. 4238 Capital Drive Monroe, NC 28110 704 261 7823 Manfred.Abel@gbo.com
### Establishment registration number: 8020040
Date prepared: Apr. 17, 2018
- 2. DEVICE
VACUETTE® EVOPROTECT Safety Blood Collection/Infusion Set Trade Name:
- Blood Collection Set Common name:
Classification: Needle, hypodermic, single lumen Name: Product Code: FMI Regulation No: 880.5570 Class: ನ Review Panel: General Hospital
### 3. PREDICATE DEVICE
VACUETTE SAFETY BLOOD COLLECTION SET & SAFETY Predicate: INFUSION SET (K121908)
### 4. DEVICE DESCRIPTION
The VACUETTE® EVOPROTECT SAFETY Blood Collection/Infusion Set is a singleuse, sterile, winged needle bonded to flexible tubing with a Luer connector and a semiautomatic release of the safety mechanism. The winged needle is designed with a safety mechanism, which allows for activation ensuring the needle is covered immediately following venipuncture to aid in the protection against accidental needlestick injury. The EVOPROTECT Safety Blood Collection/Infusion Set is individually wrapped, sterile with a luer port. The luer port can be used to connect FDA cleared accessories like luer adapter, holder, etc.
{4}------------------------------------------------
# 5. INDICATION FOR USE
The VACUETTE EVOPROTECT Safety Blood Collection/Infusion Set is indicated for venous blood collection and/or the short-term infusion of intravenous fluids. It is to be used by appropriately trained healthcare professionals in accordance with the instructions.
## 6. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICES
The subject device and the predicate device have the same fundamental technology, technological characteristics, and intended use.
Based on the device features, principles of operation and technological characteristics, the VACUETTE® EVOPROTECT Safety Blood Collection/Infusion Set is substantially equivalent to the predicate device.
| | NEW GBO Product<br>VACUETTE®<br>EVOPRTOTECT Safety<br>Blood Collect/ Infusion<br>set | Predicate Device<br>VACUETTE® Safety Blood<br>Collection / Infusion set | Comparison | |
|--------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|--|
| FDA Status | Under review | K121908 | NA | |
| Classification | Class II | Class II | same | |
| Regulation | 880.5570 | 880.5570 | same | |
| Classification<br>Product Code | FMI | FMI | same | |
| Intended Use | The VACUETTE<br>EVOPROTECT Safety<br>Blood Collection/Infusion<br>Set is indicated for<br>venous blood collection<br>and/or the short-term<br>infusion of intravenous<br>fluids. It is to be used by<br>appropriately trained<br>healthcare professionals<br>in accordance with the<br>instructions. | The SAFETY Blood<br>Collection/Infusion Set is a<br>single-use, sterile, winged<br>blood collection needle<br>bonded to a flexible tubing<br>with a luer connector. The<br>Vacuette Safety Blood<br>Collection/Infusion Set is<br>used for blood collection<br>and/or the short-term<br>infusion of intravenous<br>fluids. The winged needle is<br>designed with a safety<br>shield, which can be<br>activated to cover the<br>needle immediately<br>following venipuncture to<br>aid in the protection against<br>accidental needlestick<br>injury. | same | |
| Sterile | yes | yes | same | |
| | NEW GBO Product<br>VACUETTE®<br>EVOPRTOTECT Safety<br>Blood Collect/ Infusion<br>set | Predicate Device<br>VACUETTE® Safety Blood<br>Collection / Infusion set | Comparison | |
| Device Characteristics | | | | |
| Safety<br>mechanism | Semi-passive safety<br>mechanism.<br>One-hand activation. | Active safety mechanism.<br>One-hand activation. | equivalent | |
| | Manual operation for<br>activation (i.e. release).<br>Afterwards spring force<br>driven motion in safe<br>position. | Manual operation for<br>activation (i.e. release).<br>Afterwards manual driven<br>motion in safe position. | equivalent | |
| Cannula<br>Gauges | 21/23/25 | 21/23/25 | same | |
| Cannula Length | 3/4" | 3/4" | same | |
| Sterile | Yes | Yes | same | |
| Storage | 4 - 25 C° | 4 - 36 C° | equivalent | |
| Use | Single use only | Single use only | same | |
| Tubing Lengths | 4"; 7½"; 12" | 4"; 7½"; 12" | same | |
| Materials:<br>All data as detailed and as known: | | | | |
| Needle Cap | PP | PP | same | |
| Needle | Stainless Steel AISI 304 | Stainless Steel AISI 304 | same | |
| Wing | TPE | PVC | equivalent | |
| Protector | PC | PC | same | |
| Hub | PMMA | PMMA | same | |
| Activation<br>spring / manual | Stainless Steel Spring<br>AISI 301 | manual | equivalent | |
| Stopper | ABS | ABS | same | |
| Tubing | PVC with plasticizer | PVC with plasticizer | same | |
| Connector | ABS | ABS | same | |
| Connector Cap | PP | PP | same | |
| Adhesive | UV cured adhesive | Solvent bonding | equivalent | |
| Packaging | Blister | Pouch | equivalent | |
{5}------------------------------------------------
The above table shows that the VACUETTE® EVOPROTECT Safety Blood Collection/Infusion Set is equivalent to the predicate device. The differences in storage, wing material, activation, adhesive and packaging is equivalent as demonstrated in the performance testing.
