Clinical patients (implied by retrospective clinical data)
Not applicable for this study
Verification of diffusion and perfusion analysis tools
Indications for Use
ProFuse CAD is a post-processing software tool intended for viewing, reviewing and reporting imaging files such as Magnetic Resonance Imaging (MR), Computed Tomography (CT) and Positron Emission Tomography (PET) studies. The software is able to process time series modality data acquired before, during, and after a contrast agent has been administered to the patient. The software allows a physician to evaluate tissue characteristics based on contrast enhancement visible on the time series data can also be processed to get subtraction image time -series with any reference time point. ProFuse CAD also provides capability to process a specialized MRI series called diffusion weighted imaging to evaluate tissue characteristics based on water diffusion. Some of the other posts -processing features include multi-planar reformats and registration between images. ProFuse CAD has a marking feature that allows the user to outline organs with minimal input and also annotate regions in the image in 3D. The software also allows for grading the annotated regions as defined by the user. After review, the software automatically generates a patient report that provides information like; organ volume, annotations; and grading along with automatically generated screenshots of the annotations in different images. Planned data from ProFuse CAD can be used for interventional procedures like MR - TRUS fusion biopsies. The data can be displayed on medical device data systems. ProFuse CAD is used as a tool to review and add to the results of interventional procedures like; adding pathology information when reviewing MR-TRUS fusion biopsy. When interpreted by a skilled physician, this device provides information that is intended to be used for screening, analysis, and interventional planning. Patient management decisions should not be made based solely on the results of ProFuse CAD. ProFuse CAD is intended to be used as an image viewer of multi-modality, digital images, including Ultrasound, CT, PET. ProFuse CAD is intended to be used in a variety of setting such as; medical offices, clinics, hospital, and home office.
Device Story
ProFuse CAD is standalone software for PC-based workstations; used by radiologists/physicians in clinics, hospitals, or offices. Inputs include multi-modality DICOM images (MRI, CT, PET, Ultrasound). Software performs organ segmentation, 3D annotation, multi-planar reformatting, and image registration. Features include time-series analysis (pharmacokinetic modeling for contrast-enhanced MRI) and diffusion-weighted imaging analysis (mathematical model fitting for parametric maps). Output includes 3D visualizations, parametric maps, and automated patient reports (organ volume, annotations, grading). Used for interventional planning (e.g., MR-TRUS fusion biopsies) and reviewing/adding pathology data. Physician interprets output to support clinical decision-making; device is not for sole diagnostic use.
Clinical Evidence
Bench testing only. Verification and validation of all input/output functions and operational modes performed. Volume estimation accuracy validated using test phantoms with simulated prostates. Diffusion and perfusion analysis tools verified using simulated and retrospective clinical data. No prospective clinical trials reported.
Technological Characteristics
Standalone PC-based software. Processes DICOM-format medical images. Features include multi-planar reformatting, image registration, pharmacokinetic modeling, and mathematical model fitting for diffusion-weighted imaging. Supports PIRADS or customizable reporting standards. Connectivity via standard DICOM interfaces for image storage/retrieval.
Indications for Use
Indicated for use by physicians as a post-processing tool for viewing, analyzing, and reporting multi-modality medical images (MRI, CT, PET, Ultrasound) to assist in screening, analysis, and interventional planning, including MR-TRUS fusion biopsies.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 21, 2018
Eigen William Mandel Director of Operations, Regulatory Affairs, and Quality Assurance 13366 Grass Valley Ave. GRASS VALLEY, CA 95945
Re: K173744
Trade/Device Name: ProFuse CAD Regulation Number: 21 CFR 892.2050 Regulation Name: Picture Archiving And Communications System Regulatory Class: Class II Product Code: LLZ Dated: September 28, 2018 Received: October 10, 2018
Dear William Mandel:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Hse 2. Nils
for Robert A. Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K173744
Device Name ProFuse CAD
### Indications for Use (Describe)
ProFuse CAD is a post-processing software tool intended for viewing, reviewing and reporting imaging files such as Magnetic Resonance Imaging (MR), Computed Tomography (CT) and Positron Emission Tomography (PET) studies.
The software is able to process time series modality data acquired before, during, and after a contrast agent has been administered to the patient. The software allows a physician to evaluate tissue characteristics based on contrast enhancement visible on the time series data can also be processed to get subtraction image time -series with any reference time point.
ProFuse CAD also provides capability to process a specialized MRI series called diffusion weighted imaging to evaluate tissue characteristics based on water diffusion. Some of the other posts -processing features include multi-planar reformats and registration between images.
ProFuse CAD has a marking feature that allows the user to outline organs with minimal input and also annotate regions in the image in 3D. The software also allows for grading the annotated regions as defined by the user. After review, the software automatically generates a patient report that provides information like; organ volume, annotations; and grading along with automatically generated screenshots of the annotations in different images.