{6}------------------------------------------------
# 7. PERFORMANCE DATA
The VACUETTE® EVOPROTECT Safety Blood Collection/Infusion Set is manufactured and tested in conformity with a variety of recognized consensus standards. Performance data show that the device performs as intended and in an equivalent manner as the predicate devices.
| Requirement | Acceptance | Results pass/fail |
|-----------------------------------------------------------------------------|----------------------------------------------------------------------------------|----------------------|
| Penetration resistance Cap | Penetration resistance >3N for 21G, 23G and 25G. Equivalent to predicate device. | pass |
| Cannula cleanliness | Tested in accordance with to ISO 7864:2014 | pass |
| Cannula radial orientation | Equivalent to predicate device. | pass |
| Cannula puncture & friction force 21G | Equivalent to predicate device | pass |
| Cannula puncture & friction force 23G | Equivalent to predicate device | pass |
| Cannula puncture & friction force 25G | Equivalent to predicate device | pass |
| Cannula bond force 21G | Tested in accordance with ISO 7864:2016 | pass |
| Cannula bond force 23G | Tested in accordance with ISO 7864:2016 | pass |
| Cannula bond force 25G | Tested in accordance with ISO 7864:2016 | pass |
| Flow rate water 10cm tubing, 21G (extreme: max flow rate over all variants) | Equivalence, internal Specification | pass |
| Flow rate water 19 cm tubing, 21G | Equivalence, internal Specification | pass |
| Flow rate water 19 cm tubing, 23G | Equivalence, internal Specification | pass |
| Flow rate water 30 cm tubing, 23G | Equivalence, internal Specification | pass |
| Flow rate water 30cm tubing, 25G (extreme: min flow rate over all variants) | Equivalence, internal Specification | pass |
| Breaking compressive force cannula before activation | Tested in accordance with ISO 23908:2011 | pass |
| Size designation | Tested in accordance with ISO 9626:2016 | pass |
| Dimensional | Tested in accordance with ISO 9626:2016 | pass |
| Resistance for stiffness | Tested in accordance with ISO 9626:2016 | pass |
| Resistance for breakage | Tested in accordance with ISO 9626:2016 | pass |
| Resistance for corrosion | Tested in accordance with ISO 9626:2016 | pass |
| Requirement | Acceptance | Results<br>pass/fail |
| Activation force safety<br>mechanism | Tested in accordance with ISO<br>23908:2016 | pass |
| Breaking force safety mechanism<br>compression | Tested in accordance with ISO<br>23908:2011 | pass |
| Breaking force safety mechanism<br>tension | Tested in accordance with ISO<br>23908:2011 | pass |
| Bond strength Hub to Stopper | Equivalence,<br>ISO 8536-4:2010 + Amd 1:2013 | pass |
| Bond strength Tubing to Stopper | Equivalence,<br>ISO 8536-4:2010 + Amd 1:2013 | pass |
| Bond strength Tubing to<br>Connector | Equivalence,<br>ISO 8536-4:2010 + Amd 1:2013 | pass |
| Tubing breaking force | Equivalence,<br>ISO 8536-4:2010 + Amd 1:2013 | pass |
| Detaching torque Luer Adapter<br>(LA) to Connector | Equivalent to predicate device | pass |
| Detaching force Luer Adapter to<br>Connector | Equivalent to predicate device | pass |
| Spring characteristic | Equivalent to ISO 23908:2011,<br>simulated use | pass |
| Leakage of product (pressure &<br>vacuum) | Equivalent to predicate device | pass |
| Simulated use | According to intended use and<br>specification | pass |
| Tightness single packaging<br>(Bubble test) | According to ASTM F2096-04 | pass |
| Tightness single packaging (Dye<br>penetration) | According to ASTM F1929-15, ISO<br>11607-2 and EN 868-5 | pass |
| Sealing strength and max.<br>opening force single packaging | According to ISO 11607-1 and EN 868-<br>5 | pass |
| Biocompatibility testing | Testing according to ISO 10993-1 and<br>FDA guidance | pass |
| Sterility testing | Testing according to 11137-1 | pass |
{7}------------------------------------------------
# Sterility testing
VACUETTE® EVOPROTECT Safety Blood Collection/ Infusion Sets are sterilized by irradiation for a final Sterility Assurance Level (SAL) of 10-6 in accordance with ISO 11137-1 Sterilization of health care products. Bacterial endotoxin testing was performed by according to ANSI/AAMI ST72:2011.
### Biocompatibility testinq
Biocompatibility for the VACUETTE® EVOPROTECT Safety Blood Collection/lnfusion Set was performed in accordance with standards of the ISO 10993 series and others to demonstrate that the subject device meets requirements of the following standards:
- Cytotoxicity (ISO 10993-5, USP 38: 2016) ●
- . Sensitization (ISO 10993-10:2010, ISO 10993-12 2012)
- Irritation (ISO 10993-10) .
- . Acute Systemic Toxicity (ISO 10993-11, USP 39 and ASTM F 750-87)
- Pyrogenicity (USP 39<151>) ●
{8}------------------------------------------------
- Hemocompatibility (ISO 10993-4, ASTM F756) .
- . Particulate Matter (USP 788)
- Genotoxicity characterization of extractable (ISO 10993-3) .
### 8. CONCLUSION
Results from the performance verification and validation show that the EVOPROTECT Safety Blood Collection/Infusion Set meets established performance criteria and performance is as intended and substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.