Planned data from ProFuse CAD can be used for interventional procedures like MR - TRUS fusion biopsies. The data can be displayed on medical device data systems.
ProFuse CAD is used as a tool to review and add to the results of interventional procedures like; adding pathology information when reviewing MR-TRUS fusion biopsy.
When interpreted by a skilled physician, this device provides information that is intended to be used for screening, analysis, and interventional planning. Patient management decisions should not be made based solely on the results of ProFuse CAD.
ProFuse CAD is intended to be used as an image viewer of multi-modality, digital images, including Ultrasound, CT, PET.
ProFuse CAD is intended to be used in a variety of setting such as; medical offices, clinics, hospital, and home office.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the text "leading innovation in medical imaging". The text is arranged in two lines, with "leading innovation" on the first line and "in medical imaging" on the second line. The text is in a sans-serif font and is gray in color. The background is white.
Image /page/4/Picture/1 description: The image shows the word "eigen" in a sans-serif font. The word is white and is set against a teal background. The letters are evenly spaced and the overall design is simple and modern.
### 510(k) Summary
| Submitter's Name: | Eigen |
|----------------------------|------------------------------------------------------|
| Submitter's Address: | 13366 Grass Valley Avenue, Grass Valley, CA 95945 |
| Submitter's Telephone: | 530-274-1240 x237 |
| Contact Name: | William Mandel |
| Date Summary was Prepared: | 3 Aug 2017 |
| Common or Usual Name: | ProFuse Computer Aided Detection |
| Trade or Proprietary Name: | ProFuse CAD |
| Classification Name: | System, Image Processing, Radiological, LLZ |
| | Picture Archiving and Communications, 21CFR 892.2050 |
#### Device Description
ProFuse CAD is standalone Computer Aided Detection (CAD) software that is used by radiologists to visualize, analyze, plan and interpret medical images using tools available in the software. Some of the tools in the software are
- Visualization: The software allows simultaneous visualization of different images with the ability to view them in different orientations e.g. transverse, sagittal, coronal and 3D.
- Organ segmentation: The software allows user to outline the organ of interest (e.g. prostate)
- Image annotation: The software allows user to mark regions of interest (ROI) on images
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the word "eigen" in a sans-serif font. The word is white and is set against a teal background. The letters are evenly spaced and the word is centered in the image.
- Time-series analysis: The software allows user to view the time-curve plots for a single location or average over a region (ROI). The user could also calculate different parametric maps based on pharmacokinetic modeling. The parametric maps could be viewed as a separate series or be overlaid as color on another series. The software also allows users to calculate subtracting images from a customizable user defined time point.
- Diffusion series analysis: Diffusion weighted series is a special type of magnetic resonance imaging (MRI) sequence that measures diffusion of water molecules in the body. A set of different diffusion weighted images are obtained and from the set of images the software allows the user to fit different mathematical models to extract different parametric maps. The parametric maps could be viewed as a separate series or be overlaid as color on another series.
- Report: The software allows the user to create reports following either the PIRADS (see definition) standard or a customizable standard.
- Export: The software allows the planned images to be exported for different procedures e.g. biopsy, therapy, etc. Information exported by ProFuse CAD could then be used in other software solutions like ProFuse Bx, ProFuse FP.
- Review: Review, add and modify data from the 3D visualization software (Artemis). The data includes acquired volume, organ segmentation, planned and recorded biopsy location and pathology information.
{6}------------------------------------------------
leading innovation in medical imaging
Image /page/6/Picture/1 description: The image features the word "eigen" in a sans-serif font. The word is white and is set against a teal background. The letters are evenly spaced and the overall design is simple and modern.
### Primary and Referenced Predicate Devices
| | DynaCAD<br>V1.0<br>(K041286) | iCAD<br>(K091529)<br>(K121517) | ProFuse<br>(K120187) | |
|-----------------------------|------------------------------|--------------------------------|----------------------|-------------|
| Properties | Primary | Referenced | Referenced | ProFuse CAD |
| Visualization | X | X | X | X |
| Organ segmentation | X | X | X | X |
| Image annotation | X | X | X | X |
| Time-series analysis | X | X | | X |
| Diffusion - series analysis | X | X | | X |
| Report | X | X | X | X |
| Export | X | X | X | X |
| Review | X | X | X | X |
#### Intended Use
ProFuse CAD is a post-processing software tool intended for viewing, analyzing, reviewing and reporting imaging files such as Magnetic Resonance Imaging (MRI), Computed Tomography (CT) and Positron Emission Tomography (PET) studies.
The software is able to process Dynamic Contrast Enhanced (DCE) MRI dataset to estimate the pharmacokinetic modeling parameters estimated allows a physician to evaluate tissue characteristics. The DCE data can also be processed to get subtraction image time –series with any reference time point.
ProFuse CAD also provides capability to process a specialized MRI series called diffusion weighted imaging to evaluate tissue characteristics based on water diffusion. Some of the other posts –processing features include multi-planar reformats and registration between images. EIGEN PREMARKET NOTIFICATION – PROFUSE CAD CONFIDENTIAL ર-૩
{7}------------------------------------------------
leading innovation in medical imaqinq
Image /page/7/Picture/1 description: The image features the word "eigen" in a sans-serif font. The word is rendered in a light blue or cyan color, contrasting with the background, which is a darker shade of blue or teal. The letters are evenly spaced and appear to be well-defined, creating a clean and legible visual.
ProFuse CAD has a marking feature that allows the user to outline organs and also annotate regions in the image in 3D. The software also allows for grading the annotated regions as defined by the user. After review, the software generates a patient report that provides information like organ volume and region of interest assessment categories along with screenshots of annotations in different images.
Planned data from ProFuse CAD can be used for interventional procedures like MR – TRUS fusion biopsies. The data can be displayed on medical device data systems.
ProFuse CAD is used as a tool to review and add to the results of interventional procedures like; adding lab pathology information when reviewing MR-TRUS fusion biopsy.
When interpreted by a skilled physician, this device provides information that is intended to be used for screening, analysis, and interventional planning. Patient management decisions should not be made based solely on the results of ProFuse CAD.
ProFuse CAD is intended to be used as an image viewer of multi-modality, digital images, including Ultrasound, CT, PET.
ProFuse CAD is intended to be used in a variety of setting such as; medical offices, clinics, hospital, and home office.
#### Substantial Equivalence:
The product's technical features are substantially equivalent to the DynaCAD V1.0 (K041286) and iCAD (K091529) (K121517). All the features of the ProFuse (K120187) software are included in the ProFuse CAD software. ProFuse CAD is a software product that runs on PC-based workstations. Image data is input to the devices and used to generate 3-D views and perform image processing. Like three primary and referenced predicate devices, the software has image measurement, multi-planar reformatting, segmentation and image registration abilities, fusion of images from different modalities, image storage and retrieval, as well as patient information management functions.
DynaCAD V1.0 and iCAD is software products that accept multiple image data types including magnetic resonance, computed tomography, and ultrasound. The ProFuse has been designed for processing medical images in standard DICOM format.
{8}------------------------------------------------
leading innovation in medical imaging
Image /page/8/Picture/1 description: The image shows the word "eigen" in a sans-serif font. The word is light blue and is set against a teal background. The letters are evenly spaced and the word is horizontally oriented.
ProFuse CAD is build off our previous version, which was 510(k) cleared and contains all of the features in that version. The previous version had the trade name of ProFuse and was all so know as Multi-Modality Image Fusion (MMIF).
## Testing and Performance Data:
All product and engineering specifications were verified and validated. Test phantoms incorporating simulated prostates were used only for volume estimation accuracy studies. Simulated and retrospective clinical data were used for verification of the diffusion and perfusion analysis tools.
# Safety and Effectiveness
ProFuse CAD labeling contains instructions for use and necessary cautions, warnings and notes to provide for safe and effective use of the device. Risk Management is ensured via ProFuse CAD IRisk Management procedure, which is used to identify potential hazards. These potential hazards are controlled via the product (software) development process, verification and validation testing.
# Nonclinical Testing and Performance Information
Nonclinical and performance testing has been performed by designated individuals as required by Eigen's quality procedures. Verification & Validation Test Plans were designed to evaluate all input functions, output functions, and actions performed by ProFuse CAD in each operational mode. ProFuse CAD has been assessed and tested at the manufacturer's facility and has passed all in-house testing criteria including validating design, function and specifications. Measurement validation using, phantoms, clinical images were used to show that ProFuse CAD preforms as well as or better than the other primary and referenced predicate devices and furthermore shows that ProFuse CAD was safe and effective. Nonclinical and performance testing results are provided in the 510(k) and demonstrate that the predetermined acceptance criteria are met. ProFuse CAD has been designed to comply with applicable standards.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the words "leading innovation in medical imaging" in gray font. To the right of the words is a teal square with the word "eigen" in a light teal font. The image is a logo for a medical imaging company.
Image /page/9/Picture/1 description: The image shows the word "eigen" in a light blue font against a teal background. The letters are sans-serif and appear to be a clean, modern typeface. The word is horizontally oriented and fills a significant portion of the frame.
# Conclusion:
The results of comparing the intended use, function, technological characteristics, mode of operation and specifications of ProFuse CAD with those of the other primary and referenced predicate devices demonstrates that ProFuse CAD is substantially equivalent to existing products on the market today.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.